← Product Code [DQA](/productcode/DQA) · K201384

# Pulse Oximeter (K201384)

_Shenzhen Raysintone Technology Co., Ltd. · DQA · Dec 7, 2020 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K201384

## Device Facts

- **Applicant:** Shenzhen Raysintone Technology Co., Ltd.
- **Product Code:** [DQA](/productcode/DQA.md)
- **Decision Date:** Dec 7, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2700
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Fingertip Pulse Oximeter is a reusable non-invasive device intended for the spot checking of oxygen saturation of arterial hemoglobin (SpO2) and the pulse rate of adult patients in hospital and other healthcare environments. And it is not intended to be used under motion or low perfusion scenarios.

## Device Story

Fingertip pulse oximeter; reusable; non-invasive. Uses red and infrared LEDs and a photodetector to measure light absorption through finger tissue. Microprocessor calculates SpO2 and pulse rate based on plethysmographic waveforms derived from varying light absorption of oxygenated/deoxygenated hemoglobin. Operated by clinicians in hospital/healthcare settings for spot checks. Displays results (SpO2, pulse rate, bar graph, waveform) on LED or OLED screen. Provides immediate physiological data to assist clinical decision-making regarding patient oxygenation status. Powered by 2x1.5V AAA batteries.

## Clinical Evidence

Clinical testing conducted per ISO 80601-2-61:2011 Annex EE to evaluate SpO2 accuracy in human subjects. Bench testing confirmed compliance with ISO 80601-2-61, IEC 60601-1, and IEC 60601-1-2. Biocompatibility (cytotoxicity, sensitization, irritation) per ISO 10993-1. Cleaning validation performed per FDA guidance.

## Technological Characteristics

Reusable fingertip pulse oximeter; 2-wavelength optical absorption sensing; LED/OLED display; 2x1.5V AAA battery power. Complies with ISO 80601-2-61, IEC 60601-1, IEC 60601-1-2, and ISO 10993-1. Dimensions vary by model (57.5x32.5x32mm or 64x38.5x36.5mm).

## Regulatory Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

## Predicate Devices

- Shenzhen Yimi Life Technology Co., Ltd., YM101/YM201 Pulse Oximeter ([K191430](/device/K191430.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 7, 2020

Shenzhen Raysintone Technology CO., Ltd Xingbao Yang General Manager Zone C, 4th Floor, F Building, Xinwei Second Industrial Zone Guangming District Shenzhen, Guangdong 518106 China

Re: K201384

Trade/Device Name: Fingertip Pulse Oximeter - models M130, M130A, M130B, M130C, M160A, M160C M170A, M170C, M230, M230A, M230B and M230C Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: November 1, 2020 Received: November 9, 2020

Dear Xingbao Yang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K201384

#### Device Name

Fingertip Pulse Oximeter, models M130A, M130C, M160A, M160C M170A, M170C, M230, M230A, M230B and M230C

#### Indications for Use (Describe)

The Fingertip Pulse Oximeter is a reusable non-invasive device intended for the spot checking of oxygen saturation of arterial hemoglobin (SpO2) and the pulse rate of adult patients in hospital and other healthcare environments. And it is not intended to be used under motion or low perfusion scenarios.

| Type of Use (Select one or both, as applicable)                                       |                                                                                                                                                                                                                              |                                                                                       |                                                                                      |
|---------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
|                                                                                       | <table><tr><td><span style="font-size: 1.2em;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D)</td></tr><tr><td><span style="font-size: 1.2em;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</td></tr></table> | <span style="font-size: 1.2em;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 1.2em;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
| <span style="font-size: 1.2em;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) |                                                                                                                                                                                                                              |                                                                                       |                                                                                      |
| <span style="font-size: 1.2em;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)  |                                                                                                                                                                                                                              |                                                                                       |                                                                                      |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

# Prepared in accordance with the requirements of 21 CFR Part 807.92

| 1. Submitter:                        | Shenzhen Raysintone Technology Co., Ltd.<br>Zone C, 4th Floor, F Building, Xinwei Second Industrial Zone,<br>Guangming New District, Shenzhen, 518106, P.R. China<br>TEL: +86 755 27407409                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                              |                                                                |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|----------------------------------------------------------------|
| Contact Person:                      | Xingbao Yang                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                              |                                                                |
| Prepare date:                        | 2020-11-01                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                              |                                                                |
| 2. Device name<br>and classification | Device Name: Fingertip Pulse Oximeter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                              |                                                                |
|                                      | Models: M130, M130A, M130B, M130C, M160A, M160C, M170A,<br>M170C, M230, M230A, M230B and M230C                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                              |                                                                |
|                                      | Classification Name: 21 CFR 870.2700 Oximeter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                              |                                                                |
|                                      | Product code: DQA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                              |                                                                |
|                                      | Regulatory Class: Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                              |                                                                |
| 3. Reason for<br>Submission          | New Application. No prior submission associated with the current<br>submission.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                              |                                                                |
| 4. Predicate<br>Device(s)            | Shenzhen Yimi Life Technology Co., Ltd., YM101/YM201 Pulse<br>Oximeter / K191430                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                              |                                                                |
| 5. Device<br>Description             | The oximeter consists of probe, electronic circuits, and display and<br>plastic enclosures. And one side of probe is designed to locate light<br>emitting diodes and a light detector (called a photo-detector). Red and<br>Infrared lights are shone through the tissues from one side of the probe<br>to the other. Then parts of the light emitted absorbed by blood and<br>tissues. The light absorbed by the blood varies with the oxygen<br>saturation of haemoglobin. After that, the photo-detector detects the light<br>volume transmitted through the tissues which depends on blood pulse,<br>Hereafter, the microprocessor calculates a value for the oxygen<br>saturation (SpO2).<br>The subject device is a reusable device, and need to reprocess as<br>suggested in the user manual after each use. And the device is intended<br>to be used on the finger, and powered by 2*1.5V AAA battery.<br>M230, M230A, M230B and M230C display the measuring results on 1.3"<br>LED screen, and the screen of M130, M130A, M130B, M130C, M160A,<br>M160C, M170A and M170C are 0.96" OLED. The configuration of the<br>models is shown in the table below. |                              |                                                                |
|                                      | Function                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | M230, M230A,<br>M230B, M230C | M130, M130A,<br>M130B, M130C,<br>M160A, M160C,<br>M170A, M170C |
|                                      | Display                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | LED                          | OLED                                                           |
| measurement                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                              |                                                                |
| Pulse rate measurement               | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Yes                          |                                                                |
| Bar graph display                    | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Yes                          |                                                                |
| Battery display                      | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Yes                          |                                                                |
| Automatic power-off function         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Yes                          |                                                                |
| Pulse beat sound                     | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                          |                                                                |
| Sound On/Off function                | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                          |                                                                |
| Pulse waveform display               | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                          |                                                                |
| Direction display                    | 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 4                            |                                                                |
| Data Record                          | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                          |                                                                |

Yes

Spo2 parameter

Yes

{4}------------------------------------------------

#### Scientific concepts that form the basis for the device

Pulse oximeter sensors have red and infrared low voltage light emitting diodes which serve as light sources. The emitted light is transmitted through the tissue, then detected by the photodetector and sent to the microprocessor of the pulse oximeter. All constituents of the human body, venous and arterial blood, and tissue absorb light. The pulsating of arterial blood results in changes in the absorption to added hemoglobin (Hb) and oxygenated hemoglobin (HbO2) in the path of the light. Since HbO2 and Hb absorb light to varving degrees, this varying absorption is translated into plethysmographic waveforms at both red and infrared wavelengths. The relationship of red and infrared plethysmographic signal amplitude can be directly related to arterial oxygen saturation.

Significant physical and performance characteristics of the device, please refer to the column Proposed Device in Table 1 for details.

#### 6. Indications for The Fingertip Pulse Oximeter is a reusable non-invasive device intended Use for the spot checking of oxygen saturation of arterial hemoglobin (SpO2) and the pulse rate of adult patients in hospital and other healthcare environments. And it is not intended to be used under motion or low perfusion scenarios.

## 7. Predicate Device Comparison

Please refer to following table to find differences between the subject device and predicate device.

| ITEM                                    | Predicate Device<br>YM 101/YM201 Pulse<br>Oximeter<br>K191430                                                                                                                                                                                                                          | Proposed Device<br>M130, M130A, M130B, M130C,<br>M160A, M160C M170A,<br>M170C,M230, M230A, M230B,<br>M230C                                                                                                                                                                                                                                    | Comparison<br>Result  |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|
| Manufacture                             | Shenzhen Yimi<br>Life Technology Co., Ltd.                                                                                                                                                                                                                                             | Shenzhen Raysintone<br>Technology Co., Ltd.                                                                                                                                                                                                                                                                                                   | ---                   |
| Indications for<br>Use                  | The pulse oximeter is a<br>reusable device and<br>intended for spot-checking<br>of oxygen saturation and<br>pulse rate for use with the<br>finger of adult patients in<br>healthcare environments.<br>And it is not intended to be<br>used under motion or low<br>perfusion scenarios. | The Fingertip Pulse Oximeter is<br>a reusable non-invasive device<br>intended for the spot checking<br>of oxygen saturation of arterial<br>hemoglobin (SpO2) and the<br>pulse rate of adult patients in<br>hospital and other healthcare<br>environments. And it is not<br>intended to be used under<br>motion or low perfusion<br>scenarios. | Different             |
| Contraindications                       | Not intended to be used<br>under motion or low<br>perfusion scenarios.                                                                                                                                                                                                                 | Not intended to be used under<br>motion or low perfusion<br>scenarios.                                                                                                                                                                                                                                                                        | Same                  |
| Operational Specifications              |                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                               |                       |
| Intended patient<br>population          | Adult                                                                                                                                                                                                                                                                                  | Adult                                                                                                                                                                                                                                                                                                                                         | Same                  |
| Intended<br>application site            | Finger                                                                                                                                                                                                                                                                                 | Finger                                                                                                                                                                                                                                                                                                                                        | Same                  |
| Measurement<br>Principles               | 2-wavelength Relative<br>Optical Absorption                                                                                                                                                                                                                                            | 2-wavelength Relative<br>Optical Absorption                                                                                                                                                                                                                                                                                                   | Same                  |
| Signal Detection<br>Method              | Photodetector                                                                                                                                                                                                                                                                          | Photodetector                                                                                                                                                                                                                                                                                                                                 | Same                  |
| Display content                         | YM101:<br>SpO2%, PR, battery<br>indicator, Pulse rate bar<br>graph<br><br>YM201:<br>SpO2%, PR, battery<br>indicator, Pulse rate<br>bar graph, pulse waveform                                                                                                                           | M230/M230A/M230B/M230C:<br>SpO2%, PR, battery indicator,<br>Pulse rate, bar graph<br><br>M130/M130A/M130B/M130C/M<br>160A/M160C/M170A/M170C:<br>SpO2%, PR, battery indicator,<br>Pulse rate, bar graph, pulse<br>waveform, Setting item, Data<br>Graphic                                                                                      | Same<br><br>Different |
| SpO2<br>Measurement<br>Range            | 0%~100%                                                                                                                                                                                                                                                                                | 0%~100%                                                                                                                                                                                                                                                                                                                                       | Same                  |
| SpO2 Resolution                         | 1%                                                                                                                                                                                                                                                                                     | 1%                                                                                                                                                                                                                                                                                                                                            | Same                  |
| SpO2 Accuracy                           | 70~100%: ±2%<br>0% to 69%: unspecified                                                                                                                                                                                                                                                 | 80 to 100%: ±2%<br>70 to 80%: ±3%<br>0% to 69%: unspecified.                                                                                                                                                                                                                                                                                  | Different             |
| Pulse Rate<br>Range                     | 25 bpm ~ 250 bpm                                                                                                                                                                                                                                                                       | 25 bpm ~ 254 bpm                                                                                                                                                                                                                                                                                                                              | Different             |
| Pulse Rate<br>Accuracy                  | ±2 bpm                                                                                                                                                                                                                                                                                 | ±3 bpm                                                                                                                                                                                                                                                                                                                                        | Same                  |
| Pulse<br>Rate Resolution                | 1 bpm                                                                                                                                                                                                                                                                                  | 1 bpm                                                                                                                                                                                                                                                                                                                                         | Same                  |
| Shipped Sterile                         | No                                                                                                                                                                                                                                                                                     | No                                                                                                                                                                                                                                                                                                                                            | Same                  |
| Power suppler                           | 2*1.5V AAA alkaline battery                                                                                                                                                                                                                                                            | 2*1.5V AAA alkaline battery                                                                                                                                                                                                                                                                                                                   | Same                  |
|                                         |                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                               |                       |
| Storage and<br>Transport<br>Environment | Temperature: -20 °C to 60 °C<br>Atmospheric Pressure: 50 kPa to 107.4 kPa<br>Relative Humidity:<br>10%-95% (no condensation)                                                                                                                                                           | Temperature: -20 °C to 60 °C<br>Atmospheric Pressure: 50 kPa to 107.4 kPa<br>Relative Humidity: 10%-95% (no condensation)                                                                                                                                                                                                                     | Same                  |
| Operating<br>Environment                | Temperature: 15 °C to 40 °C<br>Atmospheric Pressure: 70 kPa to 106 kPa<br>Relative Humidity:<br>15%-95% (no condensation)                                                                                                                                                              | Temperature: 15 °C to 40 °C<br>Atmospheric Pressure: 70 kPa to 106 kPa<br>Relative Humidity: 15%-95% (no condensation)                                                                                                                                                                                                                        |                       |
| Compliance Standards                    |                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                               |                       |
| Bio-compatibility                       | ISO 10993-1<br>ISO 109903-5<br>ISO 10993-10                                                                                                                                                                                                                                            | ISO 10993-1<br>ISO 109903-5<br>ISO 10993-10                                                                                                                                                                                                                                                                                                   | Same                  |
| Electrical Safety                       | IEC 60601-1<br>IEC 60601-1-11                                                                                                                                                                                                                                                          | IEC 60601-1<br>IEC 60601-1-11                                                                                                                                                                                                                                                                                                                 |                       |
| EMC                                     | IEC 60601-1-2                                                                                                                                                                                                                                                                          | IEC 60601-1-2                                                                                                                                                                                                                                                                                                                                 |                       |
| Performance                             | ISO 80601-2-61                                                                                                                                                                                                                                                                         | ISO 80601-2-61                                                                                                                                                                                                                                                                                                                                |                       |
| Physical Specifications                 |                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                               |                       |
| Dimension<br>(Width*Height*Depth)       | 57mm×30mm×30 mm                                                                                                                                                                                                                                                                        | M230,M230A,M230B,M230C,M<br>130,M130A,M130B,M130C:<br>57.5mm×32.5mm×32mm<br>M160A,M160C,M170A,M170C:<br>64mm×38.5mm×36.5mm                                                                                                                                                                                                                    | Different             |
| Weight                                  | <28 g (without the batteries)                                                                                                                                                                                                                                                          | M230,M230A,M230B,M230C,M<br>130,M130A,M130B,M130C:<br>Approx: 49g (including the<br>batteries)<br>M160A,M160C,M170A,M170C:<br>Approx: 62g (including the<br>batteries)                                                                                                                                                                        |                       |

Table 1 Comparison between the predicate and the subject devices

{5}------------------------------------------------

{6}------------------------------------------------

As seen in the comparison tables, the subject and predicate devices have same design principle, similar design features and performance specifications. The different technological characteristics between the subject and predicate devices will not raise different questions of safety or effectiveness.

## 8. Performance Testing

Performance data includes "Non-Clinical Data" and "Clinical Data", brief description of which are shown as below.

## Non-Clinical Testing:

The following performance data were provided in support of the substantial equivalence determination.

## Biocompatibility testing

The biocompatibility evaluation for the Fingertip Pulse Oximeter was conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The worst case of the whole system is considered tissue contacting for duration of less than 24 hours.

{7}------------------------------------------------

And the testing included the following tests, results of which demonstrate the biocompatibility of the subject device:

- Cytotoxicity
- Skin Sensitization
- Skin Irritation

#### Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted, and the results show that the subject device complies with the IEC 60601-1: 2005+CORR. 1 (2006)+CORR. 2 (2007)+AM1 (2012) Medical electrical equipment Part 1: General requirements for basic safety and essential performance for safety and the IEC 60601-1-2: 2007 Medical electrical equipment --Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests standard for EMC.

## Bench Testing

Bench testing was conducted and the results show that the subject device complies with the ISO 80601-2-61: 2011 Medical electrical equipment — Part 2-61: Particular requirements for basic safety and essential performance of Pulse Oximeter Equipment standard. And Pulse Rate Accuracy meets the requirements defined in ISO 80601-2-61, Clause 201.12.1.104.

#### Software Verification and Validation Testing

Software documentation including verification & validation was provided in accordance with FDA Guidance: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices for software with a moderate level of concern.

#### Cleaning Validation

Cleaning and disinfection validation testing was conducted in accordance with FDA guidance "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling" issued March 17,2015. Moreover, the performance of the subject device shows no degradation after repeated cleaning and disinfection as suggested in the manual.

## Clinical data:

Clinical testing is conducted per Annex EE Guideline for evaluating and documenting SpO2 ACCURACY in human subjects of ISO 80601-2-61:2011 Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment.

## 9. Conclusion

Verification and validation testing was conducted on the subject device Fingertip Pulse Oximeter and all testing passed pre-specified criteria. This premarket notification submission demonstrates that the subject device is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K201384](https://fda.innolitics.com/device/K201384)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
