← Product Code [EBW](/productcode/EBW) · K201192

# Impla-NX (Model: ISE-270M) (K201192)

_Micro-Nx Co., Ltd. · EBW · May 21, 2021 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K201192

## Device Facts

- **Applicant:** Micro-Nx Co., Ltd.
- **Product Code:** [EBW](/productcode/EBW.md)
- **Decision Date:** May 21, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.4200
- **Device Class:** Class 1
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The Impla-NX (Model: ISE-270M) is intended for use in dental oral surgery and dental implant. The main unit is designed to be used with a specific dental micromotor that drives dental handpieces fitted with appropriate tools to cut hard tissues in the mouth.

## Device Story

Impla-NX (Model: ISE-270M) is an implant surgical engine for dental surgery. System components include a main control unit, BLDC motor, and foot control. Device operates by converting AC power to DC to drive the BLDC motor, which provides rotational power to connected dental handpieces. Used in clinical dental settings by professionals to perform oral surgery and implant procedures. Output is rotational mechanical energy for cutting hard tissues. Healthcare providers control parameters (speed, torque) via the unit; device assists in precise surgical site preparation. Benefits include controlled, motorized tissue cutting for dental implants.

## Clinical Evidence

No clinical data. Bench testing only. Testing included ISO 3964 (coupling dimensions), ISO 7494-1 (dental units), ISO 14457:2017 (handpieces/motors), ISO 17665-1/2 (steam sterilization), IEC 60601-1/1-2 (electrical safety/EMC), IEC 80601-2-60 (dental equipment safety), and IEC 62471 (photobiological safety).

## Technological Characteristics

System includes control unit, BLDC motor, and wired foot switch. Operates on AC power converted to DC. Features include adjustable speed (200-40,000 rpm) and torque (max 70Ncm). Connectivity is via wired handpiece/foot control. Sterilization validation per ISO 17665-1, 2. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 80601-2-60.

## Regulatory Identification

A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.

## Predicate Devices

- Surgic Pro by NSK Nakanishi, Inc ([K173905](/device/K173905.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 21, 2021

MICRO-NX Co., Ltd. % Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 1150 Roosevelt STE 200 Irvine, California 92620

Re: K201192

Trade/Device Name: Impla-NX (Model: ISE-270M) Regulation Number: 21 CFR 872.4200 Regulation Name: Dental handpiece and accessories Regulatory Class: Class I, reserved Product Code: EBW Dated: May 14, 2021 Received: May 19, 2021

Dear Priscilla Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

510(k) Number (if known)

Device Name Impla-NX (Model: ISE-270M)

Indications for Use (Describe)

The Impla-NX (Model: ISE-270M) is intented for use in dental implant. The main unit is designed to be used with a specific dental micromotor that drives dental handpieces fitted with appropriate tools to cut hard tissues in the mouth.

Type of Use (Select one or both, as applicable)

× Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

# 510(k) Summary

## (K201192)

This summary of 510(k) is being submitted in accordance with requirements of 21 CFR Part 807.92

Date: May 3, 2021

#### 1. 510K Applicant / Submitter:

MICRO-NX Co., Ltd. 37, Dalseong2chadong 3-ro, Guji-myeon, Dalseong-gun, Daegu 711-891 REPUBLIC OF KOREA Tel: +82-53-611-0071 Fax: +82-53-611-0074

#### 1. Submission Contact Person

LK Consulting Group USA, Inc. 1150 Roosevelt STE 200, Irvine CA 92620 Priscilla Chung Phone: 714.202.5789 Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com

#### 2. Device

- Proprietary Name Impla-NX (Model: ISE-270M)
- Common Name Implant Surgical Engine .
- . Classification Name - Dental handpiece and accessories
- Classification 21CFR 872.4200 (Product Code: EBW) .

#### 3. Predicate Device

- Surgic Pro by NSK Nakanishi, Inc (K173905)
#### 4. Description:

As an engine to operate and control handpiece used for dental implant surgery, this device is composed of main body, BLDC motor and foot control.

With external power supply being provided, this device operates the main body through switching AC power to DC power and it performs dental implant surgery with handpiece being connected for rotation power generated through turning BLDC motor.

{4}------------------------------------------------

### 8. Indications for Use

The Impla-NX (Model: ISE-270M) is intended for use in dental oral surgery and dental implant. The main unit is designed to be used with a specific dental micromotor that drives dental handpieces fitted with appropriate tools to cut hard tissues in the mouth.

#### 9. Substantial Equivalence Discussion:

The Impla-NX (Model: ISE-270M) is substantially equivalent to Surgic pro( K173905) in the indications for use and the technological characteristics. The following comparison table is presented to demonstrate substantial equivalence.

The Impla-NX (Model: ISE-270M) does not have new Indications for Use: it is the same as the predicate device's indications for use. It shows equivalent specifications with the predicate devices in most of parameters.

The difference between the subject device and the predicate device is the number of memory set but this does not affect the performance as it is just for user convenience. The user can always adjust the setting parameters as needed.

Another difference is the torque of motor that the subject device has a slightly lower max torque than the predicate device. However, the max torque of 70Ncm is more than enough to be used in the clinical setting for the Indications for Use and it is covered by the max value of the predicate device; therefore, it would not raise a question in safety.

We conclude that the subject device is substantially equivalent to the predicate device.

|                     | Subject Device                                                                                                                                                                                                       | Predicate Device                                                                                                                                                                                            | Comparison                                                              |                |                                                                             |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|----------------|-----------------------------------------------------------------------------|
| 510(k) Number       | K201192                                                                                                                                                                                                              | K173905                                                                                                                                                                                                     | -                                                                       |                |                                                                             |
| Device Name         | Impla-NX (Model:<br>ISE-270M)                                                                                                                                                                                        | Surgic Pro                                                                                                                                                                                                  | -                                                                       |                |                                                                             |
| Common Name         | Controller, Foot,<br>Handpiece and cord                                                                                                                                                                              | Controller, Foot,<br>Handpiece And<br>Cord                                                                                                                                                                  | -                                                                       |                |                                                                             |
| Manufacturer        | MICRO-NX Co.,<br>Ltd.                                                                                                                                                                                                | NSK Nakanishi,<br>Inc                                                                                                                                                                                       | -                                                                       |                |                                                                             |
| Indications for Use | The Impla-NX<br>(Model: ISE-270M)<br>is intended for use<br>in dental oral<br>surgery and dental                                                                                                                     | The Surgic Pro+ /<br>Surgic Pro is<br>intended for use in<br>dental oral surgery<br>and dental implant.                                                                                                     | Same                                                                    |                |                                                                             |
|                     | implant. The main<br>unit is designed to<br>be used with a<br>specific dental<br>micromotor that<br>drives dental<br>handpieces fitted<br>with appropriate<br>tools to cut hard<br>tissues in the mouth.<br>Electric | The main unit is<br>designed to be used<br>with a specific<br>dental micromotor<br>that drives dental<br>handpieces fitted<br>with appropriate<br>tools to cut hard<br>tissues in the<br>mouth.<br>Electric |                                                                         |                |                                                                             |
| Micromotor drive    | Micromotor drive                                                                                                                                                                                                     | Micromotor drive                                                                                                                                                                                            | Same                                                                    |                |                                                                             |
| Package contents    | Control Unit<br>Micromotor<br>Foot switch(wired)<br>Irrigation Tube<br>Operation Manual                                                                                                                              | Control Unit<br>Micromotor<br>Foot Controller<br>Handpiece<br>Irrigation tube<br>Operation Manual                                                                                                           | Different<br>- The subject<br>device does not<br>include<br>handpieces. |                |                                                                             |
|                     | Optic                                                                                                                                                                                                                | Yes                                                                                                                                                                                                         | Yes                                                                     | Same           |                                                                             |
|                     | Memory                                                                                                                                                                                                               | ■ 6 Pre-set implant<br>systems<br>■ 9 user defined of<br>9 steps each                                                                                                                                       | ■ 8 Pre-set implant<br>systems<br>■ 8 user defined of<br>8 steps each   | Different      |                                                                             |
|                     |                                                                                                                                                                                                                      | Torque on the motor                                                                                                                                                                                         | Max.70Ncm                                                               | Max.80Ncm      | Same<br>- It is within the<br>control range of<br>the predicate<br>product. |
|                     |                                                                                                                                                                                                                      | Speed on the motor                                                                                                                                                                                          | 200-40,000 rpm                                                          | 200-40,000 rpm | Same.                                                                       |

{5}------------------------------------------------

# 10. Performance Tests (Non-clinical)

Non-clinical bench tests were performed as followings:

- ISO 3964 Dental Handpieces Coupling dimensions

{6}------------------------------------------------

- ISO 7494-1 Dentistry - Dental units -Part 1: General requirements and test methods

- ISO 14457:2017 Dentistry Handpieces and motors
- ISO 17665-1, 2 Steam Sterilization Validation
- IEC 60601-1, IEC 60601-1-2: Electrical safety and EMC
- IEC 80601-2-60: Particular Requirements For The Basic Safety And Essential Performance Of Dental Equipment
- IEC 62471 Photobiological safety of lamps and lamp systems

Along with the above tests, sterilization validation, and software validation were also conducted. None of the testing demonstrated any design characteristics that violated the requirements of the standards or resulted in any safety hazard.

#### 10. Conclusions:

Based on the information provided in this premarket notification, MICRO-NX Co., Ltd. concludes that the Impla-NX (Model: ISE-270M) is substantially equivalent to the predicate device as described herein in safety and effectiveness.

---

**Source:** [https://fda.innolitics.com/device/K201192](https://fda.innolitics.com/device/K201192)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
