Lumipulse G CA19-9-N

K200997 · Fujirebio Diagnostics,Inc. · NIG · May 14, 2020 · Immunology

Device Facts

Record IDK200997
Device NameLumipulse G CA19-9-N
ApplicantFujirebio Diagnostics,Inc.
Product CodeNIG · Immunology
Decision DateMay 14, 2020
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 866.6010
Device ClassClass 2

Intended Use

Lumipulse® G CA19-9-N is a Chemiluminescent Enzyme Immunoassay (CLEIA) for the quantitative measurement of CA 19-9 in human serum or plasma (sodium heparin, lithium heparin, or dipotassium EDTA) on the LUMIPULSE G System. The assay is to be used as an aid in the management of patients diagnosed with cancer of the exocrine pancreas who have detectable levels of CA 19-9 at some point in their disease process. Serial testing for patient CA 19-9 assay values should be used in conjunction with other clinical methods used for monitoring cancer of the exocrine pancreas.

Device Story

Lumipulse G CA19-9-N; in vitro diagnostic immunoassay for quantitative measurement of CA19-9 in human serum and plasma; used by clinical laboratory professionals; modification involves increasing tamoxifen test concentration for interference evaluation from 0.228 mg/dL to 6.0 mg/dL; fundamental scientific technology remains unchanged; device provides quantitative results to aid in clinical management of patients.

Clinical Evidence

Bench testing only. Verification and validation activities performed to assess the impact of the tamoxifen interference concentration change. No clinical data required for this modification.

Technological Characteristics

Chemiluminescent enzyme immunoassay (CLEIA) for use on the Lumipulse G system. Modification: increased tamoxifen interference test concentration to 6.0 mg/dL. Fundamental technology unchanged.

Indications for Use

Indicated for use as a tumor-associated antigen immunological test system for the quantitative measurement of CA19-9 in human serum and plasma.

Regulatory Classification

Identification

A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.

Special Controls

*Classification.* Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. May 14, 2020 Fujirebio Diagnostics, Inc. Kristin Maddaloni Regulatory Affairs Specialist 201 Great Valley Parkway Malvern, Pennsylvania 19355 Re: K200997 Trade/Device Name: Lumipulse G CA19-9-N Regulation Number: 21 CFR 866.6010 Regulation Name: Tumor-Associated Antigen Immunological Test System Regulatory Class: Class II Product Code: NIG Dated: April 15, 2020 Received: April 16, 2020 Dear Kristin Maddaloni: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Carolina Kagan Acting Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K200997 Device Name Lumipulse G CA19-9-N #### Indications for Use (Describe) WARNING: The concentration of CA19-9 in a given specimen, as determined by assays from different manufacturers, can vary due to differences in assay methods and reagent specificity. The results reported by the laboratory to the physician must include the identity of the assay for CA19-9 used. Values obtained with different assay methods cannot be used interchangeably. If, in the course of monitoring a patient, the assay method used for determining serial levels of CA 19-9 is changed, the laboratory must perform additional serial testing to confirm baseline values. Prior to changing assays, the laboratory MUST confirm baseline values for patients being serially monitored. Lumipulse & CA19-9-N should not be used for cancer screening or diagnosis. WARNING: Patients known to be genotypically negative for the Lewis blood group antigen will be unable to produce the CA 19-9 antigen even in the presence of malignant tissue. Phenotyping for the presence of the Lewis antigen may be insufficient to detect true Lewis antigen negative individuals. Even patients who are genotypically positive for the Lewis antigen may produce varying levels of CA19-9 based on gene dosage effect. Lumipulse & CA19-9-N is a Chemiluminescent Enzyme Immunoassay (CLEIA) for the quantitative measurement of CA19-9 in human serum or plasma (sodium heparin, lithium heparin, or dipotassium EDTA) on the LUMIPULSE G System. The assay is to be used as an aid in the management of patients diagnosed with cancer of the exocrine pancreas who have detectable levels of CA19-9 at some point in their disease process. Serial testing for patient CA19-9 assay values should be used in conjunction with other clinical methods used for monitoring cancer of the exocrine pancreas. Type of Use (Select one or both, as applicable) | <div> <span style="font-size:14px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | |-------------------------------------------------------------------------------------------------------| | <div> <span style="font-size:14px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image contains the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized blue and green graphic to the left of the company name. The company name is written in a blue sans-serif font, with "Diagnostics, Inc." written in a smaller, black sans-serif font below. # SPECIAL 510(k) SUMMARY # A. GENERAL INFORMATION Submission Date: April 15, 2020 Submitter Information: Submitted By: Fujirebio Diagnostics, Inc. 201 Great Valley Parkway Malvern, PA 19355 Contact Person: Kristin Maddaloni Regulatory Affairs Specialist Fujirebio Diagnostics, Inc. 201 Great Valley Pkwy Malvern, PA 19355 Office: 484-395-2126 maddalonik@fdi.com # B. PURPOSE FOR SUBMISSION Special 510(k) for a test concentration change to therapeutic interferent Tamoxifen for Lumipulse G CA19-9-N. # C. MEASURAND CA 19-9 # D. TYPE OF TEST Quantitative, Chemiluminescent Immunoassay ### E. APPLICANT Fujirebio Diagnostics, Inc. # F. PROPRIETARY AND ESTABLISHED NAMES Lumipulse® G CA 19-9-N Lumipulse® G CA19-9-N Immunoreaction Cartridges # G. REGULATORY INFORMATION | Trade Name: | Lumipulse G CA19-9-N | |------------------|-------------------------------------------------------------------------------------------------------| | Classification: | Class II | | Regulation: | 21 CFR 866.6010 | | Regulation Name: | Tumor-Associated Antigen Immunological Test System | | Product Code: | NIG – System, Test, Carbohydrate antigen (CA 19-9) for monitoring and management of pancreatic cancer | | Panel: | 82, Immunology | {4}------------------------------------------------ Image /page/4/Picture/1 description: The image contains the logo for Fujirebio Diagnostics, Inc. The logo consists of a blue and white abstract symbol on the left, followed by the company name in blue text. The word "FUJIREBIO" is in a larger, bolder font, and "Diagnostics, Inc." is in a smaller font below it. # H. INTENDED USE / INDICATIONS FOR USE #### 1. Warning Statements The concentration of CA 19-9 in a given specimen, as determined by assays from different manufacturers, can vary due to differences in assay methods and reagent specificity. The results reported by the laboratory to the physician must include the identity of the assay for CA 19-9 used. Values obtained with different assay methods cannot be used interchangeably. If, in the course of monitoring a patient, the assay method used for determining serial levels of CA 19-9 is changed, the laboratory must perform additional serial testing to confirm baseline values. Prior to changing assays, the laboratory MUST confirm baseline values for patients being serially monitored. Lumipulse G CA19-9-N should not be used for cancer screening or diagnosis. Patients known to be genotypically negative for Lewis blood group antigen are unable to produce the CA 19-9 antigen even in the presence of malignant tissue. Phenotyping for the presence of the Lewis blood group antigen may be insufficient to detect true Lewis antigen negative individuals. Even patients who are genotype positive for the Lewis antigen may produce varying levels of CA 19-9 as the result of gene dosage effect. pancreas. - 2. Intended Use / Indications for use Lumipulse® G CA19-9-N is a Chemiluminescent Enzyme Immunoassay (CLEIA) for the quantitative measurement of CA 19-9 in human serum or plasma (sodium heparin, lithium heparin, or dipotassium EDTA) on the LUMIPULSE G System. The assay is to be used as an aid in the management of patients diagnosed with cancer of the exocrine pancreas who have detectable levels of CA 19-9 at some point in their disease process. Serial testing for patient CA 19-9 assay values should be used in conjunction with other clinical methods used for monitoring cancer of the exocrine pancreas. - 3. Special instrument requirements: Lumipulse G System #### DEVICE DESCRIPTION . #### Reagents The Lumipulse G CA19-9-N Immunoreaction Cartridges (cat. # 235126) consists of 3 x 14 tests. Each kit contains the following: - 1) Antibody-Coated Particle Solution (Liquid when used, 250 µL/Immunoreaction Cartridge) Contains 150 µq/mL anti-CA19-9 monoclonal antibody (mouse)-coated particles, protein stabilizers (bovine and mouse) and chemical stabilizers in 0.15 M sodium chloride/Tris {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for FUJIREBIO Diagnostics, Inc. The logo features a stylized, abstract shape in blue and green on the left, followed by the company name in blue, bold letters. Below the company name, the words "Diagnostics, Inc." are written in a smaller, black font. buffer. This solution contains gelatin and turns into gel at 15 ℃ or lower. Preservative: sodium azide. - 2) Enzyme-Labeled Antibody Solution (Liquid, 350 µL/Immunoreaction Cartridge) Contains 0.5 µg/mL alkaline phosphatase (ALP: calf)-labeled anti-CA19-9 monoclonal antibody (mouse), protein stabilizers (bovine, calf, and mouse) and chemical stabilizers in 0.05 M sodium chloride/Bis-Tris buffer. Preservative: sodium azide. # J. DESCRIPTION OF MODIFIED DEVICE - 1. Modified device: Tamoxifen was tested at 6.0 mg/dL. Tamoxifen test concentration was originally tested at 0.228 mg/dL in the cleared device under K191973, but now it has been tested at 6.0 mg/dL for the modified device. | | Lumipulse® G CA19-9-N Cleared<br>under K191973 | Lumipulse® G CA19-9-N Modified<br>Device | |--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------| | SIMILARITIES | | | | Assay Intended Use | Lumipulse® G CA19-9-N is a<br>Chemiluminescent Enzyme<br>Immunoassay (CLEIA) for the<br>quantitative measurement of CA<br>19-9 in human serum or plasma<br>(sodium heparin, lithium heparin,<br>or dipotassium EDTA) on the<br>LUMIPULSE G System.<br><br>The assay is to be used as an aid<br>in the management of patients<br>diagnosed with cancer of the<br>exocrine pancreas who have<br>detectable levels of CA 19-9 at<br>some point in their disease<br>process. Serial testing for patient<br>CA19-9 assay values should be<br>used in conjunction with other<br>clinical methods used for<br>monitoring cancer of the exocrine<br>pancreas. | Same | | Product Code | NIG | Same | | Regulation | 21 CFR 866.6010 | Same | | Antibody Type and Source | monoclonal antibody (mouse) | Same | | Sample Type | human serum and plasma | Same | | Reagent Storage | Store at 2-10°C | Same | | Sample Size | 100 µL | Same | | Calibration Frequency | Every 30 days | Same | | Instrument | Lumipulse G System | Same | | Methodology | CLEIA | Same | | Measuring Range | 0.7-500 U/mL | Same | ## 2. Modified Device Comparison to the Cleared Device: {6}------------------------------------------------ Image /page/6/Picture/1 description: The image contains the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic on the left, featuring blue and white shapes, followed by the company name in blue, bold letters. Below the company name, the words "Diagnostics, Inc." are written in a smaller, lighter font. | | Lumipulse® G CA19-9-N Cleared<br>under K191973 | Lumipulse® G CA19-9-N Modified<br>Device | |----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------| | Capture | Monoclonal anti-CA19-9 antibody<br>(mouse)-coated particles | Same | | Conjugate Antibody | ALP (calf)-labeled anti-CA19-9<br>monoclonal mouse antibody | Same | | Calibrators | 2 level set (1 vial/level):<br>• Cal 1: 0 U/mL<br>• Cal 2: 500 U/mL | Same | | Calibration Range | 0-500 U/mL | Same | | Controls | 2 levels every 24 hours | Same | | DIFFERENCES | | | | Tamoxifen Test Concentration | Tamoxifen test concentration was<br>originally tested at <b>0.228 mg/dL</b><br>according to the study protocol;<br>(however in the 510(k), a<br>typographical error occurred, and it<br>was listed as 6.0 mg/dL.) | Tamoxifen test concentration is tested<br>at <b>6.0 mg/dL</b> per a new protocol and<br>study report. | | Samples<br>(Mean Test Concentrations<br>(U/mL) (n=3)) | Serum Pool 1 = 30.1<br>Serum Pool 2 = 189.5<br>Serum Pool 3 = 330.1 | Serum Pool 1 = 31.2<br>Serum Pool 2 = 184.4<br>Serum Pool 3 = 327.8 | | Samples<br>(Mean Control Concentrations<br>(U/mL) (n=3)) | Serum Pool 1 = 30.1<br>Serum Pool 2 = 187.5<br>Serum Pool 3 = 322.4 | Serum Pool 1 = 31.3<br>Serum Pool 2 = 187.9<br>Serum Pool 3 = 324.0 | | Test Concentration | 0.228 mg/dL | 6.0 mg/dL | | % Difference Results | 0% to 2% | -2% to 1% | # K. STANDARDS/GUIDANCE DOCUMENTS REFERENCED - . CLSI EP07 - Interference Testing in Clinical Chemistry; Approved Guideline - Third Edition. 2018 - CLSI EP37 Supplemental Tables for Interference Testing in Clinical Chemistry; . Approved Guideline - First Edition. 2018 # L. TEST PRINICIPLE Lumipulse® G CA19-9-N is an assay system, including a set of immunoassay reagents, for the quantitative measurement of CA 19-9 in human serum and plasma (sodium heparin, lithium heparin, or dipotassium EDTA) based on CLEIA technology by a two-step immunoassay method on the LUMIPULSE G1200 System. CA 19-9 in specimens specifically binds to anti-CA19-9 monoclonal antibody (mouse) on the particles, and antigenantibody immunocomplexes are formed. The particles are washed and rinsed to remove unbound materials. Alkaline phosphatase (ALP: calf)-labeled anti-CA19-9 monoclonal antibody (mouse) specifically binds to CA 19-9 of the immunocomplexes on the particles, and additional immunocomplexes are formed. The particles are washed and rinsed to {7}------------------------------------------------ Image /page/7/Picture/1 description: The image contains the logo for FUJIREBIO Diagnostics, Inc. The logo consists of a stylized graphic on the left, resembling a water droplet or a stylized flower, with blue and green vertical lines. To the right of the graphic is the company name, "FUJIREBIO" in a bold, blue sans-serif font, with "Diagnostics, Inc." in a smaller, black sans-serif font below it. remove unbound materials. Substrate Solution is added and mixed with the particles. AMPPD contained in the Substrate Solution is dephosphorylated by the catalysis of ALP indirectly coniugated to particles. Luminescence (at a maximum wavelength of 477 nm) is generate by the cleavage reaction of dephosphorylated AMPPD. The luminescent signal reflects the amount of CA 19-9. Indicated for the serial measurement of CA 19-9 to aid in the management of patients diagnosed with cancers of the pancreas. Serial testing with Lumipulse® G CA19-9-N for CA 19-9 patient values is used in conjunction with other clinical methods in the management of pancreatic cancer patients. # M. SUMMARY OF DESIGN CONTROL ACTIVITIES ### Design Control Activities Summary for Test Concentration Change for Therapeutic InterferentTamoxifen | Device<br>Change | Risks | Verification/<br>Validation Method(s) | Acceptance<br>Criteria | Summary of<br>Results | |------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------| | Tamoxifen is<br>tested at<br>concentration<br>of 6.0 mg/dL | There is no risk to<br>this device<br>change. The test<br>concentration to<br>tamoxifen is<br>being updated to<br>ensure the<br>concentration<br>listed in our PI is<br>appropriate for<br>the intended use<br>population. | Supplemental Therapeutic<br>Interference Study performed for<br>tamoxifen. (protocol and<br>acceptance criteria were the same<br>as found in K191973 019_Section<br>18_Lumipulse G CA19-9-<br>N_Bench, without any deviations)<br>The Supplemental Therapeutic<br>Interference Study was run to<br>mitigate any risk because it was<br>performed with a higher<br>concentration of Tamoxifen than<br>was originally reported in the<br>510(k) labeling. | Individual<br>samples must<br>have a percent<br>difference of ±<br>10% difference<br>from the<br>control. | -2% to 1%<br>(Met acceptance of<br>±10% difference for<br>each individual<br>sample between the<br>test and control<br>samples.) | # N. ADDITIONAL UPDATE In the original 510(k) Summary and labeling for K191973, there was a typo for Streptozotocin. It was listed as 3.96 mg/dL but should have been 28 mg/dL. This update has been made to the 510(k) Summary and labeling and is included in this submission. # O. DECLARATION OF CONFORMITY (DOC) Please see K191973, 010_Section 9_Lumipulse G CA19-9-N 510k_Decl of Conformity for the declaration of conformity.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...