← Product Code [HKI](/productcode/HKI) · K201325

# Optomed Smartscope M5 with Optomed Smartscope FA (K201325)

_Optomed Oyj · HKI · Nov 20, 2020 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K201325

## Device Facts

- **Applicant:** Optomed Oyj
- **Product Code:** [HKI](/productcode/HKI.md)
- **Decision Date:** Nov 20, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.1120
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

Optomed Smartscope M5 digital camera with Smartscope FA optics module is intended to capture fluorescein angiograms of the human eye.

## Device Story

The Optomed Smartscope M5 with FA optics module is a handheld ophthalmic camera used in medical environments to capture fluorescein angiograms. The device consists of the M5 camera body and an interchangeable FA optics module attached via bayonet connector. The FA module utilizes blue and infrared light sources and internal fixation targets to guide light through the pupil to the fundus. The camera sensor captures images/video, which are stored on a memory card or transferred to a PC via USB or WLAN. The device is operated by trained healthcare professionals. Captured images are used for documentation and clinical consultation, assisting physicians in visualizing retinal vasculature to support diagnostic decision-making. The device is powered by a rechargeable Ni-MH battery and includes an eye cup for patient contact.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing included electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), optical safety (IEC 62471, ANSI Z80.36-2016), biocompatibility (ISO 10993-5/10), optical performance (ISO 10940), and usability (IEC 62366-1).

## Technological Characteristics

Handheld ophthalmic camera with interchangeable optics modules. FA module uses blue (Osram Oslon LB H9GP) and NIR (OSRAM Oslon SFH-4715) LEDs. 5MP color CMOS sensor. 2.4" TFT-LCD display. Diopter compensation -20D to +20D. 40-degree FOV. Connectivity: USB 1.1, WLAN. Power: 4.8V 1000mAh Ni-MH battery. Eye cup material: Momentive LIM 6030AB. Software: IEC 62304 compliant. Standards: IEC 60601-1, IEC 60601-1-2, ISO 10940, ISO 15004-1/2.

## Regulatory Identification

An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

## Predicate Devices

- Optomed Smartscope M5 digital camera / Optomed Smartscope EY4 optics module ([K132186](/device/K132186.md))
- Kowa Genesis Df ([K091098](/device/K091098.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Optomed Oyj Jyri Leskelä Director Q&RA Yrttipellontie 1 Oulu, 90230 Finland

Re: K201325

Trade/Device Name: Optomed Smartscope M5 with Optomed Smartscope FA Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic Camera Regulatory Class: Class II Product Code: HKI Dated: May 7, 2020 Received: May 18, 2020

Dear Jyri Leskelä:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for Elvin Ng

Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K201325

Device Name Optomed Smartscope M5 Optomed Smartscope FA

Indications for Use (Describe)

Optomed Smartscope M5 digital camera with Smartscope FA optics module is intended to capture fluorescein angiograms of the human eye.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

| 510(k) submitter:              | Optomed Oyj                                       |                                                                              |
|--------------------------------|---------------------------------------------------|------------------------------------------------------------------------------|
| Address:                       | Yrttipellontie 1, FI-90230 Oulu, Finland          |                                                                              |
| Company phone:                 | +358 20 741 3380                                  |                                                                              |
| Contact person:                | Mr. Jyri Leskelä, Quality and Regulatory Director |                                                                              |
| Date:                          | October 15, 2020                                  |                                                                              |
| Subject device:                | Trade name:                                       | Optomed Smartscope FA                                                        |
|                                | Common/usual name:                                | Ophthalmic camera                                                            |
|                                | Classification name:                              | camera, ophthalmic, ac-powered (21 CFR 886.1120)                             |
|                                | Regulatory Class:                                 | II                                                                           |
|                                | Product Code:                                     | HKI                                                                          |
| Predicate device:<br>(Primary) | Trade name:                                       | Optomed Smartscope M5 digital camera<br>Optomed Smartscope EY4 optics module |
|                                | 510(k) number:                                    | K132186                                                                      |
|                                | Regulatory Class:                                 | II                                                                           |
|                                | Product code:                                     | HKI                                                                          |
| Predicate device:              | Trade name:                                       | Kowa Genesis Df                                                              |
|                                | 510(k) number:                                    | K091098                                                                      |
|                                | Regulatory Class:                                 | II                                                                           |
|                                | Product code:                                     | HKI                                                                          |

Image /page/3/Picture/4 description: The image shows the word "OPTOMED" in blue font. The "O" in "OPTOMED" is replaced with an eye symbol. The eye symbol is a circle with a smaller circle inside of it.

{4}------------------------------------------------

### Description of the Device

Optomed Smartscope M5 camera with Smartscope FA optics module is designed for use in a medical environment. Captured images and video are used for documentation and consultation. Optomed Smartscope M5 camera has a memory card where captured images and recorded videos are saved. Optomed Smartscope M5 camera is used with interchangeable optics modules previously cleared Smartscope EY4 and Smartscope ES2 and subject device Smartscope FA. Optics modules are attached to the camera with bayonet connectors.

Optomed Smartscope FA optics module is designed for fundus fluorescein angiography imaging of the human eye. Optomed Smartscope FA optics module includes two light sources: Blue light and infrared light, optical component that reflects light to the optical path and lenses that guide light through the pupil to the back of the eye and back to the camera sensor. Optomed Smartscope FA optics module has nine internal fixation targets for the patient to fixate on during imaging. Optomed Smartscope FA optics module receives power from Optomed Smartscope M5 digital camera.

Optomed Smartscope M5 camera has a WLAN module inside and when WLAN is used, Optomed Smartscope M5 camera transfers captured images and recorded videos to the PC automatically immediately after imaging. Images and videos can also be transferred to PC from the memory card when the camera is placed on Charging Station and the USB cable is connected between Charging Station and the PC. The image data transfer method to PC is similar as with any other USB mass storage device.

Optomed Smartscope M5 camera has a rechargeable Ni-MH battery that is charged when the camera is placed on Charging Station, which is connected to the mains by power supply cable. Charging Station can also be used as an external battery charger for the spare battery included in sales case. When Optomed Smartscope M5 camera is not used, it may be stored on Charging Station. Storing the device on Optomed Smartscope M5 camera Charging Station is not harmful for the battery.

Image /page/4/Picture/6 description: The image shows the word "OPTOMED" in blue font. The "O" in "OPTOMED" is replaced with an eye symbol. The eye symbol is a blue circle with a white dot in the center.

{5}------------------------------------------------

# Indications for Use

Optomed Smartscope M5 digital camera with Smartscope FA optics module is intended to capture fluorescein angiograms of the human eye.

Optomed Smartscope FA is classified as prescription device. Federal law restricts this device to sale by or on the order of a physician or licensed practitioner. The device may only be operated by persons who have been properly trained or who have the required knowledge and experience to do so. The device may only be used in accordance with its intended use.

Optomed Smartscope M5 camera is modular and can be used together with dedicated optics modules (Smartscope EY4 and Smartscope FA) to capture images from the fundus of the eye. Predicate device Kowa Genesis Df is used to fundus imaging as well. Changes implemented to subject device Smartscope FA do not affect safety and effectiveness of the product.

# Comparison of Technological Characteristics

Table 1 below includes a summary of the technical information used in the substantial equivalence discussion.

| Point of comparison | Optomed Smartscope M5<br>Optomed Smartscope FA                                                                                                          | Optomed Smartscope M5<br>Optomed Smartscope EY4                                                                                                                          | Kowa Genesis Df                                                                                                     |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| 510(k) number       | K201325                                                                                                                                                 | K132186                                                                                                                                                                  | K091098                                                                                                             |
| Indications for use | Optomed Smartscope M5<br>digital camera with<br>Smartscope FA optics<br>module is intended to<br>capture fluorescein<br>angiograms of the human<br>eye. | Optomed Smartscope M5<br>digital camera with<br>Smartscope EY4 optics<br>module is intended to<br>capture digital images and<br>video of the fundus of the<br>human eye. | A hand-held mydriatic<br>retinal camera which<br>captures fundus image<br>(For color and FA retinal<br>photography) |
| Usage               | Prescription use                                                                                                                                        | Prescription use                                                                                                                                                         | Prescription use                                                                                                    |
| Use condition       | Intended to be used with<br>mydriatic drops.                                                                                                            | Intended to be used<br>without mydriasis but can<br>be used also with<br>mydriatic drops.                                                                                | Intended to be used with<br>mydriatic drops                                                                         |

Image /page/5/Picture/9 description: The image shows the word "OPTOMED" in blue font. The two Os are stylized, with the second O having a circle around it and a dot in the center, resembling an eye. The font is bold and sans-serif, and the overall design is simple and modern. The background is a light gray color.

{6}------------------------------------------------

| Illumination source                     | Smartscope FA:<br>Blue: Osram Oslon LB<br>H9GP<br>NIR: OSRAM Oslon SFH-<br>4715<br>Target LEDs: OSRAM LR<br>QH9F | Smartscope EY4:<br>White: OSRAM Oslon<br>LUW-H9GP<br>NIR: OSRAM Oslon SFH-<br>4715<br>Target LEDs: OSRAM LR<br>QH9F | Observation light source:<br>white LED / blue LED<br>Photographing light source:<br>Xenon flash lamp 23WS |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Display system                          | 2.4", TFT-LCD, 240x320<br>pixels, 262 000 colors,<br>antireflective coating                                      | 2.4", TFT-LCD, 240x320<br>pixels, 262 000 colors,<br>antireflective coating                                         | Visual observation                                                                                        |
| Camera sensor<br>specification          | Color CMOS camera<br>maximum resolution 5 Mp                                                                     | Color CMOS camera<br>maximum resolution 5 Mp                                                                        | Color CCD camera<br>2,000,000 pixels                                                                      |
| Diopter compensation                    | From -20 D to +20 D                                                                                              | At least from -20 D to +20<br>D                                                                                     | -15D~+35D                                                                                                 |
| Field of view                           | 40 degrees                                                                                                       | 40 degrees                                                                                                          | Horizontal: 30 degree<br>Vertical: 25 degree                                                              |
| Storage media                           | MicroSDHC memory card                                                                                            | MicroSDHC memory card                                                                                               | Flash memory card                                                                                         |
| Image data format                       | JPEG, MPEG-1, MPEG-4                                                                                             | JPEG, MPEG-1, MPEG-4                                                                                                | JPEG, uncompressed data                                                                                   |
| Weight                                  | Smartscope M5: 400g<br>Smartscope FA: 300g                                                                       | Smartscope M5: 400g<br>Smartscope EY4: 300g                                                                         | Approx.1kg                                                                                                |
| Battery                                 | Re-chargeable Ni-MH<br>battery,<br>HR4U700AAA, 4.8V, 1000<br>mAh                                                 | Re-chargeable Ni-MH<br>battery,<br>HR4U700AAA, 4.8V, 1000<br>mAh                                                    | Not battery-operated<br>device                                                                            |
| Output terminals and data<br>collection | USB (1.1) terminal (B-<br>connector).<br>Compatible with<br>Windows® 7/8.1/10.                                   | USB (1.1) terminal (B-<br>connector).<br>Compatible with<br>Windows® 7/8.1/10.                                      | USB (1.1) terminal (B<br>connector)<br>Foot Switch connector<br>cable terminal                            |
| Exposure parameters                     | "Exempt Group" (no risk)<br>LED product according to<br>IEC 62471:2006                                           | "Exempt Group" (no risk)<br>LED product according to<br>IEC 62471:2006                                              | N/A                                                                                                       |

Image /page/6/Picture/3 description: The image shows the word "OPTOMED" in blue font. The "O" in "OPTOMED" is replaced with an eye symbol. The eye symbol is a circle with a smaller circle inside, and a white dot in the center.

{7}------------------------------------------------

|           | Group 1 instrument<br>according to ANSI Z80.36-<br>2016 | Group 2 instrument<br>according to ISO 15004-<br>2:2007 |     |
|-----------|---------------------------------------------------------|---------------------------------------------------------|-----|
| Standards | IEC 60601-<br>1:2005+A1:2012 (edition<br>3.1)           | IEC 60601-<br>1:2005+A1:2012 (edition<br>3.1)           | N/A |
|           | IEC 60601-1-2:2014<br>(edition 4.0)                     | IEC 60601-1-2:2014<br>(edition 4.0)                     |     |
|           | IEC 60601-1-<br>6:2010+A1:2013 (edition<br>3.1)         | IEC 60601-1-<br>6:2010+A1:2013 (edition<br>3.1)         |     |
|           | IEC 62471:2006                                          | IEC 62471:2006                                          |     |
|           | ISO 15004-1:2006                                        | ISO 15004-1:2006                                        |     |
|           | ANSI Z80.36-2016                                        | ISO 15004-2:2007                                        |     |
|           | ISO 10940:2009                                          | ISO 10940:2009                                          |     |
|           | ISO 10993-5:2009                                        | ISO 10993-5:2009                                        |     |
|           | ISO 10993-10:2009                                       | ISO 10993-10:2009                                       |     |
|           | IEC 62304:2006+A1:2015                                  | IEC 62304:2006+A1:2015                                  |     |
|           | IEC 62366-1:2015                                        | IEC 62366-1:2015                                        |     |
|           |                                                         |                                                         |     |

The modifications in Optomed Smartscope M5 camera with Smartscope FA optics module compared to Optomed Smartscope M5 camera with Optomed Smartscope EY4 optics module are:

Smartscope FA:

- . Illumination source blue led
- . Short pass filter in illumination path in front of blue led
- Barrier filter in imaging path
- . Optical safety Group 1 tested according to ANSI Z80.36-2016

The differences in Optomed Smartscope M5 camera with Smartscope FA optics module compared to Kowa Genesis Df are:

Smartscope FA:

- . Increased FOV (field of view)
- Higher image resolution

OPTOMED

{8}------------------------------------------------

## Performance Data

The following performance data is provided in support of the substantial equivalence determination.

### Electrical safety and electromagnetic compatibility (EMC)

Optomed Smartscope M5 camera with Smartscope FA optics module was tested according to all suitable clauses of IEC 60601-1:2005+A1:2012 (safety) and IEC 60601-1-2:2014 (EMC). Optomed Smartscope M5 camera with Smartscope FA optics module fulfills the requirements of the standards.

#### Optical safety

Optomed Smartscope M5 camera with Smartscope FA optics module was tested according to the standard IEC 62471:2006 and is classified as "Exempt Group" (NO RISK) LED products.

Optomed Smartscope M5 camera with Smartscope FA optics module was tested according to the standard ANSI Z80.36-2016 and is classified as Group 1 ophthalmic instruments.

### Software Verification and Validation

The level of concern of the software is moderate. Software verification and validation were conducted to ensure the fulfillment of the system requirements and functional specifications. Optomed Smartscope M5 camera with Smartscope FA optics module complies with the standard IEC 62304:2006+A1:2015.

### Environmental testing

Optomed Smartscope M5 camera with Smartscope FA optics module was tested according to ISO 15004-1:2006, IEC 60601-1 and IEC 60068-2 standards to verify the mechanical stress and ambient conditions for use and storage as prescribed for the device fulfills the requirements of the standard. In addition, Optomed Smartscope M5 camera with Smartscope FA optics module was tested according to IEC 60068-2 for high temperature, low temperature, shock, vibration and bump to verify transportation conditions.

#### Biocompatibility

Optomed Smartscope FA optics module has an eye cup that contacts skin around eye during normal usage of the device. The material of Optomed Smartscope FA optics module eye cup is Momentive LIM 6030AB. In vitro cytotoxicity tests according to ISO 10993-5:2009 and Irritation and skin

Image /page/8/Picture/15 description: The image shows the logo for Optomed. The logo is blue and features the word "OPTOMED" in a bold, sans-serif font. The "O" in "OPTOMED" is replaced by a stylized eye symbol, which is also blue. The background of the image is a light gray color.

{9}------------------------------------------------

sensitization tests according to ISO 10993-10:2009 were performed to the material of the eye cup, and the biocompatibility of Optomed Smartscope FA optics module is at suitable level.

#### Optical performance

Optomed M5 camera with Optomed Smartscope FA optics module fulfills the requirements of the standard ISO 10940:2009 Ophthalmic instruments - Fundus cameras.

#### Usability (Human Factors)

Optomed M5 camera with Optomed Smartscope FA optics module was designed and evaluated by following the principles depicted in the usability engineering process. The usage of Optomed M5 camera with Optomed Smartscope FA optics module was evaluated to be suitable for its intended use and the devices complies with the standards IEC 60601-1-6:2010+A1:2013 and IEC 62366-1:2015.

### Conclusions

Based on the provided performance data and the comparison, Optomed M5 camera with Optomed Smartscope FA optics module, is as safe, as effective and performs similar to the predicate devices. Optomed M5 camera with Optomed Smartscope FA optics module is substantially equivalent to the predicate devices.

Image /page/9/Picture/8 description: The image shows the word "OPTOMED" in blue. The "O" in "OPTOMED" is replaced with an eye symbol. The eye symbol is a blue circle with a white dot in the center.

---

**Source:** [https://fda.innolitics.com/device/K201325](https://fda.innolitics.com/device/K201325)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
