← Product Code [GEX](/productcode/GEX) · K201111

# GentleMax Pro Plus (K201111)

_Candela Corporation · GEX · May 26, 2020 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K201111

## Device Facts

- **Applicant:** Candela Corporation
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** May 26, 2020
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

755mm: Temporary hair reduction. Stable long-term or permanent reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing after a treatment regime. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. On all skin types (Fitzpatrick I- VI) including tanned skin. Treatment of benign pigmented lesions. Treatment of wrinkles. The photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias). 1064nm: Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The lasers are indicated on all skin types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratosis, nevi, chloasma, verrucae, skin tags, keratosis, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques. The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. Treatment of wrinkles. 1064nm: Temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes, Trichophyton rubrum and T. mentagrophytes, and/or yeast Candida Albicans, etc.)

## Device Story

Dual-wavelength laser system (755nm Alexandrite, 1064nm Nd:YAG) for dermatological procedures; delivers high-intensity light to target melanin or vascular structures; includes Dynamic Cooling Device (DCD) using cryogen bursts to protect epidermis; operated by clinicians in professional settings; system uses optical fiber delivery; output parameters (energy, spot size, pulse duration) controlled via user interface; therapeutic heat induces photocoagulation/hemostasis; benefits include hair reduction, lesion clearance, and scar improvement.

## Clinical Evidence

No clinical studies were conducted. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1, IEC 60601-2-22), electromagnetic compatibility (IEC 60601-1-2), and laser safety (IEC 60825-1). Biocompatibility established via ISO 10993-5 and ISO 10993-10 testing.

## Technological Characteristics

Class IV laser; flashlamp-excited solid-state Alexandrite (755nm) and Nd:YAG (1064nm); lens-coupled optical fiber delivery; cryogen-based DCD skin cooling; green diode aiming beam (520-550nm); repetition rate 1-10 Hz; pulse duration 0.25-100ms; max fluence 400-520 J/cm2; firmware-controlled user interface.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Candela GentleMAX Family of Lasers System ([K140122](/device/K140122.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

May 26, 2020

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is the Department of Health & Human Services logo. The right side of the logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Candela Corporation Yverre Bobay Director, Regulatory Affairs 251 Locke Dr Marlborough, Massachusetts 01752

Re: K201111

Trade/Device Name: GentleMax Pro Plus Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: April 24, 2020 Received: April 27, 2020

Dear Yverre Bobay:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin Kejing Chen Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

### K201111

Device Name GentleMAX Family of Laser Systems

# Indications for Use (Describe)

755mm:

Temporary hair reduction. Stable long-term or permanent reduction targeting of melanin in hair follicles. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing after a treatment regime. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. On all skin types (Fitzpatrick I- VI) including tanned skin.

Treatment of benign pigmented lesions.

Treatment of wrinkles.

The photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas).

1064nm:

Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment hair reduction is defined as the long-term. stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The lasers are indicated on all skin types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus and spider veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratosis, nevi, chloasma, verucae, skin tags, keratosis, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques.

The laser is indicated for pigmented lesion size. for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.

Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. Treatment of wrinkles.

Temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes, Trichophyton rubrum and T. mentagrophytes, and/or yeast Candida Albicans, etc.)

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary GentleMax Pro Plus System

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92

### 1. DATE PREPARED

April 24, 2020

### 2. APPLICANT NAME

Yverre Bobay Global Director, Regulatory Candela Corporation 251 Locke Drive Marlborough MA 01752 M: (617) 669-6181 yverreb@candelamedical.com

### 3. Official Correspondent

Greg Wallender VP, Global Quality & Regulatory Affairs Candela Corporation 530 Boston Post Road Wayland, MA 01778 M: (925) 206-5439 gregw@candelamedical.com

#### 4. DEVICE INFORMATION

| Proprietary Name:      | GentleMax Family of Laser Systems                                                                                                     |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name:     | Dermatology Laser System                                                                                                              |
| Classification Name:   | Laser surgical instrument for use in General and Plastic<br>surgery and in dermatology (21 CFR Section 878.4810,<br>Product Code GEX) |
| Product Code:          | GEX                                                                                                                                   |
| Device Classification: | Class II                                                                                                                              |
| Laser Classification:  | Class IV                                                                                                                              |

### 5. PREDICATE DEVICE

Candela GentleMAX Family of Lasers System (K140122).

{5}------------------------------------------------

#### 6. INTENDED USE/INDICATION FOR USE

There are no new Intended Uses for the GentleMax Pro Plus. The Indications for Use Statement is unchanged for the GentleMax Pro Plus is as follows:

## 755mm:

Temporary hair reduction. Stable long-term or permanent reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing after a treatment regime. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. On all skin types (Fitzpatrick I- VI) including tanned skin.

Treatment of benign pigmented lesions.

Treatment of wrinkles.

The photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias).

## 1064nm:

Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The lasers are indicated on all skin types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratosis, nevi, chloasma, verrucae, skin tags, keratosis, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques.

The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.

Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. Treatment of wrinkles.

Temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes, Trichophyton rubrum and T. mentagrophytes, and/or yeast Candida Albicans, etc.)

{6}------------------------------------------------

#### 7. DEVICE DESCRIPTION

The GentleMax Pro Plus contains two separate laser heads (Alexandrite and Nd:YAG), which produce laser light outputs of 755 nm and 1064 nm, respectively. The output of each laser head is optically combined on the laser rail, so that their beam paths are identical as they exit the laser system. This allows the use of a single delivery system which can output either 755 nm or 1064 nm wavelengths. The laser system creates a beam of high intensity light that penetrates deep into the skin tissue where it delivers a controlled amount of therapeutic heat. The Dynamic Cooling Device (DCD) protects the upper layers of the skin with a cooling burst of cryogen.

### 8. TECHNOLOGICAL CHARACTERISTICS

The new GentleMax Pro Plus system has the same design, technological characteristics, operating principles, and intended use as the Candela GentleMax Family of Laser System (K140122). The devices share the same technical features, such as calibration port, wavelengths, laser medium, delivery systems, power supply, cooling system, electronics, firmware and user display screen. The devices share the same operating principles, such as energy but the Gentle Max Pro Plus offers a faster repetition rate, and offers larger spot sizes. Any minor differences do not raise any new types of safety or effectiveness questions because the GentleMax Pro Plus parameters are similar to the predicates.

| Specifications                              | Modified Candela<br>GentleMax Pro Plus                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Current Candela<br>GentleMax Pro                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|---------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                                | Candela Corp.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Candela Corp.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| K Number                                    | Not Assigned                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | K140122                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Product Code                                | GEX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | GEX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Regulation<br>Number                        | 878.4810                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 878.4810                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Device Class                                | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Indications<br>for Use                      | The GentleMax Family of Laser Systems is<br>indicated for the following at the specified<br>wavelength:<br>755 nm<br>Temporary hair reduction. Stable long-term<br>or permanent reduction through selective<br>targeting of melanin in hair follicles.<br>Permanent hair reduction is defined as long-                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | The GentleMax Family of Laser Systems is<br>indicated for the following at the specified<br>wavelength:<br>755 nm<br>Temporary hair reduction. Stable long-term<br>or permanent reduction through selective<br>targeting of melanin in hair follicles.<br>Permanent hair reduction is defined as long-                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                             | term stable reduction in the number of hairs regrowing after a treatment regime.<br>Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. On all skin types (Fitzpatrick I- VI) including tanned skin.<br>Treatment of benign pigmented lesions.<br>Treatment of wrinkles.<br>The photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | term stable reduction in the number of hairs regrowing after a treatment regime.<br>Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. On all skin types (Fitzpatrick I- VI) including tanned skin.<br>Treatment of benign pigmented lesions.<br>Treatment of wrinkles.<br>The photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                             | 1064nm<br>Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The lasers are indicated on all skin types Fitzpatrick 1-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratosis, nevi, chloasma, verrucae, skin tags, keratosis, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques.<br>The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. | 1064nm<br>Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The lasers are indicated on all skin types Fitzpatrick 1-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratosis, nevi, chloasma, verrucae, skin tags, keratosis, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques.<br>The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. |
|                                             | Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.<br>Treatment of wrinkles.<br>1064nm<br>Temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes, Trichophyton rubrum and T. mentagrophytes, and/or yeast Candida Albicans, etc.)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.<br>Treatment of wrinkles.<br>1064nm<br>Temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes, Trichophyton rubrum and T. mentagrophytes, and/or yeast Candida Albicans, etc.)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Laser Type                                  | Flashlamp-excited, Solid state Alexandrite and Nd:YAG laser                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Flashlamp-excited, Solid state Alexandrite and Nd:YAG laser                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Aiming Beam                                 | Green diode laser<br>(520-550 nm)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Green diode laser<br>(520-550 nm)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Wavelengths<br>(nm)                         | 755 & 1064                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 755 & 1064                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Maximum                                     | 68 Joules (J) ALEX; 90 J                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 53 joules (J) ALEX; 80 J                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Energy (J)                                  | Nd:YAG                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Nd: YAG                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Accuracy of<br>Output                       | ± 20%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | ± 20%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Maximum<br>Fluence<br>(J/cm2)               | 400 J/cm2 (ALEX)<br>520 J/cm2 (YAG)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 400 J/cm2 (ALEX)<br>520 J/cm2 (YAG)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Spot Size<br>(mm)                           | 1.5 mm, 3 mm, 3x10 mm, 5 mm,<br>6 mm, 8 mm, 10 mm, 12 mm, 15<br>mm, 18 mm, 20 mm, 22 mm, 24<br>mm, 26 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 1.5 mm, 3 mm, 3x10 mm, 5 mm,<br>6 mm, 8 mm, 10 mm, 12 mm, 15<br>mm, 18 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Repetition<br>Rate (Hz)                     | 1 – 10 Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 1 - 10 Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Pulse<br>Duration (ms)                      | 0.25 - 100 ms (new 2ms)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 0.25 - 100 ms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Operating<br>Modes                          | Pulse                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Pulse                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Energy<br>Delivery                          | Footswitch & finger switch                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Footswitch & Handpiece switch                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Beam<br>Delivery                            | Lens-coupled optical fiber                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Lens-coupled optical fiber                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Voltage and<br>Power (30A<br>configuration) | 200 V to 240 V~, 50/60 Hz,<br>single phase, 4,600 VA or 20 A at<br>230 V~                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 200 V to 240 V~, 50/60 Hz,<br>single phase, 4,600 VA or 20 A at<br>230 V~                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Voltage and<br>Power (20A<br>configuration) | 200 V to 240 V~, 50/60 Hz,<br>single phase, 3,600 VA or 16 A at<br>230 V~                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 200 V to 240 V~, 50/60 Hz,<br>single phase, 3,600 VA or 16 A at<br>230 V~                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| System<br>Cooling                           | Ambient Air                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Ambient Air                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Skin Cooling                                | Yes<br>Cryogen (DCD)…

---

**Source:** [https://fda.innolitics.com/device/K201111](https://fda.innolitics.com/device/K201111)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
