← Product Code [MBH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBH) · K201084

# Physica system (K201084)

_Lima Corporate S.P.A. · MBH · Dec 14, 2020 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBH/K201084

## Device Facts

- **Applicant:** Lima Corporate S.P.A.
- **Product Code:** [MBH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBH.md)
- **Decision Date:** Dec 14, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3565
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Physica system is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions: - Non-inflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, and avascular necrosis (not applicable to Physica TT Tibial Plate); - Inflammatory degenerative joint disease including rheumatoid arthritis; - Correction of functional deformity; - Revision procedures where other treatments or devices have failed; and - Treatment of fractures that are unmanageable using other techniques. Additional indications for Physica LMC component are: - Moderate varus, valgus, or flexion deformities. In patients with preserved and well functioning collateral ligaments, Physica PS components are also indicated for: - Absent or not-functioning posterior cruciate ligament; - Severe antero-posterior instability of the knee joint. Femoral, tibial and patellar components of the Physica system are intended for cemented use, with the exception of the TT Tibial Plate that is intended for uncemented use. Tibial liners can be used with cemented tibial components.

## Device Story

Physica system is a total knee replacement prosthesis comprising femoral component, tibial plate, tibial stem, patella, and tibial liners. Components are implanted by orthopedic surgeons in clinical settings to replace damaged knee joints. System restores joint function, corrects deformities, and addresses instability or fractures. Femoral, tibial, and patellar components are generally cemented; TT Tibial Plate is uncemented. Device provides mechanical structural support to replace diseased bone/cartilage, improving patient mobility and reducing pain.

## Clinical Evidence

No clinical data. Substantial equivalence demonstrated via bench testing, including wear testing, contact area/pressure analysis, locking strength testing, and fatigue testing of tibial plate and distal features.

## Technological Characteristics

System components include femoral, tibial, and patellar elements. Materials include conventional UHMWPE and cross-linked UHMWPE with Vitamin E. Tibial plate available in cemented and uncemented versions. Mechanical fixation principle. No software or electronic components.

## Regulatory Identification

A knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement. This identification includes fixed-bearing knee prostheses where the ultra high molecular weight polyethylene tibial bearing is rigidly secured to the metal tibial base plate.

## Special Controls

*Classification.* Class II (special controls). The special control is FDA's guidance: “Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA.” See § 888.1 for the availability of this guidance.

## Predicate Devices

- PHYSICA SYSTEM ([K141934](/device/K141934.md), [K193284](/device/K193284.md))
- VANGUARD COMPLETE KNEE SYSTEM ([K113550](/device/K113550.md))
- NEXGEN TRABECULAR METAL TIBIAL TRAY ([K072160](/device/K072160.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

December 14, 2020

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

LimaCorporate S.p.A. % Stephen Peoples President Peoples & Associates Consulting LLC 5010 Lodge Pole Lane Fort Wayne, Indiana 46814

Re: K201084

Trade/Device Name: Physica system Regulation Number: 21 CFR 888.3565 Regulation Name: Knee Joint Patellofemorotibial Metal/Polymer Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: MBH, JWH, HRY Dated: November 9, 2020 Received: November 10, 2020

Dear Stephen Peoples:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ting Song, Ph.D., R.A.C. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K201084

Device Name Physica system

Indications for Use (Describe)

Physica system is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions:

- · Non-inflammatory degenerative joint disease including
	- osteoarthritis
	- traumatic arthritis, and
	- avascular necrosis (not applicable to Physica TT Tibial Plate);
- · Inflammatory degenerative joint disease including rheumatoid arthritis;
- · Correction of functional deformity;
- · Revision procedures where other treatments or devices have failed; and
- · Treatment of fractures that are unmanageable using other techniques.

Additional indications for Physica LMC component are:

- · Moderate varus, valgus, or flexion deformities.
In patients with preserved and well functioning collateral ligaments, Physica PS components are also indicated for:

- · Absent or not-functioning posterior cruciate ligament;
- · Severe antero-posterior instability of the knee joint.

Femoral, tibial and patellar components of the Physica system are intended for cemented use, with the exception of the TT Tibial Plate that is intended for uncemented use. Tibial liners can be used with cemented tibial components.

| Type of Use (Select one or both, as applicable)                                                                                                                     |                                                                       |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| <div> <span> <b>                         X                     </b>                     Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> | <div> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Summary of Safety and Effectiveness

Date: December 14, 2020

Manufacturer: Limacorporate S.p.A. Via Nazionale, 52 33038 – Villanova di San Daniele Udine - Italy

U.S. Contact Person: Dr. Stephen J. Peoples Principal Consultant SPeoplesVMD@gmail.com PEOPLES & ASSOCIATES CONSULTING, LLC 411 Auditorium Blvd. Winona Lake, IN 46590 Phone: 260-645-0327 FAX: +39 0432945512

#### Trade name: Physica system

#### Common name: Knee Prosthesis

| Classification Name: |                                                                                                                   |
|----------------------|-------------------------------------------------------------------------------------------------------------------|
| Product Code         | Regulation and Classification Name                                                                                |
| MBH                  | Knee joint patellofemorotibial metal/polymer porous-coated<br>uncemented prosthesis per 21 CFR 888.3565           |
| JWH                  | Knee joint Patellofemorotibial Polymer/Metal/Polymer Semi-<br>Constrained Cemented Prosthesis per 21 CFR 888.3560 |
| HRY                  | Knee joint Femorotibial Metal/Polymer Semi-Constrained Cemented<br>Prosthesis per 21 CFR 888.3530                 |

# Closcification Name:

#### Description:

The Physica system is a total knee replacement system consisting of a Femoral component, Tibial plate, Tibial stem and patella; the Tibial stem and Patella components are optional to be used as required for each individual patient.

Tibial liners and Patellar components made of conventional UHMWPE and cross-linked UHMWPE with Vitamin E are provided as part of the Physica system. The Tibial Plate is provided in cemented and uncemented versions.

# Indications for use:

Physica system is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions:

- · Non-inflammatory degenerative joint disease including
	- osteoarthritis
	- traumatic arthritis, and
	- avascular necrosis (not applicable to Physica TT Tibial Plate);
- · Inflammatory degenerative joint disease including rheumatoid arthritis;
- · Correction of functional deformity;
- Revision procedures where other treatments or devices have failed; and
- · Treatment of fractures that are unmanageable using other techniques.

{4}------------------------------------------------

Additional indications for Physica LMC component are:

- · Moderate varus, valgus, or flexion deformities.
In patients with preserved and well functioning collateral ligaments, Physica PS components are also indicated for:

- · Absent or not-functioning posterior cruciate ligament;
- · Severe antero-posterior instability of the knee joint.

Femoral, tibial and patellar components of the Physica system are intended for cemented use, with the exception of the TT Tibial Plate that is intended for uncemented use. Tibial liners can be used with cemented or uncemented tibial components.

#### Predicate Devices:

| Applicant                  | Device name                         | Cleared via         |
|----------------------------|-------------------------------------|---------------------|
| LIMACORPORATE S.P.A        | PHYSICA SYSTEM                      | K141934,<br>K193284 |
| BIOMET MANUFACTURING CORP. | VANGUARD COMPLETE KNEE SYSTEM       | K113550             |
| ZIMMER, INC.               | NEXGEN TRABECULAR METAL TIBIAL TRAY | K072160             |

#### Summary of technology comparison:

The intended use, design, and materials of the TT Tibial Plate and the tibial liners and patellar components manufactured from cross-linked UHMWPE with Vit. E are substantially equivalent to those of the predicate devices. Design Control Activities have been successfully completed.

# Non-clinical testing:

Mechanical testing had demonstrated the device's ability to perform substantially equivalent to the predicate devices in:

- Wear test;
- Contact areas and pressures at tibio-femoral and patello-femoral interfaces; .
- Test on the locking strength between the tibial plate and the tibial liner; ●
- Fatigue testing of the tibial plate; ●
- Fatigue testing of distal features of tibial plate. ●

Since the tibial liners and patellar components manufactured from cross-linked UHMWPE with Vit. E have an equivalent design and geometry to the predicate tibial liners and patellar components, the ROM of the subject devices when used with the previously cleared femoral components is the same as the ROM provided by the predicate Physica system tibial liners and patellar components.

# Clinical testing:

Clinical testing was not necessary to demonstrate substantial equivalence of the subject devices to the predicate devices.

{5}------------------------------------------------

# Conclusion:

Based upon a comparison of intended use, materials, summary of technological characteristics, and preclinical testing, the Physica system is substantially equivalent to the predicate devices identified in this premarket notification.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBH/K201084](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/MBH/K201084)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
