K201370 · Starkey Laboratories, Inc. · KLW · Jun 19, 2020 · Ear, Nose, Throat
Device Facts
Record ID
K201370
Device Name
Multiflex Tinnitus Technology
Applicant
Starkey Laboratories, Inc.
Product Code
KLW · Ear, Nose, Throat
Decision Date
Jun 19, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 874.3400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Multiflex Tinnitus Technology is a tool to generate sounds to be used in a Tinnitus Management Program to relieve patients suffering from tinnitus. The target population is primarily the adult population over 18 years of age. The Multiflex Tinnitus Technology is targeted for healthcare professionals, which are treating patients suffering from tinnitus, as well as conventional hearing disorders. The fitting of the Multiflex Tinnitus Technology must be done by a hearing professional participating in a Tinnitus Management Program.
Device Story
Multiflex Tinnitus Technology is a firmware-based tinnitus masker embedded within the DSP stage of Starkey hearing aids. It generates broadband white noise that can fluctuate in amplitude and frequency. Hearing professionals use proprietary 'Inspire' software to enable and adjust stimulus parameters. The device offers two modes: manual adjustment to patient needs or automatic adjustment via audiogram shaping functionality. The output is delivered through the hearing aid to provide tinnitus relief. It is intended for use by patients under the guidance of a hearing professional in a clinical setting. The device does not impact standard hearing aid functions. Benefits include customizable sound therapy to assist in tinnitus management programs.
Clinical Evidence
No clinical testing was completed or relied upon. Substantial equivalence is supported by bench testing, risk analysis, and verification/validation of the new audiogram-shaping functionality against established performance specifications.
Technological Characteristics
Firmware-based algorithm embedded in hearing aid DSP integrated circuit. Powered by 1.4V batteries (size 13, 312, 10) or rechargeable. Maximum output limited to 87 dB SPL (2cc coupler). Connectivity via proprietary Inspire programming software. No specific material standards cited.
Indications for Use
Indicated for adults >18 years suffering from tinnitus; used by hearing professionals within a Tinnitus Management Program to provide sound-based relief.
Regulatory Classification
Identification
A tinnitus masker is an electronic device intended to generate noise of sufficient intensity and bandwidth to mask ringing in the ears or internal head noises. Because the device is able to mask internal noises, it is also used as an aid in hearing external noises and speech.
Special Controls
The special control for this device is patient labeling regarding:(1) Hearing health care professional diagnosis, fitting of the device, and followup care, (2) Risks, (3) Benefits, (4) Warnings for safe use, and (5) Specifications.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Starkey Laboratories, Inc. Ka Xiong Sr. Regulatory Affairs Specialist 6700 Washington Ave S Eden Prairie, Minnesota 55344
# Re: K201370
Trade/Device Name: Multiflex Tinnitus Technology Regulation Number: 21 CFR 874.3400 Regulation Name: Tinnitus Masker Regulatory Class: Class II Product Code: KLW Dated: May 20, 2020 Received: May 22, 2020
# Dear Ka Xiong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Michael J. Ryan Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201370
Device Name Multiflex Tinnitus Technology
## Indications for Use (Describe)
The Multiflex Timitus Technology is a tool to generate sounds to be used in a Tinnitus Management Program to relieve patients suffering from tinnitus. The target population is primarily the adult population over 18 years of age.
The Multiflex Tinnitus Technology is targeted for healthcare professionals, which are treating patients suffering from timitus, as well as conventional hearing disorders. The fitting of the Multiflex Timitus Technology must be done by a hearing professional participating in a Tinnitus Management Program.
| Type of Use ( <i>Select one or both, as applicable</i> ) | |
|---------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"> <span>☒</span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div style="display:flex; align-items:center;"> <span>☐</span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| 510(k) Number | K201370 |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared | 19 June 2020 |
| Submitter Name &<br>Address | Starkey Laboratories, Inc.<br>6700 Washington Ave. South<br>Eden Prairie, MN 55344 |
| Contact Person | Ka Zoua Xiong<br>Senior Regulatory Affairs Specialist<br>Phone: 952-915-6235 ext.6235<br>Email: KaZoua_Xiong@starkey.com |
| Alternative Contact<br>Person | Yangjun Xing<br>Senior Regulatory Affairs Manager<br>Phone: 952-947-4635 ext.4635<br>Email: Yangjun Xing@starkey.com |
| Proprietary/Trade<br>Name | Multiflex Tinnitus TechnologyTM |
| Common/Usual Name | Tinnitus Masker |
| Product Classification<br>Code | KLW |
| Product Regulation<br>Number and Name | 21 CFR § 874.3400 - Tinnitus Masker |
| Device Class | II |
| Predicate Device | Predicate: Multiflex Tinnitus TechnologyTM<br>510(k) Number: K122876<br>Clearance Date: 31 Oct 2012 |
| Device Description | Multiflex Tinnitus TechnologyTM (Tinnitus Multiflex) is designed to generate noise that<br>can optionally periodically fluctuate in amplitude and frequency to provide relief for<br>patients suffering from tinnitus.<br><br>Tinnitus Multiflex is a firmware program code (algorithm) that is embedded as part of<br>the integrated circuit (IC) in the Digital Signal Processing (DSP) stage of a hearing-aid.<br>More specifically, Tinnitus Multiflex is an added branch of code that is separate from<br>the normal existing hearing-aid firmware. As such, hearing-aid operating characteristics<br>(besides tinnitus-stimulus functionalities) are not impacted by the Tinnitus Multiflex<br>branch of code. Tinnitus Multiflex may be deployed across all Starkey hearing-aid<br>models.<br><br>Tinnitus Multiflex functionalities and parameters are enabled and adjusted by a hearing-<br>care professional utilizing the Starkey proprietary hearing-aid programming software,<br>Inspire. The Tinnitus Multiflex can be manually adjusted to a patient's stimulus needs<br>or automatically adjusted via the new audiogram shaping functionality. |
| Indications for Use/<br>Intended Use | The Multiflex Tinnitus Technology is a tool to generate sounds to be used in a Tinnitus<br>Management Program to relieve patients suffering from tinnitus. The target population<br>is primarily the adult population over 18 years of age.<br><br>The Multiflex Tinnitus Technology is targeted for healthcare professionals, which are<br>treating patients suffering from tinnitus, as well as conventional hearing disorders. The<br>fitting of the Multiflex Tinnitus Technology must be done by a hearing professional<br>participating in a Tinnitus Management Program. |
{5}------------------------------------------------
| Features | Multiflex Tinnitus TechnologyTM (Subject Device) | Multiflex Tinnitus TechnologyTM (Predicate Device - K122876) | |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use and Indications for Use | Same | The Multiflex Tinnitus Technology is a tool to generate sounds to be used in a Tinnitus Management Program to relieve patients suffering from tinnitus. The target population is primarily the adult population over 18 years of age.<br><br>The Multiflex Tinnitus Technology is targeted for healthcare professionals, which are treating patients suffering from tinnitus, as well as conventional hearing disorders. The fitting of the Multiflex Tinnitus Technology must be done by a hearing professional participating in a Tinnitus Management Program. | |
| Schedule of Use | Same | This device is intended for use for a maximum of sixteen (16) hours a day when set at the maximum output level. | |
| Design Features | Same. | The Multiflex Tinnitus TechnologyTM is designed to generate broadband white noise that periodically fluctuates in amplitude and frequency to provide relief for patients suffering from tinnitus. | |
| Mechanism of Action | The Multiflex Tinnitus TechnologyTM provides a flat stimulus shape (irrespective of patient thresholds) that can be manually adjusted to a patient's stimulus needs.<br><br>The Multiflex Tinnitus TechnologyTM can also utilize the Tinnitus audiogram shaping functionality to automatically shape the tinnitus stimulus based on a patient's audiogram. | The Multiflex Tinnitus TechnologyTM provides a flat stimulus shape (irrespective of patient thresholds) that can be manually adjusted to a patient's stimulus needs. | |
| Technological Characteristics | Same | Firmware based, enabled with a hearing-aid programming software (Inspire). | |
| | Performance<br>Specifications | Same | Max overall output level - 87 dB<br>SPL |
| | | Same | Max A-weighted overall output - 87<br>dB SPL |
| | | | Max 1/3 octave output level – 87 dB<br>SPL |
| | Maximum<br>Sound<br>Loudness<br>Output | Same | Maximum output in a 2cc coupler<br>limited to 87 dB SPL. |
| | Volume<br>Control | Same | Tinnitus output levels are set to<br>specified limits by the hearing-care<br>professional. The patient<br>volume/level controls can be enabled<br>to provide adjustments that can<br>decrease or increase output levels<br>within set limits. |
| | Power Source | Same | May use any 1.4V hearing-aid<br>battery (size 13, 312 or 10) or be<br>rechargeable. |
| Summary of Non-<br>Clinical Testing | Risk analyses and verification testing were performed for the addition of the audiogram-<br>shaping functionality. Results demonstrated that Tinnitus Multiflex meets the device<br>testing acceptance criteria described in the predicate submission (K122876) and is<br>substantially equivalent to the predicate (Multiflex Tinnitus Technology™). | | |
| Summary of Clinical<br>Testing | No new clinical testing was completed, nor relied upon, in support of this Special<br>510(k). | | |
| | Changes to the subject device, Tinnitus Multiflex, do not raise new questions of safety<br>or effectiveness. The addition of the audiogram-shaping functionality was assessed by<br>appropriate risk management methodology and addressed with verification and<br>validation testing. | | |
| Statement of<br>Equivalence | Acoustic measurements of output levels, utilizing previously defined limits, were used<br>during verification to ensure that the maximum allowable stimulus level for the subject<br>device was the same as the predicate device. All results passed, and the results of the<br>verification and validation testing support that the addition of the audiogram-shaping<br>functionality does not impact safety and effectiveness. | | |
| | The subject and predicate devices are technologically identical and neither the subject<br>device safety and effectiveness, nor intended use is impacted. The results support<br>substantial equivalence of the subject device to the predicate device (K122876). | | |
{6}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.