Genius AI Detection is a computer-aided detection and diagnosis (CADe/CADx) software device intended to be used with compatible digital breast tomosynthesis (DBT) systems to identify and mark regions of interest including soft tissue densities (masses, architectural distortions and asymmetries) and calcifications in DBT exams from compatible DBT systems and provide confidence scores that offer assessment for Certainty of Findings and a Case Score. The device intends to aid in the interpretation of digital breast tomosynthesis exams in a concurrent fashion, where the interpreting physician confirms or dismisses the findings during the reading of the exam.
Device Story
Genius AI Detection is a CADe/CADx software device for digital breast tomosynthesis (DBT) exams. It analyzes standard mammographic views using deep learning networks to identify potential abnormalities; specifically soft tissue densities (masses, architectural distortions, asymmetries) and calcifications. The device outputs CAD results including lesion location, lesion outline, and confidence scores, plus a case-level score. These findings are packaged into a DICOM Mammography CAD SR object for display on compatible review workstations. Used concurrently by MQSA-qualified radiologists during exam interpretation; the physician confirms or dismisses findings. The device aims to improve diagnostic performance and assisted-read detection, potentially benefiting patients through more accurate and efficient breast cancer screening and diagnosis.
Clinical Evidence
Clinical evidence includes an MRMC reader study (n=17 readers, 390 cases: 106 cancer, 284 negative) and a standalone study (n=764 cases). MRMC results: AUC 0.825 with CAD vs 0.794 without (diff +0.031); sensitivity 75.9% with CAD vs 66.8% without (diff +9.0%); recall rate 25.8% with CAD vs 23.4% without (diff +2.4%); read-time 52.0s with CAD vs 46.3s without. Standalone study confirmed comparable performance between standard and high-resolution acquisition modes using fROC analysis.
Technological Characteristics
Software-based CADe/CADx device. Utilizes deep learning networks for image analysis. Compatible with Hologic Selenia Dimensions and 3Dimensions systems (standard 1mm, high-res 1mm, and 6mm SmartSlices). Output via DICOM Mammography CAD SR. Complies with IEC 62304 (software lifecycle) and ISO 14971 (risk management).
Indications for Use
Indicated for symptomatic and asymptomatic women undergoing mammography to aid in the detection, localization, and characterization of soft tissue densities (masses, architectural distortions, asymmetries) and calcifications in digital breast tomosynthesis (DBT) exams.
Regulatory Classification
Identification
A radiological computer-assisted detection and diagnostic software is an image processing device intended to aid in the detection, localization, and characterization of fracture, lesions, or other disease-specific findings on acquired medical images (e.g., radiography, magnetic resonance, computed tomography). The device detects, identifies, and characterizes findings based on features or information extracted from images, and provides information about the presence, location, and characteristics of the findings to the user. The analysis is intended to inform the primary diagnostic and patient management decisions that are made by the clinical user. The device is not intended as a replacement for a complete clinician's review or their clinical judgment that takes into account other relevant information from the image or patient history.
Special Controls
A radiological computer assisted detection and diagnosis software must comply with the following special controls: Design verification and validation must include: 1. i. A detailed description of the image analysis algorithm, including but not limited to a description of the algorithm inputs and outputs, each major component or block, how the algorithm and output affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide improved assisted-read detection and diagnostic performance as intended in the indicated user population(s), and to characterize the standalone device performance for labeling. Performance testing includes standalone test(s), side-by-side comparison(s), and/or a reader study, as applicable. iii. Results from standalone performance testing used to characterize the independent performance of the device separate from aided user performance. The performance assessment must be based on appropriate diagnostic accuracy measures (e.g., receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Devices with localization output must include localization accuracy testing as a component of standalone testing. The test dataset must be representative of the typical patient population with enrichment made only to ensure that the test dataset contain a sufficient number of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, concomitant disease, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment. iv. Results from performance testing that demonstrate that the device provides improved assisted-read detection and/or diagnostic performance as intended in the indicated user population(s) when used in accordance with the instructions for use. The reader population must be comprised of the intended user population in terms of but not limited to clinical training, certification, and years of experience. The performance assessment must be based on appropriate diagnostic accuracy measures (e.g., receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Test datasets must meet the requirements described in 1(iii) above. v. Appropriate software documentation, including device hazard analysis, software requirements specification document, software design specification document, traceability analysis, system level test protocol, pass/fail criteria, testing results, and cybersecurity measures. 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the device instructions for use, including the intended reading protocol and how the user should interpret the device output. iii. A detailed description of the intended user, and any user training materials as programs that addresses appropriate reading protocols for the device to ensure that the end user is fully aware of how to interpret and apply the device output. iv. A detailed description of the device inputs and outputs. v. A detailed description of compatible imaging hardware and imaging protocols. vi. Warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality or for certain subpopulations), as applicable. vii. A detailed summary of the performance testing, including: test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as anatomical characteristics, patient demographics and medical history, user experience, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the image analysis algorithm, including a description of the algorithm inputs and outputs, each major component or block, how the algorithm and output affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide improved assisted-read detection and diagnostic performance as intended in the indicated user population(s), and to characterize the standalone device performance for labeling. Performance testing includes standalone test(s), side-by-side comparison(s), and/or a reader study, as applicable.
(iii) Results from standalone performance testing used to characterize the independent performance of the device separate from aided user performance. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Devices with localization output must include localization accuracy testing as a component of standalone testing. The test dataset must be representative of the typical patient population with enrichment made only to ensure that the test dataset contains a sufficient number of cases from important cohorts (*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, concomitant disease, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment.(iv) Results from performance testing that demonstrate that the device provides improved assisted-read detection and/or diagnostic performance as intended in the indicated user population(s) when used in accordance with the instructions for use. The reader population must be comprised of the intended user population in terms of clinical training, certification, and years of experience. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Test datasets must meet the requirements described in paragraph (b)(1)(iii) of this section.(v) Appropriate software documentation, including device hazard analysis, software requirements specification document, software design specification document, traceability analysis, system level test protocol, pass/fail criteria, testing results, and cybersecurity measures.
(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use.
(ii) A detailed description of the device instructions for use, including the intended reading protocol and how the user should interpret the device output.
(iii) A detailed description of the intended user, and any user training materials or programs that address appropriate reading protocols for the device, to ensure that the end user is fully aware of how to interpret and apply the device output.
(iv) A detailed description of the device inputs and outputs.
(v) A detailed description of compatible imaging hardware and imaging protocols.
(vi) Warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality or for certain subpopulations), as applicable.(vii) A detailed summary of the performance testing, including test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as anatomical characteristics, patient demographics and medical history, user experience, and imaging equipment.
{1}------------------------------------------------
November 18, 2020
Image /page/0/Picture/1 description: The image contains two logos. On the left is the Department of Health & Human Services logo, which features a stylized human figure. To the right is the FDA (U.S. Food & Drug Administration) logo, with the letters "FDA" in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text to the right of the square.
Hologic, Inc. % Ms. Deborah Thomas Regulatory Affairs Manager 250 Campus Drive MARLBOROUGH MA 01752
Re: K201019
Trade/Device Name: Genius AI Detection Regulation Number: 21 CFR 892.2090 Regulation Name: Radiological computer assisted detection and diagnosis software Regulatory Class: Class II Product Code: QDQ Dated: October 16, 2020 Received: October 19, 2020
Dear Ms. Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{2}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201019
Device Name Genius AI Detection
#### Indications for Use (Describe)
Genius AI Detection is a computer-aided detection and diagnosis (CADe/CADx) software device intended to be used with compatible digital breast tomosynthesis (DBT) systems to identify and mark regions of interest including soft tissue densities (masses, architectural distortions and asymmetries) and calcifications in DBT exams from compatible DBT systems and provide confidence scores that offer assessment for Certainty of Findings and a Case Score. The device intends to aid in the interpretation of digital breast tomosynthesis exams in a concurrent fashion, where the interpreting physician confirms or dismisses the findings during the reading of the exam.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
| <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### Traditional 510(k) Summary K201019
This 510(k) Summary is submitted in accordance with the requirements of 21 CFR Part 807.92
| Date Prepared: | November 13, 2020 |
|----------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| Manufacturer: | Hologic, Inc. |
| | 36 Apple Ridge Road |
| | Danbury, CT 06810 USA |
| Establishment Registration #: | 1220984 |
| Contact Person: | Deborah Thomas |
| | Regulatory Affairs Manager |
| | P: 508.210.6107 |
| Identification of the Device: | |
| Proprietary/Trade Name: | Genius Al Detection |
| Classification Name: | Radiological Computer Assisted Detection/Diagnosis<br>Software for Lesions Suspicious For Cancer |
| Regulatory Number: | 21 CFR 892.2090 |
| Product Code: | QDQ |
| Device Class: | Class II |
| Review Panel: | Radiology |
| Identification of the Legally Marketed Predicate Device: | |
| Trade Name: | PowerLook Tomo Detection V2 Software |
| Classification Name: | Radiological Computer Assisted Detection/Diagnosis<br>Software for Lesions Suspicious For Cancer |
| Regulatory Number: | 21 CFR 892.2090 |
| Product Code: | QDQ |
| Device Class: | Class II |
| Review Panel: | Radiology |
| Submitter/510(k) Holder: | iCAD |
| Clearance: | K182373 (cleared December 6, 2018) |
#### Device Description:
Genius Al Detection is a software device intended to identify potential abnormalities in breast tomosynthesis images. Genius Al Detection analyzes each standard mammographic view in a digital breast tomosynthesis examination using deep learning networks. For each detected lesion, Genius Al Detection produces CAD results that include the location of the lesion, an outline of the lesion and a confidence score for that lesion. Genius Al Detection also produces a case score for the entire tomosynthesis exam.
{5}------------------------------------------------
Genius Al Detection packages all CAD findings derived from the corresponding analysis of a tomosynthesis exam into a DICOM Mammography CAD SR object and distributes it for display on DICOM compliant review workstations. The interpreting physician will have access to the CAD findings concurrently to the reading of the tomosynthesis exam. In addition, a combination of peripheral information such as number of marks and case scores may be used on the review workstation to enhance the interpreting physician's workflow by offering a better organization of the patient worklist.
## Indications for Use:
Genius Al Detection is a computer-aided detection and diagnosis (CADe/CADx) software device intended to be used with compatible digital breast tomosynthesis (DBT) systems to identify and mark regions of interest including soft tissue densities (masses, architectural distortions and asymmetries) and calcifications in DBT exams from compatible DBT systems and provide confidence scores that offer assessment for Certainty of Findings and a Case Score. The device intends to aid in the interpretation of digital breast tomosynthesis exams in a concurrent fashion, where the interpreting physician confirms or dismisses the findings during the reading of the exam.
# Standards:
- IEC 62304: 2015 – Medical device software – Software Life Cycle Processes (#13-79)
- ISO 14971: 2012 – Medical devices – Application of Risk Management to Medical Devices
- . DEN180005 Evaluation of automatic class III designation for OsteoDetect – Decision summary with special controls.
# FDA Guidance Documents:
- Guidance for Industry and FDA Staff Guidance for the Content of Premarket Submissions for ● Software Contained in Medical Devices (Issued on May 11, 2005)
- . Computer-Assisted Detection Devices Applied to Radiology Images and Radiology Device Data – Premarket Notification [510(k)] Submissions (Issued on July 3, 2012)
- . Guidance for Industry and FDA Staff - Clinical Performance Assessment: Considerations for Computer-Assisted Detection Devices Applied to Radiology Images and Radiology Device Data -Premarket Approval (PMA) and Premarket Notification [510(k)] Submissions (Issued on January 22, 2020)
- . "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices," issued on October 2, 2014
- "Off-the-Shelf Software Use in Medical Devices," issued on September 9, 1999
{6}------------------------------------------------
# Summary of Substantial Equivalence:
| Features and<br>Characteristics | Subject Device | Predicate Device | Difference and<br>comments |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| | Hologic, Inc.<br>Genius AI Detection | iCAD Inc.<br>PowerLook®<br>Tomo | |
| Regulation<br>Number/Name | 21 CFR 892.2090 /<br>Radiological Computer Assisted<br>Detection and Diagnosis Software | Same | N/A |
| Product Code | QDQ | Same | N/A |
| Regulation<br>Description | A radiological computer assisted detection<br>and diagnostic software is an image<br>processing device intended to aid in the<br>detection, localization, and<br>characterization of fracture, lesions, or<br>other disease specific findings on acquired<br>medical images (e.g. radiography, MR, CT).<br>The device detects, identifies and<br>characterizes findings based on features<br>or information extracted from images, and<br>provides information about the presence,<br>location, and characteristics of the<br>findings to the user. The analysis is<br>intended to inform the primary diagnostic<br>and patient management decisions that<br>are made by the clinical user. The device<br>is not intended as a replacement for a<br>complete clinician's review or their<br>clinical judgment that takes into account<br>other relevant information from the<br>image or patient history. | Same | N/A |
| Indications for<br>Use | Genius Al Detection is a computer-aided<br>detection and diagnosis (CADe/CADx)<br>software device intended to be used with<br>compatible digital breast tomosynthesis<br>(DBT) systems to identify and mark<br>regions of interest including soft tissue<br>densities (masses, architectural<br>distortions and asymmetries) and<br>calcifications in DBT exams from<br>compatible DBT systems and provide<br>confidence scores that offer assessment<br>for Certainty of Findings and a Case Score.<br>The device intends to aid in the<br>interpretation of digital breast<br>tomosynthesis exams in a concurrent<br>fashion, where the interpreting physician<br>confirms or dismisses the findings during<br>the reading of the exam. | PowerLook® Tomo<br>Detection V2 software<br>is a computer-<br>assisted detection and<br>diagnosis (CAD)<br>software device<br>intended to be used<br>concurrently by<br>interpreting physicians<br>while reading digital<br>breast tomosynthesis<br>(DBT) exams from<br>compatible DBT<br>systems. The system<br>detects soft tissue<br>densities (masses,<br>architectural<br>distortions and<br>asymmetries) and<br>calcifications in the 3D<br>DBT slices. The<br>detections and<br>Certainty of Finding<br>and Case Scores assist<br>interpreting physicians<br>in identifying soft<br>tissue densities and | |
| Compatible<br>DBT Systems | Hologic Selenia Dimensions<br>Hologic 3Dimensions<br><br>Supports both models in the following<br>modes:<br>standard resolution 1-mm slices high resolution 1-mm slices<br>(Clarity HD), high resolution 6-mm SmartSlices<br>(3DQuorum) | tissue densities and<br>Hologic Selenia<br>Dimensions (standard<br>resolution, 1-mm<br>slices)<br>GE Pristina | The subject<br>device and<br>predicate device<br>are compatible<br>with different<br>systems as<br>noted. |
| Type of CAD<br>Software | Radiological computer assisted<br>detection and diagnostic software. | same | N/A |
| Mode of Action | Image processing device utilizing<br>machine learning to aid in the<br>detection, localization, and<br>characterization of soft tissue densities<br>(masses, architectural distortions and<br>asymmetries) and calcifications in the<br>1-mm 3D DBT slices. Findings are co-<br>registered to 6-mm SmartSlices. | Image processing<br>device utilizing<br>machine learning<br>to aid in the<br>detection,<br>localization, and<br>characterization of soft<br>tissue densities<br>(masses, architectural<br>distortions and<br>asymmetries) and | Co-registration of<br>findings to 6-mm<br>SmartSlices. |
| Clinical Output | To inform the primary diagnostic and<br>patient management decisions that are<br>made by the clinical user. | same | N/A |
| Patient<br>Population | Symptomatic and asymptomatic<br>women undergoing mammography | same | N/A |
| End Users | MQSA-Qualified Interpreting<br>Physicians and Radiologists | same | N/A |
| Image Source<br>Modalities | Digital breast tomosynthesis slices | same | N/A |
| Output Device | Softcopy Workstation | same | N/A |
| Deployment | Stand-alone computer | same | N/A |
| Visualization<br>Features | Places mark within suspicious lesion by<br>default (Emphasize™; RightOn™) and<br>reports confidence of finding next to each<br>identified lesion in the image. CAD display<br>may be toggled on/off. Option to<br>automatically zoom into or contour the<br>suspicious region of interest (PeerView™). | Contours suspicious<br>lesions by default and<br>displays confidence of<br>finding next to each<br>identified lesion in the<br>image. CAD display may<br>be toggled on/off. No<br>other marks. | extra display<br>functions of marks |
| Method Of Use | Concurrent read | Concurrent read. FFDM<br>and 2D synthetic views<br>when available are to<br>be reviewed before | N/A |
| Supported Views | CC and MLO | same | N/A |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
# Comparison with Predicate Device:
The Summary of Substantial Equivalence Table above details the similarities and differences between the Genius Al Detection device and its predicate device, PowerLook® Tomo Detection V2. Both devices aid in the detection, localization, and characterization of disease specific findings on acquired medical images.
Genius Al Detection is the same technology as the predicate per 21 CFR 892.2090; both devices are radiological computer assisted detection and diagnostic software, intended to aid in the detection, localization, and characterization of disease specific findings on acquired medical images. The outputs of both devices serve to augment the interpretation of digital breast tomosynthesis exams as a concurrent reading tool. The output is used to inform and assist the interpreting physician, supplementing their clinical expertise and judgment. In the case of any differences that may occur between the subject device and the predicate, rationale for safety and effectiveness is provided above along with special controls established for Radiological Computer Assisted Detection and Diagnosis Software that are in place to further mitigate any risks in these differences.
## Compatible DBT Systems
The following image types have been tested and are compatible with Genius Al Detection:
- Hologic standard resolution tomosynthesis slices (1 mm)
- Hologic high resolution tomosynthesis slices (1 mm) (Clarity HD)
- Hologic high resolution SmartSlices (6 mm) (3DQuorum)
The CAD marks generated by Genius Al Detection for the above image types can also be projected on their corresponding synthesized 2D images, providing that the diagnostic review workstation supports such a feature.
## Performance Testing - Reader Study Results
The study was successfully executed with all 17 readers; the results below represent a per-protocol analysis of the 390 cases (106 cancers, and 284 negative cases) included in the MRMC where both rounds of reading were completed.
Based on analyses that do not control type I error and therefore cannot be generalized to specific comparisons outside this particular study, in this study:
- The average observed AUC was 0.825 (95% Cl: 0.783, 0.867) with CAD and 0.794 (95% Cl: 0.748, 0.840) without CAD. The difference in observed AUC was +0.031 (95% Cl: 0.012, 0.051).
{10}------------------------------------------------
- The average observed reader sensitivity for cancer cases was 75.9% with CAD and 66.8% without CAD. The difference in observed sensitivity was +9.0% (99% Cl: 6.0%, 12.1%).
- . The average observed recall rate for non-cancer cases was 25.8% with CAD and 23.4% without CAD. The observed difference in negative recall rate was +2.4% (99% Cl: 0.7%, 4.2%).
- . The average observed case read-time was 52.0s with CAD and 46.3s without CAD. The observed difference in read-time was 5.7s (95% Cl: 4.9s to 6.4s).
# Standalone Testing:
Hologic conducted an MRMC reader study to assess the safety and efficacy of Genius Al Detection on Hologic's high resolution tomosynthesis images, where 3D data sets are reconstructed at 70μm. Parallel to the MRMC study, a standalone study was conducted to establish equivalence of Genius Al Detection performance on Hologic's standard resolution tomosynthesis images, where 3D data sets are reconstructed at ~100um compared to the performance on the high resolution tomosynthesis images. The standalone study was conducted on paired high resolution and standard resolution 3D data sets, where each high-resolution reconstructed 3D volume had a counterpart standard resolution 3D volume, both acquired from a single exposure and under the same compression.
All the results of the standalone study confirmed that Genius Al Detection when operating under the Hologic's standard tomosynthesis acquisition mode performs comparably to when operating under the high-resolution mode.
Using an overall data set of 764 cases including 106 cancers and 658 non-cancer cases, Genius AI 1. Detection demonstrated comparable detection performance on both Hologic's standard and highresolution acquisition modes as observed by fROC analysis.
No significant differences were observed in the number and type of cancers detected by Genius Al 2. Detection in either acquisition mode.
3. Stratified fROC analysis using lesion type as well as breast density also showed comparable performance of Genius Al Detection operating in either acquisition mode
# Assessment of Benefit-Risk, Safety and Effectiveness, and Substantial Equivalence:
As a part of the submission, we have demonstrated the probable benefits of the device through clinical study including reader accuracy as assessed by AUC (i.e. diagnostic performance and improved assistedread detection from the sensitivity analysis. In totality, Hologic finds that the proposed device has a positive balance in terms of probable benefits vs probable risks and thus may be considered safe and effective based on the special controls established in DEN180005 such that "the device will provide improved assisted-read detection and diagnostic performance."
## Conclusion:
Based on the information submitted in this premarket notification, the Genius Al Detection device and its predicate device, PowerLook® Tomo Detection V2 both have a similar intended use and are devices which aid in the detection, localization, and characterization of disease specific findings on acquired medical images. The differences discussed are not significant to the technology and clinical application of the device. The proposed Genius Al Detection device has been found to be substantially equivalent to the predicate PowerLook® Tomo Detection V2 device (K182373).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.