K200931 · Fulwell, LLC · GEI · Nov 24, 2020 · General, Plastic Surgery
Device Facts
Record ID
K200931
Device Name
RF Surgical Generator
Applicant
Fulwell, LLC
Product Code
GEI · General, Plastic Surgery
Decision Date
Nov 24, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
RF Surgical Generator FW-120A is an electrosurgical generator containing monopolar and bipolar technology. It is intended for use with accessories during surgical procedures where the surgeon requires electrosurgical cutting and coagulating.
Device Story
RF Surgical Generator FW-120A provides high-power RF energy for cutting and coagulating human tissue. System includes generator, IEC neutral plate, and two-pedal footswitch. Operator selects waveforms and power levels via front panel push buttons; digital displays provide feedback. Device supports monopolar (via handpiece and neutral plate) and bipolar (via bipolar forceps) modes. Safety features include automatic system self-check at power-on, continuous monitoring during operation, and interlocked controls to prevent operation during faults (e.g., accessory failure, overheating, high contact resistance). Used in clinical settings by surgeons. Output power controlled by footswitch or hand switches. Benefits include precise tissue cutting and hemostasis during surgery.
Clinical Evidence
Bench testing only. Evaluated for electrical safety (ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-2) and mechanical/functional capabilities. Thermal damage testing performed on porcine muscle, liver, and kidney tissue to measure thermal zone dimensions. System performance and waveform output tests confirmed equivalence to predicate.
Technological Characteristics
RF electrosurgical generator; monopolar and bipolar output. 5 modes: Cutting (4.0MHz), Blend (4.0MHz), Hemostasis (4.0MHz), Fulguration (4.0MHz), Bipolar (1.7MHz). Duty cycle 10s on/30s off. Class I electric shock protection; Type BF defibrillation protection. Connectivity via footswitch/hand switches. Non-sterile. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2.
Indications for Use
Indicated for use in surgical procedures requiring electrosurgical cutting and coagulating. Applicable to patients undergoing general surgical procedures where electrosurgery is required.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA" being larger and bolder than the rest of the text.
November 24, 2020
Fulwell LLC % Jet Li Regulation Manager Guangzhou KEDA Biological Tech Co., Ltd. 6F, No.1 TianTai road, Science City, LuoGang District Guangzhou, Guangdong China
Re: K200931
Trade/Device Name: RF Surgical Generator Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: September 23, 2020 Received: September 28, 2020
Dear Jet Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200931
Device Name RF Surgical Generator, Model: FW-120A
Indications for Use (Describe)
RF Surgical Generator FW-120A is an electrosurgical generator containing monopolar and bipolar technology. It is intended for use with accessories during surgical procedures where the surgeon requires electrosurgical cutting and coagulating.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial, sans-serif;">□</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary (K200931)
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 878.4400, and there were no prior submissions for the subject device.
#### 1 Submitter Information
Sponsor: Fulwell LLC Address: 3411 NW 48 Street, Miami, FL 33142 USA Contact Person: Mrs. Perla Melisa Mcliberty Title: Deputy General Manager Phone: +1-305-535-3568 E-mail: register@fulwell.us.com
Application Correspondent: Jet Li Company: Guangzhou KEDA Biological Technology Co., Ltd E-mail: med-jl@foxmail.com Phone: 86-18588874857 Address: 6F, No.1 TianTai road, Science City, LuoGang District, GuangZhou City, China
#### Subject Device Information 2
Type of 510(k) submission: Traditional Trade/Device Name: RF Surgical Generator Model: FW-120A Classification Name: Electrosurgical, Cutting & Coagulation & Accessories Review Panel: General & Plastic Surgery Product Code: GEI Regulation Number: 21 CFR 878.4400 Regulation Class: 2
#### 3 Predicate Device Information
Sponsor: ELLMAN INTL., INC. Trade/Device Name: Surgitron 120 IEC 510K number: K013255 Classification Name: Electrosurgical, Cutting & Coagulation & Accessories Review Panel: General & Plastic Surgery Product Code: GEI Regulation Number: 21 CFR 878.4400 Requlation Class: 2
{5}------------------------------------------------
#### Device Description 4
RF Surgical Generator FW-120A, is a compact source of high power RF energy, enhanced capability of radiosurgery generator, which can be employed for a variety of radiosurgery procedures, such as cut and coagulate living human tissue.
The device which is consisted of a generator, an IEC Neutral Plate, and a Two Pedal Footswitch, provides 5 modes and corresponding output power ranges that is user/operator selectable based on the surgical procedure undertaken. The action is achieved by front panel selection of waveforms and power level. All selection is affected through push buttons and indicators which give the operator feedback of status. Power level for each mode is indicated by front panel digital displays
The device is provided with safety features of automatically system initialize self-check after power on and continuously monitored during the device running. The device will provide error information and/ or alerts when detect following faults, such as a) Accessary fail connection, b) Overheating of Main Unit, c) Neutral plates Contact resistance exceed the limit, d) System initialize error. The error monitoring is interlocked with the controls to prevent operation when fault condition happened in the unit.
The generator is configured for Monopolar and Biopolar output and the final output power could be controlled by pedal foot switch and/or hand switches. Recommended Monopolar Handpieces 510(k) No.: K092634. Recommended IEC Neutral Plate 510(k) No.: K102372.
#### 5 Intended Use/Indication for use
RF Surgical Generator FW-120A is an electrosurgical generator containing monopolar and bipolar technology. It is intended for use with accessories during surgical procedures where the surgeon requires electrosurgical cutting and coagulating.
#### 6 Performance data
Verification testing, including electrical safety/electromagnetic compatibility, software verification/validation were performed for RF Surgical Generator.
- RF Surgical Generator has been evaluated the safety and performance by lab bench testing a) according to the following standards:
- ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009(R)2012 and A A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)
- △ IEC 60601-1-2, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests, 2014
- > IEC 60601-2-2, Medical Electrical Equipment - Part 2-2: Particular Requirements for The Basic Safety And Essential Performance Of High Frequency Surgical Equipment And
{6}------------------------------------------------
High Frequency Surgical Accessories, Edition 6.0 2017-03
- b) RF Surgical Generator has been evaluated the basic mechanical and functional capabilities to determine substantial equivalence to the predicate device. According to the guidance document Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery, the device has been tested for applicable items such as thermal damage, CQM and so on.
In particular, tests have been carried out with respect to the following subject areas.
- 1) System Performance and waveform outputs test.
2) Thermal effects testing: the thermal damage of the tissue due to the HF current was measured in terms of size (length, width and depth) of the thermal zone in all applicable modes: porcine muscle, liver, and kidney.
- c) RF Surgical Generator has performed software verifications to ensure the device worked appropriately as a moderated concern level software according to the Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. .
#### 7 Comparison to Predicate Device
Compare with predicate device, the subject device is very similar in design principle, intended use, indications for use, functions, material and the applicable standards. The differences between subject device and predicate device do not raise any new questions of safety or effectiveness.
| Elements of<br>comparison | Subject Device | Predicate Device | Verdict |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Fulwell LLC | ELLMAN INTL., INC. | -- |
| 510K number | K200931 | K013255 | -- |
| Product Name | RF Surgical Generator<br>FW-120A | Surgitron 120 IEC<br>Also known as Surgitron 4.0 Dual RF | -- |
| Classification<br>Name | Electrosurgical, Cutting & Coagulation<br>& Accessories | Electrosurgical, Cutting & Coagulation<br>& Accessories | SE |
| Regulation<br>Class | 2 | 2 | SE |
| Regulation<br>Number | 21 CFR 878.4400 | 21 CFR 878.4400 | SE |
| OTC & Rx | Rx | Rx | SE |
| Indications for Use | | | |
| Indications for<br>Use | RF Surgical Generator FW-<br>120A is an electrosurgical<br>generator containing<br>monopolar and bipolar<br>technology. It is intended for<br>use with accessories during<br>surgical procedures where the<br>surgeon requires<br>electrosurgical cutting and<br>coagulating. | Cutting: snoring, submucosal palatal<br>shrinkage, traditional uvulopalatoplasty<br>(RAUP), myringotomy with effective<br>hemorrhage control, epistaxis<br>treatment, turbinate shrinkage, skin<br>incisions, biopsy, cysts, abscesses,<br>tumors, cosmetic repairs, development<br>of skin flaps, skin tags and<br>blepharoplasty.<br>Blend cutting and coagulation: snoring,<br>submucosal palatal shrinkage | SE,<br>Minor<br>Difference.<br>We remove<br>the specified<br>indication<br>for use for<br>subject<br>device; it do<br>not raise<br>safety and |
| | | | |
| | | traditional uvulopalatoplasty (RAUP),<br>myringotomy with effective hemorrhage<br>control, epistaxis treatment, turbinate<br>shrinkage, skin tags, papilloma keloids,<br>keratosis, verrucae, basal cell<br>carcinoma, nevi, fistulas, epithelioma,<br>cosmetic repairs, cysts, abscesses,<br>development of skin flaps.<br>Hemostasis: control of bleeding,<br>epilation, telangiectasia.<br>Fulguration: basal cell carcinoma,<br>papilloma, cyst destruction, tumors,<br>verrucae, hemostasis.<br>Bipolar: pinpoint precise coagulation,<br>pinpoint hemostasis in any field (wet or<br>dry), snoring, submucosal palatal<br>shrinkage, traditional uvulopalatoplasty<br>(RAUP), myringotomy with effective<br>hemorrhage control, epistaxis treatment<br>and turbinate shrinkage. | effectivenes<br>s issue. |
| | Device Design | | |
| Working<br>Theory | ■ Monopolar: RF generator connects<br>its accessories of a handpiece and an<br>IEC Neutral Plate (adhered to patient<br>skin) to form a cyclic circuit, the RF<br>current generated from the generator<br>and through the Monopolar handpiece<br>to achieve CUT or COAG, then return<br>to generator by the IEC Neutral Plate.<br>■ Bipolar: RF current generated from<br>the generator and the cyclic circuit<br>formed between the two tips of the<br>bipolar forceps, the RF power through<br>the two tips to work on patient obtaining<br>COAG. | ■ Monopolar: RF generator connects<br>its accessories of a handpiece and an<br>IEC Neutral Plate (adhered to patient<br>skin) to form a cyclic circuit, the RF<br>current generated from the generator<br>and through the Monopolar handpiece<br>to achieve CUT or COAG, then return<br>to generator by the IEC Neutral Plate.<br>■ Bipolar: RF current generated from<br>the generator and the cyclic circuit<br>formed between the two tips of the<br>bipolar forceps, the RF power through<br>the two tips to work on patient obtaining<br>COAG. | SE |
| Appearance | Image: FW-120A | Image: Ellman Surgitron Dual 11 | SE<br>Note 1 |
| Interface<br>Accessories | Two Pedal Footswitch, Handpieces, IEC<br>Neutral Plate, Power Cord | Triple Footswith, Handpieces, Neutral<br>Pad, Power Cord | SE<br>Note 1 |
| Operation<br>Mode | Cutting, Blend cutting and coagulation,<br>Hemostasis, Fulguration, Bipolar | Cutting, Blend cutting and coagulation,<br>Hemostasis, Fulguration, Bipolar | SE |
| Waveforms | Cutting:4.0MHz Sinusoid<br>Blend cutting and coagulation: 4.0MHz<br>with fully rectified envelope<br>Hemostasis: 4.0MHz with partially<br>rectified envelope<br>Fulguration: 4.0MHz with modulated<br>Bipolar: 1.7MHz | Cutting:4.0MHz CW Sin-Wave<br>Blend cutting and coagulation: 4.0MHz<br>with rectified full-wave envelope<br>Hemostasis: 4.0MHz with square wave<br>rectified envelope<br>Fulguration: 4.0MHz Sin-Wave<br>Modulated<br>Bipolar: 1.7MHz for Fulgurating Spark-<br>Gap | SE<br>Note 2 |
| Activation | Cutting: via footswitch or finger switch<br>Blend cutting and coagulation: via<br>footswitch or finger switch<br>Hemostasis: via footswitch or finger<br>switch<br>Fulguration: via footswitch or finger<br>switch | Cutting: via footswitch or finger switch<br>Blend cutting and coagulation: via<br>footswitch or finger switch<br>Hemostasis: via footswitch or finger<br>switch<br>Fulguration: via footswitch or finger<br>switch | SE |
{7}------------------------------------------------
{8}------------------------------------------------
| K200931 | |
|---------|--|
|---------|--|
| Rated Duty<br>Cycle | 10s on, 30s off | 10s on, 30s off | SE |
|-----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Energy type | RF | RF | SE |
| Feedback<br>system | Indicators, audio alarm, Error codes | Indicators, audio alarm, Error codes | SE |
| Neutral<br>Electrode<br>monitor<br>resistance | MAX 1000Ω | MAX 1000Ω | SE |
| Input Pow er | Nominal line voltage and frequency:<br>120Vac~ ±10%, 60 Hz.<br>/230Vac~ ±10%, 50 Hz.<br>Input current at max output pow er:<br>3.3/1.5A. | Voltage: 240/220/120/100V, ±10%<br>240/230/120/100V, ±10%<br>Frequency: 50Hz<br>Current: For 240~220V: 3.3A,<br>For 120~100V: 1.5A | SE<br>Note 1 |
| Output Pow er | Cutting: 120W@500Ω<br>Blend cutting and coagulation:<br>80W@500Ω<br>Hemostasis: 60W@500Ω<br>Fulguration: 40W@500Ω<br>Bipolar: 120W@200Ω | Cutting: 120W@500Ω<br>Blend cutting and coagulation:<br>80W@500Ω<br>Hemostasis: 40W@500Ω<br>Fulguration: 40W@500Ω<br>Bipolar: 120W@200Ω | SE<br>Note 2 |
| Fuses | T 2.0A, 250V, Tw o pieces for 230Vac<br>T 4.0A, 250V, Two pieces for 120Va | 2* T 2.0AL, for 250V<br>2* T 4.0AL, for 110V | SE<br>Note 1 |
| Electric shock<br>protection | Class I | Class Ilb | SE<br>Note 1 |
| Defibrillation<br>protection | Type BF | Type BF | SE |
| Dimensions | 235mm*120mm*340mm(W*H*D) | 9"*5"*13" (W*H*D)<br>≈228mm*127mm*330mm | SE<br>Note 1 |
| Weight | 11Kg | 18lbs, ≈8.16kg | SE<br>Note 1 |
| Working<br>condition | -10°C- +40°C<br>30% - 75% | -10°C- +40°C<br>30% - 75% | SE |
| Transport and<br>Storage | -10°C- +50°C, 10% - 95%<br>500hPa-1060hPA | -10°C- +50°C, 10% - 95%<br>500hPa-1060hPA | SE |
| Sterility | Non-sterile | Non-sterile | SE |
| FDA-Recognized Standards | | | |
| Safety<br>Performance | ANSI AAMI ES60601-1<br>IEC 60601-2-2 | UL2601, IEC 601-1, IEC 601-2-2<br>BS15724:2.2 | SE<br>Note 2 |
| EMC<br>Performance | IEC 60601-1-2 | / | SE<br>Note 3 |
## Note 1
Although the subject device design and specification parameters, such as appearance / dimensions / weight / power supply / electric shock protection, have a little difference to the predicate device, these minor differences do not affect the device' mainly function, which are has been testing the electrical safety via the IEC 60601 standards tests, and these differences will not create any new safety risks to the performance and safety.
## Note 2
Although the waveform and output power of subject device has a little different to predicate device, the performance test has proved that the performance of subject device is substantial equivalence
{9}------------------------------------------------
to the predicate device. So the differences do not affect the safety and effectiveness.
## Note 3
Although the complied standards of electrical safety and EMC performance are different to the predicate device, it's only because the version of predicate device's complied standards are invalidated and been updated. The subject device has complied to the necessary standards as existing similar products, so the differences do not affect the safety and effectiveness.
#### 9 Conclusion
The subject device RF Surgical Generator has all features of the predicate device for intended use. Thus, the subject device is substantially equivalent to the predicate device.
#### 10 Summary Prepared Date
22 Nov. 2020
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.