K201347 · DePuy Orthopaedics, Inc. · OIY · Jun 18, 2020 · Orthopedic
Device Facts
Record ID
K201347
Device Name
DePuy ATTUNE™ Total Knee System
Applicant
DePuy Orthopaedics, Inc.
Product Code
OIY · Orthopedic
Decision Date
Jun 18, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DePuy Attune Total Knee System is intended for cemented and noncemented use as a total knee replacement system.
Device Story
Total knee replacement system comprising polyethylene tibial inserts and patella components; mates with existing cemented/noncemented femoral and tibial base components. Used by orthopedic surgeons in clinical settings to replace damaged knee joints. Device restores joint function and reduces pain for patients with severe arthritis or failed prior implants. This submission specifically extends the shelf life of AOX polyethylene components from 5 to 8 years; no changes to design, materials, or principle of operation.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including Material Oxidation Index/Resistance Testing (ASTM F2102-17, ASTM F2003) and Wear Testing with Particle Analysis (ISO 14243-2, ISO 14243-3, ASTM F1877-05).
Technological Characteristics
Components manufactured from AOX ultra-high molecular weight polyethylene. Fixed bearing tibial inserts (CR and PS) and patella components. Gamma irradiated. Shelf life extended to 8 years. No changes to materials, sterilization, or design.
Indications for Use
Indicated for patients with severely painful and/or severely disabled joint resulting from osteoarthritis, post-traumatic arthritis, rheumatoid arthritis, or a failed previous implant.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
DePuy ATTUNE Medialized Dome and Anatomic Patellae Components (K103756)
ATTUNE Total Knee System – Cementless CR & PS Femoral Components (K140881)
Submission Summary (Full Text)
{0}------------------------------------------------
June 18, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services. To the right, there is a blue square with the letters "FDA" in white. Next to the square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
Mary Wood Regulatory Affairs Associate DePuy Orthopaedics, Inc. 700 Orthopaedic Drive Warsaw. Indiana 46582
Re: K201347
Trade/Device Name: DePuy ATTUNE™ Total Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: OIY, JWH, MBH Dated: May 13, 2020 Received: May 21, 2020
Dear Mary Wood:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**Ting Song -S**
Ting Song, Ph.D., R.A.C. Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201347
Device Name ATTUNE Total Knee System
#### Indications for Use (Describe)
The DePuv Attune Total Knee System is intended for cemented use as a total knee replacement system.
Candidates for total knee replacement include patients with and/or severely disabled joint resulting from osteoarthritis, post-traumatic arthritis, rheumatoid arthritis, or a failed previous implant.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K201347
Page 1 of 5
# 5. 510k Summary
{4}------------------------------------------------
## 510(K) SUMMARY
#### (As required by 21 CFR 807.92 and 21 CFR 807.93)
| Submitter Information | |
|---------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | DePuy Orthopaedics, Inc. |
| Address | 700 Orthopaedic Drive |
| | Warsaw, Indiana 46582 |
| Phone number | 812-219-1298 |
| Fax number | N/A |
| Establishment Registration # | 1818910 |
| Name of contact person | Mary Wood |
| Date prepared | May 8, 2020 |
| Name of device | |
| Trade or proprietary name | ATTUNE™ Total Knee System |
| Common or usual name | Total Knee Arthroplasty Prosthesis |
| Classification name | Knee joint patellofemorotibial polymer/metal/polymer semi-constrained<br>cemented prosthesis |
| | Knee joint patellofemorotibial metal/polymer porous-coated<br>uncemented prosthesis. |
| Class | II |
| Classification panel | 87 Orthopedics |
| Regulation | 21 CFR 888.3560 |
| | 21 CFR 888.3565 |
| Product Code(s) | OIY, JWH, MBH |
| Legally marketed device(s) to which<br>equivalence is claimed | Primary Predicate:<br>DePuy ATTUNE Knee System, K101433 |
| | Reference Predicates:<br>DePuy ATTUNE PS Knee System, K111433 |
| | DePuy ATTUNE Medialized Dome and Anatomic Patellae<br>Components, K103756 |
| | ATTUNE Total Knee System – Cementless CR & PS Femoral<br>Components, K140881 |
| Reason for 510(k)<br>submission | The purpose of this submission is to extend the current approved shelf<br>life of 5 years to 8 years. |
| Device description | The subject device is a polyethylene tibial inserts and patella component<br>made from AOX Polyethylene that mates with existing cemented and<br>noncemented femoral and tibial base components |
| Intended use of the device | The DePuy Attune Total Knee System is intended for cemented and noncemented use as a total knee replacement system. |
| Indications for use | Candidates for total knee replacement include patients with a severely painful and/or severely disabled joint resulting from osteoarthritis, post-traumatic arthritis, rheumatoid arthritis, or a failed previous implant. |
{5}------------------------------------------------
#### Summary of Technologies/ Substantial Equivalence Discussion
The purpose of this submission is to propose a change in shelf life from 5 years for the ATTUNE AOX Fixed Bearing Tibial Inserts for both CR (Cruciate Retaining) and PS (Posterior Stabilized) as well as the Medialized Dome and Anatomic Patellae components
There are no other modifications associated with the currently marketed products. The intended use, product design, principle of operation and materials are not impacted by this change.
| Non-Clinical Testing: | The following testing was conducted (in compliance with FDA Guidance<br>Class II Special Controls Guidance Document: Knee Joint<br>Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated<br>Uncemented Prostheses: Guidance for Industry and FDA) to demonstrate<br>substantial equivalence of safety and efficacy with the predicate devices:<br>Material Oxidation Index/Resistance Testing (ASTM F2102-17 ASTM F2003) Wear Testing with Particle Analysis (ISO 14243-2, ISO 14243-3 ASTM F1877-05) |
|-------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Clinical Testing: | Clinical testing was not necessary to determine substantial equivalence<br>between the ATTUNE Polyethylene Fixed Bearing Tibia Inserts and<br>Patella components and the predicate devices. |
| Conclusions drawn from<br>non-clinical and clinical<br>data | The subject ATTUNE Polyethylene Fixed Bearing Tibia Inserts and Patella<br>components with 8 year shelf life are substantially equivalent to the<br>predicate ATTUNE Polyethylene Fixed Bearing Tibia Inserts and Patella<br>components with 5 year shelf life. |
The testing provided below was used to support equivalence to the predicate devices.
{6}------------------------------------------------
Note: There have been no modifications made to the intended use, design, biocompatibility, materials, sterilization or packaging of the existing DePuy ATTUNE Knee System.
The only modification being sought via this Special 510(k) submission is a change in the shelf life of the ATTUNE C/R Fixed Bearing Insert, ATTUNE P/S Fixed Bearing Insert and the ATTUNE Medialized Dome and Anatomic Patellae components which are made from AOX Polyethylene that is gamma irradiated.
The ATTUNE Knee System is not an invitro-diagnostic device.
# Submission Purpose
The purpose of this submission is to propose a change in shelf life from 5 years to 8 years for the ATTUNE AOX Fixed Bearing Tibial Inserts for both CR (Cruciate Retaining) and PS (Posterior Stabilized) as well as the Medialized Dome and Anatomic Patellae components. A complete list of the affected devices is provided as Exhibit A. The change will require an update to the shelf life indicated on the Labeling. Draft labeling is provided as Exhibit B.
There are no other modifications associated with the currently marketed product predicate ATTUNE Total Knee System. The intended use, product design, principle of operation and materials are not impacted by this change.
## Device Description
The ATTUNE CR Fixed Bearing Tibial inserts are available in sizes 1-10 and in thicknesses of 5, 6, 7, 8, 10, 12, 14, 16, and 18mm. They are manufactured from AOX ultra high molecular weight polyethylene.
The ATTUNE PS Fixed Bearing Tibial Inserts are available in sizes 1-10 and in thicknesses of 5, 6, 7, 8, 10, 12, 14, 16, 18, 20, and 22mm. They are manufactured from AOX ultra high molecular weight polyethylene.
The ATTUNE Medialized Dome and Medialized Anatomic Patellae are available in sizes 29, 32, 35, 38. and 41mm. They are manufactured from AOX ultra high molecular weight polyethylene.
The proposed change to the shelf life of the subject AOX components will have no impact to the predicate, ATTUNE Knee System (K101433, K111433, K103756, K140881) which are intended for both cemented and noncemented use.
{7}------------------------------------------------
## Materials
The subject ATTUNE Fixed Bearing Tibial Inserts and Patellae components are manufactured from AOX Polyethylene. This change proposes a shelf life change from 5 to 8 years; no material or sterilization changes are proposed.
### Design
No design changes are proposed for this submission. The change is to extend the shelf life of the existing ATTUNE AOX Fixed Bearing Tibial Inserts and Patellae components from 5 years to 8 years.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.