K201044 · Chirana T. Injecta · FMF · Jan 18, 2022 · General Hospital
Device Facts
Record ID
K201044
Device Name
CHIRANA Insulin Syringes
Applicant
Chirana T. Injecta
Product Code
FMF · General Hospital
Decision Date
Jan 18, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CHIRANA® Insulin Syringes are intended for subcutaneous injection of U100 insulin or U40 insulin.
Device Story
CHIRANA Insulin Syringes are sterile, single-use, manual piston syringes; available in 0.3mL, 0.5mL, and 1.0mL volumes; feature graduated barrel, plunger rod, and needle/hub assembly. Needle gauges range 27G-31G; lengths 6mm-12.7mm. Used for subcutaneous insulin delivery; operated by patients or healthcare providers. Device functions via manual aspiration and injection; output is precise insulin dose delivery. Benefits include reliable subcutaneous administration of insulin for diabetes management.
Clinical Evidence
Bench testing only. Conformance demonstrated with ISO 8537, ISO 7886-1, ISO 9626, and ISO 7864. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, sensitization, irritation, systemic toxicity, material-mediated pyrogenicity, hemocompatibility). Particulate matter testing met USP <788> criteria. Sterilization validated per ISO 11135. Package integrity and shelf-life validated per ASTM D4169-16 and ASTM F1980-16.
Technological Characteristics
Piston syringe design; materials: Stainless Steel 304 (needle), Polypropylene (barrel/plunger), Polystyrene (hub tip), Polyethylene (caps). Lubricant: Medical Grade Silicone. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Complies with ISO 8537, ISO 7886-1, ISO 9626, ISO 7864, and ISO 10993-1.
Indications for Use
Indicated for subcutaneous injection of U100 or U40 insulin in patients requiring insulin therapy.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
Sterile Insulin Syringe for Single use, with needle (K190002)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
January 18, 2022
CHIRANA T. Injecta % Nathan Wright Engineer & Regulatory Specialist Empirical Testing Corp. 4628 Northpark Drive Colorado Springs, Colorado 80918
Re: K201044
Trade/Device Name: CHIRANA Insulin Syringes Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF, FMF, FMI Dated: December 14, 2021 Received: December 15, 2021
## Dear Nathan Wright:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K201044
#### Device Name CHIRANA® Insulin Syringes
Indications for Use (Describe)
The CHIRANA® Insulin Syringes are intended for subcutaneous injection of U100 insulin or U40 insulin.
Type of Use (Select one or both, as applicable) □ Prescription Use (Part 21 CFR 801 Subpart D)
D Over-The Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection including suggestions for reducing this burden to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K201044 510(K) SUMMARY
| Preparation Date: | January 18, 2022 |
|--------------------|------------------------------------------------------------------------------------------------------------|
| Submitter Name: | CHIRANA T. Injecta, a.s.<br>Nám. Dr. Schweitzera 194<br>916 01 Stará Turá<br>Slovak |
| Contact Person: | Nathan Wright<br>Engineer & Regulatory Specialist, Empirical Testing Corp. |
| Telephone Number: | 719-351-0248 |
| Fax Number: | N/A |
| E-mail Address: | nwright@empiricaltech.com |
| Trade Name: | CHIRANA® Insulin Syringes |
| Regulation Name: | Piston Syringe |
| Regulation Number: | 21 CFR 880.5860 |
| Product Code: | FMF |
| Device Class: | Class II |
| Regulation Name: | Hypodermic Single Lumen Needle |
| Regulation Number: | 21 CFR 880.5570 |
| Product Code: | FMI |
| Device Class: | Class II |
| Predicate Device: | K190002, Sterile Insulin Syringe for Single use, with needle (Shanghai<br>Kohope Medical Devices Co., Ltd) |
# Device Description
The CHIRANA® Insulin Syringes are a 0.3mL, 0.5mL, or 1.0mL syringes designed for subcutaneous injection of a desired dose of insulin. The CHIRANA® Insulin Syringes consist of a graduated barrel, plunger rod and needle/hub assembly. It is available in various needle gauge sizes (27G, 29G, 30G and 31G) and various needle lengths (6mm, 8mm, and 12.7mm). The CHIRANA® Insulin Syringes are sterile, single use, and non-toxic. These devices operate on the principles of a piston syringe. The following are the list of syringe and needle combinations offered.
| Insulin Type | Syringe Volume | Needle Gauge | Needle Diameter |
|--------------|----------------|--------------|-----------------|
| Insulin U40 | 1 mL | 29G | 12.7 mm |
| Insulin U100 | 1 mL | 27G | 12.7 mm |
| | | 29G | 8 mm |
| | | 29G | 12.7 mm |
| | | 30G | 12.7 mm |
| | | 30G | 8 mm |
| | | 31G | 8 mm |
| | | 31G | 6 mm |
| Insulin 100 | 0.5 mL | 29G | 12.7 mm |
| | | 30G | 12.7 mm |
| | | 30G | 8 mm |
| | | 31G | 8 mm |
| | | 31G | 6 mm |
{5}------------------------------------------------
| Insulin Type | Syringe Volume | Needle Gauge | Needle Diameter |
|--------------|----------------|--------------|-----------------|
| Insulin 100 | 0.3 mL | 29G | 12.7 mm |
| | | 30G | 12.7 mm |
| | | 31G | 8 mm |
| | | 31G | 6 mm |
# Indications for Use
The CHIRANA® Insulin Syringes are intended for subcutaneous injection of U100 insulin or U40 insulin.
| Characteristic | Predicate Device<br>Shanghai Kohope Medical Devices Co., Ltd<br>Syringes for Single use, with needle<br>K190002 | Subject Device<br>CHIRANA® Insulin Syringes<br>K201044 |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| Indications for Use | The sterile Insulin Syringe for Single use<br>with needle, with the calibration unit of<br>insulin for U-100, is a device intended for<br>medical purposes for the manual aspiration<br>of insulin, and for the injection of insulin<br>into parts of insulin into parts of the body<br>below the surface skin. | The CHIRANA® Insulin Syringes are<br>intended for subcutaneous injection of<br>U100 insulin or U40 insulin. |
| Indications for Use | The sterile Insulin Syringe for Single use<br>with needle, with the calibration unit of<br>insulin for U-40, is a device intended for<br>medical purposes for the manual aspiration<br>of insulin, and for the injection of insulin<br>into parts of insulin into parts of the body<br>below the surface skin. | |
| Prescription Only or<br>Over the Counter | Over the Counter | Over the Counter |
## Discussions of differences in Indications for Use statement
There are only editorial differences to the indications for use statement between the subject device which do not change the indications.
# Technological Characteristics
The table below includes a comparison of the technological characteristics between the new device and those of the predicate device:
| | | CHIRANA® Insulin Syringes | Sterile Insulin<br>Syringe for Single use, with needle |
|--------|---------------|-------------------------------------------------------|--------------------------------------------------------|
| Design | Manufacturer | CHIRANA T. Injecta | Shanghai Kohope Medical Devices Co., Ltd |
| | 510(k) Number | Subject | K190002 |
| | Product Code | FMF, FMI | FMF, FMI |
| | Design | graduated barrel, plunger rod and needle/hub assembly | graduated barrel, plunger rod and needle/hub assembly |
| | Volume | 0.3mL, 0.5mL, 1mL | 0.3mL, 0.5mL, 1mL |
| | Gauge1 | 27G, 29G, 30G, 31G | 29G, 30G, 31G |
| | Length2 | 6mm, 8mm, 12.7mm | 8mm, 13mm |
{6}------------------------------------------------
| | CHIRANA® Insulin Syringes | Sterile Insulin<br>Syringe for Single use, with needle |
|-----------|-------------------------------------------------|--------------------------------------------------------|
| Materials | Needle<br>Stainless Steel 304 | Stainless Steel 304 |
| | Barrel<br>Polypropylene | Polypropylene |
| | Plunger<br>Polypropylene | Polypropylene |
| | Hub Tip3<br>Polystyrene | Isoprene rubber |
| | Needle Cap<br>Polyethylene | Polypropylene |
| | Protective End Cap<br>Polyethylene | Polypropylene |
| | Cannula Lubricant3<br>Medical Grade Silicone | Unknown |
| | Biocompatibility<br>Conforming to ISO 10993-1 | Conforming to ISO 10993-1 |
| | Sterilization4<br>EO (ethylene gas) to SAL=10-6 | SAL=10-6 |
'The subject offers additional needle gauges than the predicate to accommodate a wider range of patient needs. This difference does not affect the safety and effectiveness of the subject needle gauges are standard syringe needle gauges. ?The subject offers additional needle lengths than the predicate to accommodate a wider range of patient needs. This difference does not affect the safety and effectiveness of the subject needle lengths are standard syring needle lengths. 3The subject hub tip and cannula lubricant materials are similar but not identical to that of the predicate components. Biocompatibility testing and performance testing confirmed that all materials of subject are safe and effective for use.
4The subject and predicate are provided sterile to the SAL of 10 . The sterilization method of the predicate is not publically known. but this does not introduce any questions for safety and effectiveness because the subject sterilization method has been validated per ISO 11135.
## Performance Testing
The sterile CHIRANA® Insulin Syringes described in this summary were tested and demonstrated to be in conformance with the following FDA recognized standards:
- ISO 8537 Third edition: Sterile single-use syringes, with or without needle, for insulin
- ISO 7886-1 Second edition: Sterile hypodermic syringes for single use Part 1: Syringes for manual use ●
- ISO 9626 Second edition: Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods
- ISO 7864 Fourth edition: Sterile hypodermic needles for single use - Requirements and test methods
### Biocompatibility
In accordance with ISO 10993-1, the needle is classified as: Externally Communicating Device, Blood Path Indirect, Prolonged Contact (24 hours to 30 days). The following testing was conducted:
- Cytotoxicity ●
- Sensitization ●
- Irritation and Intracutaneous Reactivity ●
- Subacute/Sub-chronic Toxicity
- Acute Systemic Toxicity ●
- Material-Mediated Pyrogenicity ●
- Hemocompatibility
Particulate matter testing was conducted in accordance with USP <788> Particulate Matter in Injections and met the USP acceptance criteria.
### Sterility, Shipping and Shelf-Life
{7}------------------------------------------------
- Sterilization validation per ISO 11135 Second edition: Sterilization of health-care products Ethylene oxide -● Requirements for the development, validation and routine control of a sterilization process for medical devices
- Package integrity testing, after environmental conditioning and simulated transportation in accordance with ● ASTM D4169-16: Standard Practice for Performance Testing of Shipping Containers and Systems was conducted on the final, packaged, and sterile devices
- . Shelf life of 5 years for single unit packaging is validated using the FDA recognized standard ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
- Shelf life of 2 years and four months for multi-unit packaging is validated with closure integrity test
## Conclusions
The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The CHIRANA® Insulin Syringes are substantially equivalent to the Syringe for Single use, with needle (K190002) with respect to the indications for use, target populations, treatment method, and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.