K201153 · Shanghai Suokang Medical Implants Co., Ltd. · OWR · Jun 13, 2023 · General, Plastic Surgery
Device Facts
Record ID
K201153
Device Name
Tisuthes e-PTFE Surgical Patch
Applicant
Shanghai Suokang Medical Implants Co., Ltd.
Product Code
OWR · General, Plastic Surgery
Decision Date
Jun 13, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
TisuthesTM e-PTFE Surgical Patches are intended for facial plastic and reconstruction surgery.
Device Story
Tisuthes e-PTFE Surgical Patch is an implantable device for soft tissue augmentation or repair in facial plastic and reconstructive surgery. Provided as sterile sheets, blocks, or preformed 3D shapes (nasal, chin, malar, temporal, piriform). Used by surgeons in clinical settings to provide structural support or contouring. Device acts as a permanent implant; biocompatible e-PTFE material integrates with host tissue. Benefits include restoration of facial aesthetics and structural integrity. No electronic components or software.
Clinical Evidence
Bench testing only. No clinical data provided. Compliance with ASTM F754-08 and ISO 10993 series (biocompatibility, cytotoxicity, sensitization, irritation, systemic toxicity, genotoxicity, implantation effects) and ISO 14630 (general requirements for non-active surgical implants) demonstrated.
Technological Characteristics
Material: Expanded Polytetrafluoroethylene (e-PTFE). Form factor: Sheets, blocks, and preformed 3D implants. Sterilization: Ethylene Oxide (ISO 11135). Standards: ASTM F754-08, ISO 14630, ISO 10993 series. Non-active, mechanical implant.
Indications for Use
Indicated for facial plastic and reconstructive surgery. Contraindicated for cardiovascular defects, TMJ reconstructions, cosmetic lip filler, and dermal placement.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font.
June 13, 2023
Shanghai Suokang Medical Implants Co., Ltd. Lixinhua Lixinhua Deputy General Manager Building 2, No. 613, Suide Rd., Putuo Region, Shanghai, 200311. P.R. China Shanghai, Shanghai 200000 China
Re: K201153
Trade/Device Name: Tisuthes e-PTFE Surgical Patch Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: OWR, FTL Dated: Mav 22, 2021 Received: May 27, 2021
Dear Lixinhua Lixinhua:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# David Krause -S
David Krause, Ph.D. Deputy Director OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K201153
Device Name TisuthesTM e-PTFE Surgical Patch
#### Indications for Use (Describe)
TisuthesTM e-PTFE Surgical Patches are indicated for facial plastic and reconstructive surgery.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### Section 5 - 510(k) Summary
Date of Summary Preparation: 2020.11.19
#### 1. Submitter's Identifications
Submitter's Name: Shanghai Suokang Medical Implants Co., Ltd. Address: Building 2, No. 613, Suide Rd., Putuo District,Shanghai,China
Contact Person: Lixinhua Contact Email: surgithes@vip.sina.com
Phone number: +86 21 6608 2766 ext.812
Fax number: +86 66082138
#### 2. Correspondent's Identifications
Correspondent's Name: Shanghai Suokang Medical Implants Co., Ltd. Address: Building 2, No. 613, Suide Rd., Putuo District, Shanghai,China Contact Person: Lixinhua Contact Email: surgithes@vip.sina.com Phone number: +86 21 6608 2766 ext.812 Fax number: +86 66082138
#### 3. Name of the Device
Device Classification Name: Mesh, Surgical, Polymeric Product Code: FTL , OWR
Product Name: Tisuthes™ e-PTFE Surgical Patch Trade Name: Tisuthes™ e-PTFE Surgical Patch Common Name: Surgical mesh Classification Panel: General & Plastic Surgery Product Code: FTL , OWR Device Classification: Class II
#### 4. The Predicate Devices
K021889
Trade/Device Name: Surgiform Augmentation Material
K091011
Trade/Device Name: PureForm ePTFE Facial Implants (Nasal, Chin, Malar)
#### 5. Device Description
{5}------------------------------------------------
5.1 Tisuthes™ e-PTFE Surgical Patches are intended for facial plastic and reconstruction surgery. The device is provided sterile.
The device is available in the forms of sheets, blocks and preformed three-dimensional forms (chin, nasal, temporal piriform and malar) with multiple sizes.
5.2. Symbol
TP: Expanded Polytetrafluoroethylene (e-PTFE) Surgical Patch (Brand Name:
Tisuthes)
5.3 Product numbers definition
5.3.1 Marking method
TP:
Image /page/4/Figure/9 description: The image shows a diagram with labels for different codes. The top line is labeled "Serial number" and is connected to "X...XX". The middle line is labeled "Category number" and is connected to "X". The bottom line is labeled "Product code" and is connected to "TP".
Category number: S: Soft Patch
R: Block
N: Nasal implants
C: Chin implants
M: Malar implants
F: Piriform implants
T: Temporal implants
#### 5.4 Model and Specification
| Spc. | a<br>(mm) | b<br>(mm) | c<br>(mm) | d<br>(mm) | Tolerance | Diagram |
|-------|-----------|-----------|-----------|-----------|-----------|--------------------------------|
| TPS01 | 60 | 40 | 1 | / | ±10% | Image: Diagram of a soft patch |
| TPS02 | 60 | 40 | 2 | / | ±10% | |
| TPS03 | 60 | 40 | 3 | / | ±10% | |
| TPS04 | 60 | 40 | 4 | / | ±10% | |
| TPS05 | 80 | 60 | 1 | / | ±10% | |
| TPS06 | 80 | 60 | 2 | / | ±10% | |
| TPS07 | 80 | 60 | 3 | / | ±10% | |
| TPS08 | 80 | 60 | 4 | / | ±10% | |
{6}------------------------------------------------
| TPS09 | 140 | 60 | 1 | / | ±10% | |
|-------------------------|-----------|-----------|-----------|-----------|-----------|-----------------------------------------------------|
| TPS10 | 140 | 60 | 2 | / | ±10% | |
| TPS11 | 140 | 60 | 3 | / | ±10% | |
| TPS12 | 140 | 60 | 4 | / | ±10% | |
| Block (TPR) | | | | | | |
| | a<br>(mm) | b<br>(mm) | c<br>(mm) | d<br>(mm) | Tolerance | |
| Spc. | | | | | | Diagram |
| TPR01 | 70 | 20 | 5 | / | ±10% | |
| TPR02 | 70 | 20 | 8 | / | ±10% | |
| TPR03 | 60 | 50 | 3 | / | ±10% | |
| TPR04 | 60 | 40 | 5 | / | ±10% | |
| TPR05 | 70 | 15 | 8 | / | ±10% | |
| TPR06 | 70 | 15 | 6 | / | ±10% | |
| TPR07 | 70 | 15 | 5 | / | ±10% | |
| TPR08 | 100 | 50 | 5 | / | ±10% | |
| TPR09 | 100 | 50 | 8 | / | ±10% | |
| Nasal implants (TPN) | | | | | | |
| Spc. | a<br>(mm) | b<br>(mm) | c<br>(mm) | d<br>(mm) | Tolerance | Diagram |
| TPN11 | 55 | 16 | 5.5 | 20 | ±10% | |
| TPN12 | 65 | 11 | 5 | 20 | ±10% | |
| TPN13 | 55 | 11 | 5 | 20 | ±10% | |
| TPN14 | 55 | 11 | 6.5 | 20 | ±10% | |
| TPN15 | 50 | 12 | 5 | 10 | ±10% | |
| TPN16 | 50 | 12 | 6.5 | 10 | ±10% | |
| TPN17 | 60 | 12 | 5 | 10 | ±10% | |
| Chin implants (TPC) | | | | | | |
| Spc. | a<br>(mm) | b<br>(mm) | c<br>(mm) | d<br>(mm) | Tolerance | Diagram |
| TPC01 | 55 | 40 | 8 | 15 | ±10% | |
| TPC02 | 55 | 40 | 12 | 15 | ±10% | |
| | | | | | | |
| TPC03 | 55 | 40 | 10 | 15 | ±10% | Image: Diagram of a lens with labels a, b, c, and d |
| TPC04 | 60 | 25 | 8 | 15 | ±10% | Image: Diagram of a lens with labels a, b, c, and d |
| TPC05 | 60 | 25 | 10 | 15 | ±10% | Image: Diagram of a lens with labels a and d |
| TPC06 | 55 | 40 | 8 | 15 | ±10% | Image: Diagram of a lens with labels b and c |
| TPC07 | 62 | 42 | 8 | 15 | ±10% | |
| TPC08 | 62 | 42 | 10 | 15 | ±10% | Image: Diagram of a lens with labels a and d |
| TPC09 | 60 | 40 | 8 | 15 | ±10% | Image: Diagram of a lens with labels a, b, c, and d |
| TPC10 | 50 | 18 | 5.5 | 15 | ±10% | Image: Diagram of a lens with labels a, b, c, and d |
| TPC11 | 42 | 10 | 5 | 15 | ±10% | Image: Diagram of a lens with labels a, c, and d |
| TPC12 | 55 | 28 | 8 | 15 | ±10% | |
| TPC13 | 55 | 28 | 10 | 15 | ±10% | |
| TPC14 | 55 | 28 | 12 | 15 | ±10% | Image: Diagram of TPC14 |
| TPC15 | 60 | 28 | 8 | 15 | ±10% | Image: Diagram of TPC15 |
| TPC16 | 60 | 28 | 10 | 15 | ±10% | Image: Diagram of TPC16 |
| TPC17 | 60 | 28 | 12 | 15 | ±10% | Image: Diagram of TPC17 |
| Malar implants(TPM) | | | | | | |
| Spc. | a (mm) | b (mm) | c (mm) | d (mm) | Tolerance | Diagram |
| TPM01 | 60 | 28 | 5 | / | ±10% | Image: Diagram of TPM01 |
| TPM02 | 60 | 28 | 5 | / | ±10% | Image: Diagram of TPM02 |
| Piriform Implants (TPF) | | | | | | |
| Spc. | a (mm) | b (mm) | c (mm) | d (mm) | Tolerance | Diagram |
| TPF01 | 25 | 22 | 5 | / | ±10% | Image: Diagram of TPF01 |
| TPF03 | 25 | 22 | 8 | / | ±10% | Image: Diagram of TPF03 |
| TPF02 | 25 | 22 | 5 | / | ±10% | Image: Diagram of TPF02 |
| TPF04 | 25 | 22 | 8 | / | ±10% | Image: Diagram of TPF04 |
{7}------------------------------------------------
{8}------------------------------------------------
### 6. Intended Use of Device
Tisuthes™ e-PTFE Surgical Patches are intended for facial plastic and reconstruction surgery.
{9}------------------------------------------------
## 7. Summary of Substantial Equivalence
| Product | | Proposed device | Predicate device | Predicate device |
|---------|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| | | TisuthesTM ePTFE Surgical Patch K201153 | Surgiform Augmentation Material K021889 | Surgiform's PureForm ePTFE Impants K091011 |
| | Indication for use | TisuthesTM e-PTFE Surgical Patches are intended for facial plastic and reconstructive surgery. | For Plastic and Reconstructive Surgery. | PureForm ePTFE Facial Implants are indicated for use in facial plastic and reconstructive surgery. |
| | Contraindications | Cardiovascular defects Tempormandibular joint (TMJ reconstructions) Cosmetic lip filler Dermal placement | Cardiovascular defects Tempormandibular joint (TMJ reconstructions) Cosmetic lip filler Dermal placement | Tempormandibular joint (TMJ reconstructions) Dermal placement |
| | Material | e-PTFE | e-PTFE | e-PTFE |
| | Design principle | Augmentation or repair of the soft tissue | Augmentation or repair of the soft tissue | Augmentation or repair of the soft tissue |
| | Shapes | The device is available in the forms of sheets, blocks, and preformed three-dimensional forms (chin, nasal, temporal piriform and malar) with multiple sizes. | Flat sheets and Carvable block | Pre-formed 3-D shapes (chin, nasal, malar and para-nasal) in multiple sizes. |
| | | | | |
| | | | | |
| | | | | |
| | Physical characteristics | Dimension (mm) | Sheets:<br>Length: 60, 80,140;<br>Width: 40, 60;<br>Thickness: 1, 2, 3, 4.<br><br>Blocks:<br>Length: 60, 70, 100;<br>Width: 15, 20, 40, 50; | Sheets:<br>Length: 80, 190;<br>Width: 30, 70, 170;<br>Thickness: 0.35, 1, 2, 3.<br><br>Blocks:<br>Length: 60, 70;<br>Width: 15, 30, 50, 100; |
{10}------------------------------------------------
| | | Thickness: 3, 5, 6, 8. | Thickness: 5, 6, 8, 10. | | | | | | | |
|--|--|----------------------------------|---------------------------|---------------------------------|---------------|---------------|---------------|---------------|---------------|--|
| | | Nasal implants[1]: (TPN11~TPN17) | | ASIAN STRAIGHT NASAL DORSUM | | | | | | |
| | | a: 50, 55, 60; | | | Item<br>Code | Dim A<br>(mm) | Dim B<br>(mm) | Dim C<br>(mm) | Dim D<br>(mm) | |
| | | b: 11, 12, 16; | | | 4AND03 | 50.0 | 7.5 | 9.5 | 4.0 | |
| | | c: 5, 5.5,6.5; | | | 4AND04 | 50.0 | 8.0 | 10.0 | 4.0 | |
| | | d: 10, 20. | | | 5AND02 | 50.0 | 7.5 | 9.5 | 5.0 | |
| | | | | | 6AND02 | 50.0 | 8.0 | 10.0 | 6.0 | |
| | | | ASIAN CURVED NASAL DORSUM | | | | | | | |
| | | | | Item<br>Code | Dim A<br>(mm) | Dim B<br>(mm) | Dim C<br>(mm) | Dim D<br>(mm) | | |
| | | | | 4AND01 | 50.0 | 9.5 | 7.5 | 4.0 | | |
| | | | | 4AND02 | 50.0 | 10.0 | 8.0 | 4.0 | | |
| | | | | 5AND01 | 50.0 | 9.5 | 7.5 | 5.0 | | |
| | | | | 6AND01 | 50.0 | 10.0 | 8.0 | 6.0 | | |
| | | | | | | | | | | |
| | | Chin implants: | | CHIN IMPLANTS | | | | | | |
| | | a: 42, 55, 60, 62; | | Item<br>Code | Dim A<br>(mm) | Dim B<br>(mm) | Dim C<br>(mm) | Dim D<br>(mm) | | |
| | | b: 10, 18, 25, 28, 40, 42; | | 1EC01 | 52.9 | 38.1 | 7.7 | 13.5 | | |
| | | c: 5, 5.5, 8, 10, 12; | | 1EC02 | 59.3 | 42.9 | 8.2 | 14.8 | | |
| | | d: 12. | | 1EC03 | 66.0 | 47.4 | 9.5 | 16.9 | | |
| | | | | | | | | | | |
| | | Malar implants: | | MALAR IMPLANTS (RIGHT AND LEFT) | | | | | | |
| | | a: 60; | | Item<br>Code | Dim A<br>(mm) | Dim B<br>(mm) | Dim C<br>(mm) | | | |
| | | b: 28; | | 1ML01 | 35.0 | 18.0 | 3.5 | | | |
| | | c: 5. | | 1ML02 | 45.4 | 23.0 | 5.0 | | | |
| | | | | 1ML03 | 49.4 | 24.0 | 5.7 | | | |
| | | | | 1ML04 | 55.0 | 27.1 | 5.8 | | | |
| | | Piriform implants | | ASIAN PARA-NASAL | | | | | | |
| | | a: 25; | | Item<br>Code | Dim A<br>(mm) | Dim B<br>(mm) | Dim C<br>(mm) | | | |
| | | b: 22; | | 4PN01 | 20.0 | 18.0 | 4.0 | | | |
| | | c: 5, 8. | | 5PN01 | 20.0 | 18.0 | 5.0 | | | |
| | | | | 6PN01 | 21.0 | 19.0 | 6.0 | | | |
| | | | | | | | | | | |
{11}------------------------------------------------
| Pore Size (microns) | >15 | 20-40 | 20-40 |
|---------------------|-----------------|---------------------------------|---------------------------------|
| Provided sterile | Yes | Yes | Yes |
| Re-sterilization | Not recommended | May be sterilized up to 3 times | May be sterilized up to 3 times |
#### 8. Substantial Equivalence:
Substantial Equivalence discussion:
The proposed devices and the predicated device the same classification information, same specifications, same performance effectiveness. The Model of Proposed Device is included in the Model of Predicate for use of proposed device is included in the predicated devices are as safe, as effective and perform as well as the predicate device. Therefore the proposed devices are Substantially Equivalent (SE) to the predicate device which is US legally market device.
{12}------------------------------------------------
## 9. Non-Clinical Tests Performed:
- A F754-08 (Reapproved 2015) Standard Specification For Implantable Polytetrafluoroethylene (PTFE) Sheet, Tube, And Rod Shapes Fabricated From Granular Molding Powders
- > F881-94 (Reapproved 2014) Standard Specification For Silicone Elastomer Facial Implants
- > ISO 14630 Fourth Edition 2012-12-01 Non-Active Surgical Implants -- General Requirements
- > ISO 10993-1 Fifth Edition 2018-08 Biological Evaluation Of Medical Devices -Part 1: Evaluation And Testing Within A Risk Management Process
- > ISO 10993-3 Third Edition 2014-10-1 Biological Evaluation Of Medical Devices - Part 3: Tests For Genotoxicity, Carcinogenicity And Reproductive Toxicity
- > ISO 10993-4 Third Edition 2017-04 Biological Evaluation Of Medical Devices--Part 4: Selection Of Tests For Interactions With Blood
- > ISO10993-5 Third Edition 2009-06-01 Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity
- > ISO 10993-6 Third Edition 2016-12-01 Biological Evaluation Of Medical Devices -- Part 6: Tests For Local Effects After Implantation
- > ISO 10993-7:2008(R)2012 Biological Evaluation Of Medical Devices - Part 7: Ethylene Oxide Sterilization Residuals
- > ISO 10993-10 Third Edition 2010-08-01 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization
- > ISO 10993-11 Third Edition 2017-09 Biological Evaluation Of Medical Devices -Part 11: Tests For Systemic Toxicity
- > ISO 11135:2014 Sterilization Of Health-Care Products - Ethylene Oxide -Requirements For The Development, Validation And Routine Control Of A Sterilization Process For Medical Devices
#### 10. Conclusion:
The proposed devices of Tisuthes™ e-PTFE Surgical Patch are determined to be Substantially Equivalent (SE) to the predicate devices.
--- End of this section ---
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.