K201361 · W.O.M. World of Medicine GmbH · HIF · Jul 29, 2020 · Obstetrics/Gynecology
Device Facts
Record ID
K201361
Device Name
PNEUMOCLEAR
Applicant
W.O.M. World of Medicine GmbH
Product Code
HIF · Obstetrics/Gynecology
Decision Date
Jul 29, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1730
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The device PNEUMOCLEAR™ is a CO2 insufflator intended for use during diagnostic and/or therapeutic endoscopic procedures to distend a cavity by filling it with CO2 gas. The Standard, High Flow/Bariatric, Pediatric and Advanced Flow operating mode of the device are indicated to fill and distend a perioneal cavity with gas during a laparoscopic procedure. The Pediatric operating mode is indicated for pediatic laparoscopic procedures. The Vessel Harvest operating mode is indicated for use during endoscopic vessel harvesting procedures to create a cavity along the saphenous vein or radial artery. The TAMIS operating mode is indicated to fill and distend the rectum and colon using CO2 gas during transanal minimally invasive surgery.
Device Story
Microprocessor-controlled CO2 insufflator; manages cavity distension during endoscopic procedures. Inputs: CO2 gas source; pressure/flow settings via GUI. Operation: Electronic-mechanical system monitors/maintains pressure; features six modes (Standard, High Flow/Bariatric, Pediatric, Advanced Flow, Vessel Harvest, TAMIS). Outputs: Controlled CO2 delivery; optional smoke evacuation/filtration. Used in clinical settings by physicians/trained staff. Enhances visualization/access by creating/maintaining working space; smoke evacuation improves visibility. Integrates with single-use sterile tube sets (with optional heating/humidification).
Clinical Evidence
Bench testing only. No clinical data presented. Verification included electrical safety (AAMI ANSI ES60601-1), EMC (IEC 60601-1-2), packaging/shelf-life (ASTM D642, D999, D4169, D4728, D5276, D6344, F88, F1886, F1929, F1980, F2096, ISO 11607-1, ISO 2233), and simulated use testing for smoke evacuation and sensor sealing.
Indicated for patients undergoing diagnostic/therapeutic endoscopic procedures requiring cavity distension with CO2 gas, including laparoscopic procedures (Standard, High Flow/Bariatric, Pediatric, Advanced Flow modes), pediatric laparoscopic procedures, endoscopic vessel harvesting (saphenous vein or radial artery), and transanal minimally invasive surgery (TAMIS).
Regulatory Classification
Identification
A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 29, 2020
W.O.M. WORLD OF MEDICINE GmbH Soeren Markworth Head of Regulatory Affairs Salzufer 8 Berlin, Berlin 10587 Germany
Re: K201361 Trade/Device Name: PNEUMOCLEAR™ Regulation Number: 21 CFR§ 884.1730 Regulation Name: Laparoscopic Insufflator Regulatory Class: II Product Code: HIF, OSV Dated: Mav 22, 2020 Received: May 22, 2020
Dear Soeren Markworth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jason R. Roberts, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K201361
Device Name PNEUMOCLEARTM
#### Indications for Use (Describe)
The device PNEUMOCLEAR™ is a CO2 insufflator intended for use during diagnostic and/or therapeutic endoscopic procedures to distend a cavity by filling it with CO2 gas. The Standard, High Flow/Bariatric, Pediatric and Advanced Flow operating mode of the device are indicated to fill and distend a perioneal cavity with gas during a laparoscopic procedure. The Pediatric operating mode is indicated for pediatic laparoscopic procedures. The Vessel Harvest operating mode is indicated for use during endoscopic vessel harvesting procedures to create a cavity along the saphenous vein or radial artery. The TAMIS operating mode is indicated to fill and distend the rectum and colon using CO2 gas during transanal minimally invasive surgery.
| Type of Use (Select one or both, as applicable) | <span> <input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
PRAStaff@fda.hhs.gov
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image contains a logo with a circular design and a stylized "WOM" text below it. Underneath the logo and text, the words "PNEUMOCLEAR™" are written. The phrase "510(k) Premarket Notification" is written below the brand name.
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
#### General Information:
| Submitter: | W.O.M. WORLD OF MEDICINE GmbH<br>Salzufer 8<br>10587 Berlin<br>Germany<br>Phone: +49 30 39981 594<br>Fax: +49 30 39981 593 |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Registration Number: | 3001556604 |
| Contact Person: | Dr. Soeren Markworth<br>Head of Regulatory Affairs<br>Salzufer 8<br>10587 Berlin, Germany<br>Phone: +4930-399 81-594<br>Fax: +4930-399 81-593<br>e-mail: soeren.markworth@wom.group |
| Date Prepared: | July 24, 2020 |
| Proposed Device: | |
| Trade Name: | PNEUMOCLEAR™ |
| Common Name: | Carbon Dioxide Insufflator for Laparoscopy and Vessel<br>Harvesting |
| Classification Name: | Insufflator, Laparoscopic |
| Regulation Number: | 21 CFR 884.1730 |
| Regulation Name: | Laparoscopic Insufflator |
| Regulatory Class: | II |
| Product Code: | HIF, OSV |
| Predicate Device: | |
| Trade Name: | PNEUMOCLEAR™ |
| 510(k) Number: | K170784 |
| Applicant: | W.O.M. WORLD OF MEDICINE GmbH |
| Regulation Number: | 21 CFR 884.1730 |
| Regulation Name: | Laparoscopic Insufflator |
| Regulatory Class: | II |
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for WOM, which is a stylized circle with a knot in the middle and a small triangle at the bottom. Below the logo, the text "WOM" is written in a bold sans-serif font. Below that, the text "PNEUMOCLEAR™" is written in a smaller font, followed by "510(k) Premarket Notification".
HIF, OSV Product Code:
The PNEUMOCLEAR™ has not been subject to a design related recall.
#### Device Description:
The PNEUMOCLEAR™ is a microprocessor controlled device that consists of the following major components and features: a casing, a world power supply, pressure reducers, pressure sensors, venting systems, various safety valves, a suction pump, a fluid sensor and a software controlled graphical user interface (GUI) touch screen with various setting keys and display elements. The proposed device offers six operating modes: Standard, High Flow/Bariatric, Pediatric, Advanced Flow, Vessel Harvest, and TAMIS. The PNEUMOCLEAR™is not intended to enter the sterile field and cannot be sterilized. It is to be used with one of the following specially designed, single-use, sterile insufflation tube sets: (1) insufflation tube set with integrated filter, (2) insufflation tube set with integrated filter and heating wire; (3) insufflation tube set with integrated filter, heating wire and humidification media; (4) insufflation and smoke evacuation tube set with integrated filter, heating wire. and humidification media; and (5) insufflation and smoke evacuation tube set with integrated filter. When a smoke evacuation tube set is connected, the PNEUMOCLEAR™ allows for removal and filtration of CO2 and surgical smoke from the abdomen. rectum or colon during laparoscopic and transanal minimally invasive procedures and at the end of a procedure.
### Intended Use / Indication for Use:
The device PNEUMOCLEAR™ is a CO2 insufflator intended for use during diagnostic and/or therapeutic endoscopic procedures to distend a cavity by filling it with CO2 gas. The Standard, High Flow/Bariatric, Pediatric and Advanced Flow operating mode of the device are indicated to fill and distend a peritoneal cavity with gas during a laparoscopic procedure. The Pediatric operating mode is indicated for pediatric laparoscopic procedures. The Vessel Harvest operating mode is indicated for use during endoscopic vessel harvesting procedures to create a cavity along the saphenous vein or radial artery. The TAMIS operating mode is indicated to fill and distend the rectum and colon using CO2 gas during transanal minimally invasive surgery.
### Comparison of Technological Characteristics:
The PNEUMOCLEAR™ that is the subject of this 510(k) is substantially equivalent to its predicate, predecessor K170784 PNEUMOCLEAR™. This is a catch-up 510(k) submission intended to update the PNEUMOCLEAR™ record at FDA with the changes that have been implemented for the device. As such, the intended use and indications remain identical to those for the cleared device. Further, the
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for WOM PNEUMOCLEAR. The logo consists of a circular design with a stylized "W" inside, above the text "WOM". Below that is the text "PNEUMOCLEAR™" and "510(k) Premarket Notification".
implemented changes were primarily to tweak the optional features based on user feedback or internal, routine, developmental updates. These are captured in the comparison table below.
| Substantial Equivalence Comparison Table | | |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Cleared Device: | Subject Device: |
| | PNEUMOCLEAR™ | PNEUMOCLEAR™ |
| | (K170784) | (this submission) |
| Manufacturer | W.O.M.World of Medicine GmbH | Identical |
| Regulation Number | 884.1730 | Identical |
| Regulation Name | Laparoscopic Insufflator | Identical |
| Procodes | HIF, OSV | Identical |
| Indications for Use | The device PNEUMOCLEAR™ is a CO2<br>insufflator intended for use during<br>diagnostic and/or therapeutic<br>endoscopic<br>procedures to distend a cavity by filling it<br>with gas. The Standard, High<br>Flow/Bariatric, Pediatric and Advanced<br>Flow operating mode of the device are<br>indicated to fill and distend a peritoneal<br>cavity with gas during a laparoscopic<br>procedure. The Pediatric operating<br>mode is indicated for pediatric<br>laparoscopic procedures. The Vessel<br>Harvest operating mode is indicated for<br>use during endoscopic vessel harvesting<br>procedures to create a cavity along the<br>saphenous vein or radial artery. The<br>TAMIS operating mode is indicated to fill<br>and distend the rectum and colon using<br>CO2 gas during trans anal minimal<br>invasive surgery. | The device PNEUMOCLEAR™ is a CO₂<br>insufflator intended for use during<br>diagnostic and/or therapeutic endoscopic<br>procedures to distend a cavity by filling it<br>with CO₂ gas. The Standard, High<br>Flow/Bariatric, Pediatric and Advanced<br>Flow operating modes of the device are<br>indicated to fill and distend a peritoneal<br>cavity with gas during a laparoscopic<br>procedure. The Pediatric operating mode is<br>indicated for pediatric laparoscopic<br>procedures. The Vessel Harvest operating<br>mode is indicated for use during<br>endoscopic vessel harvesting procedures<br>to create a cavity along the saphenous vein<br>or radial artery. The TAMIS operating mode<br>is indicated to fill and distend the rectum<br>and colon using CO₂ gas during trans anal<br>minimal invasive surgery. |
| Substantial Equivalence Comparison Table | | |
| | Cleared Device: | Subject Device: |
| | PNEUMOCLEAR™ | PNEUMOCLEAR™ |
| | (K170784) | (this submission) |
| Target Population | Patients undergoing the procedures<br>indicated for the device. | Identical |
| Target Users | For physicians and trained hospital<br>staff. | Identical |
| Power source | Line powered | Identical |
| Mechanism of Action | Insufflation of the target area, which is<br>monitored and maintained by a<br>microprocessor-controlled, electronic-<br>mechanical system | Identical |
| Maximum Flow | 50 l/min | Identical |
| Maximum Pressure | 30 mmHg | Identical |
| Heating Option | Yes | Identical |
| Humidification<br>Option | Yes | Identical |
| Smoke Evacuation<br>Option | Yes | Identical |
| Tube Sets | | |
| Substantial Equivalence Comparison Table | | |
| | Cleared Device: | Subject Device: |
| | PNEUMOCLEAR™<br>(K170784) | PNEUMOCLEAR™<br>(this submission) |
| Model ST295 | Insufflation tube set with integrated filter | Identical |
| Model ST296 | Insufflation tube set with integrated filter and heating wire | Identical<br>Packaging change to address a shipping issue. |
| Model ST297 | Insufflation tube set with integrated filter, heating wire and humidification | Implemented a manufacturing change to better seal a component. |
| Model ST298 | Insufflation and smoke evacuation tube set with integrated filter, heating wire and humidification | • Implemented a manufacturing change to better seal a component.<br>• Implemented a design change to a gas line. |
| Model ST299 | Insufflation and smoke evacuation tube set with integrated filter | Implemented a design change to a gas line. |
| Sterilization Process | EO | Identical |
| Materials | MABS, PVC, Hydrophobic Glass Fiber, Polyolefin, Silicone, FEP, Polyester, Nylon. | Additional Terlux 2802 and fiber materials for the design change to a gas line. |
| Single-Use | Yes | Identical |
| Shelf-life | 3 years | Identical |
| Software | Version 1.0.27 | Version 1.0.34 |
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image contains a logo with a circular design and a stylized shape inside, along with the text "WOM" below it. Underneath "WOM", the text "PNEUMOCLEAR™" is displayed, followed by "510(k) Premarket Notification". The text is arranged in a clear, hierarchical manner, suggesting a product name and its regulatory status.
{8}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for WOM, which is a stylized circle with a design inside and the letters "WOM" below it. Below the logo is the text "PNEUMOCLEAR™" and "510(k) Premarket Notification". The text is centered below the logo. The image appears to be a product or company logo with a notification about premarket approval.
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for WOM PNEUMOCLEAR. The logo consists of a circular design with a stylized figure-eight shape inside, above the text "WOM PNEUMOCLEAR™". Below this is the text "510(k) Premarket Notification".
The differences between the two versions of PNEUMOCLEAR™ do not raise different questions of safety and effectiveness and have been confirmed through testing.
#### Performance Data:
The changes implemented for the PNEUMOCLEAR™ were confirmed using the same guidance/recognized standards that were used for the cleared device. These include:
- Electrical Safety and Electromagnetic Compatibility ●
- o AAMI ANSI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012
- IEC 60601-1-2 Edition 4.0 2014-02 O
#### Packaging and Shelf-Life Testing ●
The effectiveness of the tube set barrier provided by the modified packaging was confirmed using the following transportation, packaging and shelf-life standards:
- o ASTM D642 (2015)
- ASTM D999 (2008) O
- ASTM D4169 (2016) O
- ASTM D4728 (2006) O
- ASTM D5276 (1998) O
- ASTM D6344 (2004) O
- ASTM F88 (2015) O
- ASTM F1886 (2016) O
- ASTM F1929 (2015) O
- ASTM F1980 (2016) O
- ASTM F2096 (2011) O
- ISO 11607-1 First Edition 2006-04-15 O
- O ISO 2233 (2001)
#### Software/Bench Testing ●
W.O.M. developed and verified the software in accordance with a major level of concern described in the FDA "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and following the IEC 62304 Edition 1.1 2015-06 standard.
The sensor sealing process for tube sets ST297 and ST298 was modified and confirmed through submerged testing. The gas line for the ST297 and ST298 tube sets was modified to improve the surgical smoke evacuation and verified through simulated use testing.
{10}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for WOM, which is a circle with a design inside and a small red triangle at the bottom. Below the logo is the word "WOM" in large, bold letters. The text "PNEUMOCLEAR™ 510(k) Premarket Notification" is written below the logo and the word WOM.
The complete Version 1.0.34 software and all other changes were bench-tested to confirm that they met their requirements before implementation.
### Conclusion:
The provided performance testing supports a substantial equivalence determination. The subject PNEUMOCLEAR™ is substantially equivalent to its predicate, predecessor K170784 PNEUMOCLEAR™
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.