SAFECARE Multi-Drug Urine Test Cup, SAFECARE Multi-Drug Urine Test Dip Card
Applicant
Safecare Biotech (Hangzhou) Co., Ltd.
Product Code
NFT · Clinical Toxicology
Decision Date
May 27, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3100
Device Class
Class 2
Intended Use
SAFECARE® Multi-Drug Urine Test Dip Card is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline d-Propoxyphene and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP) in human urine at the cutoff concentrations of: [Table of cutoffs]. Configuration of SAFECARE® Multi-Drug Urine Test Dip Card can consist of any combination of the above listed drug analytes. The test may yield positive results for the prescription drugs Buprenorphine, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. The tests are intended for over-the-counter use. SAFECARE® Multi-Drug Urine Test Cup is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Marijuana, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline d-Propoxyphene and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP) in human urine at the cutoff concentrations of: [Table of cutoffs]. Configuration of SAFECARE® Multi-Drug Urine Test Cup can consist of any combination of the above listed drug analytes. The test may yield positive results for the prescription drugs Buprenorphine, Nortriptyline, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. The tests are intended for over-the-counter use.
Device Story
Single-use in vitro diagnostic devices (Dip Card or Cup formats) for qualitative detection of drugs of abuse in human urine. Operates via competitive binding, lateral flow immunochromatographic assay; urine sample migrates via capillary action. If drug concentration is below cutoff, monoclonal antibody-coated particles bind to immobilized drug-conjugate, forming a visible colored line. If drug concentration exceeds cutoff, binding sites are saturated, preventing line formation. Used by lay users in home or point-of-care settings. Results are visual; positive results are preliminary and require confirmation via GC/MS or LC/MS. Provides rapid screening to assist in identifying potential drug presence.
Clinical Evidence
No clinical studies performed. Evidence consists of analytical performance (precision, interference, specificity, pH/specific gravity effects) and a lay-user study. Lay-user study (n=310 per format) evaluated performance across various drug concentrations relative to cutoffs, confirming high accuracy (typically 85-100% correct results) compared to LC/MS. Flesch-Kincaid analysis confirmed 7th-grade reading level for instructions.
Technological Characteristics
Lateral flow immunochromatographic assay; competitive binding principle. Single-use dip card or cup format. Materials include monoclonal mouse antibodies, drug-conjugates, and desiccant. No electronic components or software. Stable at 4-30°C for 24 months.
Indications for Use
Indicated for qualitative, simultaneous detection of drugs of abuse (Amphetamine, Oxazepam, Marijuana, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, MDMA, Phencyclidine, Methadone, Nortriptyline, Propoxyphene, EDDP) in human urine. Intended for over-the-counter use. Not intended to distinguish between prescription use and abuse.
Regulatory Classification
Identification
An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Related Devices
K182654 — SAFECARE Multi-Drug Urine Test Cup, SAFECARE Multi-Drug Urine Test Dip Card · Safecare Biotech (Hangzhou) Co., Ltd. · Nov 19, 2018
K181790 — CLUNGENE Multi-Drug Test Dip Card, CLUNGENE Multi-Drug Test Easy Cup · Hangzhou Clongene Biotech Co., Ltd. · Jul 26, 2018
K170049 — AssureTech Panel Dip Test, AssureTech Quick Cup Test · Assure Tech. (Hangzhou) Co, Ltd. · May 2, 2017
K182530 — BIOEASY Multi-Drug Test Cup · Shenzhen Bioeasy Biotechnology Co., Ltd. · Nov 9, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
May 27, 2020
Safecare Biotech (Hangzhou) Co., Ltd. % Joe Shia Manager LSI International 504 E Diamond Ave., Suite I Gaithersburg, MD 20877
Re: k201120
Trade/Device Name: SAFECARE® Multi-Drug Urine Test Dip Card SAFECARE® Multi-Drug Urine Test Cup Regulation Number: 21 CFR 862.3100 Regulation Name: Amphetamine Test System Regulatory Class: Class II Product Code: NFT, NFW, NFY, NGG, NGI, NFV, NGL, PTH, NGM, PTG, QAW, QBF Dated: April 23, 2020 Received: April 27, 2020
Dear Joe Shia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Marianela Perez-Torres, Ph.D. Acting Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) k201120
#### Device Name
SAFECARE® Multi-Drug Urine Test Dip Card SAFECARE® Multi-Drug Urine Test Cup
#### Indications for Use (Describe)
SAFECARE® Multi-Drug Urine Test Dip Card is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline d-Propoxyphene and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP) in human urine at the cutoff concentrations of:
| Drug(Identifier) | Cut-off level |
|-----------------------------------------------------|---------------|
| Amphetamine | 1000 ng/mL |
| Oxazepam | 300 ng/mL |
| Cocaine | 300 ng/mL |
| Marijuana | 50 ng/mL |
| Methamphetamine | 1000 ng/mL |
| Morphine | 2000 ng/mL |
| Oxycodone | 100 ng/mL |
| Secobarbital | 300 ng/mL |
| Buprenorphine | 10 ng/mL |
| Methylenedioxy-methamphetamine | 500 ng/mL |
| Phencyclidine | 25 ng/mL |
| Methadone | 300 ng/mL |
| Nortriptyline | 1000 ng/mL |
| d-Propoxyphene | 300 ng/mL |
| 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine | 300 ng/mL |
Configuration of SAFECARE® Multi-Drug Urine Test Dip Card can consist of any combination of the above listed drug analytes.
The test may yield positive results for the prescription drugs Buprenorphine, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. The tests are intended for over-the-counter use.
SAFECARE® Multi-Drug Urine Test Cup is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Marijuana, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline d-Propoxyphene and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP) in human urine at the cutoff concentrations of:
| Drug(Identifier) | Cut-off level |
|----------------------|---------------|
| Amphetamine | 1000 ng/mL |
| Oxazepam | 300 ng/mL |
| FORM FDA 3881 (7/17) | Page 1 of 2 |
{3}------------------------------------------------
| Cocaine | 300 ng/mL | |
|-----------------------------------------------------|------------|--|
| Marijuana | 50 ng/mL | |
| Methamphetamine | 1000 ng/mL | |
| Morphine | 2000 ng/mL | |
| Oxycodone | 100 ng/mL | |
| Secobarbital | 300 ng/mL | |
| Buprenorphine | 10 ng/mL | |
| Methylenedioxy-methamphetamine | 500 ng/mL | |
| Phencyclidine | 25 ng/mL | |
| Methadone | 300 ng/mL | |
| Nortriptyline | 1000 ng/mL | |
| d-Propoxyphene | 300 ng/mL | |
| 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine | 300 ng/mL | |
Configuration of SAFECARE® Multi-Drug Urine Test Cup can consist of any combination of the above listed drug analytes.
The test may yield positive results for the prescription drugs Buprenorphine, Nortriptyline, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.
The tests are intended for over-the-counter use.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) SUMMARY
#### 510(k) Number: k201120
- May 27, 2020 1. Date: Safecare Biotech (Hangzhou) Co. Ltd. 2. Submitter: 18 Haishu Road, Yuhang District Hangzhou, China 3. Contact person: Joe Shia LSI International Inc. 504E Diamond Ave., Suite I
- Email: shiajl@yahoo.com SAFECARE® Multi-Drug Urine Test Dip Card 4. Device Name: SAFECARE® Multi-Drug Urine Test Cup
Gaithersburg, MD 20877 Telephone: 240-505-7880
Classification:
Class 2
| Product Code | Classification | Regulation Section | Panel |
|---------------------------------------|----------------|----------------------------------------------------------------|-----------------|
| NFT<br>Amphetamine | II | 21 CFR § 862.3100, Amphetamine Test System | Toxicology (91) |
| NFW<br>Cannabinoids | II | 21 CFR § 862.3870, Cannabinoids Test System | Toxicology (91) |
| NFY<br>Cocaine | II | 21 CFR § 862.3250, Cocaine and Cocaine Metabolites Test System | Toxicology (91) |
| NGG<br>Methamphetamine | II | 21 CFR § 862.3610, Methamphetamine Test System | Toxicology (91) |
| NGI<br>Morphine | II | 21 CFR § 862.3640, Morphine Test System | Toxicology (91) |
| NFV<br>Oxazepam | II | 21 CFR § 862.3170, Benzodiazepine Test System | Toxicology (91) |
| NGL<br>Oxycodone | II | 21 CFR § 862.3650, Opiate Test System | Toxicology (91) |
| PTH<br>Secobarbital | II | 21 CFR § 862.3150, Barbiturate Test System | Toxicology (91) |
| NGL<br>Buprenorphine | II | 21 CFR § 862.3650, Opiate Test System | Toxicology (91) |
| NGG<br>Methylenedioxy-methamphetamine | II | 21 CFR § 862.3610, Methamphetamine Test System | Toxicology (91) |
| NGM<br>Phencyclidine | unclassified | Enzyme Immunoassay Phencyclidine | Toxicology (91) |
| PTG<br>Methadone | II | 21 CFR § 862.3620, Methadone Test System | Toxicology (91) |
| QAW<br>Nortriptyline | II | 21 CFR, 862.3910 Tricyclic Antidepressant Drugs Test System | Toxicology (91) |
| QBF<br>Propoxyphene | II | 21 CFR, 862.3700 Propoxyphene Test System | Toxicology (91) |
{5}------------------------------------------------
| Product Code | Classification | Regulation Section | Panel |
|----------------------------------------------------------------|----------------|------------------------------------------|--------------------|
| PTG<br>2-ethylidene-1, 5-dimethyl-<br>3, 3-diphenylpyrrolidine | II | 21 CFR § 862.3620, Methadone Test System | Toxicology<br>(91) |
- 5. Predicate Device
The SAFECARE® Multi-Drug Urine Test Dip Card and SAFECARE® Multi-Drug Urine Test Cup (K182654)
- 6. Intended Use
SAFECARE® Multi-Drug Urine Test Dip Card is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline d-Propoxyphene and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP) in human urine at the cutoff concentrations of:
| Drug (Identifier) | Cut-off level |
|-----------------------------------------------------|---------------|
| Amphetamine | 1000 ng/mL |
| Oxazepam | 300 ng/mL |
| Cocaine | 300 ng/mL |
| Marijuana | 50 ng/mL |
| Methamphetamine | 1000 ng/mL |
| Morphine | 2000 ng/mL |
| Oxycodone | 100 ng/mL |
| Secobarbital | 300 ng/mL |
| Buprenorphine | 10 ng/mL |
| Methylenedioxy-methamphetamine | 500 ng/mL |
| Phencyclidine | 25 ng/mL |
| Methadone | 300 ng/mL |
| Nortriptyline | 1000 ng/mL |
| d-Propoxyphene | 300 ng/mL |
| 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine | 300 ng/mL |
Configuration of SAFECARE® Multi-Drug Urine Test Dip Card can consist of any combination of the above listed drug analytes.
The test may yield positive results for the prescription drugs Buprenorphine, Nortriptyline, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. The tests are intended for over-the-counter use.
SAFECARE® Multi-Drug Urine Test Cup is competitive binding, lateral flow immunochromatographic assays for qualitative and simultaneous detection of Amphetamine, Oxazepam, Cocaine, Marijuana, Methamphetamine, Morphine, Oxycodone, Secobarbital, Buprenorphine, Methylenedioxy-methamphetamine, Phencyclidine, Methadone, Nortriptyline d
{6}------------------------------------------------
Propoxyphene and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP) in human urine at the cutoff concentrations of:
| Drug (Identifier) | Cut-off level |
|-----------------------------------------------------|---------------|
| Amphetamine | 1000 ng/mL |
| Oxazepam | 300 ng/mL |
| Cocaine | 300 ng/mL |
| Marijuana | 50 ng/mL |
| Methamphetamine | 1000 ng/mL |
| Morphine | 2000 ng/mL |
| Oxycodone | 100 ng/mL |
| Secobarbital | 300 ng/mL |
| Buprenorphine | 10 ng/mL |
| Methylenedioxy-methamphetamine | 500 ng/mL |
| Phencyclidine | 25 ng/mL |
| Methadone | 300 ng/mL |
| Nortriptyline | 1000 ng/mL |
| d-Propoxyphene | 300 ng/mL |
| 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine | 300 ng/mL |
Configuration of SAFECARE® Multi-Drug Urine Test Cup can consist of any combination of the above listed drug analytes.
The test may yield positive results for the prescription drugs Buprenorphine. Nortriptyline, Oxazepam, Secobarbital, Propoxyphene and Oxycodone when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method. The tests are intended for over-the-counter use.
## 7. Device Description
The SAFECARE® Dip Card Tests and SAFECARE® Cup Tests are immunochromatographic assays that use a lateral flow system for the qualitative detection of Amphetamine, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Morphine, Secobarbital, Methadone, Methylenedioxymethamphetamine, Oxycodone, Buprenorphine, Phencyclidine, Nortriptyline, Propoxyphen and 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (target analytes) in human urine. The products are single-use in vitro diagnostic devices, which come in the formats of Dip Cards or Cups. Each test kit contains a Test Device (in one of the two formats), a package insert and a urine cup for sample collection. Each test device is sealed with a desiccant in an aluminum pouch.
### 8. Substantial Equivalence Information
A summary comparison of features of the SAFECARE® Dip Card Tests and SAFECARE® Cup Tests and the predicate devices is provided in following tables.
{7}------------------------------------------------
### Table 1: Features Comparison of SAFECARE® Dip Card Tests and the Predicate Device
| Item | Device | Predicate - K182654 |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Indication(s)<br>for Use | For the qualitative determination of drugs of<br>abuse in human urine. | Same (but the number of<br>drugs detected is<br>different) |
| | Similarities | |
| Methodology | Competitive binding, lateral flow<br>immunochromatographic assays based on<br>the principle of antigen antibody<br>immunochemistry. | Same |
| Type of Test | Qualitative | Same |
| Specimen Type | Human Urine | Same |
| Intended Use | For over-the-counter | Same |
| Configurations | Dip Card | Same |
| | Differences | |
| Calibrator and Cut-Off<br>Values | Amphetamine (AMP): 1,000 ng/ml<br>Oxazepam (BZO):300 ng/ml<br>Cocaine (COC): 300 ng/ml<br>11-Nor-△9-Tetrahydrocannabinol-9-COOH<br>(THC):50 ng/ml<br>Methamphetamine (MET): 1,000 ng/ml<br>Morphine (OPI): 2000ng/ml<br>Secobarbital (BAR): 300 ng/ml<br>Methadone (MTD): 300 ng/ml<br>Methylenedioxymethamphetamine<br>(MDMA): 500 ng/ml<br>Oxycodone (OXY): 100 ng/ml<br>Buprenorphine (BUP): 10 ng/ml<br>Phencyclidine (PCP): 25 ng/ml<br>Nortriptyline (TCA): 1000 ng/ml<br>Propoxyphene (PPX): 300 ng/ml<br>2-ethylidene-1, 5-dimethyl-3, 3-<br>diphenylpyrrolidine (EDDP): 300 ng/mL | Same as candidate device<br>with exclusion of 2-<br>ethylidene-1, 5-dimethyl-3,<br>3-diphenylpyrrolidine<br>(EDDP): 300 ng/mL |
## Table 2: Features Comparison of SAFECARE® Cup Tests and the Predicate Devices
| Item | Device | Predicate - K182654 |
|--------------------------|------------------------------------------------------------------------|---------------------------------------------------------|
| Indication(s)<br>for Use | For the qualitative determination of<br>drugs of abuse in human urine. | Same (but the number of<br>drugs detected is different) |
| Similarities | | |
{8}------------------------------------------------
| Methodology | Competitive binding, lateral flow<br>immunochromatographic assays based on<br>the principle of antigen antibody<br>immunochemistry. | Same |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Type of Test | Qualitative | Same |
| Specimen Type | Human Urine | Same |
| Intended Use | For over-the-counter | Same |
| Configurations | Cup | Cup |
| | Differences | |
| Calibrator and Cut-Off<br>Values | Amphetamine (AMP): 1,000 ng/ml<br>Oxazepam (BZO):300 ng/ml<br>Cocaine (COC): 300 ng/ml<br>11-Nor-△9-Tetrahydrocannabinol-9-COOH<br>(THC):50 ng/ml<br>Methamphetamine (MET): 1,000 ng/ml<br>Morphine (OPI): 2000ng/mL<br>Secobarbital (BAR): 300 ng/ml<br>Methadone (MTD): 300 ng/ml<br>Methylenedioxymethamphetamine<br>(MDMA): 500 ng/ml<br>Oxycodone (OXY): 100 ng/ml<br>Buprenorphine (BUP): 10 ng/ml<br>Phencyclidine (PCP): 25 ng/ml<br>Nortriptyline (TCA): 1000 ng/ml | Same as candidate device<br>with exclusion of 2-<br>ethylidene-1, 5-dimethyl-3,<br>3-diphenylpyrrolidine<br>(EDDP): 300 ng/mL |
| | Propoxyphene (PPX): 300 ng/ml<br>2-ethylidene-1, 5-dimethyl-3, 3-<br>diphenylpyrrolidine (EDDP): 300<br>ng/mL | |
### 9. Test Principle
The SAFECARE® Dip Card Tests, and SAFECARE® Cup Tests are rapid tests for the qualitative detection of Amphetamine, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Morphine, Secobarbital, Methadone, Methylenedioxymethamphetamine, Oxycodone, Buprenorphine, Phencyclidine, Nortriptyline, Propoxyphen and 2-ethylidene-1, 5-dimethyl-3, 3diphenylpyrrolidine in urine samples. The tests are lateral flow chromatographic immunoassays. During testing, a urine specimen migrates upward by capillary action. If target drugs present in the urine specimen are below the cut-off concentration, it will not saturate the binding sites of its specific monoclonal mouse antibody coated on the particles. The antibody-coated particles will then be captured by immobilized drug-conjugate and a visible colored line will show up in the test line region. The colored line will not form in the test line region if the target drug level exceeds its cutoff-concentration because it will saturate all the binding sites of the antibody
{9}------------------------------------------------
coated on the particles. A band should form in the control region of the devices regardless of the presence of drug or metabolite in the sample to indicate that the tests have been performed properly.
- 10. Performance Characteristics
- 1. Analytical Performance
- a. Precision
Precision studies were carried out for samples with concentrations of -100% cut off, -75% cut off. -50% cut off. -25% cut off. cut off. +25% cut off. +50% cut off. +75% cut off and +100% cut off. These samples were prepared by spiking drug in negative samples. Each drug concentration was confirmed by LC/MS. All sample aliquots were blindly labeled by the person who prepared the samples and didn't take part in the sample testing. For each concentration, tests were performed two runs per day for 25 days per device in a randomized order. The results obtained are summarized in the following tables for 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine. Please refer to k182654 for precision data for Methylenedioxy-Methamphetamine, Oxycodone, Buprenorphine, Phencyclidine, Nortriptyline and Propoxyphene, and to k181968 for precision data for Oxazepam, Methamphetamine, Morphine, Secobarbital and Methadone, and to k153646 for precision data for Amphetamine, Cocaine, and Cannabinoids.
| Dip Card | | | | | | | | | |
|-----------------------|---------------|--------------|--------------|--------------|---------|--------------|--------------|--------------|---------------|
| Results<br>Lot Number | -100% Cut-off | -75% Cut-off | -50% Cut-off | -25% Cut-off | Cut-off | Cut-off +25% | Cut-off +50% | Cut-off +75% | Cut-off +100% |
| Lot 1 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 24-/26+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 2 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 25-/25+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 3 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 26-/24+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine
| Cup | | | | | | | | | |
|---------------|------------------|-----------------|-----------------|-----------------|---------|-----------------|-----------------|-----------------|------------------|
| Results | | | | | | | | | |
| Lot<br>Number | -100%<br>Cut-off | -75%<br>Cut-off | -50%<br>Cut-off | -25%<br>Cut-off | Cut-off | Cut-off<br>+25% | Cut-off<br>+50% | Cut-off<br>+75% | Cut-off<br>+100% |
| Lot 1 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 27-/23+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 2 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 24-/26+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 3 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 26-/24+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
The cut-off value of 300 ng/mL for 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine is verified.
b. Linearity
Not applicable.
- c. Stability
The devices are stable at 4-30 ℃ for 24 months based on the accelerated stability study at 50 °C and real time stability studies at 4°C and 30 °C.
- d. Interference
Potential interfering substances found in human urine of physiological or pathological conditions were added to drug-free urine and target drugs urine with concentrations at 25% below and 25% above Cut-Off levels. These urine samples were tested using three batches of each device. Compounds that showed no interference at a concentration
{10}------------------------------------------------
of 100ug/mL. with the exception of albumin with no interference at a concentration of 100mg/dL and ethanol at 1% volume, are summarized in the following tables. There were no differences observed between the SAFECARE® Cup and Dip Card formats.
| Acetominophen (4-<br>Acetamidophenol) | Ecgonine methyl ester | D,L-Octopamine |
|---------------------------------------|------------------------|--------------------------------------------|
| Acetophenetidin | Erythromycin | Oxalic acid |
| N-Acetylprocainamide | ẞ-Estradiol | Oxolinic acid |
| Acetylsalicylic acid | Ethanol (1% vol) | Oxymetazoline |
| Albumin (100mg/dL) | Fenoprofen | Papaverine |
| Aminopyrine | Furosemide | Penicillin-G |
| Amoxicillin | Gentisic acid | Perphenazine |
| Ampicillin | Hemoglobin | Phenelzine |
| Apomorphine | Hydralazine | Prednisone |
| Ascorbic acid | Hydrochlorothiazide | DL-Propranolol |
| Aspartame | Hydrocortisone | D-Pseudoephedrine |
| Atropine | O-Hydroxyhippuric acid | Quinine |
| Benzilic acid | 3-Hydroxytyramine | Ranitidine |
| Benzoic acid | Ibuprofen | Salicylic acid |
| Bilirubin | D,L-Isoproterenol | Serotonin (5- Hydroxytyramine) |
| Chloralhydrate | Isoxsuprine | Sulfamethazine |
| Chloramphenicol | Ketamine | Sulindac |
| Chlorothiazide | Ketoprofen | Tetrahydrocortisone, 3-acetate |
| Chlorpromazine | Labetalol | Tetrahydrocortisone 3-(β-<br>Dglucuronide) |
| Cholesterol | Loperamide | Tetrahydrozoline |
| Clonidine | Meperidine | Thiamine |
| Cortisone | Meprobamate | Thioridazine |
| (-) Cotinine | Methoxyphenamine | Triamterene |
| Creatinine | Nalidixic acid | DL-Tyrosine |
| Deoxycorticosterone | Naloxone | Trifluoperazine |
| Dextromethorphan | Naltrexone | Trimethoprim |
| Diclofenac | Naproxen | D,L-Tryptophan |
| Diflunisal | Niacinamide | Tyramine |
| Digoxin | Nifedipine | Uric acid |
| Diphenhydramine | Norethindrone | Verapamil |
| Disopyramide | Noscapine | Zomepirac |
e. Specificity
To test specificity, drug metabolites and other structure related compounds that are likely to cross-react in urine samples were tested using three batches of each device. The lowest concentration that caused a positive result for each compound are listed below for 2-ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine. The rest of the data were in k182654 for for Methylenedioxy-Methamphetamine, Oxycodone, reported Buprenorphine, Phencyclidine, Nortriptyline and Propoxyphene, and k181968 for Amphetamine, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Morphine, Secobarbital and Methadone. There were no differences observed between the SAFECARE® Cup and Dip Card formats.
| Compounds | Result<br>Positive at(ng/ml) | % Cross-Reactivity |
|-----------|------------------------------|--------------------|
| Methadone | 300000 | 0.1% |
{11}------------------------------------------------
| EMDP | 300000 | 0.1% |
|--------------------------------------|----------|-------|
| Doxylamine | >100,000 | <0.3% |
| Disopyramide | >100,000 | <0.3% |
| LAAM (Levo-alpha-acetylmethadol) HCl | >100,000 | <0.3% |
| Alpha Methadol | >100,000 | <0.3% |
f. Effect of Urine Specific Gravity and Urine pH
To investigate the effect of urine specific gravity and urine pH, urine samples, with 1.000 to 1.035 specific gravity or urine samples with pH 4 to 9 were spiked with target drugs at 25% below and 25% above Cut-Off levels. These samples were tested using three lots of each device. Results were all positive for samples at and above +25% Cut-Off and all negative for samples at and below -25% Cut-Off. There were no differences observed between the SAFECARE® Cup and Dip Card formats.
## 2. Comparison Studies
Method comparison studies for the SAFECARE® Dip Card Tests and the SAFECARE® Cup Tests were performed in-house with three laboratory assistants for each device. Operators ran 80 (40 negative and 40 positive) unaltered clinical samples for each drug. The samples were blind labeled and compared to LC/MS results are presented in the tables below for 2ethylidene-1, 5-dimethyl-3, 3-diphenylpyrrolidine (EDDP). The rest data were reported in k182654 for Methylenedioxy-Methamphetamine, Oxycodone, Buprenorphine, Phencyclidine, Nortriptyline and Propoxyphene, and in k181968 for Amphetamine, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Morphine, Secobarbital and Methadone.
| Dip<br>Card | | Negative | Low<br>Negative by<br>LC/MS<br>(less than<br>-50%) | Near Cutoff<br>Negative by<br>LC/MS<br>(Between<br>-50% and<br>cutoff) | Near Cutoff<br>Positive by<br>LC/MS<br>(Between the<br>cutoff and<br>+50%) | High Positive<br>by LC/MS<br>(greater than<br>+50%) |
|-------------|----------|----------|----------------------------------------------------|------------------------------------------------------------------------|----------------------------------------------------------------------------|-----------------------------------------------------|
| Viewer<br>A | Positive | 0 | 0 | 1 | 20 | 20 |
| | Negative | 10 | 10 | 19 | 0 | 0 |
| Viewer<br>B | Positive | 0 | 0 | 0 | 19 | 20 |
| | Negative | 10 | 10 | 20 | 1 | 0 |
| Viewer<br>C | Positive | 0 | 0 | 0 | 20 | 20 |
| | Negative | 10 | 10 | 20 | 0 | 0 |
## EDDP
### Discordant Results
| Viewer | Sample Number | LC/MS Result | Dip Card Viewer Results |
|----------|---------------|--------------|-------------------------|
| Viewer A | CM2632 | 291 | Positive |
| Viewer B | CM9474 | 309 | Negative |
{12}------------------------------------------------
| Cup | | Negative | Low<br>Negative by<br>LC/MS<br>(less than<br>-50%) | Near Cutoff<br>Negative by<br>LC/MS<br>(Between<br>-50% and<br>cutoff) | Near Cutoff<br>Positive by<br>LC/MS<br>(Between the<br>cutoff and<br>+50%) | High Positive<br>by LC/MS<br>(greater than<br>+50%) |
|--------|----------|----------|----------------------------------------------------|------------------------------------------------------------------------|----------------------------------------------------------------------------|-----------------------------------------------------|
| Viewer | Positive | 0 | 0 | 1 | 20 | 20 |
| A | Negative | 10 | 10 | 19 | 0 | 0 |
| Viewer | Positive | 0 | 0 | 0 | 19 | 20 |
| B | Negative | 10 | 10 | 20 | 1 | 0 |
| Viewer | Positive | 0 | 0 | 0 | 20 | 20 |
| C | Negative | 10 | 10 | 20 | 0 | 0 |
#### Discordant Results
| Viewer | Sample Number | LC/MS Result | Cup<br>Viewer Results |
|----------|---------------|--------------|-----------------------|
| Viewer A | CM1684 | 285 | Positive |
| Viewer B | CM2850 | 372 | Negative |
### Lay-user study:
A lay user study was performed at three intended user sites with 310 lay persons for each device format. The lay users had diverse educational and professional backgrounds and ranged in age from 18 to > 50 years. Urine samples were prepared at the following concentrations; negative, +/-75%, +/-50%, +/-25% of the cutoff by spiking drugs into drug free-pooled urine specimens. The concentrations of the samples were confirmed by LC/MS. Each sample was aliquoted into individual containers and blind-labeled. Each participant was provided with the package insert, 1 blind labeled sample and a device. Each device was tested. Typical results are shown below.
| % of Cutoff | Number of<br>samples | Drug<br>Concentration by<br>LC/MS (ng/mL) | Lay person Results | No. of<br>Positive | | No. of<br>Negative | The percentage<br>of correct results<br>(%) |
|--------------|----------------------|-------------------------------------------|--------------------|--------------------|-----|--------------------|---------------------------------------------|
| -100% Cutoff | 20 | 0 | 0 | 20 | 100 | | |
| -75% Cutoff | 20 | 253 | 0 | 20 | 100 | | |
| -50% Cutoff | 170 | 496 | 0 | 170 | 100 | | |
| -25% Cutoff | 20 | 753 | 2 | 18 | 90 | | |
| +25% Cutoff | 20 | 1249 | 17 | 3 | 85 | | |
| +50% Cutoff | 40 | 1498 | 40 | 0 | 100 | | |
| +75% Cutoff | 20 | 1758 | 20 | 0 | 100 | | |
{13}------------------------------------------------
# COC:
| % of Cutoff | Number of<br>samples | Drug<br>Concentration by<br>LC/MS(ng/mL) | Lay person Results | | The percentage<br>of correct<br>results (%) |
|--------------|----------------------|------------------------------------------|--------------------|--------------------|---------------------------------------------|
| | | | No. of<br>Positive | No. of<br>Negative | |
| -100% Cutoff | 20 | 0 | 0 | 20 | 100 |
| -75% Cutoff | 20 | 76 | 0 | 20 | 100 |
| -50% Cutoff | 170 | 147 | 0 | 170 | 100 |
| -25% Cutoff | 20 | 226 | 1 | 19 | 95 |
| +25% Cutoff | 20 | 372 | 18 | 2 | 90 |
| +50% Cutoff | 40 | 449 | 40 | 0 | 100 |
| +75% Cutoff | 20 | 524 | 20 | 0 | 100 |
# THC:
| % of Cutoff | Number of<br>samples | Drug<br>Concentration by<br>LC/MS(ng/mL) | Lay person Results | | The percentage |
|--------------|----------------------|------------------------------------------|--------------------|--------------------|---------------------------|
| | | | No. of<br>Positive | No. of<br>Negative | of correct results<br>(%) |
| -100% Cutoff | 20 | 0 | 0 | 20 | 100 |
| -75% Cutoff | 20 | 12 | 0 | 20 | 100 |
| -50% Cutoff | 170 | 24 | 0 | 170 | 100 |
| -25% Cutoff | 20 | 36 | 1 | 19 | 95 |
| +25% Cutoff | 20 | 63 | 18 | 2 | 90 |
| +50% Cutoff | 40 | 75 | 40 | 0 | 100 |
| +75% Cutoff | 20 | 88 | 20 | 0 | 100 |
## BAR:
| % of Cutoff | Number of<br>samples | Drug<br>Concentration by<br>LC/MS(ng/mL) | Lay person Results | | The percentage<br>of correct results<br>(%) |
|--------------|----------------------|------------------------------------------|--------------------|--------------------|---------------------------------------------|
| | | | No. of<br>Positive | No. of<br>Negative | |
| -100% Cutoff | 20 | 0 | 0 | 20 | 100 |
| -75% Cutoff | 20 | 76 | 0 | 20 | 100 |
| -50% Cutoff | 170 | 147 | 0 | 170 | 100 |
| -25% Cutoff | 20 | 225 | 1 | 19 | 95 |
| +25% Cutoff | 20 | 374 | 18 | 2 | 90 |
| +50% Cutoff | 40 | 449 | 40 | 0 | 100 |
| +75% Cutoff | 20 | 524 | 20 | 0 | 100 |
{14}------------------------------------------------
| % of Cutoff | Number of<br>samples | Drug<br>Concentration by<br>LC/MS(ng/mL) | Lay person Results | | The percentage<br>of correct results<br>(%) |
|--------------|----------------------|------------------------------------------|--------------------|--------------------|---------------------------------------------|
| | | | No. of<br>Positive | No. of<br>Negative | |
| -100% Cutoff | 20 | 0 | 0 | 20 | 100 |
| -75% Cutoff | 20 | 76 | 0 | 20 | 100 |
| -50% Cutoff | 170 | 147 | 0 | 170 | 100 |
| -25% Cutoff | 20 | 227 | 1 | 19 | 95 |
| +25% Cutoff | 20 | 374 | 19 | 1 | 95 |
| +50% Cutoff | 40 | 450 | 40 | 0 | 100 |
| +75% Cutoff | 20 | 526 | 20 | 0 | 100 |
# BZO:
## MET:
| % of Cutoff | Number of<br>samples | Drug<br>Concentration by<br>LC/MS(ng/mL) | Lay person Results | | The percentage |
|--------------|----------------------|------------------------------------------|--------------------|--------------------|---------------------------|
| | | | No. of<br>Positive | No. of<br>Negative | of correct results<br>(%) |
| -100% Cutoff | 20 | 0 | 0 | 20 | 100 |
| -75% Cutoff | 20 | 248 | 0 | 20 | 100 |
| -50% Cutoff | 170 | 497 | 0 | 170 | 100 |
| -25% Cutoff | 20 | 749 | 1 | 19 | 95 |
| +25% Cutoff | 20 | 1251 | 18 | 2 | 90 |
| +50% Cutoff | 40 | 1499 | 40 | 0 | 100 |
| +75% Cutoff | 20 | 1754 | 20 | 0 | 100 |
## MTD:
| % of Cutoff | Number of<br>samples | Drug<br>Concentration by<br>LC/MS(ng/mL) | Lay person Results | | The percentage |
|--------------|----------------------|------------------------------------------|--------------------|--------------------|---------------------------|
| | | | No. of<br>Positive | No. of<br>Negative | of correct results<br>(%) |
| -100% Cutoff | 20 | 0 | 0 | 20 | 100 |
| -75% Cutoff | 20 | 76 | 0 | 20 | 100 |
| -50% Cutoff | 170 | 147 | 0 | 170 | 100 |
| -25% Cutoff | 20 | 223 | 1 | 19 | 95 |
| +25% Cutoff | 20 | 374 | 17 | 3 | 85 |
| +50% Cutoff | 40 | 450 | 40 | 0 | 100 |
| +75% Cutoff | 20 | 526 | 20 | 0 | 100 |
{15}------------------------------------------------
| % of Cutoff | Number of samples | Drug Concentration by LC/MS(ng/mL) | Lay person Results | | The percentage of correct results (%) |
|--------------|-------------------|------------------------------------|--------------------|-----------------|---------------------------------------|
| | | | No. of Positive | No. of Negative | |
| -100% Cutoff | 20 | 0 | 0 | 20 | 100 |
| -75% Cutoff | 20 | 501 | 0 | 20 | 100 |
| -50% Cutoff | 170 | 999 | 0 | 170 | 100 |
| -25% Cutoff | 20 | 1501 | 1 | 19 | 95 |
| +25% Cutoff | 20 | 2499 | 19 | 1 | 95 |
| +50% Cutoff | 40 | 2999 | 40 | 0 | 100 |
| +75% Cutoff | 20 | 3501 | 20 | 0 | 100 |
## MDMA:
| % of Cutoff | Number of<br>samples | Drug<br>Concentration by<br>LC/MS(ng/mL) | Lay person Results | | The percentage<br>of correct results<br>(%) |
|--------------|----------------------|------------------------------------------|--------------------|--------------------|---------------------------------------------|
| | | | No. of<br>Positive | No. of<br>Negative | |
| -100% Cutoff | 20 | 0 | 0 | 20 | 100 |
| -75% Cutoff | 20 | 126 | 0 | 20 | 100 |
| -50% Cutoff | 170 | 248 | 0 | 170 | 100 |
| -25% Cutoff | 20 | 375 | 1 | 19 | 95 |
| +25% Cutoff | 20 | 624 | 18 | 2 | 90 |
| +50% Cutoff | 40 | 748 | 40 | 0 | 100 |
| +75% Cutoff | 20 | 877 | 20 | 0 | 100 |
# OXY:
| % of Cutoff | Number of<br>samples | Drug<br>Concentration by<br>LC/MS(ng/mL) | Lay person Results | | The percentage<br>of correct results<br>(%) |
|--------------|----------------------|------------------------------------------|--------------------|--------------------|---------------------------------------------|
| | | | No. of<br>Positive | No. of<br>Negative | |
| -100% Cutoff | 20 | 0 | 0 | 20 | 100 |
| -75% Cutoff | 20 | 25 | 0 | 20 | 100 |
| -50% Cutoff | 170 | 48 | 0 | 170 | 100 |
| -25% Cutoff | 20 | 74 | 1 | 19 | 95 |
| +25% Cutoff | 20 | 126 | 18 | 2 | 90 |
| +50% Cutoff | 40 | 150 | 40 | 0 | 100 |
| +75% Cutoff | 20 | 176 | 20 | 0 | 100 |
# OPI:
{16}------------------------------------------------
# BUP:
| % of Cutoff | Number of<br>samples | Drug<br>Concentration by<br>LC/MS(ng/mL) | Lay person Results | | The percentage<br>of correct results<br>(%) |
|--------------|----------------------|------------------------------------------|--------------------|--------------------|---------------------------------------------|
| | | | No. of<br>Positive | No. of<br>Negative | |
| -100% Cutoff | 20 | 0 | 0 | 20 | 100 |
| -75% Cutoff | 20 | 3 | 0 | 20 | 100 |
| -50% Cutoff | 170 | 5 | 0 | 170 | 100 |
| -25% Cutoff | 20 | 8 | 2 | 18 | 90 |
| +25% Cutoff | 20 | 13 | 18 | 2 | 90 |
| +50% Cutoff | 40 | 15 | 40 | 0 | 100 |
| +75% Cutoff | 20 | 18 | 20 | 0 | 100 |
## PCP:
| % of Cutoff | Number of<br>samples | Drug<br>Concentration by<br>LC/MS(ng/mL) | Lay person Results | | The percentage<br>of correct results<br>(%) |
|--------------|----------------------|------------------------------------------|--------------------|--------------------|---------------------------------------------|
| | | | No. of<br>Positive | No. of<br>Negative | |
| -100% Cutoff | 20 | 0 | 0 | 20 | 100 |
| -75% Cutoff | 20 | 6 | 0 | 20 | 100 |
| -50% Cutoff | 170 | 12 | 0 | 170 | 100 |
| -25% Cutoff | 20 | 19 | 2 | 18 | 90 |
| +25% Cutoff | 20 | 32 | 18 | 2 | 90 |
| +50% Cutoff | 40 | 37 | 40 | 0 | 100 |
| +75% Cutoff | 20 | 43 | 20 | 0 | 100 |
# TCA:
| % of Cutoff | Number of<br>samples | Drug<br>Concentration by<br>LC/MS(ng/mL) | Lay person Results | | The percentage<br>of correct results<br>(%) |
|--------------|----------------------|------------------------------------------|--------------------|--------------------|---------------------------------------------|
| | | | No. of<br>Positive | No. of<br>Negative | |
| -100% Cutoff | 20 | 0 | 0 | 20 | 100 |
| -75% Cutoff | 20 | 250 | 0 | 20 | 100 |
| -50% Cutoff | 170 | 497 | 0 | 170 | 100 |
| -25% Cutoff | 20 | 750 | 1 | 19 | 95 |
| +25% Cutoff | 20 | 1249 | 18 | 2 | 90 |
| +50% Cutoff | 40 | 1500 | 40 | 0 | 100 |
| +75% Cutoff | 20 | 1749 | 20 | 0 | 100 |
{17}------------------------------------------------
| % of Cutoff | Number of<br>samples | Drug<br>Concentration by<br>LC/MS(ng/mL) | Lay person Results | | The percentage<br>of correct results<br>(%) |
|--------------|----------------------|------------------------------------------|--------------------|--------------------|---------------------------------------------|
| | | | No. of<br>Positive | No. of<br>Negative | |
| -100% Cutoff | 20 | 0 | 0 | 20 | 100 |
| -75% Cutoff | 20 | 76 | 0 | 20 | 100 |
| -50% Cutoff | 170 | 148 | 0 | 170 | 100 |
| -25% Cutoff | 20 | 223 | 3 | 17 | 85 |
| +25% Cutoff | 20 | 374 | 18 | 2 | 90 |
| +50% Cutoff | 40 | 448 | 40 | 0 | 100 |
| +75% Cutoff | 20 | 525 | 20 | 0 | 100 |
#### PPX:
#### EDDP:
| % of Cutoff | Number of<br>samples | Drug<br>Concentration by<br>LC/MS(ng/mL) | Lay person Results | | The percentage |
|--------------|----------------------|------------------------------------------|--------------------|--------------------|---------------------------|
| | | | No. of<br>Positive | No. of<br>Negative | of correct results<br>(%) |
| -100% Cutoff | 20 | 0 | 0 | 20 | 100 |
| -75% Cutoff | 20 | 76 | 0 | 20 | 100 |
| -50% Cutoff | 170 | 152 | 0 | 170 | 100 |
| -25% Cutoff | 20 | 226 | 1 | 19 | 95 |
| +25% Cutoff | 20 | 375 | 18 | 2 | 90 |
| +50% Cutoff | 40 | 452 | 40 | 0 | 100 |
| +75% Cutoff | 20 | 526 | 20 | 0 | 100 |
Lay-users were also given surveys on the ease of understanding the package insert instructions. All lay users indicated that the device instructions can be easily followed. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.
- 3. Clinical Studies
Not applicable.
#### 11. Conclusion
Based on the test principle and acceptable performance characteristics including precision, cut-off, interference, specificity, method comparison, and lay-user studies of the devices, it's concluded that the SAFECARE® Dip Card Tests and SAFECARE® Cup Tests are substantially equivalent to the predicate.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.