The AltaTrack equipment is an imaging device with Fiber Optic RealShape (FORS) technology intended to aid the positioning and navigation of a connected AltaTrack Catheter and/or AltaTrack Guidewire during endovascular procedures of the peripheral, aortic and aortic side branch vasculature, by creating a 3D image in real time of an AltaTrack Catheter and/or AltaTrack Guidewire. The AltaTrack Guidewire is an angiographic guidewire with Fiber Optic RealShape (FORS) technology, intended to direct a catheter during navigation in endovascular procedures of the peripheral, aortic and aortic side branch vasculature. The AltaTrack Catheter is an angiographic catheter with Fiber Optic RealShape (FORS) technology, intended to deliver radiopaque media or lead a guidewire in endovascular procedures of the peripheral, aortic and aortic side branch vasculature.
Device Story
AltaTrack system uses Fiber Optic RealShape (FORS) technology to visualize shape of invasive devices (guidewire/catheter) over entire length; inputs include optical signals from fiber-optic sensors embedded in devices; system transforms inputs into 3D real-time images; output overlaid on 2D fluoroscopy or pre-operative 3D CT images; used in interventional suites or hybrid ORs by vascular surgeons/interventional radiologists; provides enhanced anatomical context and device visualization; aids navigation during endovascular procedures; benefits patient by potentially reducing radiation exposure or improving procedural efficiency.
Clinical Evidence
No clinical data. Evidence consists of bench testing, animal/phantom testing in simulated clinical environments, and human factors summative usability evaluation. Animal/phantom testing confirmed intended use and safety measures. Usability study with interventional radiologists/vascular surgeons and OR staff demonstrated safe/effective performance for critical and non-critical tasks.
Technological Characteristics
System uses Fiber Optic RealShape (FORS) technology. Components include lasers, optical components, computer hardware, and sterile single-use guidewires/catheters. Guidewire: 0.035" diameter, 120 cm length. Catheter: 5.5 Fr, 80 cm length. Connectivity: integrated with interventional X-ray systems. Sterilization: Ethylene Oxide (EO). Standards: IEC 62304, IEC 62366-1, IEC 60601-1, IEC 60825-1, ISO 10993 series, ISO 10555-1.
Indications for Use
Indicated for patients eligible for endovascular procedures of the peripheral, aortic, and aortic side branch vasculature. Used to aid positioning and navigation of catheters and guidewires.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
Predicate Devices
MediGuide™ Technology System Version 17.0 (K162643)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 17, 2020
Philips Medical System Nederland B.V. Gert De Vries Senior Regulatory Affairs Manager Veenpluis 4-6 Best, Noord-Brabant 5684 PC Netherlands
Re: K201081
Trade/Device Name: AltaTrack equipment, AltaTrack Guidewire, AltaTrack Catheter Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable diagnostic computer Regulatory Class: Class II Product Code: DQK, DQX, DQX, DQO Dated: November 10, 2020 Received: November 16, 2020
Dear Gert De Vries:
We have reviewed your Section 510(k) premarket notification of intent to market the devices referenced above and have determined the devices are substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Carmen Gacchina Johnson, Ph.D. Assistant Director DHT2B: Division of Circulatory Support, Structural & Vascular Devices Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201081
Device Name AltaTrack equipment; AltaTrack Guidewire: AltaTrack Catheter
#### Indications for Use (Describe)
The AltaTrack equipment is an imaging device with Fiber Optic RealShape (FORS) technology intended to aid the positioning and navigation of a connected AltaTrack Catheter and/or AltaTrack Guidewire during endovascular procedures of the peripheral, aortic side branch vasculature, by creating a 3D image in real time of an AltaTrack Catheter and/or AltaTrack Guidewire.
The AltaTrack Guidewire is an angiographic guidewire with Fiber Optic RealShape (FORS) technology, intended to direct a catheter during navigation in endovascular procedures of the and aortic side branch vasculature.
The AltaTrack Catheter is an angiographic catheter with Fiber Optic RealShape (FORS) technology, intended to deliver radiopaque media or lead a guidewire in endovascular procedures of the peripheral, aortic side branch vasculature.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
This 510(k) summary has been prepared in accordance with 21 CFR 807.92.
510(k) Number: K201081
### SUBMITTER INFORMATION
Submitted by: Philips Medical Systems Nederland B.V. Veenpluis 4-6, 5684 PC Best, The Netherlands
Contact Person: Gert de Vries, Senior Regulatory Affairs Manager
Telephone Number: +31 (0)6 182 06 281 Email: gert.de.vries@philips.com Date Prepared: April 16 2020
### DEVICE NAME
| Device Trade Name: | AltaTrack equipment |
|----------------------|----------------------------------------------------------------------------------|
| Common Name: | Computer, diagnostic, programmable |
| Classification Name: | Programmable diagnostic computer, 21 CFR 870.1425: Class II<br>Product code DQK |
| Device Trade Name: | AltaTrack Guidewire |
| Common Name: | Wire, guide, catheter |
| Classification Name: | Catheter guide wire, 21 CFR 870.1330: Class II Product code DQX |
| Device Trade Name: | AltaTrack Catheter |
| Common Name: | Catheter, intravascular, diagnostic |
| Classification Name: | Diagnostic intravascular catheter, 21 CFR 870.1200: Class II<br>Product code DQO |
| PREDICATE DEVICES | |
### MediGuide™ Technology System Version 17.0 (K162643) AltaTrack equipment Radifocus Glidewire® (K152740) AltaTrack Guidewire
| AltaTrack Catheter | Radifocus® Optitorque™ Angiographic Catheter (K150232) |
|--------------------|--------------------------------------------------------|
|--------------------|--------------------------------------------------------|
# DEVICE DESCRIPTION
| AltaTrack equipment | The AltaTrack equipment is a visualization device with Fiber Optic RealShape (FORS) technology. Its function is to create a 3D image in real time of an AltaTrack Guidewire and/or AltaTrack Catheter and overlay it on real-time or pre-recorded 2D fluoroscopy images and/or on pre-operative 3D CT images, if available.<br>The AltaTrack equipment comprises software and hardware components (such as lasers, optical components, computer hardware, electrical and optical cabling), and a single-use, sterile, detachable component. |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{4}------------------------------------------------
- AltaTrack Guidewire The AltaTrack Guidewire is a single-use, sterile, hydrophilic guidewire with FORS technology. The primary function of the AltaTrack Guidewire is to direct an AltaTrack Catheter or other compatible catheter to a desired anatomical location. It can be visualized in 3D in real time by the AltaTrack equipment using FORS technology. AltaTrack Catheter The AltaTrack Catheter is a single-use, sterile, angiographic catheter with FORS technology. Its primary function is to deliver radiopaque media or lead an AltaTrack Guidewire or other compatible quidewire to a desired anatomical location. It can be visualized in 3D in real time by the AltaTrack
# INDICATIONS FOR USE
AltaTrack equipment The AltaTrack equipment is an imaging device with Fiber Optic RealShape (FORS) technology intended to aid the positioning and navigation of a connected AltaTrack Catheter and/or AltaTrack Guidewire during endovascular procedures of the peripheral, aortic and aortic side branch vasculature, by creating a 3D image in real time of an AltaTrack Catheter and/or AltaTrack Guidewire.
equipment using FORS technology.
- The AltaTrack Guidewire is an angiographic guidewire with Fiber Optic AltaTrack Guidewire RealShape (FORS) technology, intended to direct a catheter during navigation in endovascular procedures of the peripheral, aortic and aortic side branch vasculature.
The AltaTrack Catheter is an angiographic catheter with Fiber Optic AltaTrack Catheter RealShape (FORS) technology, intended to deliver radiopaque media or lead a quidewire in endovascular procedures of the peripheral, aortic and aortic side branch vasculature.
# TECHNOLOGICAL CHARACTERISTICS
The proposed AltaTrack equipment and the predicate MediGuide™ Technology AltaTrack equipment System Version 17.0 (K162643) have fundamentally the same intended use. The AltaTrack equipment and the predicate MediGuide™ Technology System both enable real time positioning and navigation for (minimally) invasive device(s) in peripheral endovascular interventions. Both support their own technology equipped quidewire(s) and catheter(s). The tracking technology deployed with AltaTrack equipment is Fiber Optic RealShape (FORS) technology to visualize the shape of the invasive devices over their entire length. The predicate device deploys Electromagnetic (E/M) sensing technology, which is typically used to visualize the distal tip of the invasive devices. The technological difference does not raise any new questions regarding safety and effectiveness
> Table below provides a device comparison between the subject AltaTrack equipment and the predicate MediGuide™ Technology System Version 17.0 (K162643).
{5}------------------------------------------------
| Attribute | Proposed device | Predicate device | | | |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|-----------------------------------------------|--|
| Name | AltaTrack equipment | MediGuide™ Technology<br>System Version 17.0<br>(K162643) | | | |
| Manufacturer | Philips Medical Systems | St. Jude Medical | | | |
| Product code | DQK | DQK | | | |
| Classification | II | II | | | |
| Indications for use | The AltaTrack equipment is<br>an imaging device with<br>Fiber Optic RealShape<br>(FORS) technology<br>intended to aid the<br>positioning and navigation<br>of a connected AltaTrack<br>Catheter and/or AltaTrack<br>Guidewire during<br>endovascular procedures of<br>the peripheral, aortic and<br>aortic side branch<br>vasculature, by creating a<br>3D image in real time of an<br>AltaTrack Catheter and/or<br>AltaTrack Guidewire. | The MediGuide™<br>Technology system is<br>intended for the evaluation<br>of vascular and cardiac<br>anatomy. It is intended to<br>enable real time tip<br>positioning and navigation<br>of a MediGuide Enabled™<br>/ Sensor Enabled ™<br>(equipped with a magnetic<br>sensor) invasive device<br>used in vascular or cardiac<br>interventions in the Cath<br>Lab environment, on both<br>live fluoroscopy or<br>recorded background. The<br>System is indicated for use<br>as an adjunct to<br>fluoroscopy. | | | |
| Patient<br>population | Patients eligible for<br>endovascular procedures. | Patients eligible for<br>vascular or cardiac<br>interventional procedures. | | | |
| Application<br>area | Peripheral, aortic and aortic<br>side branch | Vascular or cardiac<br>interventions | | | |
| Clinical<br>environment | In the control room and in<br>the exam room of an<br>interventional suite or<br>hybrid operating room. | In the control room and in<br>the exam room of an<br>interventional suite or<br>hybrid operating room. | | | |
| Adjunct<br>devices | Used in conjunction with<br>fluoroscopy | Used in conjunction with<br>fluoroscopy | | | |
| Summary of<br>device<br>description | AltaTrack equipment<br>consists of hardware and<br>software elements, which<br>are installed in conjunction<br>with an interventional X-ray<br>system. The interventional<br>system continues to<br>perform safely and<br>effectively per its intended<br>use, while enabling<br>enhanced anatomical<br>context visualization and<br>invasive device<br>visualization. | MediGuide™ Technology<br>System consists of<br>hardware and software<br>elements, which are<br>installed in conjunction<br>with the existing<br>fluoroscopy Imaging<br>System in a Cath Lab. The<br>conventional fluoroscopy<br>Imaging System, equipped<br>with MediGuide TM<br>Technology elements,<br>continues to perform<br>safely and effectively per<br>its intended use as<br>fluoroscopic imaging<br>device, while enabling<br>device tracking and<br>enhanced visualization<br>tools supplied by<br>MediGuide™ Technology<br>capabilities. | | | |
| Technology<br>for invasive<br>device | visualization | Fiber-Optic RealShape<br>(FORS) technology. | | Electro-magnetic (E/M)<br>sensing technology. | |
| Tracked<br>invasive | AltaTrack Catheter<br>AltaTrack Guidewire | Variety of MediGuide<br>Enabled™ /Sensor | | | |
{6}------------------------------------------------
| device | Enabled™ catheters, guidewires |
|--------|--------------------------------|
|--------|--------------------------------|
AltaTrack Guidewire The AltaTrack Guidewire and the predicate Radiofocus Glidewire (K152740) have the same intended use. The AltaTrack Guidewire and the predicate device are both designed for use in peripheral endovascular procedures, with similarities including mechanical handling properties, radiopacity, sterilization method and packaging design. Contrary to the predicate device, the AltaTrack Guidewire contains a fiberoptic sensor, which enables it to be visualized using FORS technology. The technological difference does not raise any new questions regarding safety and effectiveness.
> The table below provides a device comparison between the subject AltaTrack Guidewire and the predicate Radifocus Glidewire (K152740).
| Attribute | Proposed device<br>AltaTrack Guidewire | Predicate device<br>Radifocus Glidewire<br>(K152740) |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | AltaTrack Guidewire | Radifocus Glidewire<br>(K152740) |
| Manufacturer | Philips Medical Systems | Terumo Corporation |
| Product code | DQX | DQX |
| Classification | II | II |
| Indications for<br>Use | The AltaTrack Guidewire is<br>an angiographic guidewire<br>with Fiber Optic<br>RealShape (FORS)<br>technology, intended to<br>direct a catheter during<br>navigation in endovascular<br>procedures of the<br>peripheral, aortic and<br>aortic side branch<br>vasculature. | The Glidewire is<br>designed to direct a<br>catheter to the desired<br>anatomical location in<br>the peripheral<br>vasculature during<br>diagnostic or<br>interventional<br>procedures.<br>This device is not<br>intended for<br>neurovascular or<br>coronary interventions. |
| Site in the<br>body | Peripheral, aortic and<br>aortic side branch<br>vasculature | Peripheral vasculature |
| Operation<br>principle | Manual | Manual |
| Specifications | Wire diameter: 0.035"<br>Device length (in-body<br>section): 120 cm | Wire diameter: 0.035"<br>Device length: 260, 300,<br>350, 400, and 450 cm.<br><br>Note that K152740<br>extends the working<br>length of its predicate<br>(K863138), which has<br>lengths from 30-300<br>cm.<br>Lengths of the flexible<br>part of the distal tip: 3<br>and 5 cm.<br>Distal tip shape: straight<br>and angled<br>Shaft configurations:<br>standard and stiff |
| | Length of the flexible part<br>of the distal tip: 3 cm<br>Distal tip shape: angled<br>Shaft configuration:<br>standard | |
| Packaging<br>Configuration | Individual package, unit<br>box, shipping carton. | Individual package, unit<br>box, shipping carton. |
| Sterilization<br>Method | Ethylene Oxide (EO) | EO |
| Attribute | Proposed device | Predicate device |
| Name | AltaTrack Catheter | Radifocus® Optitorque™ |
| | | Angiographic Catheter |
| Manufacturer | Philips Medical Systems | Terumo Corporation |
| Product code | DQO | DQO |
| Classification | II | II |
| Indications for | The AltaTrack Catheter is | The Radifocus Optitorque |
| use | an angiographic catheter | Angiographic Catheter is |
| | with Fiber Optic RealShape | indicated for use in cardiac |
| | (FORS) technology, | and vascular procedures. It |
| | intended to deliver | is designed to deliver |
| | radiopaque media or lead a | radiopaque media, guide |
| | guidewire in endovascular | wires, catheters, and |
| | procedures of the | therapeutic agents to |
| | peripheral, aortic and aortic | selected sites in the vascular |
| | side branch vasculature. | system. The different shapes |
| | | are designed to selectively |
| | | engage arteries from access |
| | | sites such as the femoral, |
| | | radial, and brachial artery. |
| Application | Peripheral, aortic and aortic | Selected sites in the vascular |
| area | side branch | system. |
| Operation | Manual | Manual |
| principle | | |
| Specifications | Outer / Inner Diameter: | Outer / Inner Diameter: |
| | 5.5 Fr / 0.99mm | • 4 Fr / 1.05mm |
| | | • 5 Fr / 1.22mm |
| | | • 6 Fr / 1.32mm |
| | Length: 80 cm | Lengths: 65-120 cm |
| | Guidewire compatibility: | Guidewire compatibility: |
| | 0.035" | ≤0.038 " |
| | Maximum Labeled Injection | Maximum Labeled Injection |
| | Pressure: | Pressure: |
| | Not applicable for manual | • 4 Fr: 750 psi |
| | injection | • 5 Fr and 6 Fr: 1000 psi |
| Packaging | Plastic tray | Paperboard mount |
| | Individual package | Individual package |
| | Unit box | Unit box |
| | Shipping carton | Shipping carton |
| | | |
| | Pouch | Pouch |
| Sterilization | Ethylene Oxide (EO) | EO |
AltaTrack Catheter
The AltaTrack Catheter and the predicate Radifocus® Optitorque™ Angiographic Catheter (K150232) have fundamentally the same intended use. Both the AltaTrack
{7}------------------------------------------------
Catheter and the predicate device are designed for use in peripheral endovascular procedures with similarities including mechanical handling properties, radiopacity, sterilization method and packaging design.
Contrary to the predicate device, the AltaTrack Catheter contains a fiber-optic sensor, which enables it to be visualized using FORS technology. The technological difference does not raise any new questions regarding safety and effectiveness.
The table below provides a device comparison between the subject AltaTrack Catheter and the predicate Radifocus® Optitorque™ Angiographic Catheter (K150232).
### PERFORMANCE TESTING
The performance testing for AltaTrack equipment, AltaTrack Guidewire and AltaTrack Catheter consists of norm-compliance testing, design verification and validation testing. For norm-compliance and design verification, the testing was performed per subject device, while the validation, such as non-clinical validation and usability validation, were performed at system level representing a logical clinical workflow following the intended use.
{8}------------------------------------------------
Performance testing data of the proposed devices demonstrate that the subject devices are substantially equivalent to their predicate devices, and that the design input requirements.
# Norm-compliance and verification
| AltaTrack<br>equipment | Norm-compliance performance testing has been performed on the AltaTrack equipment according to the following FDA recognized consensus standards, and were all passed: |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | [Rec. Number: 13-79], IEC 62304 Edition 1.1 2015 Medical device software - Software life cycle processes (Consolidated version) [Rec. Number: 5-114], IEC 62366-1:2015 Medical devices – Application of usability engineering to medical devices [Rec. Number: 5-117], ISO 15223-Thid Edition 2016, Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied IEC 60601-1:2005 + A1:2012, Medical electrical equipment edition 3.1 - Part 1: General requirements for basic safety and essential performance1 [Rec. Number: 19-8], IEC 60601-1-2, Edition 4.0, 2014 Medical electrical equipment -- Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -- Requirements and tests [Rec. Number: 12-273], IEC 60825-1:2007 Safety of laser products - Part 1: Equipment classification and requirements [Rec. Number: 2-220], ISO 10993-1, Fourth Edition, 2009/A1:2010, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process [Including: Technical Corrigendum 1 (2010)] [Rec. Number: 2-245], ISO 10993-5 Third edition 2009, Biological evaluation of medical devices. Part 5: tests for in vitro cytotoxicity. [Rec. Number: 14-408], ANSI AAMI ISO 10993-7:2008(R) 2012, Biological evaluation of medical devices. Part 7: Ethylene oxide sterilization residuals [Rec. Number: 2-174], ANSI AAMI ISO 10993-10:2010/(R)2014, Biological evaluation of medical devices. Part 10: Tests for irritation and skin sensitization [Rec. Number: 14-529], ANSI AAMI ISO 11135:2014, Sterilization of health care products - Ethylene oxide -Requirements for development, validation and routing control of a sterilization process for medical devices [Rec. Number: 14-454], ANSI AAMI ISO 11607-1:2006/(R)2010, Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barriers systems and packaging systems [Including: Amendment 1 (2014)] [Rec. Number: 14-455], ANSI AAMI ISO 11607-2:2006/(R)2010, Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes [Including: Amendment 1 (2014)] [Rec. Number:14-314], AAMI ANSI ST67:2011/(R)2017, Sterilization of healthcare products - Requirements and guidance for selecting a sterility assurance level (SAL) for products labeled "sterile" |
| | Non-clinical performance testing has been performed to verify the implementation of requirements according to the system requirement specifications, as well as the implementation of identified risk control measures from risk management. The requirements categories include: "Functional Requirements", "Integration &<br>Interoperability Requirements", "Configurations, Components and Options", "User Interface Requirements", "Service Requirements", "Manufacturing Requirements" and "Security Requirements". |
Test results demonstrated all executed verification tests have been passed.
<sup>1</sup> Note!: the test report reflects the compliance with [Rec. Number: 19-4], AAMI ANSI 60601-1:2005/R2012 and A1:2012, c1:2009/(r) and A2:2010/(r)2012 Medical electrical equipment edition 3.1 – Part 1: General requirements for basic safety and essential performance.
{9}------------------------------------------------
AltaTrack Guidewire and AltaTrack Catheter
Both AltaTrack Guidewire and AltaTrack Catheter are sterile, single-use devices to be used in peripheral vascular applications. There are commonalities in the list of FDA recognized consensus standards. The following standards for norm-compliance testing are applicable for both AltaTrack Guidewire and AltaTrack Catheter, and all relevant tests for both AltaTrack Guidewire and AltaTrack Catheter have been passed:
- [Rec. Number: 5-114], IEC 62366-1:2015, Medical devices Application ● of usability engineering to medical devices
- . [Rec. Number: 5-117], ISO 15223-Thid Edition 2016, Medical devices -Symbols to be used with medical device labels, labelling and information to be supplied
- [Rec. Number: 2-220], ISO 10993-1, Fourth Edition, 2009/A1:2010, ● Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process [Including: Technical Corrigendum 1 (2010)]
- [Rec. Number: 2-248]. ISO 10993-4 Third edition 2017. Biological ● evaluation of medical devices. Part 4: Selection of tests for interactions with blood
- . [Rec. Number: 2-245], ISO 10993-5 Third edition 2009, Biological evaluation of medical devices. Part 5: tests for in vitro cytotoxicity.
- [Rec. Number: 2-250]. ASTM F756-17. Standard Practice for Assessment . of Hemolytic Properties of Materials
- . [Rec. Number: 14-408], ANSI AAMI ISO 10993-7:2008(R) 2012, Biological evaluation of medical devices. Part 7: Ethylene oxide sterilization residuals
- [Rec. Number:2-174], ANSI AAMI ISO 10993-10:2010/(R)2014, Biological ● evaluation of medical devices. Part 10: Tests for irritation and skin sensitization
- [Rec. Number: 2-255], ISO 10993-11 Third edition 2017, Biological evaluation of medical devices. Part 11: tests for systemic toxicity.
- [Rec. Number: 14-529], ANSI AAMI ISO 11135:2014, Sterilization of . health care products - Ethylene oxide -Requirements for development, validation and routing control of a sterilization process for medical devices
- [Rec. Number: 14-454], ANSI AAMI ISO 11607-1:2006/(R)2010, . Packaging for terminally sterilized medical devices - Part 1: Requirements for material…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.