K201338 · Lifenet Health · FMF · Jul 13, 2020 · General Hospital
Device Facts
Record ID
K201338
Device Name
Allograft MIS Delivery System
Applicant
Lifenet Health
Product Code
FMF · General Hospital
Decision Date
Jul 13, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Allograft MIS Delivery System is intended to be used for the delivery of hydrated allograft bone graft material to an orthopedic surgical site.
Device Story
The Allograft MIS Delivery System is a sterile, single-use, disposable device used in orthopedic surgery to deliver hydrated allograft bone graft material. The system comprises a 5 cc cannula (containing the graft), a plunger, a dispenser, and two end caps. The surgeon operates the dispenser with one hand to advance the plunger, which expresses the graft material from the cannula directly into the surgical site. The device is designed to be preloaded with graft material at a registered graft facility before being shipped to the surgical facility. By providing a controlled, incremental delivery mechanism, the device assists the surgeon in placing bone graft material accurately, potentially improving surgical efficiency and patient outcomes during orthopedic procedures.
Clinical Evidence
Bench testing only. Testing included biocompatibility, dimensional and fit verification, functional verification (mechanical extrusion and simulated use), and package performance/stability. All specified acceptance criteria were met.
Technological Characteristics
Cannula material: medical grade polypropylene. Mechanism: ratchet-actuated handle for incremental plunger advancement. Capacity: 5 cc. Sterilization: Gamma irradiation (dispenser/plunger) and Ethylene Oxide (cannula/end caps). Single-use, disposable. No software or electronic components.
Indications for Use
Indicated for delivery of hydrated allograft bone graft material to orthopedic surgical sites in patients requiring bone grafting procedures.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
Graftgun Universal Graft Delivery System (K180937)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square are the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" in a smaller font below.
July 13, 2020
LifeNet Health Brittany Beasley Global Regulatory Affairs Program Manager 1864 Concert Drive Virginia Beach, Virginia 23453
Re: K201338
Trade/Device Name: Allograft MIS Delivery System Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Dated: May 19, 2020 Received: May 20, 2020
Dear Brittany Beasley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201338
Device Name Allograft MIS Delivery System
Indications for Use (Describe)
The Allograft MIS Delivery System is intended to be used for the delivery of hydrated allograft bone graft material to an orthopedic surgical site.
| Type of Use (Select one or both, as applicable) | <table><tr><td><div> <input type="checkbox"/> </div></td><td><div> <input type="checkbox"/> </div></td></tr></table> | <div> <input type="checkbox"/> </div> | <div> <input type="checkbox"/> </div> |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|---------------------------------------|---------------------------------------|
| <div> <input type="checkbox"/> </div> | <div> <input type="checkbox"/> </div> | | |
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is the logo for LifeNet Health. The logo features a blue and green circular icon with a white figure inside, followed by the words "LifeNet Health" in blue and green. Below the name is the tagline "Saving Lives. Restoring Health. Giving Hope."
# 510(k) Summary
## Allograft MIS Delivery System
| Submitter Information: | |
|------------------------|--------------------------------------------------------------------|
| Name: | LifeNet Health |
| Address: | 1864 Concert Drive<br>Virginia Beach, Virginia 23453 |
| Contact: | Brittany Beasley, RAC<br>Global Regulatory Affairs Program Manager |
| Phone: | (757) 609-4201 |
| Email: | brittany_beasley@lifenethealth.org |
| Date Prepared: | July 10, 2020 |
| Device Information: | |
| Trade Name: | Allograft MIS Delivery System |
| Common Name: | Graft Delivery Device |
| Regulation Name: | Piston Syringe |
| Regulation Number: | 21 CFR 880.5860 |
| Regulatory Class: | Class II |
| Product Code: | FMF |
| Proposed Panel: | General Hospital |
| Predicate Device: | |
| Name: | Graftgun Universal Graft Delivery System (K180937 |
| Manufacturer: | SurGenTec LLC |
## Device Description:
The Allograft MIS Delivery System is a sterile, single-use, disposable graft delivery device intended for the delivery of hydrated allograft bone graft material to an orthopedic surgical site. The delivery system consists of: a cannula for containing and delivering the allograft material to the surgical site; a plunger to express the allograft material from the cannula; a dispenser to advance the plunger down the length of the cannula; and two end caps to retain the allograft material in the cannula prior to system assembly.
The cannula component is a 5 cc straight, open bore tube with a double threaded interface that mates with two end caps which are removed prior to cannula attachment to the dispenser. The dispenser is designed for single-hand, incremental delivery of graft materials and may be used to dispense multiple cannulas for a single patient.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a logo for LifeNet Health. The logo features a blue and green circular icon with a stylized human figure in the center. To the right of the icon is the text "LifeNet Health" in blue and green. Below the text is the tagline "Saving Lives. Restoring Health. Giving Hope." in blue.
## Indications for Use:
The Allograft MIS Delivery System is intended to be used for the delivery of hydrated allograft bone graft material to an orthopedic surgical site.
## Comparison of Technological Characteristics to the Predicate:
| | | Graftgun Universal Graft Delivery<br>System<br>(predicate device) | Allograft MIS<br>Delivery System<br>(subject device) |
|-----------|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| General | Applicant | SurGenTec | LifeNet Health |
| | Regulation No. &<br>Name | 21 CFR 880.5860,<br>Piston Syringe | 21 CFR 880.5860,<br>Piston Syringe |
| | Procode | FMF | FMF |
| | 510(K) Number | K180937 | K201338 |
| | Labeling | Indications for use | The Graftgun Universal Graft<br>Delivery System is intended to be<br>used for the delivery of hydrated<br>allograft, autograft, or synthetic bone<br>graft material to an orthopedic<br>surgical site. |
| | Prescription /<br>OTC Use | Prescription Use | Prescription Use |
| | Type of Use | Single, Disposable | Single, Disposable |
| Design | Mechanism of<br>operation | Graft material expressed from graft<br>tube via a plunger, operated by a<br>ratchet-actuated handle. Material<br>expressed from graft tube directly to<br>graft site. | Graft material expressed from<br>cannula via a plunger, operated by a<br>ratchet-actuated handle. Material<br>expressed from graft cannula<br>directly to graft site. |
| | Patient-contact<br>materials | Graft tube - Medical grade<br>polypropylene<br>Loading Device/Syringe Tube -<br>Medical grade polycarbonate<br>Radiopaque ring marker - Medical<br>grade stainless steel - 316L, 316F,<br>304H, 304HC and ABS | Graft cannula - Medical grade<br>polypropylene |
| | Graft<br>Tube/Cannula | Option of having the 5cc graft tube<br>provided clean to a graft facility,<br>preloaded with graft, packaged,<br>sterilized and shipped to the surgical<br>facility prior to the surgical<br>procedure. | Cannula provided sterile to a graft<br>facility, preloaded with graft,<br>packaged, and shipped to the<br>surgical facility prior to the surgical<br>procedure. |
| | Radiopaque<br>marker | Yes | No |
| | Volume capacity | Up to 5 cc | Up to 5 cc |
| | Volume markings | Yes | No |
| | | Graftgun Universal Graft Delivery<br>System<br>(predicate device) | Allograft MIS<br>Delivery System<br>(subject device) |
| Sterility | Sterility Claim /<br>SAL | Sterile, SAL 1 x 10-6<br>(see Note 1) | Sterile, SAL 1 x 10-6<br>(see Note 2) |
| | Sterilization<br>Method | Gamma irradiation | Dispenser & Plunger:<br>Gamma irradiation<br>Cannula & End caps:<br>Ethylene Oxide |
| | | Note 1: Graft tubes are provided to a<br>registered graft facility. Their<br>sterilization process is validated as<br>per their internal accepted standards<br>prior to the prefilled tubes being<br>shipped. | Note 2: Cannulas and end caps are<br>provided to a registered graft<br>facility. Their manufacturing steps<br>conform to Current Good Tissue<br>Practice requirements under 21<br>CFR 1271, Subpart D. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is the logo for LifeNet Health. The logo consists of a blue and green circle with a person in the center, followed by the words "LifeNet Health" in blue and green. Below the words "LifeNet Health" is the tagline "Saving Lives. Restoring Health. Giving Hope."
## Performance Data:
The following testing was performed on the subject device to demonstrate substantial equivalence:
- Biocompatibility
- 트 Dimensional and fit verification
- I Functional verification
- -Mechanical extrusion
- -Simulated use
- 트 Package performance and stability
Testing was performed on samples that represented the finished device and all specified acceptance criteria were met.
## Conclusion:
The Allograft MIS Delivery System is substantially equivalent to the Graftgun dispensing unit of the Graftgun Universal Graft Delivery System in intended use, mechanism of operation, patientcontacting materials, and intent to supply the device to a registered graft facility prior to the surgical facility. The differences between the subject device and the predicate device are minor and do not impact the safety and effectiveness of the subject device as demonstrated by the provided performance testing.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.