EPIQ Diagnostic Ultrasound System Series, Affiniti Diagnostic Ultrasound System Series, CX50 Diagnostic Ultrasound System, Sparq Diagnostic Ultrasound System, Lumify Diagnostic Ultrasound System
K201012 · Philips Ultrasound, Inc. · IYN · May 1, 2020 · Radiology
Device Facts
Record ID
K201012
Device Name
EPIQ Diagnostic Ultrasound System Series, Affiniti Diagnostic Ultrasound System Series, CX50 Diagnostic Ultrasound System, Sparq Diagnostic Ultrasound System, Lumify Diagnostic Ultrasound System
Applicant
Philips Ultrasound, Inc.
Product Code
IYN · Radiology
Decision Date
May 1, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
AI/ML, Pediatric
Indications for Use
The intended use of the Affiniti 30, Affiniti 70 Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow analysis of the human body with the following Indications for Use: Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Transesophageal (Cardiac), Transrectal, Transvaginal. The clinical environments where the Affiniti Diagnostic Ultrasound Systems can be used include Clinics, and clinical point-of-care for diagnosis of patients. When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live Echo Guidance. The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information, and only for the purposes for which it was designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgment and best clinical procedure.
Device Story
Philips Diagnostic Ultrasound Systems (Lumify, EPIQ, Affiniti, CX50, Sparq) and QLAB software are reusable capital equipment for high-resolution imaging and fluid flow analysis. Systems use transducers with piezoelectric elements to generate and receive ultrasound waves; console processes echoes into images for display. Used in clinics, hospitals, and point-of-care by trained professionals. QLAB software views/quantifies image data on PCs or on-board systems. Devices assist interventionalists/surgeons via EchoNavigator for cardiovascular procedures using live X-ray and Echo guidance. Expanded labeling provides guidance for lung and cardiac ultrasound in COVID-19 patients. Benefits include diagnostic visualization and clinical decision support for cardiovascular and general imaging.
Clinical Evidence
No clinical data. Bench testing only; substantial equivalence supported by design control measures and comparison to predicate devices.
Technological Characteristics
Ultrasound imaging systems using piezoelectric transducers (curved, linear, sector, phased array, xMatrix). Biocompatibility per ISO 10993-1. Systems include primary console, software, and transducers. Connectivity via standard interfaces. Sterilization via compatible probe covers.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in patients across various clinical applications including abdominal, cardiac (adult/pediatric/fetal/transesophageal), vascular (cerebral/peripheral/intraoperative), obstetric/gynecological, musculoskeletal, urological, and small organ (breast/thyroid/testicle). Used in clinics, hospitals, and point-of-care settings by healthcare professionals.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Philips Lumify Diagnostic Ultrasound System (K192226)
Philips EPIQ Diagnostic Ultrasound System (K182857)
Philips Affiniti Diagnostic Ultrasound System (K182857)
Philips CX50 Diagnostic Ultrasound System (K162329)
Philips Sparq Diagnostic Ultrasound System (K162329)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Philips Ultrasound, Inc. % Deep Pal Director, Regulatory Affairs 22100 Bothell Everett Highway Bothell WA 98021
May 1, 2020
## Re: K201012
Trade/Device Name: Affiniti Diagnostic Ultrasound System Series Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX, QIH Dated: April 17, 2020 Received: April 20, 2020
Dear Deep Pal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{2}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Number (if known) K201012
### Device Name
Affiniti Diagnostic Ultrasound System Series
### Indications for Use (Describe)
The intended use of the Affiniti 30, Affiniti 70 Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow analysis of the human body with the following Indications for Use: Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Transesophageal (Cardiac), Transrectal, Transvaginal.
The clinical environments where the Affiniti Diagnostic Ultrasound Systems can be used include Clinics, and clinical point-of-care for diagnosis of patients.
When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live Echo Guidance. The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information, and only for the purposes for which it was designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgment and best clinical procedure.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{4}------------------------------------------------
510(k) Number (if known) K201012
Device Name CX50 Diagnostic Ultrasound System
### Indications for Use (Describe)
Philips CX50 Diagnostic Ultrasound Systems is intended for diagnostic ultrasound imaging in B (or 2- D), M-mode (including Anatomical M-mode), Pulse Wave Doppler, Continuous Wave Doppler, Color Doppler, Tissue Doppler Imaging and Harmonics (Tissue and Contrast) modes.
It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications: Ophthalmic Intraoperative Laparoscopic Fetal Abdominal Pediatric Small Organ Adult Cephalic Neonatal Cephalic Trans-vaginal Musculo-skeletal Gynecological Cardiac Pediatric Trans-Esoph. (Cardiac) Intracardiac echo Peripheral Vessel Other (Carotid)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{5}------------------------------------------------
510(k) Number (if known) K201012
### Device Name
EPIQ Diagnostic Ultrasound System Series
### Indications for Use (Describe)
The intended use of the EPIQ, EPIQ 7 is diagnostic ultrasound imaging and fluid flow analysis of the human body with the following Indications for Use: Abdominal, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal.
The clinical environments where the EPIQ Diagnostic Ultrasound Systems can be used include Clinics, Hospitals, and clinical point-of-care for diagnosis of patients.
When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live Echo Guidance. The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information, and only for the purposes for which it was designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgment and best clinical procedure.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{6}------------------------------------------------
510(k) Number (if known) K201012
Device Name Lumify Diagnostic Ultrasound System
### Indications for Use (Describe)
Philips Lumify Diagnostic Ultrasound System is intended for diagnostic ultrasound imaging in B (2D), Color Doppler, Combined (B+Color), and M modes. It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications: Fetal/Obstetric, Abdominal, Pediatric, Cephalic, Urology, Gynecological, Cardiac Fetal Echo, Small Organ, Musculoskeletal, Peripheral Vessel, Carotid, Cardiac. Lumify is a transportable ultrasound system intended for use in environments where healthcare is provided by healthcare professionals.
Type of Use (Select one or both, as applicable)
| <span style="font-size:12px">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size:12px">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{7}------------------------------------------------
510(k) Number (if known) K201012
Device Name
QLAB Advanced Quantification Software
Indications for Use (Describe)
QLAB Advanced Quantification Software is a software application package. It is designed to view and quantify image data acquired on Philips ultrasound systems.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 201. Subpart D) </span> |
|---------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 201. Subpart C) </span> |
| X Prescription Use (Part 21 CFR 801 Subpart D)
|__ | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{8}------------------------------------------------
510(k) Number (if known) K201012
Device Name Sparq Diagnostic Ultrasound System
### Indications for Use (Describe)
Philips Sparq Diagnostic Ultrasound System is intended for diagnostic ultrasound imaging in B (or 2- D), M-mode (including Anatomical M-mode), Pulse Wave Doppler, Continuous Wave Doppler, Color Doppler, Tissue Doppler Imaging and Harmonics (Tissue and Contrast) modes. It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications: Ophthalmic Fetal Abdominal Pediatric Small Organ Adult Cephalic Trans-vaginal Trans-rectal Musculo-skeletal Gynecological Cardiac Adult Trans-Esoph. (Cardiac) Peripheral Vessel.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{9}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, which makes the blue letters stand out.
I
# 510(k) Summary
This summary of safety and effectiveness information is submitted in accordance with 21 CFR § 807.92. Date Prepared: April 30, 2020
| I. Submitter | |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer Name<br>and Address | Philips Ultrasound, Inc.<br>22100 Bothell Everett Hwy<br>Bothell, WA 98021-8431 |
| Contact<br>Information | Deep Pal<br>Head of Regulatory Affairs<br>TEL: +1 (425) 219-3412<br>EMAIL: deep.pal@philips.com |
| Secondary Contact<br>Information | Benny Lam<br>Principal Regulatory Affairs Specialist<br>TEL: +1 (425)-215-3496<br>EMAIL: benny.lam@philips.com |
| II. Device | |
| Trade Name | Philips Lumify Diagnostic Ultrasound System<br>Philips EPIQ Diagnostic Ultrasound System Series<br>Philips Affiniti Diagnostic Ultrasound System Series<br>Philips CX50 Diagnostic Ultrasound System<br>Philips Sparq Diagnostic Ultrasound System<br>Philips QLAB Advanced Quantification Software |
| Common Name | Diagnostic ultrasound system and transducers<br>Automated Radiological Image Processing Software |
| Regulation Description | Ultrasonic pulsed doppler imaging system<br>Ultrasonic pulsed echo imaging system<br>Diagnostic ultrasonic transducer<br>Picture archiving and communications system |
| Regulation Number | 892.1550<br>892.1560<br>892.1570<br>892.2050 |
| Product Code | IYN |
| Secondary Product Codes | IYO |
{10}------------------------------------------------
## ITX OIH
| Device Class | Class II |
|----------------------|-----------|
| Classification Panel | Radiology |
#### III. Predicate Device
- K192226 Philips Lumify Diagnostic Ultrasound System ■
- K182857 Philips EPIQ Diagnostic Ultrasound System ■
- l K182857 - Philips Affiniti Diagnostic Ultrasound System
- K162329 Philips CX50 Diagnostic Ultrasound System ■
- I K162329 - Philips Sparq Diagnostic Ultrasound System
- K191647 - QLAB Advanced Quantification Software
#### IV. Device Description
Philip Diagnostic Ultrasound Systems are durable, reusable capital equipment medical devices which are intended for high-resolution general imaging, interventional radiology, cardiology, vascular and OB/GYN applications and fluid flow analysis. They are intended to be used by trained professionals at various settings of patient care such as clinical admission, periodic evaluations, prior to hospitalization discharge, and/or academic research, via maneuverable caster wheels or mobile handheld components.
Philips QLAB Advanced Quantification software (QLAB) is designed to view and quantify image data acquired on Philips ultrasound systems. QLAB is available either as a stand-alone product that can function on a standard PC, a dedicated workstation, and on-board Philips' Diagnostic Ultrasound Systems.
Software modes/applications, scanning protocols, and pre-installed settings or functionality to create dedicated settings for imaging of specific anatomy are available with the subject Philips Diagnostic Ultrasound Systems and QLAB software and may vary among model configurations.
The diagnostic ultrasound systems are manufactured with hardware components which consist of:
- a primary console (e.g., workstation, tablet) with built-in software components, features, and 1) various clinical applications, and
- 2) a range of compatible ultrasound transducers.
A suite of compatible transducer types are offered such as transesophageal echocardiography (TEE) transducers, non-imaging (pencil) probes, curved array, linear array and sector/ohased array transducer models which have a prefix "C" are designed to provide larger fields of view and penetration for e.g., consolidation between bone spaces while linear array transducer models, which can be identified with a "L" prefix model name are designed to provide shallow visualization which are optimal for tissue layer interfaces. Sector/phased array transducer models which usually have a "S" prefix model names are designed to be applied for difficult anatomical sites (e.g., cardiac intercostal space (ICS)). Other Philips ultrasound transducers may also start with the prefix "X" which is referring to the xMatrix technology for two full-resolution planes of imaging. Volumetric three-dimensional (3D) imaging functionality is also available when specific transducer models are connected to the primary console. Philips ultrasound
{11}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "PHILIPS" in a bold, sans-serif font. The color of the text is a bright blue. The letters are evenly spaced and the word is centered in the image.
### SPECIAL 510(k) Expanded Labeling for Lumify/EPIQ/Affiniti/CX50/Sparq/QLAB Devices
transducers may be bagged into compatible transducer/probe cover sheaths, designed by other manufacturers, for each procedure to prevent cross-contamination and reduce the risk of healthcareassociated infections.
Some configurations may have additional previously-cleared accessories, components and software features which are manufactured by Philips Ultrasound, Inc. or other manufacturers.
{12}------------------------------------------------
| Device | Indications for Use |
|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Philips Lumify<br>Diagnostic<br>Ultrasound<br>System | Philips Lumify Diagnostic Ultrasound System is intended for diagnostic ultrasound imaging in B (2D), Color Doppler, Combined<br>(B+Color), and M modes. It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications:<br>Fetal/Obstetric, Abdominal, Pediatric, Cephalic, Urology, Gynecological, Cardiac Fetal Echo, Small Organ, Musculoskeletal,<br>Peripheral Vessel, Carotid, Cardiac. Lumify is a transportable ultrasound system intended for use in environments where<br>healthcare is provided by healthcare professionals. |
| Philips EPIQ<br>Diagnostic<br>Ultrasound<br>System | The intended use of the EPIQ, EPIQ 5, EPIQ 7 Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow<br>analysis of the human body with the following Indications for Use: Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac<br>Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular),<br>Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral<br>Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal.<br>The clinical environments where the EPIQ Diagnostic Ultrasound Systems can be used include Clinics, Hospitals, and clinical<br>point-of-care for diagnosis of patients.<br>When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with image guidance during<br>treatment of cardiovascular disease in which the procedure uses both live X-ray and live Echo Guidance.<br>The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating<br>instructions given in the product user information, and only for the purposes for which it was designed. However, nothing stated<br>in the user information reduces your responsibility for sound clinical judgement and best clinical procedure. |
| Philips Affiniti<br>Diagnostic<br>Ultrasound<br>System | The intended use of the Affiniti 30, Affiniti 50 and Affiniti 70 Diagnostic Ultrasound Systems is diagnostic ultrasound imaging<br>and fluid flow analysis of the human body with the following Indications for Use: Abdominal, Cardiac Adult, Cardiac other<br>(Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic<br>(Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional),<br>Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle),<br>Transesophageal (Cardiac), Transrectal, Transvaginal.<br>The clinical environments where the Affiniti Diagnostic Ultrasound Systems can be used include Clinics, Hospitals, and clinical<br>point-of-care for diagnosis of patients.<br>When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with image guidance during<br>treatment of cardiovascular disease in which the procedure uses both live X-ray and live Echo Guidance. |
| Device | Indications for Use |
| Philips CX50<br>Diagnostic<br>Ultrasound<br>System | The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating<br>instructions given in the product user information, and only for the purposes for which it was designed. However, nothing stated<br>in the user information reduces your responsibility for sound clinical judgement and best clinical procedure |
| | Philips CX50 Diagnostic Ultrasound Systems is intended for diagnostic ultrasound imaging in B (or 2- D), M-mode (including<br>Anatomical M-mode), Pulse Wave Doppler, Continuous Wave Doppler, Color Doppler, Tissue Doppler Imaging and Harmonics<br>(Tissue and Contrast) modes.<br>It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications: Ophthalmic Intraoperative<br>Laparoscopic Fetal Abdominal Pediatric Small Organ Adult Cephalic Neonatal Cephalic Trans-vaginal Musculo-skeletal<br>Gynecological Cardiac Adult Cardiac Pediatric Trans-Esoph. (Cardiac) Intracardiac echo Peripheral Vessel Other (Carotid) |
| Philips Sparq<br>Diagnostic<br>Ultrasound<br>System | Philips Sparq Diagnostic Ultrasound System is intended for diagnostic ultrasound imaging in B (or 2- D), M-mode (including<br>Anatomical M-mode), Pulse Wave Doppler, Continuous Wave Doppler, Color Doppler, Tissue Doppler Imaging and Harmonics<br>(Tissue and Contrast) modes. It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications:<br>Ophthalmic Fetal Abdominal Pediatric Small Organ Adult Cephalic Trans-vaginal Trans-rectal Musculo-skeletal Gynecological<br>Cardiac Adult Trans-Esoph. (Cardiac) Peripheral Vessel |
| QLAB<br>Advanced<br>Quantification<br>software | QLAB Advanced Quantification Software is a software application package. It is designed to view and quantify image data<br>acquired on Philips ultrasound systems. |
{13}------------------------------------------------
{14}------------------------------------------------
Page 6 of 19
#### VI. Comparison of Technological Characteristics with the Predicate Device
Table 1: Technological Comparison of Subject Device (i.e., Philips Lumify Diagnostic Ultrasound System) & Predicate Device
| Standard Feature | Philips Lumify Ultrasound System<br>K# Pending<br>(Subject Device) | Philips Lumify Diagnostic<br>Ultrasound System<br>K192226<br>(Predicate Device) | Comparison |
|-----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Indications for Use | Philips Lumify Diagnostic Ultrasound<br>System is intended for diagnostic<br>ultrasound imaging in B (2D), Color<br>Doppler, Combined (B+Color), and M<br>modes. It is indicated for diagnostic<br>ultrasound imaging and fluid flow<br>analysis in the following applications:<br>Fetal/Obstetric, Abdominal, Pediatric,<br>Cephalic, Urology, Gynecological,<br>Cardiac Fetal Echo, Small Organ,<br>Musculoskeletal, Peripheral Vessel,<br>Carotid, Cardiac. Lumify is a<br>transportable ultrasound system<br>intended for use in environments where<br>healthcare is provided by healthcare<br>professionals. | Philips Lumify Diagnostic Ultrasound<br>System is intended for diagnostic<br>ultrasound imaging in B (2D), Color<br>Doppler, Combined (B+Color), and M<br>modes. It is indicated for diagnostic<br>ultrasound imaging and fluid flow<br>analysis in the following applications:<br>Fetal/Obstetric, Abdominal, Pediatric,<br>Cephalic, Urology, Gynecological,<br>Cardiac Fetal Echo, Small Organ,<br>Musculoskeletal, Peripheral Vessel,<br>Carotid, Cardiac. Lumify is a<br>transportable ultrasound system<br>intended for use in environments where<br>healthcare is provided by healthcare<br>professionals. | Identical |
| Reusable? | Yes | Yes | Identical |
| Duration of Use | Limited (≤ 24 hours) | Limited (≤ 24 hours) | Identical |
| Scientific<br>Technology | Ultrasound Imaging | Ultrasound Imaging | Identical |
| Operating<br>principles | Compatible device generates electrical<br>current and sends to a connected,<br>compatible transducer to stimulate its<br>piezoelectric elements at the distal end. | Compatible device generates electrical<br>current and sends to a connected,<br>compatible transducer to stimulate its<br>piezoelectric elements at the distal end. | Identical |
| Standard Feature | Philips Lumify Ultrasound System<br>K# Pending<br>(Subject Device) | Philips Lumify Diagnostic<br>Ultrasound System<br>K192226<br>(Predicate Device) | Comparison |
| | Stimulation causes the elements to<br>expand and contract which creates a<br>high-pressured wave (i.e., soundwave). A<br>series of soundwaves then propagate<br>toward tissue medium (e.g., mucus<br>membrane, bone). The transducer then<br>receives echoed soundwaves that are<br>reflected from the tissue medium and<br>transmits it to the system. The system<br>processes the echoed soundwaves into<br>an image which is displayed on the<br>display monitor screen of the system for<br>user interpretation. | Stimulation causes the elements to<br>expand and contract which creates a<br>high-pressured wave (i.e., soundwave). A<br>series of soundwaves then propagate<br>toward tissue medium (e.g., mucus<br>membrane, bone). The transducer then<br>receives echoed soundwaves that are<br>reflected from the tissue medium and<br>transmits it to the system. The system<br>processes the echoed soundwaves into<br>an image which is displayed on the<br>display monitor screen of the system for<br>user interpretation. | |
| Type of Previously-<br>cleared<br>Transducers | Curved Array<br>Linear Array<br>Sector Array | Curved Array<br>Linear Array<br>Sector Array | Identical |
| Acoustic Outputs<br>Within Range? | Yes | Yes | Identical |
| Previously cleared<br>Imaging Modes? | Yes | Yes | Identical |
| Biocompatibility | ISO 10993-1 | ISO 10993-1 | Identical |
{15}------------------------------------------------
# Expanded Labeling for Lumify/EPIQ/Affiniti/CX50/Sparq/QLAB Devices
Page 7 of 19
{16}------------------------------------------------
Page 8 of 19
Table 2: Technological Comparison of Subject Device (i.e., Philips EPIQ Diagnostic Ultrasound System Series) & Predicate Device
| Standard Feature | Philips EPIQ Diagnostic Ultrasound<br>System Series<br>K# Pending | Philips EPIQ Diagnostic Ultrasound<br>System Series<br>K182857 | Comparison |
|-----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| | (Subject Device) | (Predicate Device) | |
| Indications for Use | The intended use of the EPIQ, EPIQ 5,<br>EPIQ 7 Diagnostic Ultrasound Systems<br>is diagnostic ultrasound imaging and<br>fluid flow analysis of the human body<br>with the following Indications for Use:<br>Abdominal, Cardiac Adult, Cardiac other<br>(Fetal), Cardiac Pediatric, Cerebral<br>Vascular, Cephalic (Adult), Cephalic<br>(Neonatal), Fetal/Obstetric,<br>Gynecological, Intraoperative (Vascular),<br>Intraoperative (Cardiac), Musculoskeletal<br>(Conventional), Musculoskeletal<br>(Superficial), Other: Urology, Pediatric,<br>Peripheral Vessel, Small Organ (Breast,<br>Thyroid, Testicle), Transesophageal<br>(Cardiac), Transrectal, Transvaginal.<br>The clinical environments where the<br>EPIQ Diagnostic Ultrasound Systems<br>can be used include Clinics, Hospitals,<br>and clinical point-of-care for diagnosis of<br>patients.<br>When integrated with Philips<br>EchoNavigator, the systems can assist<br>the interventionalist and surgeon with<br>image guidance during treatment of | The intended use of the EPIQ, EPIQ 5,<br>EPIQ 7 Diagnostic Ultrasound Systems<br>is diagnostic ultrasound imaging and<br>fluid flow analysis of the human body<br>with the following Indications for Use:<br>Abdominal, Cardiac Adult, Cardiac other<br>(Fetal), Cardiac Pediatric, Cerebral<br>Vascular, Cephalic (Adult), Cephalic<br>(Neonatal), Fetal/Obstetric,<br>Gynecological, Intraoperative (Vascular),<br>Intraoperative (Cardiac), Musculoskeletal<br>(Conventional), Musculoskeletal<br>(Superficial), Other: Urology, Pediatric,<br>Peripheral Vessel, Small Organ (Breast,<br>Thyroid, Testicle), Transesophageal<br>(Cardiac), Transrectal, Transvaginal.<br>The clinical environments where the<br>EPIQ Diagnostic Ultrasound Systems<br>can be used include Clinics, Hospitals,<br>and clinical point-of-care for diagnosis of<br>patients.<br>When integrated with Philips<br>EchoNavigator, the systems can assist<br>the interventionalist and surgeon with<br>image guidance during treatment of | Identical |
| Standard Feature | Philips EPIQ Diagnostic Ultrasound<br>System Series<br>K# Pending<br>(Subject Device) | Philips EPIQ Diagnostic Ultrasound<br>System Series<br>K182857<br>(Predicate Device) | Comparison |
| | cardiovascular disease in which the<br>procedure uses both live X-ray and live<br>Echo Guidance.<br>The systems are intended to be installed,<br>used, and operated only in accordance<br>with the safety procedures and operating<br>instructions given in the product user<br>information, and only for the purposes<br>for which it was designed. However,<br>nothing stated in the user information<br>reduces your responsibility for sound<br>clinical judgement and best clinical<br>procedure. | cardiovascular disease in which the<br>procedure uses both live X-ray and live<br>Echo Guidance.<br>The systems are intended to be installed,<br>used, and operated only in accordance<br>with the safety procedures and operating<br>instructions given in the product user<br>information, and only for the purposes<br>for which it was designed. However,<br>nothing stated in the user information<br>reduces your responsibility for sound<br>clinical judgement and best clinical<br>procedure. | |
| Reusable? | Yes | Yes | Identical |
| Duration of Use | Limited (≤ 24 hours) | Limited (≤ 24 hours) | Identical |
| Scientific<br>Technology | Ultrasound Imaging | Ultrasound Imaging | Identical |
| Operating<br>principles | System console generates electrical<br>current and sends to a connected,<br>compatible transducer to stimulate its<br>piezoelectric elements at the distal end.<br>Stimulation causes the elements to<br>expand and contract which creates a<br>high-pressured wave (i.e., soundwave). A<br>series of soundwaves then propagate<br>toward tissue medium (e.g., mucus<br>membrane, bone). The transducer then<br>receives echoed soundwaves that are | System console generates electrical<br>current and sends to a connected,<br>compatible transducer to stimulate its<br>piezoelectric elements at the distal end.<br>Stimulation causes the elements to<br>expand and contract which creates a<br>high-pressured wave (i.e., soundwave). A<br>series of soundwaves then propagate<br>toward tissue medium (e.g., mucus<br>membrane, bone). The transducer then<br>receives echoed soundwaves that are | Identical |
| Standard Feature | Philips EPIQ Diagnostic Ultrasound<br>System Series<br>K# Pending<br>(Subject Device) | Philips EPIQ Diagnostic Ultrasound<br>System Series<br>K182857<br>(Predicate Device)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.