← Product Code [QJP](/productcode/QJP) · K200910

# BOSS Balloon Guide Catheter (K200910)

_Marblehead Medical · QJP · May 2, 2020 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K200910

## Device Facts

- **Applicant:** Marblehead Medical
- **Product Code:** [QJP](/productcode/QJP.md)
- **Decision Date:** May 2, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The BOSS Balloon Guide Catheter is indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neurovascular systems. The balloon provides temporary vascular occlusion during these procedures. The Balloon Guide Catheter is also indicated for use as a conduit for retrieval devices.

## Device Story

Sterile, single-use intravascular catheter (90 cm length); variable stiffness shaft with stainless steel coil and braid reinforcement; features bifurcated dual-port luer hub (central port for device introduction, inflation port for balloon) and distal radiopaque marker band. Distal hydrophilic coating reduces friction. Used in peripheral and neurovascular procedures; operated by physicians. Device inserted over guidewire (or via direct access with included 6F dilator) into blood vessel. Balloon inflated/deflated via syringe to provide temporary vascular occlusion. Acts as conduit for interventional/retrieval devices. Output is physical access and vessel occlusion; enables physician to perform neurovascular/peripheral interventions. Benefits include improved device delivery and procedural control.

## Clinical Evidence

No clinical studies performed. Substantial equivalence supported by non-clinical bench testing (physical integrity, functionality, performance) and acute porcine animal model study. Bench testing included ISO 10555-1/4 compliance, biocompatibility (ISO 10993), and simulated use/usability testing.

## Technological Characteristics

Materials: medical grade plastics, stainless steel, polyurethane balloon. Construction: coaxial lumen, stainless steel coil/braid reinforcement, hydrophilic coating. Dimensions: 90 cm length, 3.1mm (9.4Fr) max OD. Connectivity: N/A. Sterilization: 100% Ethylene Oxide (EtO) to SAL 10^-6. Software: None.

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- FlowGate2™ Balloon Guide Catheter ([K153729](/device/K153729.md))
- Neuron™ MAX System ([K111380](/device/K111380.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 2, 2020

Regulatory Technology Services LLC Prithul Bom Responsible Third Party Official 1000 Westgate Drive, Suite #510k Saint Paul, Minnesota 55114

Re: K200910

Device Name: BOSS™ Balloon Guide Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: OJP, DOY Dated: April 4, 2020 Received: April 6, 2020

Dear Prithul Bom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Xiaolin Zheng, Ph.D., M.S. Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K200910

Device Name BOSS Balloon Guide Catheter

#### Indications for Use (Describe)

The BOSS Balloon Guide Catheter is indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neurovascular systems. The balloon provides temporary vascular occlusion during these procedures. The Balloon Guide Catheter is also indicated for use as a conduit for retrieval devices.

| Type of Use (Select one or both, as applicable)            |                                                           |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# BOSS™ Balloon Guide Catheter 510(k) Summary

#### SUBMITTER [807.92(a)(1)]

| Submitter's Name:           | Marblehead Medical, LLC                       |
|-----------------------------|-----------------------------------------------|
| Address:                    | 11600 96th Ave N Maple Grove, MN 55369        |
| Contact Person:             | Kristin Mortenson                             |
| Telephone:                  | (612) 202-1142                                |
| email                       | Mortenson.Kristin@MarbleheadMedical.com       |
| Date Prepared:              | April 24, 2020                                |
| DEVICE [807.92(a)(2)]       |                                               |
| Device Trade Name:          | BOSS™ Balloon Guide Catheter                  |
| Common Name:                | Balloon Guide Catheter                        |
| Classification Name:        | Percutaneous Catheter                         |
| Product Code:               | QJP, DQY                                      |
| Regulatory Class:           | Class II                                      |
| Regulation Number:          | 21 CFR 870.1250                               |
| PREDICATE [807.92(a)(3)]    |                                               |
| Primary Predicate Device:   | 8F FlowGate2™ Balloon Guide Catheter, K153729 |
| Secondary Predicate Device: | Neuron™ MAX System, K111380                   |

## DEVICE DESCRIPTION [807.92(a)(4)]

The BOSS Balloon Guide Catheter system is a sterile, single-use intravascular catheter. The BOSS Balloon Guide Catheter is a 90 cm long, variable stiffness catheter utilizing a bifurcated dual port luer hub on the proximal end and a radiopaque marker band at the tip distal to the balloon. The catheter shaft has an annular inflation lumen and a coaxial central lumen with stainless steel coil and is braid reinforced. The hub central port leads to the central lumen to facilitate introduction of interventional devices through the central lumen. The inflation port is positioned at an angle to the central port, connecting to the annular inflation lumen, and is used to facilitate inflating and deflating the balloon with a syringe. The BOSS Balloon Guide Catheter uses a distal hydrophilic coating to provide lubricity and reduce friction between the catheter shaft and the vessel wall. The BOSS Balloon Guide Catheter is packaged with the 6F dilator to facilitate the option of direct access into the blood vessel without the use of an introducer sheath.

### INDICATION FOR USE [807.92(a)(5)]

The BOSS Balloon Guide Catheter is indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neurovascular

{4}------------------------------------------------

systems. The balloon provides temporary vascular occlusion during these procedures. The Balloon Guide Catheter is also indicated for use as a conduit for retrieval devices.

# PREDICATE DEVICE COMPARISON [807.92(a)(6)]

The following table provides a comparison of the key characteristics of the BOSS Balloon Guide Catheter System to the predicate devices.

|                                         | Primary Predicate device                                                                                                                                                                                                                                                                                                                                                                                                                      | Secondary Predicate device                                                                                                               | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature                                 | FlowGate2TM Balloon Guide Catheter (K153720)                                                                                                                                                                                                                                                                                                                                                                                                  | NeuronTM MAX System (K111380)                                                                                                            | BOSS Balloon Guide Catheter System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| FDA Classification                      | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                      | Class II                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Product Code(s)                         | DQY                                                                                                                                                                                                                                                                                                                                                                                                                                           | DQY                                                                                                                                      | QJP, DQY                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Regulation Number                       | 870.1250                                                                                                                                                                                                                                                                                                                                                                                                                                      | 870.1250                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Regulation Name                         | Percutaneous Catheter                                                                                                                                                                                                                                                                                                                                                                                                                         | Percutaneous Catheter                                                                                                                    | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Indications for Use                     | FlowGateTM Balloon Guide Catheters are indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for retrieval devices.                                        | The NeuronTM MAX System is indicated for the introduction of interventional devices into the peripheral, coronary and neuro vasculature. | The primary predicate and subject device have the same indications for use, except the primary predicate includes "and other angiographic" and the subject device does not include angiographic procedures. The primary predicate and subject device do not include coronary vasculature like the secondary predicate. The secondary predicate is included to support the direct vascular access and the hydrophilic coating. The secondary predicate does not include a balloon, therefore does not include temporary vascular occlusion. In addition, the secondary predicate does not indicate use as conduit for retrieval devices, a more specific use within the intended use. |
|                                         | Primary Predicate device                                                                                                                                                                                                                                                                                                                                                                                                                      | Secondary Predicate device                                                                                                               | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Feature                                 | FlowGate2TM Balloon Guide Catheter (K153720)                                                                                                                                                                                                                                                                                                                                                                                                  | NeuronTM MAX System (K111380)                                                                                                            | BOSS Balloon Guide Catheter System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                          | These differences in wording between the subject device and the predicates are not critical to the intended therapeutic or surgical use of the device, the introduction and guidance of intravascular devices into selected vasculature. The differences in wording do not affect the safety and effectiveness of the device when the device is used as labeled.                                                                                                                                                                                                                                                                                                                     |
| Anatomical Locations                    | Peripheral and neuro vasculature                                                                                                                                                                                                                                                                                                                                                                                                              | Peripheral, coronary and neuro vascular systems                                                                                          | Same as primary predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Principle of Operation                  | The Balloon Guide Catheter is used to facilitate the selective placement of interventional devices. After needle access and insertion of up to a 0.038" guidewire into the femoral artery, the device is inserted into the artery through an introducer sheath over the guidewire. The lumen of device is used for insertion and guidance of an intravascular catheter and / or retrieval devices into a selected blood vessel for treatment. | Same as primary predicate except includes the direct access use option without an introducer sheath                                      | Same as secondary predicate. Subject device includes a direct access use option whereas the primary predicate does not.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Material                                | Commonly used medical grade plastics and stainless steel                                                                                                                                                                                                                                                                                                                                                                                      | Not listed                                                                                                                               | Same as primary predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Reinforced Catheter Shaft Reinforcement | Stainless Steel coil and braid reinforced                                                                                                                                                                                                                                                                                                                                                                                                     | Stainless steel braid reinforced                                                                                                         | Same as primary predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Marker Band /                           | 0.02-in (0.5 mm) wide                                                                                                                                                                                                                                                                                                                                                                                                                         | Radiopaque marker is                                                                                                                     | 0.08-in (2 mm) wide                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                                         | Primary Predicate device                                                                                                                                                                                                                                                                                                                                                                                                                      | Secondary Predicate<br>device                                                                                                            | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Feature                                 | FlowGate2™ Balloon<br>Guide Catheter<br>(K153720)                                                                                                                                                                                                                                                                                                                                                                                             | Neuron™ MAX System<br>(K111380)                                                                                                          | BOSS Balloon Guide<br>Catheter System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Location                                | 0.03-in (0.75 mm) from the<br>distal tip edge                                                                                                                                                                                                                                                                                                                                                                                                 | included on the distal<br>end, width and<br>location not listed                                                                          | 0.02-in (0.5mm) from the<br>distal tip edge distal to<br>the balloon. Similar,<br>difference does not raise<br>new questions regarding<br>safety and efficacy, all<br>devices utilize<br>radiopaque marker<br>bands                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Compliant<br>Balloon                    | Yes, silicone                                                                                                                                                                                                                                                                                                                                                                                                                                 | N/A                                                                                                                                      | Yes, polyurethane.<br>Similar to primary<br>predicate, material<br>differences do not raise<br>new questions of safety<br>and efficacy, both<br>materials are used for<br>intravascular use.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Labeled Shaft<br>Outer<br>Diameter      | 0.108-in (max)<br>(2.7mm) 8Fr                                                                                                                                                                                                                                                                                                                                                                                                                 | 0.110-in (max)<br>(2.8mm) 8F                                                                                                             | 0.123-in (max)<br>(3.1mm) 9.4Fr<br>Similar, differences do<br>not raise new questions<br>of safety and efficacy.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Labeled Shaft<br>Inner<br>Dimension     | 0.084-in (2.21mm) 6.4Fr                                                                                                                                                                                                                                                                                                                                                                                                                       | 0.088-in (2.24 mm)<br>6.7F                                                                                                               | Same as secondary<br>predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Effective<br>length                     | 85 and 95 cm                                                                                                                                                                                                                                                                                                                                                                                                                                  | 80 and 90 cm                                                                                                                             | 90 cm, same as<br>secondary predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Tip Shape                               | Straight                                                                                                                                                                                                                                                                                                                                                                                                                                      | Straight                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Maximum<br>Balloon<br>Volume            | 0.6ml                                                                                                                                                                                                                                                                                                                                                                                                                                         | N/A                                                                                                                                      | Same as primary<br>predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Radiopaque                              | Distal Tip has radiopaque<br>marker bands, stainless<br>steel reinforcement in the<br>catheter shaft renders the<br>shaft visible on<br>fluoroscopy                                                                                                                                                                                                                                                                                           | Distal Tip has<br>radiopaque zone                                                                                                        | Same as primary<br>predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Coating                                 | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                           | Hydrophilic Coating –<br>distal shaft                                                                                                    | Same as the secondary<br>predicate. Coating is on                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Feature                                 | Primary Predicate device                                                                                                                                                                                                                                                                                                                                                                                                                      | Secondary Predicate device                                                                                                               | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                         | FlowGate2TM Balloon<br>Guide Catheter<br>(K153720)                                                                                                                                                                                                                                                                                                                                                                                            | NeuronTM MAX System<br>(K111380)                                                                                                         | BOSS Balloon Guide<br>Catheter System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Internal<br>Construction                | Coaxial Lumen                                                                                                                                                                                                                                                                                                                                                                                                                                 | Single lumen                                                                                                                             | the distal portion<br>proximal to the balloon<br>Same as primary<br>predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Supplied items<br>/ accessories         | Dilator, Rotating<br>Hemostasis Valve, Tuohy<br>Borst Valve with Sideport,<br>Peel Away Sheaths, Luer-<br>Activated Valves                                                                                                                                                                                                                                                                                                                    | Dilator, Rotating<br>Hemostasis Valve<br>(RHV), Extension<br>tubing                                                                      | Dilator is provided, same<br>as both predicates. The<br>additional accessories<br>are commercially<br>available devices.<br>Differences do not raise<br>new questions of safety<br>and efficacy.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| How Supplied                            | Sterile, single use                                                                                                                                                                                                                                                                                                                                                                                                                           | Sterile, single use                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Sterilization<br>Method                 | EtO                                                                                                                                                                                                                                                                                                                                                                                                                                           | EtO                                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Sterility<br>Assurance<br>Level         | 10-6                                                                                                                                                                                                                                                                                                                                                                                                                                          | 10-6                                                                                                                                     | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

The differences between the subject device and predicates described in the comparison table above are not critical to the intended therapeutic or surgical use of the device, do not raise questions of safety and effectiveness, and as shown through testing and analysis, do not affect the safety and effectiveness of the device when used as labeled.

## NONCLINICAL PERFORMANCE TESTING SUMMARY [807.92(b)]

Determination of substantial equivalence is based on an assessment of non-clinical performance bench test data. The BOSS Balloon Guide Catheter System has successfully completed the following relevant performance testing to demonstrate substantial equivalence. Testing was performed to evaluate physical integrity, functionality, and performance of the BOSS Balloon Guide Catheter. A summary of the tests performed is provided in the table below:

{8}------------------------------------------------

| Performance Bench Testing Summary                         |                                                                                                                                                                                                                                                |                                                                                                                                                                                |                                                                                |
|-----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| Study Name                                                | Description                                                                                                                                                                                                                                    | Reference Standard                                                                                                                                                             | Results                                                                        |
| Visual<br>Inspection and<br>Dimensional<br>Verification   | To demonstrate that the<br>product meets the<br>dimensional<br>specifications                                                                                                                                                                  | ISO 10555-1:2013<br>Intravascular catheters -<br>Sterile and single-use<br>catheters - Part 1: General<br>requirements                                                         | Pass<br>All samples met the<br>pre-determined<br>acceptance criteria           |
| Surface<br>Inspection                                     | To demonstrate the<br>product satisfies the<br>visual surface<br>requirements                                                                                                                                                                  | ISO 10555-1:2013<br>Intravascular catheters -<br>Sterile and single-use<br>catheters - Part 1: General<br>requirements                                                         | Pass<br>All samples met the<br>pre-determined<br>acceptance criteria           |
| Torque Testing                                            | To demonstrate that the<br>product is capable of<br>360 degrees of rotation<br>about the central lumen<br>axis without failure.                                                                                                                | FDA guidance PTCA<br>Catheters:2010 §VIII.A.10                                                                                                                                 | Pass<br>All samples met<br>the pre-determined<br>acceptance criteria           |
| Tensile                                                   | To demonstrate the<br>product satisfies the<br>tensile strength<br>requirements for bonds<br>and tip pull test                                                                                                                                 | ISO 10555-1:2013<br>Intravascular catheters -<br>Sterile and single-use<br>catheters - Part 1: General<br>requirements, FDA<br>guidance PTCA<br>Catheters:2010 §VIII.A.7,<br>8 | Pass<br>All samples met the<br>pre-determined<br>acceptance criteria           |
| Kink Resistance                                           | To demonstrate that the<br>product has acceptable<br>kink resistance                                                                                                                                                                           | FDA Guidance PTCA:2010<br>§VIII.A.9 Kink Test                                                                                                                                  | Pass<br>All samples met the<br>pre-determined<br>acceptance criteria           |
| Catheter<br>Lubricity                                     | Pad friction test to<br>compare coated to<br>uncoated samples                                                                                                                                                                                  | Characterization only                                                                                                                                                          | Results show a 97%<br>reduction in friction<br>compared to<br>uncoated samples |
| Particulates,<br>Coating<br>Integrity                     | This study was<br>conducted to determine<br>the quantity and size of<br>particles generated<br>during simulated use                                                                                                                            | AAMI TIR42:10 Evaluation<br>of particulates associated<br>with vascular medical<br>devices, USP <788><br>Particulate Matter in<br>Injections                                   | Pass<br>All samples met the<br>pre-determined<br>acceptance criteria           |
| Radiopacity                                               | To determine the<br>radiopaque<br>characteristics of the<br>device.                                                                                                                                                                            | ISO 10555-1:2013<br>Intravascular catheters -<br>Sterile and single-use<br>catheters - Part 1: General<br>requirements                                                         | Pass<br>All samples met the<br>pre-determined<br>acceptance criteria           |
| Performance Bench Testing Summary                         |                                                                                                                                                                                                                                                |                                                                                                                                                                                |                                                                                |
| Study Name                                                | Description                                                                                                                                                                                                                                    | Reference Standard                                                                                                                                                             | Results                                                                        |
| Push/Track,<br>Device<br>Compatibility                    | To demonstrate that the<br>device is deliverable<br>through tortuous path<br>model without kinking<br>or buckling, is able<br>inflate, deflate balloon<br>at target, is compatible<br>with treatment devices<br>and removed without<br>damage. | ISO 10555-1:2013<br>Intravascular catheters -<br>Sterile and single-use<br>catheters - Part 1: General<br>requirements                                                         | Pass<br>All samples met the<br>pre-determined<br>acceptance criteria           |
| System Leak -<br>Liquid Leak                              | To demonstrate that the<br>product meets the liquid<br>leakage under pressure<br>requirements                                                                                                                                                  | ISO 10555-1:2013<br>Intravascular catheters -<br>Sterile and single-use<br>catheters - Part 1: General<br>requirements                                                         | Pass<br>All samples met the<br>pre-determined<br>acceptance criteria           |
| System Leak -<br>Aspiration                               | To demonstrate that the<br>product meets the hub<br>aspiration air leakage<br>requirements                                                                                                                                                     | ISO 10555-1:2013<br>Intravascular catheters -<br>Sterile and single-use<br>catheters - Part 1: General<br>requirements                                                         | Pass<br>All samples met the<br>pre-determined<br>acceptance criteria           |
| Tip Stiffness                                             | To demonstrate that the<br>stiffness of the distal<br>end of the product is<br>similar to other<br>marketed devices.                                                                                                                           | N/A                                                                                                                                                                            | Pass<br>All samples met<br>the pre-determined<br>acceptance criteria           |
| Corrosion<br>Resistance                                   | To demonstrate the<br>product satisfies the<br>corrosion resistance<br>requirements                                                                                                                                                            | ISO 10555-1:2013<br>Intravascular catheters -<br>Sterile and single-use<br>catheters - Part 1: General<br>requirements                                                         | Pass<br>All samples met the<br>pre-determined<br>acceptance criteria           |
| Catheter Burst<br>Pressure Under<br>Static Conditions     | To demonstrate the<br>catheter does not leak or<br>rupture up to 300 psi of<br>internal pressure.                                                                                                                                              | ISO 10555-1:2013<br>Intravascular catheters -<br>Sterile and single-use<br>catheters - Part 1: General<br>requirements                                                         | Pass<br>All samples met the<br>pre-determined<br>acceptance criteria           |
| Balloon Inflation /<br>Deflation                          | To demonstrate balloon<br>meets the inflation and<br>deflation time<br>specifications                                                                                                                                                          | ISO 10555-4:2013<br>Intravascular Catheters -<br>Sterile and Single-Use<br>Catheters - Part 4: Balloon<br>Dilatation Catheters                                                 | Pass<br>All samples met the<br>pre-determined<br>acceptance criteria           |
| Performance Bench Testing Summary                         |                                                                                                                                                                                                                                                |                                                                                                                                                                                |                                                                                |
| Study Name                                                | Description                                                                                                                                                                                                                                    | Reference Standard                                                                                                                                                             | Results                                                                        |
| Balloon Fatigue                                           | To demonstrate that<br>there is no degradation<br>of the balloon after 10<br>inflation cycles.                                                                                                                                                 | ISO 10555-4:2013<br>Intravascular Catheters -<br>Sterile and Single-Use<br>Catheters - Part 4: Balloon<br>Dilatation Catheters                                                 | Pass<br>All samples met the<br>pre-determined<br>acceptance criteria           |
| Balloon<br>Diameter /<br>Volume                           | To demonstrate that the<br>product meets the<br>inflation volume vs<br>balloon diameter<br>specifications                                                                                                                                      | ISO 10555-4:2013<br>Intravascular Catheters -<br>Sterile and Single-Use<br>Catheters - Part 4: Balloon<br>Dilatation Catheters                                                 | Pass<br>All samples met the<br>pre-determined<br>acceptance criteria           |
| Balloon Rated<br>Burst Volume                             | To demonstrate that the<br>balloon is capable of<br>withstanding an<br>injection volume of 1.5<br>ml.                                                                                                                                          | ISO 10555-4:2013<br>Intravascular Catheters -<br>Sterile and Single-Use<br>Catheters - Part 4: Balloon<br>Dilatation Catheters                                                 | Pass<br>All samples met the<br>pre-determined<br>acceptance criteria           |
| Small Bore<br>Connector<br>Compliance<br>with Standard    | To demonstrate that the<br>product meets the<br>requirements for small<br>bore connectors                                                                                                                                                      | ISO 80369-7 2016 Small-<br>bore connectors for<br>liquids and gases in<br>healthcare applications -<br>Part 7, Connectors for<br>intravascular or<br>hypodermic applications.  | Pass<br>All samples met the<br>pre-determined<br>acceptance criteria           |
| Usability Study /<br>Simulated Use                        | Evaluation of device<br>performance to meet<br>the user needs under<br>simulated use<br>conditions with<br>accessories and<br>treatment devices                                                                                                | FDA Guidance Human<br>Factors and Usability<br>Engineering 2016, IEC<br>62366:2015                                                                                             | Pass<br>All samples met the<br>pre-determined<br>acceptance criteria           |
| Conditioning,<br>Distribution,<br>and Shelf life<br>Aging | To demonstrate the<br>device met all<br>specifications at both<br>baseline (T=0) and<br>following accelerated<br>aging to a 12-month<br>shelf life equivalent<br>(T=12)                                                                        | Evaluation of device<br>following accelerated<br>aging to a 12-month shelf<br>life equivalent                                                                                  | Pass<br>All samples met the<br>pre-determined<br>acceptance criteria           |
| Performance Bench Testing Summary                         |                                                                                                                                                                                                                                                |                                                                                                                                                                                |                                                                                |
| Study Name                                                | Description                                                                                                                                                                                                                                    | Reference Standard                                                                                                                                                             | Results                                                                        |
| Package Testing                                           |                                                                                                                                                                                                                                                |                                                                                                                                                                                |                                                                                |
| Conditioning,<br>Distribution,<br>and Shelf Life<br>Aging | Evaluation of packaging<br>strength and integrity at<br>both baseline (T=0) and<br>following accelerated<br>aging to a 12-month<br>shelf life equivalent<br>(T=12)                                                                             | ISTA procedure 3A (2018),<br>ASTM D4169-16, ASTM<br>F1980-16,<br>ASTM F88-15, ASTM<br>F2096-11                                                                                 | Pass<br>All samples met the<br>pre-determined<br>acceptance criteria           |
| Visual<br>Inspection<br>Packaging and<br>Labeling         | To demonstrate that the<br>product meets the<br>packaging and labeling<br>visual and adherence<br>requirements                                                                                                                                 | N/A                                                                                                                                                                            | Pass<br>All samples met the<br>pre-determined<br>acceptance criteria           |

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

The results of these tests provide reasonable assurance that the BOSS Balloon Guide Catheter has been designed and tested to assure conformance to the requirements for its intended use and indications for use. No new safety or performance issues were raised during the testing; therefore, this device is considered to be substantially equivalent to the predicate devices.

## Biocompatibility Testing Summary

The device is categorized as Externally Communicating Device, Circulating Blood, Limited Contact (≤ 24 hours), per ISO 10993-1, the following testing was conducted:

| Test Name                           | Test Method                                                                                                                                                                                                                                                                                                        | Results                                                                                                                                            |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Cytotoxicity                        | Tested in accordance with ISO 10993-<br>5:2009, Biological Evaluation of Medical<br>Devices – Part 5: Tests for <i>in vitro</i> toxicity,<br>Neutral Red Uptake Method                                                                                                                                             | Pass<br>Noncytotoxic according to<br>the<br>predetermined<br>acceptance criteria                                                                   |
| Intracutaneous<br>Irritation        | Tested in accordance with ISO 10993-<br>10:2010, Biological Evaluation of Medical<br>Devices – Part 10: Tests for Irritation and<br>Skin Sensitization                                                                                                                                                             | Pass<br>Test requirements for<br>intracutaneous reactivity<br>were met according to the<br>predetermined<br>acceptance criteria                    |
| Sensitization                       | Tested in accordance with ISO 10993-10,<br>Biological Evaluation of Medical Devices –<br>Part 10 Tests for<br>Irritation and Skin Sensitization, Kligman<br>Maximization Test                                                                                                                                      | Pass<br>Did not elicit a<br>sensitization<br>response according to the<br>predetermined<br>acceptance criteria                                     |
| Test Name                           | Test Method                                                                                                                                                                                                                                                                                                        | Results                                                                                                                                            |
| Systemic Toxicity                   | Tested in accordance with ISO 10993-<br>11:2017, Biological Evaluation of Medical<br>Devices - Part 11: Tests for Systemic<br>Toxicity                                                                                                                                                                             | Pass<br>Test requirements for<br>systemic toxicity were<br>met according to the<br>predetermined<br>acceptance criteria                            |
| Material Mediated<br>Pyrogenicity   | Tested in accordance with ISO 10993-<br>11:2017, Biological<br>Evaluation of Medical Devices - Part 11:<br>Tests for Systemic Toxicity and USP 40<br><151> Pyrogen Test                                                                                                                                            | Pass<br>Nonpyrogenic, met the<br>predetermined<br>acceptance criteria                                                                              |
| Hemolysis                           | Tested in accordance with ASTM F756-17,<br>Standard Practice for Assessment of<br>Hemolytic Properties of Materials and ISO<br>10993-4, Biological Evaluation of Medical<br>Devices - Part 4: Selection of Tests for<br>Interactions with Blood, Tests for<br>Hemolytic Properties, Direct and Indirect<br>Methods | Pass<br>Non-hemolytic, met the<br>predetermined<br>acceptance criteria                                                                             |
| Complement<br>Activation            | Tested in accordance with ISO 10993-4,<br>Biological Evaluation of Medical Devices -<br>Part 4:2017: Selection of Tests for<br>Interactions with Blood, SC5b-9<br>Complement Activation                                                                                                                            | Pass<br>Demonstrates similar<br>complement activation<br>characteristics as the<br>control device, met the<br>predetermined<br>acceptance criteria |
| <i>In vivo</i><br>Thromboresistance | Tested in accordance with ISO 10993-<br>4:2017, Biological Evaluation of Medical<br>Devices - Part 4: Selection of Tests for<br>Interactions with Blood                                                                                                                                                            | Pass<br>Demonstrates similar<br>thromboresistance<br>characteristics as the<br>control device, met the<br>predetermined<br>acceptance criteria     |

{12}------------------------------------------------

# Sterilization and Shelf Life

The BOSS Balloon Guide Catheter sterilization process using 100% Ethylene Oxide (EO) has been validated in accordance with ISO 11135-1:2014 to achieve a SAL of 10 °. EO and ECH residuals were below the limits specified in ISO 10993-7:2008. Bacterial Endotoxin Levels (BET) were below the level of 2.15 EU/device in accordance with ANSI AAMI ST72:2011/(R)2016. Both baseline and accelerated shelf life testing were conducted (T = 0 and T = 12-month AA (accelerated aging)) demonstrating the device will perform as intended to support the proposed 12-month shelf-life.

{13}----…

---

**Source:** [https://fda.innolitics.com/device/K200910](https://fda.innolitics.com/device/K200910)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
