← Product Code [NEY](/productcode/NEY) · K201262

# Microwave Therapeutic System (K201262)

_Nanjing Eco Microwave System Co., Ltd. · NEY · Jul 9, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K201262

## Device Facts

- **Applicant:** Nanjing Eco Microwave System Co., Ltd.
- **Product Code:** [NEY](/productcode/NEY.md)
- **Decision Date:** Jul 9, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Microwave Therapeutic System is indicated for the coagulation (ablation) of soft tissue. The Microwave Therapeutic System is not intended for cardiac use.

## Device Story

Microwave Therapeutic System (SABERWAVE ECO-200G) is a software-controlled microwave generator delivering energy at 2450 MHz to a disposable ablation needle. System components include microwave source, control system, display, and temperature measurement module. Operates via microwave transistor and power amplifier; energy radiates from applicator tip to target tissue, causing thermal coagulation/ablation. Includes water circulation module for radiator cooling. Used in clinical settings by physicians. Three user-selectable operational modes (timing methods, foot switch/manual control, continuous/intermittent output) allow customization based on tissue type/size. Output affects clinical decision-making by enabling targeted tissue destruction. Benefits include reduced injury/risk compared to multi-needle puncture procedures.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), performance testing (IEC 60601-2-6), thermal effects, temperature monitoring accuracy, and software validation per FDA guidance.

## Technological Characteristics

Microwave generator, 2450 MHz frequency, 0-100W output. Single-channel architecture. Cooling via pumped normal saline. Includes temperature monitoring. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-6. Software-controlled.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Microwave Ablation System ([K183153](/device/K183153.md))

## Reference Devices

- Emprint™ Ablation System ([K133821](/device/K133821.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 9, 2021

Nanjing ECO Microwave System Co., Ltd Hong Wei. RA Manager Third & Fourth Floors, J5 Building, No. 15 Wanshou Road, Pukou District Nanjing, Jiangsu 211800 China

Re: K201262

Trade/Device Name: Microwave Therapeutic System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: NEY Dated: July 1, 2021 Received: July 6, 2021

Dear Hong Wei:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpm/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 53 1 -542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Long H. Chen
Long H. Chen -S-s
Date: 2021.07.09 08:54:29 -04'00'

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K201262

Device Name Microwave Therapeutic System

Indications for Use (Describe)

The Microwave Therapeutic System is indicated for the coagulation (ablation) of soft tissue. Therapeutic System is not intended for cardiac use.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) Summary

Prepared in accordance with the requirements of 21 CFR Part 807.92

## Prepared Date: July 8, 2021

## 1. Submitter's Information

The submitter of this pre-market notification is:

| Name:           | Nanjing ECO Microwave System Co., Ltd.                                                                                                                        |  |  |
|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Address:        | Third & Fourth Floors, J5 Building, NJUT Science & Technology<br>Industrial Park, No.15 Wanshou Road, Pukou District, 211800,<br>Nanjing, Jiangsu, P.R.China. |  |  |
| Contact person: | Hong Wei                                                                                                                                                      |  |  |
| Title:          | RA                                                                                                                                                            |  |  |
| E-mail:         | weihong@njeco.com.cn                                                                                                                                          |  |  |
| Tel:            | +86-025-58872663 Ext: 8020                                                                                                                                    |  |  |

# 2. Device Identification

| Trade/Device Name:     | Microwave Therapeutic System                                      |
|------------------------|-------------------------------------------------------------------|
| Models:                | SABERWAVE ECO-200G                                                |
| Regulation Number:     | 21 CFR 878.4400                                                   |
| Device Common<br>Name: | Microwave ablation system and accessories                         |
| Regulation Name:       | Electrosurgical cutting and coagulation device and<br>accessories |
| Regulation Class:      | Class II                                                          |
| Product Code:          | NEY                                                               |

## 3. Predicate Device

| Predicate device:                  |                                                                |
|------------------------------------|----------------------------------------------------------------|
| 510(K) number:                     | K183153                                                        |
| Device Common<br>Name:             | Microwave ablation system and accessories                      |
| Device Trade<br>/Proprietary Name: | Microwave Ablation System                                      |
| Model:                             | M150E                                                          |
| Manufacturer:                      | Surgnova Healthcare Technologies (Zhejiang) Co., Ltd.          |
| Regulation Number:                 | 21 CFR 878.4400                                                |
| Regulation Name:                   | Electrosurgical cutting and coagulation device and accessories |
| Regulation Class:                  | Class II                                                       |
| Product Code:                      | NEY                                                            |

{4}------------------------------------------------

| Reference device:  |                                                                |
|--------------------|----------------------------------------------------------------|
| 510(K) number:     | K133821                                                        |
| Device Name:       | Emprint™ Ablation System                                       |
| Manufacturer:      | Covidien LLC                                                   |
| Regulation Number: | 21 CFR 878.4400                                                |
| Regulation Name:   | Electrosurgical cutting and coagulation device and accessories |
| Regulation Class:  | Class II                                                       |
| Product Code:      | NEY                                                            |

## 4. Device Description

The Microwave Therapeutic System is a host machine will be used together with Disposable Microwave Ablation Needle for ablation of soft tissue. The Microwave Therapeutic System consists of microwave source, control system, display system and temperature measurement system. It is a software-controlled microwave generator that deliver the microwave energy at a working frequency of 2450 MHz.

The generator of the microwave therapeutic system is composed of a casinq, a power module, a control feedback unit, a display module, a microwave transmitting module, a water circulation module, and a temperature measurement module.

### 5. Indication for use

The Microwave Therapeutic System is indicated for the coagulation (ablation) of soft tissue. The Microwave Therapeutic System is not intended for cardiac use.

## 6. Comparison to Predicate Device

Comparison to the predicate devices, the subject device has same intended use, similar product design, same performance effectiveness, performance safety as the predicate device as summarized in the following table.

| Feature                                    | Subject device<br>K201262                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Predicate device<br>K183153                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Comments                                                                                                                                                                                                                                                                                   |
|--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer/K#                            | Nanjing ECO<br>Microwave System Co.,<br>Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Surgnova Healthcare<br>Technologies (Zhejiang)<br>Co., Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | --                                                                                                                                                                                                                                                                                         |
| Trade name                                 | Microwave Therapeutic<br>System<br>Model: SABERWAVE<br>ECO-200G                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Microwave ablation<br>system<br>Model: M150E                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | --                                                                                                                                                                                                                                                                                         |
| Classification name and<br>Regulation Name | Electrosurgical cutting<br>and coagulation device<br>and accessories<br>Class : II<br>Product code : NEY                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Electrosurgical cutting<br>and coagulation device<br>and accessories<br>Class : II<br>Product code : NEY                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                                       |
| Indications for use                        | The<br>Microwave<br>Therapeutic System is<br>indicated for the<br>coagulation (ablation) of<br>soft tissue. The<br>Microwave Therapeutic<br>System is not intended<br>for cardiac use.                                                                                                                                                                                                                                                                                                                                                                                                                      | The Microwave Ablation<br>System is intended for<br>the coagulation<br>(ablation) of soft tissues.<br>The MW Ablation<br>System is not intended<br>for use in cardiac<br>procedures.                                                                                                                                                                                                                                                                                                                                                                                                          | The microwave<br>generator instrument<br>and antenna of the<br>predicate device are<br>combined to submit<br>510k together, the<br>microwave generator<br>instrument and antenna<br>of subject device are<br>submitted 510k<br>separately. The intended<br>use description is the<br>same. |
| Intended purpose                           | coagulation and ablation<br>of tissue                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | coagulation and ablation<br>of tissue                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same                                                                                                                                                                                                                                                                                       |
| Operating principle                        | Microwave oscillating<br>signals are generated<br>by a microwave<br>transistor and amplified<br>by a microwave power<br>amplifier. Generator<br>delivers microwave<br>energy to the applicator<br>tip to thermally target<br>tissue, resulting in<br>coagulation and<br>ablation.<br>A microwave<br>therapeutic system<br>refers to equipment for<br>treating diseases with<br>microwave energy at a<br>working frequency of<br>2450 MHz.<br>The liquid in the silicone<br>tube is driven by the<br>rotation of the motor to<br>flow through the radiator<br>to realize the purpose of<br>radiator cooling. | During the surgery, the<br>microwave ablation<br>antenna is accurately<br>placed in the tumor<br>target area by imaging<br>techniques (such as CT,<br>US, etc.). The<br>microwave energy<br>generated by the<br>microwave generator<br>transmits to the<br>microwave ablation<br>antenna through the<br>coaxial cable, and then<br>it is radiated out through<br>the microwave<br>antenna and absorbed<br>by water molecules in<br>the tumor tissue. The<br>microwave energy<br>transform into heat, and<br>the temperature<br>rises rapidly result in<br>tumor tissue losing<br>bioactivity. | Equivalent.<br>The microwave<br>generator instrument<br>and antenna of the<br>predicate device are<br>combined to submit<br>510k together, the<br>subject device only<br>include generator<br>instrument, the principle<br>of generator instrument<br>is equivalent.                       |
| Design                                     | Single channel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Double channels                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Different, please refer to<br>comment 1.                                                                                                                                                                                                                                                   |
|                                            | cooling-water cycle,<br>thermal ablation,<br>the probe of TEMP, foot<br>switch                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | cooling-water cycle,<br>thermal ablation,<br>the probe of TEMP, foot<br>switch                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                                                                                       |
|                                            | AC input Voltage                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | AC100-240V, 50/60Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 100-240VAC 50-60 Hz                                                                                                                                                                                                                                                                        |
| Output Impedance                           | 50Ω nominal                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 50Ω nominal                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                                                                                       |
| Output parameters                          | 2450MHz±20MHz,                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 2450MHz±10MHz,                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                                                                                       |
|                                            | 0-100W                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 0-100W                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                            |
| Device<br>Temperature<br>Monitoring        | Temperature monitoring<br>features used to ensure<br>system safety                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Temperature monitoring<br>features used to ensure<br>system safety                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                                       |
| Device cooling                             | Pumped, normal saline<br>is used to cool the<br>antenna.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Pumped, normal saline<br>is used to cool the<br>antenna.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                                       |
| Operational mode                           | Three modes can be<br>selected by the user.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Only one operational<br>mode                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Different, please refer to<br>comment 2.                                                                                                                                                                                                                                                   |

{5}------------------------------------------------

{6}------------------------------------------------

Comment 1: Different, predicate device support double channels and two ablation needles can be inserted at the same time for tissue ablation. Our product only supports single channel and only one ablation needle for tissue ablation, Double-needle ablation speeds up the time of a single ablation, but there are injury and risks due to multiple punctures to patient. Single-needle ablation can also achieve the intended use and reduce the injury and risk of multiple punctures to the patient. The subject device has performed the Thermal effects test, performance test according to IEC 60601-2-6, temperature measurement accuracy test, the test result prove that single needle ablation can also achieve the intended use, the performance meet the requirements of IEC 60601-2-6, the temperature measurement accuracy meet the requirements of intended use, the difference does not affect the product's safety and performance.

Comment 2: Different, predicate devices have only a single operation mode. We have designed three modes according to the expected usage scenarios, which are convenient for users to operate according to different tissue types and sizes. The main difference between these three modes is that timing method is countdown or forward timing, foot switch control or manual control, output is continuous or intermittent, these are just to give users more operating options, and do not change the intended use and function of the product, the subject device has been conducted software verification according to the FDA software quidance, the test result proves the functions in each operational mode can be achieved. This difference does not raise a new risk.

All the differences don't affect the safety and effectiveness which is concluded after all the required testing, so no safety and effectiveness issues relating to the system come into conclusion.

### 7. Performance Data

### Clinical test:

Clinical testing is not required.

### Non-clinical data

Microwave Therapeutic System and predicate device are substantially equivalent in design concepts, technologies, which have been designed and tested in accordance with:

### Electrical Safety:

- -ES60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)
Electromagnetic Compatibility:

- -IEC 60601-1-2 Medical electrical equipment -- Part 1-2: General requirements for basic safety and essential performance -- Collateral Standard: Electromagnetic disturbances --Requirements and tests
### Performance:

{7}------------------------------------------------

### 510(k) Summary

- IEC 60601-2-6 Medical electrical equipment Part 2-6: Particular requirements for the basic safety and essential performance of microwave therapy equipment
### Software validation:

- Software validation is in compliance with FDA guidance for the content of premarket submissions for software contained in Medical devices.
#### Shelf Life:

- Accelerated aging tests were conducted to confirm the validity of the 8 years shelf life for l machine.
Thermal Effects test and Temperature monitoring test:

- FDA Guidance Premarket Notification (510(K)) Submissions for Electrosurgical Devices for -General Surgery
### 8. Conclusion

The proposed device is equivalent with respect to the basic system design and function to that of the predicate device. The proposed device isn't the implants and high-risk device. And it doesn't have new intended purposes, new medical, new target populations, and new users and so on. What's more. The differences between the predicate and proposed device do not raise new questions of safety or effectiveness.

---

**Source:** [https://fda.innolitics.com/device/K201262](https://fda.innolitics.com/device/K201262)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
