K201202 · Siemens Medical Solutions USA, Inc. · LLZ · Jun 2, 2020 · Radiology
Device Facts
Record ID
K201202
Device Name
MI View&GO
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
LLZ · Radiology
Decision Date
Jun 2, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
MI View&GO is a medical diagnostic application for viewing, manipulation, quantification, analysis and comparison of medical images with one or more time-points. MI View&GO supports functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR). MI View&GO is intended to be utilized by appropriately trained health care professionals to aid in the management of diseases associated with oncology, cardiology, neurology, and organ function. The images and results produced by MI View&GO can also be used by the physician to aid in radiotherapy treatment planning.
Device Story
MI View&GO is a software-only diagnostic application for viewing, manipulating, quantifying, and analyzing medical images (PET, NM, CT, MR). Used by healthcare professionals in clinical settings, it integrates with Siemens SPECT/CT and PET/CT scanners. Key features include auto-streaming of data post-acquisition, dynamic layout generation, cardiac masking for statistical calculations, and configurable user presets. It supports cardiology data filtering and basic viewer operations like slice correction and rotation. The software processes multi-modal and multi-time-point datasets to assist in disease management and radiotherapy planning. It does not change the fundamental technological characteristics of the predicate systems; it serves as an upgraded software package for existing imaging configurations.
Clinical Evidence
Bench testing only. Verification and validation activities confirmed that all functions perform as designed, specifications were met, and hazard mitigations were implemented. Testing included component integration, software validation, and system testing against predetermined acceptance values.
Technological Characteristics
Software-only medical device. Operates as an image processing application for PET, SPECT, CT, and MR data. Complies with IEC 62304 (software lifecycle), ISO 14971 (risk management), IEC 62366-1 (usability), and NEMA PS 3.1-3.20 (DICOM). Features include automated layout generation, cardiac masking, and data streaming. Cybersecurity controls implemented for access and data transfer protection.
Indications for Use
Indicated for appropriately trained healthcare professionals to view, manipulate, quantify, analyze, and compare medical images (PET, NM, CT, MR) to aid in managing oncology, cardiology, neurology, and organ function diseases, and for radiotherapy treatment planning.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol is a stylized representation of a caduceus, a traditional symbol of medicine, with intertwined snakes and wings. To the right of the symbol is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue, with "FDA" in a larger font size than the rest of the text.
Siemens Medical Solutions USA, Inc. % Clayton Ginn Regulatory Technical Specialist 2501 N Barrington Road HOFFMAN ESTATES IL 60192
Re: K201202
Trade/Device Name: MI View&GO VA10A Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: May 1, 2020 Received: May 4, 2020
Dear Clayton Ginn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
June 2, 2020
{2}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
#### K201202
Device Name MIView&GO VA10A
#### Indications for Use (Describe)
MI View&GO is a medical diagnostic application for viewing, manipulation, quantification, analysis and comparison of medical images with one or more time-points. MI View&GO supports functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR).
MI View&GO is intended to be utilized by appropriately trained health care professionals to aid in the management of diseases associated with oncology, neurology, neurology, and organ function. The images and results produced by MI View&GO can also be used by the physician to aid in radiotherapy treatment planning.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid 0MB number."
Form Approved: 0MB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{4}------------------------------------------------
# Section 05
## 510(k) Summary
K201202
# as required by 21 CFR Part 807.87(h)
## ldentification of the Submitter
| | Primary Contact: | Alternate Contact: |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Clayton Ginn<br>Regulatory Technical<br>Specialist<br>Siemens Medical Solutions<br>USA, Inc.<br>810 Innovation Drive<br>Knoxville, TN 37932 | Veronica Padharia<br>Regulatory Technical<br>Specialist<br>Siemens Medical Solutions<br>USA, Inc.<br>2501 N. Barrington Road<br>Hoffman Estates, IL 60192 |
| Telephone Number: | (865) 898-2692 | (630) 877-5761 |
| Fax Number: | (865) 218-3227 | (847) 304-6023 |
| Name / Address of<br>Manufacturer | Siemens Medical Solutions USA, Inc<br>Molecular Imaging<br>2501 N. Barrington Road<br>Hoffman Estates, IL 60192<br>USA | |
| Date of Submission: | May 1st, 2020 | |
| Identification of the product | | |
| Device Proprietary Name: | MI View&GO VA10A | |
| Common Name: | Image Processing Software | |
| Classification Name: | Picture Archiving and Communication System per 21 CFR<br>892.2050 | |
| Product Code: | LLZ | |
| Classification Panel: | Radiology | |
| Device Class: | Class II | |
| Device Proprietary Name: | syngo.via MI Workflows VB40A | |
| Common Name: | Image Processing Software | |
| Classification Name: | Picture Archiving and Communication System per 21 CFR<br>892.2050 | |
| Product Code: | LLZ | |
| Classification Panel: | Radiology | |
| Device Class: | Class II | |
| Manufacturer: | Siemens Medical Solutions<br>USA, inc. | |
| 510(k) Number: | K191309 (July 2019) | |
| Reference Predicate Device | | |
| Device Proprietary Name: | syngo.CT View&GO VB20A | |
| Common Name: | Image Processing Software | |
| Classification Name: | Picture Archiving and Communication System per 21 CFR<br>892.2050 | |
| Product Code: | LLZ | |
| Classification Panel: | Radiology | |
| Device Class: | Class II | |
| Manufacturer: | Siemens Medical Solutions<br>USA, Inc. | |
| 510(k) Number: | K170952 (April 2017) | |
| Reference Predicate Device | | |
| Device Proprietary Name: | syngo.CT Extended Functionality VB40A | |
| Common Name: | Image Processing Software | |
| Classification Name: | System, X-ray, Tomography, Computed per 21 CFR | |
| Product Code: | JAK | |
| Classification Panel: | Radiology | |
| Device Class: | Class II | |
| Manufacturer: | Siemens Medical Solutions<br>USA, Inc. | |
| 510(k) Number: | K192402 (September 2019) | |
{5}------------------------------------------------
## Primary Predicate Device
Classification Name:
892.1750
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{6}------------------------------------------------
syngo.via MI Workflows is deemed the primary predicate device due to it being the most similar to the device under review of this submission with respect to indications for use and technical characteristics. syngo.CT View&GO and syngo.CT Extended Functionality are considered reference predicate devices because MI View&GO has integrated technical characteristics initially cleared within these devices.
{7}------------------------------------------------
#### Device Description
MI View&GO is a software-only medical device which will be delivered with the new generation of Siemens SPECT/CT and PET/CT scanners.
MI View&GO is a medical diagnostic application for viewing, manipulation, quantification, analysis and comparison of medical images with one or more time-points. MI View&GO supports functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR)
MI View&GO is intended to be utilized by appropriately trained health care professionals to aid in the management of diseases associated with oncology, cardiology, neurology, and organ function.
| Feature | New<br>Functionality | Modification Description |
|------------------------------------------------------------------------------------------|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Auto Viewing | Yes | Stream data into the application immediately<br>following acquisition. |
| Layout Gallery with<br>Dynamic Summing,<br>Automatic Layouts, and<br>Remove Images | Yes | Generate Automatic Layouts at runtime,<br>Automatically sum frames for review in the<br>layout gallery, and remove images from a<br>display |
| Configurable User Presets | Yes | Configuration UI for preferred presets and<br>layouts |
| Tool Gallery | No | Functionality Cleared in K170952 |
| Organ Processing | No | Functionality Cleared in K191309 |
| MI Cardiology Reading and<br>Third Party Software<br>(Cedars 2017, Corridor<br>4DM 2018) | Yes | Restart applications and filter the cardiology<br>data passed to an application. |
| Series Panel | No | Functionality Cleared in K170952, K191309 |
| Basic Operation in the<br>Viewer | Yes | Correct slice positions, panning, rotation on<br>synchronized segments |
| Automatic Reconstruction,<br>Auto Ranges, and Cardiac<br>Auto Ranges | Yes | Generate parallel ranges with Auto Ranges and<br>generate a series of images in cardiac planes<br>Cardiac Auto Ranges |
| Display Types and Image<br>Fusion | No | Functionality Cleared in K170952, K191309 |
| 3D Visualization and<br>Orientation | No | Functionality Cleared in K170952, K191309 |
| Restricting Volumes and<br>Cardiac Masking | Yes | Mask the heart and exclude background regions<br>that can have an impact on statistical<br>calculations in an examination with Cardiac<br>Masking. |
| Image Optimization and<br>Preparation | No | Functionality Cleared in K170952, K191309 |
| Toggle Image Overlay | Yes | Configure customized image text |
| Spine and Rib Labeling | No | Functionality Cleared in K191309 |
MI View&GO includes the following features:
{8}------------------------------------------------
| Neuro DSA | No | Functionality Cleared in K192402 |
|--------------------|----|----------------------------------|
| Osteo Extension | No | Functionality Cleared in K192402 |
| Oncology Extension | No | Functionality Cleared in K192402 |
| Vessel Extension | No | Functionality Cleared in K192402 |
| ROI HU Threshold | No | Functionality Cleared in K192402 |
| Dual Energy ROI | No | Functionality Cleared in K192402 |
| Endoscopic View | No | Functionality Cleared in K192402 |
| Printing of Images | No | Functionality Cleared in K170952 |
## Technological Characteristics
The MI View&GO software modifications are based on the commercially available syngo.via MI Workflows software (K191309), syngo.CT View&GO software (K170952), and syngo.CT Extended Functionality (K192402). The features into this software device had no impact on the technological characteristics already present in the commercially available predicate system.
## Intended Use
An individual software program, or group of programs, routines, or algorithms that add specific image processing and/or analysis capabilities to a positron emission tomography (PET) and Single Photon Emission Computed Tomography (SPECT) imaging system configuration. A basic set of application programs and routines is included with such computer controlled imaging systems and they can be upgraded to correct programming errors or to add new system capabilities. Some application software routines or groups of routines (packages) must be combined with specific hardware or firmware accessories or configurations in order to function as intended. Application program packages are typically identified by a proprietary name and "version" or "upgrade" number.
The intended use for MI View&GO compared to the primary and reference device has not changed.
#### Indications for Use
MI View&GO is a medical diagnostic application for viewing, manipulation, quantification, analysis and comparison of medical images with one or more time-points. MI View&GO supports functional data, such as positron emission tomography (PET) or nuclear medicine (NM), as well as anatomical datasets, such as computed tomography (CT) or magnetic resonance (MR)
MI View&GO is intended to be utilized by appropriately trained health care professionals to aid in the management of diseases associated with oncology, cardiology, neurology, and organ function. The images and results produced by MI View&GO can also be used by the physician to aid in radiotherapy treatment planning.
{9}------------------------------------------------
## Performance Testing / Safety and Effectiveness
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device.
Risk Management has been ensured via risk analyses in compliance with ISO 14971 to identify and provide mitigation to potential hazards beginning early in the design cycle and continuing throughout the development of the product. Siemens Medical Solutions, USA Inc. adheres to recognized and established industry standards for development including EN ISO 13485 and IEC 62304.
Cybersecurity information in accordance with FDA Guidance documents issued October 2, 2014 has been provided. The software has specific cybersecurity controls to prevent unauthorized access, modifications, misuse or denial of use. Additionally, controls are enabled to prevent the unauthorized use of information that is stored, accessed or transferred between the software and external devices.
Verification and Validation activities have been successfully performed on the software package, including assurance that functions work as designed, performance requirements and specifications have been met, and that all hazard mitigations have been fully implemented. All testing has met the predetermined acceptance values. Traceability of the requirements specified in the requirement specifications and functional specifications is ensured during component integration, software validation and system testing.
The device is designed and manufactured in accordance with Quality System Regulations as outlined in 21 CFR 820. The FDA recognized standards are listed as follows:
- . Recognition Number 13-79: IEC 62304 Edition 1.1 2015-06
- . Recognition Number 12-300: NEMA PS 3.1 – 3.20 (2016)
- Recognition Number 5-40: ISO 14971:2007 Second Edition ●
- . Recognition Number 5-114: IEC 62366-1 Edition 1.0 2015
- . Recognition Number 5-117: ISO 15223-1 Third Edition 2016
#### Statement Regarding Substantial Equivalence:
There are no differences in the Intended Use or Fundamental Technological Characteristics of the MI View&GO software as compared to the currently commercially available syngo.via MI Workflows software (K191309). The Indications for Use for MI View&GO is a condensed version of the Indications for Use of syngo.via MI Workflows and does not expand the intended use from that of the predicate device. Both the current and predicate devices are used for viewing, manipulation, 3D- visualization and comparison of medical images from multiple imaging modalities and/or multiple time-points
Additionally, there have been no changes that raise any new issues of safety and effectiveness as compared to the predicate device. Based on this information, as well as the documentation in support of the modifications, it is Siemens opinion that the MI View&GO software-with the modifications outlined in this application-is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.