APAS Independence with IC Chromogenic MRSA BD Analysis Module; APAS Independence with IC Chromogenic MRSA TFS/S Analysis Module
K200839 · Clever Culture Systems · QQY · Oct 28, 2021 · Microbiology
Device Facts
Record ID
K200839
Device Name
APAS Independence with IC Chromogenic MRSA BD Analysis Module; APAS Independence with IC Chromogenic MRSA TFS/S Analysis Module
Applicant
Clever Culture Systems
Product Code
QQY · Microbiology
Decision Date
Oct 28, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.2190
Device Class
Class 2
Attributes
AI/ML, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K200839 · Oct 28, 2021
APAS Independence with IC Chromogenic MRSA BD Analysis Module; APAS Independence with IC Chromogenic MRSA TFS/S Analysis Module
Clever Culture Systems
Remnant clinical swab specimens from routine standard-of-care testing
Remnant clinical swab specimens were used to evaluate the clinical performance of the APAS Independence with MRSA Analysis Modules by comparing APAS-generated results to a panel of microbiologists' interpretations.
Clinical performance evaluation; Remnant clinical specimens; Standard of care
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Clinical Study Part 1 and Part 2; Retrospective evaluation of remnant clinical swab specimens; Follow-up/Duration: Not applicable
Anterior nares specimens from routine MRSA surveillance; Sample Size: 1590 samples (1573 BD plates and 1580 TFS/S plates analyzed); Number of Sites: 1 clinical site (oUS) for Part 1; 1 clinical site (US) for Part 2
Panel of microbiologists (majority result)
Clinical sensitivity and agreement of APAS-generated results compared to panel interpretation
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
MRSA colonization screening
—
Positive Percent Agreement > 98%
99.7% (US Panel) / 100.0% (AU Panel)
Analytical studies and internal software validation
The APAS Independence is an in vitro diagnostic system comprised of an instrument and software analysis module(s) for specific indications that are used to automate imaging and interpretation of microbial colonies on plates of solid culture media.
Device Story
APAS Independence is an automated system for screening microbial culture plates. It uses an imaging station with LED illumination and a CCD camera to capture digital images of chromogenic agar plates inoculated with anterior nares swabs. Software analysis modules process these images to categorize pixels, identify colony arrangements, and differentiate between MRSA and non-MRSA colony types based on color. The system operates in a laboratory setting, managed by trained personnel. It provides automated screening results (Presumptive MRSA, Presumptive non-MRSA, or Negative) which are transmitted to the LIS. The device acts as an aid to the microbiologist; all presumptive positive results require manual review by a trained microbiologist to confirm findings and determine follow-up testing. This automation reduces the manual workload for screening negative plates, allowing microbiologists to focus on plates with potential growth.
Clinical Evidence
Clinical performance was evaluated in a prospective study using 1,590 samples (clinical and simulated). Performance was compared against reference panels of three microbiologists interpreting digital images. For the BD module, PPA for presumptive MRSA was 99.7%-100.0%. For the TFS/S module, PPA for presumptive MRSA was 99.2%-99.5%. Analytical studies included reproducibility, limit of detection (colony size), and interference testing, all demonstrating performance meeting the 95% target.
Technological Characteristics
System includes automated plate handling, LED illumination, and CCD camera imaging. Software runs on APAS Controller PC. Connectivity via LIS interface. Daily calibration via Color Check Tool and System Check Tool. Software modules are specific to agar type (BBL CHROMagar MRSA II or Thermo-Fisher Spectra MRSA). Complies with IEC 61010-1, IEC 62304, IEC 62366-1, and IEC 61326-1/2-6 standards.
Indications for Use
Indicated for automated imaging and interpretation of anterior nares swab specimens inoculated on BBL CHROMagar MRSA II or Thermo-Fisher Spectra MRSA chromogenic agar plates to aid in screening for MRSA colonization. Requires review by a trained microbiologist for presumptive positive results.
Regulatory Classification
Identification
An automated image assessment system for microbial colonies on solid culture media is a system that is intended to assess the presence or absence of microbial colonies on solid microbiological culture medium, and to interpret their number, and phenotypic and morphologic characteristics through analysis of two dimensional digital images as an aid in diagnosis of infectious disease.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include a detailed description of the device, including the technology employed, components and software modules, as well as a detailed explanation of the result algorithms and any expert rules that are used to assess colony characteristics and enumerate colonies from image capture through end result.
(2) Premarket notification submissions must include detailed documentation of the analytical studies performed to characterize device performance to support the intended use, as appropriate.
(3) Premarket notification submissions must include detailed documentation from clinical studies performed on a population that is consistent with the intended use population.
(i) The clinical studies must establish the device performance based on comparison to results obtained by an acceptable reference method, as appropriate.
(ii) The clinical study documentation must include the study protocol with a predefined statistical analysis plan and the final report documenting support for the Indications for Use and the results of the statistical analysis, as appropriate.
(4) Premarket notification submissions must include detailed documentation for device software, including but not limited to software applications and hardware based components that incorporate software, and any decision-making thresholds used to generate results for the device. If a part of a Total Laboratory Automation System, the premarket notification submission must include detailed documentation addressing the instrument and software system integration.
(5) Premarket notification submissions must include detailed documentation of appropriate instructions for use regarding the intended user's device quality control procedures for the instrument system and components, as appropriate.
(6) The 21 CFR 809.10 compliant device labeling must include:
(i) Detailed user instructions to mitigate the risk of failure to operate the instrument correctly.
(ii) A detailed explanation of the interpretation of results and limitations regarding the need for review of culture plates by a qualified microbiologist, as appropriate.
(iii) A summary of performance data obtained from the analytical studies used to support device performance, as appropriate.
(iv) A summary of performance data obtained from clinical studies performed on a population that is consistent with the intended use population, as appropriate.
(7) Under 21 CFR 820.10(c) design and development, device manufacturers must, as appropriate:
(i) Conduct human factors/usability validation testing with the final version of the labeling and related materials to adequately mitigate the risk of failure to operate the instrument correctly.
(ii) Document a device training program that will be offered to the end user to adequately mitigate the risk of failure to operate the instrument correctly.
Predicate Devices
APAS Independence with Urine Analysis Module (K183648)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K200839
B Applicant
Clever Culture Systems
C Proprietary and Established Names
APAS Independence with IC Chromogenic MRSA BD Analysis Module; APAS Independence with IC Chromogenic MRSA TFS/S Analysis Module
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QQY | Class II | 21 CFR 866.2190 - Automated Image Assessment System For Microbial Colonies On Solid Culture Media | MI - Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
To obtain a substantial equivalence determination for the APAS Independence with two independent MRSA Analysis Modules
B Measurand:
Digital images of colonies cultured on chromogenic agar plates to determine presence or absence of methicillin-resistant Staphylococcus aureus (MRSA)
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
C Type of Test:
The APAS Independence when using its MRSA Analysis Modules is an in vitro diagnostic test system for automated assessment of microbial colonies on chromogenic culture media. The system is for use with anterior nares specimens inoculated on chromogenic agar to aid in screening for MRSA.
III Intended Use/Indications for Use:
A Intended Use(s):
The APAS Independence is an in vitro diagnostic system comprised of an instrument and software analysis module(s) for specific indications that are used to automate imaging and interpretation of microbial colonies on plates of solid culture media.
B Indication(s) for Use:
The APAS Independence is an in vitro diagnostic system comprised of an instrument for automated imaging of agar culture plates and a software analysis module for the following use:
1. The APAS Independence, when using its IC MRSA Chromogenic BD analysis module, automates culture plate imaging and interpretation to detect the presence or absence of colonies with colors suggestive of methicillin-resistant Staphylococcus aureus (MRSA) growth on Beckton Dickson BBL CHROMagar MRSA II agar that has been inoculated with anterior nares swabs and incubated at 36°C ± 1°C for 24 hours.
The APAS Independence, when using its IC MRSA Chromogenic BD analysis module, provides an aid in routine screening for colonization with MRSA. It provides one of two screening results: Presumptive MRSA or Negative. All culture plates that are identified as Presumptive MRSA by the APAS Independence, when using the IC MRSA Chromogenic BD analysis module require review by a trained microbiologist.
2. The APAS Independence, when using its IC MRSA Chromogenic TFS/S analysis module, automates culture plate imaging and interpretation to detect the presence or absence of colonies with colors suggestive of methicillin-resistant Staphylococcus aureus (MRSA) growth on Thermo-Fisher Spectra MRSA agar that has been inoculated with anterior nares swabs and incubated at 36°C ± 1°C for 24 hours.
The APAS Independence, when using its IC MRSA Chromogenic TFS/S analysis module, provides an aid in routine screening for colonization with MRSA. It provides one of three screening results: Presumptive MRSA, Presumptive non-MRSA, or Negative. All culture plates that are identified as Presumptive MRSA or Presumptive non-MRSA by the APAS Independence, when using the IC MRSA Chromogenic TFS/S analysis module, require review by a trained microbiologist.
K200839 - Page 2 of 30
{2}
K200839 - Page 3 of 30
C Special Conditions for Use Statement(s):
- Rx - For Prescription Use Only
- The APAS Independence with MRSA Analysis Module is only for use with BBL CHROMagar MRSA II Spectra MRSA chromogenic agar plates inoculated with anterior nares swab specimens.
- The APAS Independence with IC MRSA Chromogenic BD is qualified for use with BBL CHROMagar MRSA II agar from Becton Dickinson (product numbers 215228 and 215229). It has not been evaluated for use with non-chromogenic plates or other chromogenic plates.
- The APAS Independence with IC MRSA Chromogenic TFS/S is qualified for use with Spectra MRSA agar from Thermo-Fisher Scientific (product numbers R01821(A) and R01822(A). It has not been evaluated for use with non-chromogenic plates or other chromogenic plates.
- Organisms other than MRSA may produce a mauve color on this media and produce a presumptive positive result. Laboratories should employ confirmation procedures for all Presumptive MRSA results.
- Some MRSA strains may not produce characteristic chromogenic reactions on the media. Laboratories should consider alternative procedures when considering such strains, or employ the APAS flag functions to assist.
D Special Instrument Requirements:
APAS Independence
IV Device/System Characteristics:
A Device Description:
The APAS Independence is an automated system for the screening of culture plates to assess the presence of microbial growth. The APAS Independence system consists of the following:
- an automated plate handling mechanism that moves culture plates through the instrument,
- an imaging station to capture images of a culture plate, and
- software for analysis of the image, determination of growth and presentation of reports.
A list of the major sub-components of the APAS Independence is provided in Table 1. With the exception of the MRSA module software, these components remain the same as those used in the predicate (APAS Independence with Urine Analysis Module, K183648).
Table 1. Major sub-systems of the APAS Independence
| Component | | Function |
| --- | --- | --- |
| Hardware | Imaging Station | LED illumination of culture plates and image capture using a CCD camera |
| | Automated Plate Handling System | Movement, positioning and sorting of culture plates within the instrument |
| Software | APAS Controller PC | Controls image capture, analysis, report generation and result storage |
| | Plate Controller PC | Controls movement of culture plates between the input carriers, imaging station and output carriers or stacks |
{3}
| Component | | Function |
| --- | --- | --- |
| | Instrument Controller Software | Provides the user interface for operation of the instrument and coordinates the functions of the APAS and Plate Controller PCs |
| | Analysis Module Software | Installed on the APAS Controller PC to provide the configuration and instructions for image capture and analysis |
| | LIMS Interface Software | Installed on the Instrument Controller PC to import diagnostic details and provide context for interpretation of MRSA culture results. Imported information may be applied to the system: “LIMS Force Flag”: automatically forces an APAS result to “Review” irrespective of the growth characteristics observed “LIMS Complementary Test Flag”: automatically changes a “Negative” designation to “Review” based on user defined rules applied to additional diagnostic information |
| Quality Control | Color Calibration Tool | Multicolored disk for calibration/checking of system optics |
| | System Check | Simulated culture plates used to confirm instrument function |
The APAS Independence is intended to be installed with multiple, independent software (analysis) modules, each of which will provide an assessment of growth for a specific clinical indication. More than one analysis module may be developed for the same Indication for Use to allow APAS to assess growth on culture plates from multiple agar manufacturers. In this application, two independent MRSA analysis modules have been designed to interpret microbial growth as follows:
- APAS Independence with IC Chromogenic MRSA BD Analysis Module is designed to interpret microbial growth on BBL CHROMagar MRSA II agar from Becton Dickinson
- APAS Independence with IC Chromogenic MRSA TFS/S Analysis Module is designed to interpret microbial growth on Remel Spectra MRSA agar from Thermo Fisher Scientific.
The APAS Independence with the IC Chromogenic MRSA BD and TFS/S Analysis Modules images and evaluates chromogenic agar plates (BBL CHROMagar MRSA II and Remel Spectra MRSA, respectively) to determine and report plate results. Each module has its own reporting scheme (described in Table 2) based on the analytical and clinical performance data and corresponding risk mitigations.
Table 2. Reporting results for the IC Chromogenic MRSA BD and TFS/S analysis modules
| Chromogenic Agar Result | Chromogenic Agar Description | IC Chromogenic Analysis Modules | |
| --- | --- | --- | --- |
| | | BD Result | TFS/S Result |
| Presumptive MRSA Growth | Colonies or growth suggestive of MRSA | Presumptive MRSA | Presumptive MRSA |
K200839 - Page 4 of 30
{4}
| Chromogenic Agar Result | Chromogenic Agar Description | IC Chromogenic Analysis Modules | |
| --- | --- | --- | --- |
| | | BD Result | TFS/S Result |
| Presumptive non-MRSA Growth | Colonies or growth not suggestive of MRSA | Negative | Presumptive non-MRSA |
| No Growth | No colonies or growth | Negative | Negative |
B Principle of Operation:
APAS Independence when using a MRSA analysis module is designed to detect the presence or absence of presumptive methicillin-resistant Staphylococcus aureus (MRSA) growth on chromogenic MRSA agar plates. No quantification of growth is required since it is not used to inform clinical decisions. The system comprises an imaging station to capture images of chromogenic MRSA agar plates and a plate handling system that moves plates between the input carriers, imaging station and output carriers or stacks. The software conducts analysis of the images, assessment of microbial colonies (if present), and result designation.
The APAS Independence with MRSA Analysis Modules is indicated for screening MRSA chromogenic agar plates (BBL CHROMagar MRSA II and Remel Spectra MRSA) inoculated with anterior nares swab specimens and incubated for 24 hours. The system takes digital images of each plate that are analyzed automatically to determine if presumptive MRSA colonies are present. For each plate, the APAS Independence performs the following tasks:
a. digitally combines several images of the plate for analysis,
b. categorizes each pixel of the image according to the visually different areas on the plate (e.g., type of growth or agar background),
c. determines how pixels are arranged in relation to each other (e.g., clumps of related pixels may represent a colony),
d. counts the number of colonies per colony type (although colony counts are not reported in the MRSA Analysis Modules),
e. differentiates between MRSA and non-MRSA colony types and reports presumptive results accordingly,
f. incorporates data from the Laboratory Information System (LIS) to provide a final report; when applicable based on clinical context ("flagged"), a "negative" result may be modified to "presumptive MRSA growth" to alert the microbiologist that further investigation may be warranted.
When using the IC MRSA Chromogenic BD Analysis module (BD module), all plates with presumptive MRSA growth are designated for review by a trained microbiologist for determination of status and follow up testing according to conventional laboratory practice. Plates with presumptive non-MRSA growth or no growth are not designated for review and may be discarded according to the standard operating procedures of the laboratory.
When using the IC MRSA Chromogenic TFS/S Analysis module (TFS/S module), all plates with presumptive MRSA growth or presumptive non-MRSA growth are designated for review by a trained microbiologist for determination of status and follow up testing according to conventional laboratory practice. Plates with no growth are not designated for review and may be discarded according to the standard operating procedures of the laboratory.
K200839 - Page 5 of 30
{5}
K200839 - Page 6 of 30
1. Instrument Name:
APAS Independence
2. Specimen Identification:
Plates are identified using an integrated barcode scanner within the Imaging Module of the APAS Independence. The identity of each plate is displayed on the results screen and transmitted automatically to the LIS.
3. Specimen Sampling and Handling:
The operator loads carriers containing culture plates into the Input Module of the APAS Independence from which they are moved automatically, one plate at a time, to the Imaging Module where the barcode is read, the lid is removed, and images of the agar surface are captured. Upon completion of image capture and analysis, the plate lid is replaced, and the system automatically sorts each plate based on its designation into the appropriate carrier in the Output Module.
4. Calibration:
The APAS Independence requires daily color calibration prior to analysis of culture plates. Calibration is performed using the Color Check Tool provided with the instrument, a multicolored disk for checking and correcting color of the system optics. Instruction for the daily Color Check is included in the APAS Independence User Manual. In the event of a Color Check failure, the check should be repeated. If the repeated failure occurs, the instrument should not be used until it has been serviced.
5. Quality Control:
After daily calibration (Color Check), a System Check must be performed followed by biological Quality Control testing.
The System Check Tool comprises a pair of disks with small dots replicating colonies for checking overall system functionality and associated software. Instruction for daily System Check is provided in the User Manual for the APAS Independence.
Daily biological Quality Control testing is performed using a reference culture plate inoculated with a MRSA positive strain (e.g., Staphylococcus aureus ATCC 43300). Instructions for testing is provided in the User Manuals for the APAS Independence with IC MRSA Chromogenic BD Analysis Module and APAS Independence with IC MRSA Chromogenic TFS/S Analysis Module.
In the event of a System Check or Quality Control testing failure, the check/testing should be repeated. If the repeated failure occurs, the instrument should not be used until it has been serviced. In the event of a System Check fail, the instrument should be locked to prevent processing of additional culture plates until service has been performed.
V Substantial Equivalence Information:
A Predicate Device Name(s):
APAS Independence with Urine Analysis Module
{6}
B Predicate 510(k) Number(s):
K183648
C Comparison with Predicate(s):
Table 3. Comparison with Predicate
| Device & Predicate Device(s): | DEVICE: K200839 | PREDICATE: K183648 |
| --- | --- | --- |
| Device Trade Name | APAS Independence with IC Chromogenic MRSA BD Analysis Module; APAS Independence with IC Chromogenic MRSA TFS/S Analysis Module | APAS Independence with Urine Analysis Module |
| General Device Characteristic Similarities | | |
| Intended Use | The APAS Independence is an in vitro diagnostic system comprised of an instrument and software analysis module(s) for specific indications that are used to automate imaging and interpretation of microbial colonies on plates of solid culture media. | Same |
| Imaging Station | Light Emitting Diode (LED) illumination of culture plates and image capture using a Charge Coupled Device (CCD) camera | Same |
| Plate Handling | Automated | Same |
| APAS Controller PC | Controls image capture, analysis, report generation and result storage | Same |
| Instrument Controller PC | Provides the user interface for the APAS Independence and controls plate movement | Same |
| Analysis Module | Installed on the APAS Controller PC to provide the configuration and instructions for image capture and analysis | Same |
| Laboratory Information System (LIS) Interface | Analysis result for each plate sent to the LIS. Sample ID details retrieved from the LIS. | Same |
| Calibration | Performed daily using a manufacturer-provided Color Check Tool | Same |
| System Check | Performed daily using the manufacturer-provided System Check Tool to verify instrument and software functionality | Same |
| General Device Characteristic Differences | | |
K200839 - Page 7 of 30
{7}
| Device & Predicate Device(s): | DEVICE: K200839 | PREDICATE: K183648 |
| --- | --- | --- |
| Indications For Use | The APAS Independence is an in vitro diagnostic system comprised of an instrument for automated imaging of agar culture plates and a software analysis module for the following use:
1. The APAS Independence, when using its IC MRSA Chromogenic BD analysis module, automates culture plate imaging and interpretation to detect the presence or absence of colonies with colors suggestive of methicillin-resistant Staphylococcus aureus (MRSA) growth on Beckton Dickson BBL™ CHROMagar™ MRSA II agar that has been inoculated with anterior nares swabs and incubated at 36°C ± 1°C for 24 hours.
The APAS Independence, when using its IC MRSA Chromogenic BD analysis module, provides an aid in routine screening for colonization with MRSA. It provides one of two screening results: Presumptive MRSA or Negative. All culture plates that are identified as Presumptive MRSA by the APAS Independence, when using the IC MRSA Chromogenic BD analysis module require review by a trained microbiologist.
2. The APAS Independence, when using its IC MRSA Chromogenic TFS/S analysis module, automates culture plate imaging and interpretation to detect the presence or absence of colonies with colors suggestive of methicillin-resistant Staphylococcus aureus (MRSA) growth on Thermo-Fisher Spectra™ MRSA agar that has been inoculated with anterior nares swabs and incubated at 36°C ± 1°C for 24 hours.
The APAS Independence, when using its IC MRSA Chromogenic TFS/S analysis module, provides an aid in routine screening for colonization with MRSA. It provides one of three screening results: Presumptive MRSA, Presumptive non-MRSA, or Negative. All culture plates that are identified as Presumptive MRSA or | The APAS Independence is an in vitro diagnostic system comprised of an instrument for automated imaging of agar culture plates and a software analysis module for the following use:
The APAS Independence, when using its urine analysis module, automates urine culture plate imaging and interpretation to detect the presence or absence of microbial growth on sheep blood and MacConkey agar culture plates that are inoculated with a 1μL sample volume. The APAS Independence, when using its urine analysis module, provides a semiquantitative assessment of colony counts that are used as an aid in the diagnosis of urinary tract infection. All urine culture plates that are identified as positive for growth by the APAS Independence, when using its urine analysis module, must be reviewed by a trained microbiologist. |
K200839 - Page 8 of 30
{8}
| Device & Predicate Device(s): | DEVICE: K200839 | PREDICATE: K183648 |
| --- | --- | --- |
| | Presumptive non-MRSA by the APAS Independence, when using the IC MRSA Chromogenic TFS/S analysis module, require review by a trained microbiologist. | |
| IFU Target | Presumptive MRSA colonies | Microbial growth |
| Sample type | Anterior nares specimens | Urine samples |
| Biological Quality Control | Performed daily using standardized suspensions of Staphylococcus aureus ATCC 43300 (MRSA positive strain) | Performed daily using standardized suspensions of Escherichia coli and Enterococcus faecalis |
# VI Standards/Guidance Documents Referenced:
- ISO 14971: Medical devices - Application of Risk Management to Medical Devices; 2007.
- IEC 61010-1: Safety requirements for electrical equipment for measurement, control and laboratory use - Part 1: General requirements; 2010, 3rd edition.
- IEC 62304: Medical device software - Software life cycle processes; 2006.
- IEC 62366-1: Medical devices - Part 1: Application of usability engineering to medical devices; 2015.
- IEC 61326-1: Electrical equipment for measurement, control and laboratory use - EMC Requirements - Part 1: General requirements; 2012.
- IEC 61326-2-6: Electrical equipment for measurement, control and laboratory use - EMC Requirements - Part 2-6: Particular Requirements - In vitro diagnostic (IVD) medical Equipment; 2012.
# VII Performance Characteristics (if/when applicable):
# A Analytical Performance:
The organisms listed in Table 4 were used to evaluate the analytical performance of the APAS Independence with MRSA Analysis Modules on both on BBL CHROMagar MRSA II (referred throughout as "BD") and Remel Spectra MRSA (referred throughout as "TFS/S") chromogenic agar plates. The organism codes were internal and used for the purpose of tracking during the studies.
Table 4. Organisms tested in analytical studies
| Organism Name (Strain) | Code | Colony Appearance | | Interpretation / Designation | Used in Study |
| --- | --- | --- | --- | --- | --- |
| | | BD | TFS/S | | |
| Bacillus cereus(Stock) | MISC6 | Pink | Blue | Negative | Accuracy, Digital Image Quality |
| Bacillus licheniformis(NCTC 10341) | MISC7 | Purple | Blue | Negative | Accuracy, Digital Image Quality |
| Bacillus spp.(Stock) | MISC2 | Purple | Blue | Negative | Accuracy, Digital Image Quality |
K200839 - Page 9 of 30
{9}
| Organism Name (Strain) | Code | Colony Appearance | | Interpretation / Designation | Used in Study |
| --- | --- | --- | --- | --- | --- |
| | | BD | TFS/S | | |
| Enterococcus faecalis (ATCC 29212) | EF8 | Fine blue smear | Fine blue smear | Negative | Accuracy |
| Staphylococcus aureus (Stock) | MRSA3 | Mauve | Denim blue | Presumptive Positive | Accuracy, Reproducibility, LoD, Interference, Range of Assay |
| Staphylococcus aureus (Stock) | MRSA5 | Mauve | Denim blue | Presumptive Positive | Accuracy, Range of Assay |
| Staphylococcus aureus (ATCC 43300) | MRSA9 | Mauve | Denim blue | Presumptive Positive | Accuracy, Reproducibility, LoD, Biological QC, Digital Image Quality |
| Staphylococcus aureus (ATCC 1707) | MRSA26 | Mauve | Denim blue | Presumptive Positive | Accuracy, Digital Image Quality |
| Staphylococcus aureus (ATCC 13277) | MRSA29 | Mauve | Denim blue | Presumptive Positive | Accuracy, Digital Image Quality |
| Staphylococcus aureus (ATCC 13435) | MRSA31 | Mauve | Denim blue | Presumptive Positive | Accuracy, Digital Image Quality |
| Staphylococcus aureus (ATCC 13626) | MRSA33 | Mauve | Denim blue | Presumptive Positive | Accuracy, Digital Image Quality |
| Staphylococcus haemolyticus (Stock) | SH3 | Small white | Small white | Negative | Accuracy, Digital Image Quality |
| Staphylococcus haemolyticus (Stock) | SH6 | Small white | Small white | Negative | Accuracy, Reproducibility, LoD, Interference, Range of Assay, Digital Image Quality |
| Staphylococcus haemolyticus (Stock) | STSP2 | Small white | Small white | Negative | Reproducibility |
| Staphylococcus warneri (Stock) | STSP4 | Small white | Small white | Negative | Range of Assay |
BD: BBL CHROMagar MRSA II chromogenic agar
TFS/S: Remel Spectra MRSA chromogenic agar
## 1. Precision/Reproducibility:
## Reproducibility
The reproducibility of presumptive MRSA and non-presumptive MRSA colony detection by the APAS Independence with MRSA Analysis Modules was evaluated with three instruments. Saline was used to evaluate no growth, three dilutions of two MRSA strains (MRSA3 and MRSA9) were used to evaluate presumptive MRSA growth (i.e., mainly confluent, partial confluent, and single colony growth), and three dilutions of two *S. haemolyticus* strains (SH3 and SH6) were used to evaluate non-presumptive MRSA growth (i.e., mainly confluent, partial confluent, and single colony growth) on BD and TFS/S chromogenic agar plates. Each dilution was plated in triplicate and incubated at $35^{\circ}\mathrm{C} \ 2 \pm^{\circ}\mathrm{C}$ for 24 hours. Five replicate images of each plate were taken at three different orientations $(0^{\circ}, 120^{\circ}$ and $270^{\circ})$ on three instruments for a total of 135 images per dilution (3 replicate plates x 5 replicate images x 3 orientations x 3 instruments = 135 images per dilution).
Results of the reproducibility study with the BD and TFS/S MRSA analysis modules are shown in Table 5 and Table 6, respectively. The results demonstrate that the reproducibility between instruments is acceptable.
K200839 - Page 10 of 30
{10}
Table 5. Reproducibility of the APAS Independence with the BD MRSA Analysis Module
| Organism Code1 | Dilution | Percent Agreement | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | Instrument 1 | | Instrument 2 | | Instrument 3 | | Combined | |
| | | Growth2 | Color3 | Growth2 | Color3 | Growth2 | Color3 | Growth2[95% CI] | Color3[95% CI] |
| MRSA3 | 1 | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (135/135) [97.2-100] | 100% (135/135) [97.2-100] |
| | 2 | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (135/135) [97.2-100] | 100% (135/135) [97.2-100] |
| | 3 | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (135/135) [97.2-100] | 100% (135/135) [97.2-100] |
| MRSA9 | 1 | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (135/135) [97.2-100] | 100% (135/135) [97.2-100] |
| | 2 | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (135/135) [97.2-100] | 100% (135/135) [97.2-100] |
| | 3 | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (135/135) [97.2-100] | 100% (135/135) [97.2-100] |
| SH3 | 1 | 100% (45/45) | 80% (36/45) | 100% (45/45) | 93.3% (42/45) | 100% (45/45) | 75.6% (34/45) | 100% (135/135) [97.2-100] | 83.0% (112/135) [75.7-88.4] |
| | 2 | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 97.8% (44/45) | 100% (135/135) [97.2-100] | 99.3% (134/135) [95.9-99.9] |
| | 3 | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 93.3% (42/45) | 100% (135/135) [97.2-100] | 97.8% (132/135) [93.7-99.2] |
| SH6 | 1 | 100% (45/45) | 80% (36/45) | 100% (45/45) | 75.6% (34/45) | 100% (45/45) | 55.6% (25/45) | 100% (135/135) [97.2-100] | 70.4% (95/135) [62.2-77.4] |
| | 2 | 100% (45/45) | 97.8% (44/45) | 100% (45/45) | 88.9% (40/45) | 100% (45/45) | 82.2% (37/45) | 100% (135/135) [97.2-100] | 89.6% (121/135) [83.3-93.7] |
| | 3 | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 95.6% (43/45) | 100% (135/135) [97.2-100] | 98.5% (133/135) [94.8-99.6] |
| Saline | - | 100% (45/45) | 100% (45/45) | 97.8% (44/45) | 100% (45/45) | 95.6% (43/45) | 100% (45/45) | 97.8% (132/135) [93.7-99.2] | 100% (135/135) [97.2-100] |
1 Organism code, as noted in Table 4.
2 Value calculated by dividing the number of images with the expected growth (i.e., growth or no growth detected) by the total number of images.
3 Value calculated by dividing the number of images with the expected color (i.e., at least one colored colony or no color detected) by the total number of images.
K200839 - Page 11 of 30
{11}
Table 6. Reproducibility of the APAS Independence with the TFS/S MRSA Analysis Module
| Organism Code1 | Dilution | Percent Agreement | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | Instrument 1 | | Instrument 2 | | Instrument 3 | | Combined | |
| | | Growth2 | Color3 | Growth2 | Color3 | Growth2 | Color3 | Growth2[95% CI] | Color3[95% CI] |
| MRSA3 | 1 | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (135/135) [97.2-100] | 100% (135/135) [97.2-100] |
| | 2 | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (135/135) [97.2-100] | 100% (135/135) [97.2-100] |
| | 3 | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (135/135) [97.2-100] | 100% (135/135) [97.2-100] |
| MRSA9 | 1 | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (135/135) [97.2-100] | 100% (135/135) [97.2-100] |
| | 2 | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (135/135) [97.2-100] | 100% (135/135) [97.2-100] |
| | 3 | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (135/135) [97.2-100] | 100% (135/135) [97.2-100] |
| SH3 | 1 | 100% (45/45) | 75.6% (34/45) | 100% (45/45) | 93.3% (42/45) | 100% (45/45) | 95.6% (43/45) | 100% (135/135) [97.2-100] | 88.1% (119/135) [81.6-92.6] |
| | 2 | 100% (45/45) | 97.8% (44/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 84.4% (38/45) | 100% (135/135) [97.2-100] | 94.1% (127/135) [88.7-97.0] |
| | 3 | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (135/135) [97.2-100] | 100% (135/135) [97.2-100] |
| SH6 | 1 | 100% (45/45) | 82.2% (37/45) | 100% (45/45) | 80.0% (36/45) | 100% (45/45) | 82.2% (37/45) | 100% (135/135) [97.2-100] | 81.5% (110/135) [74.1-87.1] |
| | 2 | 100% (45/45) | 95.6% (43/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 91.1% (41/45) | 100% (135/135) [97.2-100] | 95.6% (129/135) [90.6-97.9] |
| | 3 | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (45/45) | 100% (135/135) [97.2-100] | 100% (135/135) [97.2-100] |
| No Organism (Saline) | - | 66.7% (30/45) | 100% (45/45) | 66.7% (30/45) | 100% (45/45) | 68.9% (31/45) | 100% (45/45) | 67.4% (91/135) [59.1-74.7] | 100% (135/135) [97.2-100] |
1 Organism code, as noted in Table 4.
2 Value calculated by dividing the number of images with the expected growth (i.e., growth or no growth detected) by the total number of images.
3 Value calculated by dividing the number of images with the expected color (i.e., at least one colored colony or no color detected) by the total number of images.
K200839 - Page 12 of 30
{12}
# Accuracy
The accuracy in which the APAS Independence with MRSA Analysis Modules is able to correctly assign "presumptive MRSA" and "presumptive non- MRSA" colonies was evaluated with a range of organisms noted in Table 4. Organisms were serially diluted in saline and streaked onto BBL CHROMagar MRSA II (BD) and Remel Spectra MRSA (TFS/S) chromogenic agar plates in 2 replicates or 7 replicates to produce either light or confluent growth, respectively, after 24-hour incubation at $36^{\circ}\mathrm{C} \pm 1^{\circ}\mathrm{C}$ . At least 100 isolated colonies per plate were digitally labelled by a microbiologist and compared to the corresponding APAS result.
Results of the accuracy study with the BD and TFS/S analysis modules are shown in Table 7 and Table 8, respectively. The results demonstrate that the APAS Independence can accurately detect presumptive MRSA colonies when grown on BD or TFS/S chromogenic agar plates. However, it cannot accurately detect presumptive non-MRSA colonies. This reduced specificity is acceptable since plates with "no growth" or "presumptive non-MRSA growth" do not pose a risk for patient management and are therefore discarded.
Table 7. Accuracy of the APAS Independence with the BD MRSA Analysis Module
| | Microbiologist Assigned | | | |
| --- | --- | --- | --- | --- |
| | | Presumptive MRSA Colony | Presumptive Non-MRSA Colony | Total |
| APAS Assigned | Presumptive MRSA Colony | 793 | 23 | 816 |
| | Presumptive Non-MRSA Colony | 3 | 208 | 211 |
| | Colony Not Detected | 2 | 77 | 79 |
| | Total | 798 | 308 | 1106 |
| Percent Agreement | | 99.4% (793/798) | 67.5% (208/308) | 90.5% (1101/1106) |
Table 8. Accuracy of the APAS Independence with the TFS/S MRSA Analysis Module
| All Colonies Combined | Microbiologist Assigned | | | |
| --- | --- | --- | --- | --- |
| | | Presumptive MRSA Colony | Presumptive Non-MRSA Colony | Total |
| APAS Assigned | Presumptive MRSA Colony | 772 | 262 | 1034 |
| | Presumptive Non-MRSA Colony | 2 | 226 | 228 |
| | Colony Not Detected | 2 | 56 | 58 |
| | Total | 776 | 544 | 1320 |
| Percent Agreement | | 99.5% (772/776) | 41.5% (226/544) | 75.5% (997/1320) |
K200839 - Page 13 of 30
{13}
# Digital Image Quality Study
A Digital Image Quality study was performed to evaluate the accuracy of microbiologist's interpretation of a digital image compared to reading the agar plate manually (visual read). The data generated in this study were analyzed in two different ways: microbiologist interpretation of a digital image compared to a manual plate read (Table 9-Table 15) and reproducibility of microbiologists' interpretation of a digital image read (Table 16).
a) Microbiologist interpretation of APAS-generated digital image compared to manual plate read
The purpose of this study was to assess whether an APAS-generated digital image of a chromogenic agar plate read by a microbiologist can be interpreted equivalent to a manual reading of the chromogenic agar plate (i.e., manual plate read). Three hundred clinical samples (received as swabs and cultured on appropriate agar plates for routine MRSA surveillance) and 100 contrived samples were evaluated. Contrived samples were diluted such that two dilutions were evaluated: medium growth (100-250 CFU/plate) and low growth (1-50 CFU/plate). A panel of three microbiologists blindly interpreted all 400 agar plates and the corresponding 400 digital images that were randomized after a two week "wash out" period to avoid any potential for bias based on prior readings.
A summary of the results with the BD and TFS/S analysis modules is provided in Table 9. Data shown by each microbiologist are provided in Table 10, Table 11 and Table 12 for the BD analysis module as well as in Table 13, Table 14 and Table 15 for the TFS/S analysis module.
Table 9. Digital Image Quality Summary
| | No. of Results | Percent Agreement1 | | | |
| --- | --- | --- | --- | --- | --- |
| | | No Growth | Presumptive Non-MRSA Growth | Presumptive MRSA Growth | |
| BD | Microbiologist 1 | 400 | 95.8% (252/263) | 89.6% (69/77) | 91.7% (55/60) |
| | Microbiologist 2 | 400 | 98.1% (264/269) | 98.7% (74/75) | 98.2% (55/56) |
| | Microbiologist 3 | 400 | 95.4% (250/262) | 97.6% (82/84) | 98.1% (53/54) |
| | Combined | 1200 | 96.4% (766/794) | 95.3% (225/236) | 95.9% (163/170) |
| TFS/S | Microbiologist 1 | 400 | 93.8% (182/194) | 87.1% (88/101) | 92.4% (97/105) |
| | Microbiologist 2 | 400 | 96.0% (193/201) | 90.8% (89/98) | 92.1% (93/101) |
| | Microbiologist 3 | 400 | 94.2% (180/191) | 86.8% (92/106) | 96.1% (99/103) |
| | Combined | 1200 | 94.7% (555/586) | 88.2% (269/305) | 93.5% (289/309) |
1Percent agreement determined by dividing number of digital image read results by the number of manual plate read results from the same microbiologist for each designation (i.e., "no growth", "presumptive non-MRSA growth", "presumptive MRSA growth").
The digital image quality study determined that an APAS-generated digital image using the BD module can be equivalently interpreted by a microbiologist compared to reading the plate manually (visual read). This was achieved $>95\%$ of the time for each of the three result designations. Importantly, MRSA growth agreement was $>95\%$ , which is acceptable to support use of a two-rule designation set (i.e., presumptive MRSA growth and negative) in that only plates with presumptive MRSA growth are required for microbiologist review.
K200839 - Page 14 of 30
{14}
The digital image quality study determined that an APAS-generated digital image using the TFS/S module cannot be equivalently interpreted by a microbiologist compared to reading the plate manually. Agreement was $< 95\%$ of the time for each result designation. Importantly, presumptive MRSA growth agreement was $93.5\%$ , which is not acceptable to support use of a two-rule designation set. As such, the TFS/S module will use a three-rule designation set (i.e., presumptive MRSA growth, presumptive non-MRSA growth, and negative) in that plates with any growth (presumptive MRSA or presumptive non-MRSA) are required for microbiologist review.
Table 10. APAS Independence with BD MRSA analysis module - Microbiologist 1 Digital Image vs Manual Plate Interpretation
| BD analysis module Microbiologist 1 | Manual Plate | | | | |
| --- | --- | --- | --- | --- | --- |
| | | No Growth | Presumptive Non-MRSA Growth | Presumptive MRSA Growth | Total |
| Digital Image | No Growth | 252 | 8 | 2 | 262 |
| | Presumptive Non-MRSA Growth | 11 | 69 | 3 | 83 |
| | Presumptive MRSA Growth | - | - | 55 | 55 |
| | Total | 263 | 77 | 60 | 400 |
| | No Growth Agreement: 95.8% (252/263) Presumptive Non-MRSA Agreement: 89.6% (69/77) Presumptive MRSA Agreement: 91.7% (55/60) | | | | |
Table 11. APAS Independence with BD MRSA analysis module - Microbiologist 2 Digital Image vs Manual Plate Interpretation
| BD analysis module Microbiologist 2 | Manual Plate | | | | |
| --- | --- | --- | --- | --- | --- |
| | | No Growth | Presumptive Non-MRSA Growth | Presumptive MRSA Growth | Total |
| Digital Image | No Growth | 264 | 1 | - | 265 |
| | Presumptive Non-MRSA Growth | 5 | 74 | 1 | 80 |
| | Presumptive MRSA Growth | - | - | 55 | 55 |
| | Total | 269 | 75 | 56 | 400 |
| | No Growth Agreement: 98.1% (264/269) Presumptive Non-MRSA Agreement: 98.7% (74/75) Presumptive MRSA Agreement: 98.2% (55/56) | | | | |
K200839 - Page 15 of 30
{15}
Table 12. APAS Independence with BD MRSA analysis module – Microbiologist 3 Digital Image vs Manual Plate Interpretation
| BD analysis module
Microbiologist 3 | Manual Plate | | | | |
| --- | --- | --- | --- | --- | --- |
| | | No Growth | Presumptive
Non-MRSA Growth | Presumptive
MRSA Growth | Total |
| Digital Image | No Growth | 250 | 1 | - | 251 |
| | Presumptive
Non-MRSA Growth | 10 | 82 | 1 | 93 |
| | Presumptive
MRSA Growth | 2 | 1 | 53 | 56 |
| | Total | 262 | 84 | 54 | 400 |
| | No Growth Agreement: 95.4% (250/262)
Presumptive Non-MRSA Agreement: 97.6% (82/84)
Presumptive MRSA Agreement: 98.1% (53/54) | | | | |
Table 13. APAS Independence with TFS/S MRSA analysis module – Microbiologist 1 Digital Image vs Manual Plate Interpretation
| TFS/S analysis module
Microbiologist 1 | Manual Plate | | | | |
| --- | --- | --- | --- | --- | --- |
| | | No Growth | Presumptive
Non-MRSA Growth | Presumptive
MRSA Growth | Total |
| Digital Image | No Growth | 182 | 7 | 3 | 192 |
| | Presumptive
Non-MRSA Growth | 10 | 88 | 5 | 103 |
| | Presumptive
MRSA Growth | 2 | 6 | 97 | 105 |
| | Total | 194 | 101 | 105 | 400 |
| | No Growth Agreement: 93.8% (182/194)
Presumptive Non-MRSA Agreement: 87.1% (88/101)
Presumptive MRSA Agreement: 92.4% (97/105) | | | | |
Table 14. APAS Independence with TFS/S MRSA analysis module – Microbiologist 2 Digital Image vs Manual Plate Interpretation
| TFS/S analysis module
Microbiologist 2 | Manual Plate | | | | |
| --- | --- | --- | --- | --- | --- |
| | | No Growth | Presumptive
Non-MRSA Growth | Presumptive
MRSA Growth | Total |
| Digital Image | No Growth | 193 | 8 | 4 | 205 |
| | Presumptive
Non-MRSA Growth | 4 | 89 | 4 | 97 |
| | Presumptive
MRSA Growth | 4 | 1 | 93 | 98 |
| | Total | 201 | 98 | 101 | 400 |
| | No Growth Agreement: 96.0% (193/201)
Presumptive Non-MRSA Agreement: 90.8% (89/98)
Presumptive MRSA Agreement: 92.1% (93/101) | | | | |
K200839 - Page 16 of 30
{16}
Table 15. APAS Independence with TFS/S MRSA analysis module – Microbiologist 3 Digital Image vs Manual Plate Interpretation
| TFS/S analysis module
Microbiologist 3 | Manual Plate | | | | |
| --- | --- | --- | --- | --- | --- |
| | | No Growth | Presumptive
Non-MRSA Growth | Presumptive
MRSA Growth | Total |
| Digital Image | No Growth | 180 | 12 | 3 | 195 |
| | Presumptive
Non-MRSA Growth | 10 | 92 | 1 | 103 |
| | Presumptive
MRSA Growth | 1 | 2 | 99 | 102 |
| | Total | 191 | 106 | 103 | 400 |
| | No Growth Agreement: 94.2% (180/191)
Presumptive Non-MRSA Agreement: 86.8% (92/106)
Presumptive MRSA Agreement: 96.1% (99/103) | | | | |
b) Reproducibility of the APAS-generated digital image interpretation
The data obtained from the digital image quality study was further analyzed to evaluate the ability of an APAS-generated digital image to be reproducibly interpreted by a panel of microbiologists. Here, each microbiologist's interpretation of a digital image was compared to the final panel interpretation of the digital image (i.e., majority result). The study determined that a digital image generated with both the BD and TFS/S analysis modules can be reproducibly interpreted by a microbiologist. This was achieved >95% of the time for each of the three result designations (summarized in Table 16), which is acceptable.
Table 16. Digital Image Reproducibility Summary
| | No. of Results | Percent Agreement¹ | | | |
| --- | --- | --- | --- | --- | --- |
| | | No Growth | Presumptive Non-MRSA Growth | Presumptive MRSA Growth | |
| BD analysis module | Microbiologist 1 | 400 | 98.9% (260/263) | 97.6% (80/82) | 100% (55/55) |
| | Microbiologist 2 | 400 | 100% (263/263) | 97.6% (80/82) | 100% (55/55) |
| | Microbiologist 3 | 400 | 95.4% (251/263) | 100% (82/82) | 100% (55/55) |
| | Combined | 1200 | 98.1% (774/789) | 98.4% (242/246) | 100% (165/165) |
| TFS/S analysis module | Microbiologist 1 | 400 | 98.0% (192/196) | 95.1% (98/103) | 98.0% (99/101) |
| | Microbiologist 2 | 400 | 100% (196/196) | 93.2% (96/103) | 97.0% (98/101) |
| | Microbiologist 3 | 400 | 99.0% (194/196) | 97.1% (100/103) | 98.0% (99/101) |
| | Combined | 1200 | 99.0% (582/588) | 95.1% (294/309) | 97.7% (296/303) |
¹Percent agreement determined by dividing number of individual microbiologist digital read results by the number of panel digital image microbiologist results for each designation (i.e., "no growth", "presumptive non-MRSA growth", "presumptive MRSA growth").
2. Linearity:
Not applicable
3. Analytical Specificity/Interference:
K200839 - Page 17 of 30
{17}
# Microbial Interference
A study was conducted to assess the ability of APAS Independence with MRSA Analysis Modules to accurately detect MRSA in the presence of non-MRSA growth within the recommended incubation time for each respective plate (i.e., 20 - 26 hours for BD and 24 hours for TFS/S chromogenic agar plates).
Two MRSA (MRSA3 and MRSA5) isolates and two non-MRSA (Staphylococcus haemolyticus, SH6 and Staphylococcus warneri, STSP4) organisms were prepared as pure and mixed cultures and diluted accordingly to produce culture plates with mainly confluent growth ( $>100$ colonies per plate) or light growth ( $1 - 100$ colonies per plate) in five replicates. After 24 hours of incubation, a panel of three microbiologists evaluated the presence or absence of growth and color indicative of MRSA. The majority panel result (i.e., final panel interpretation) was compared to the APAS result to determine agreement for growth and color.
The data provided in Table 17 and Table 18 supports interpretation of the BD and TFS/S chromogenic agar plates after 24 hours of incubation.
Table 17. Microbial Interference - APAS Independence with the BD MRSA Analysis Module
| Organism(s) Plated (Organism Code2) | Expected Growth | | Percent Agreement1 | |
| --- | --- | --- | --- | --- |
| | MRSA | Other | Growth | Color |
| MRSA (MRSA3) | light | - | 100.0% (5/5) | 100.0% (5/5) |
| MRSA (MRSA3) | confluent | - | 100.0% (5/5) | 100.0% (5/5) |
| MRSA (MRSA5) | light | - | 100.0% (3/3) | 100.0% (3/3) |
| MRSA (MRSA5) | confluent | - | 100.0% (7/7) | 100.0% (7/7) |
| S. haemolyticus (SH6) | - | light | 100.0% (5/5) | 100.0% (5/5) |
| S. haemolyticus (SH6) | - | confluent | 100.0% (5/5) | 60.0% (3/5) |
| S. warneri (STSP4) | - | light | 100.0% (5/5) | 100.0% (5/5) |
| S. warneri (STSP4) | - | confluent | 100.0% (5/5) | 80.0% (4/5) |
| MRSA (MRSA3) and S. haemolyticus (1:1) | confluent | confluent | 100.0% (5/5) | 100.0% (5/5) |
| MRSA (MRSA5) and S. warneri (1:1) | confluent | confluent | 100.0% (5/5) | 100.0% (5/5) |
| MRSA (MRSA3) and S. haemolyticus (1:100) | light | confluent | 100.0% (5/5) | 100.0% (5/5) |
| MRSA (MRSA5) and S. warneri (1:100) | light | confluent | 100.0% (5/5) | 100.0% (5/5) |
1 Agreement between majority panel microbiologist result (i.e., final panel interpretation) and APAS result.
2 Organism code, as noted in Table 4.
Table 18. Microbial Interference - APAS Independence with the TFS/S MRSA Analysis Module
| Organism(s) Plated (Organism Code2) | Expected CFU | | Percent Agreement1 | |
| --- | --- | --- | --- | --- |
| | MRSA | Other | Growth | Color |
| MRSA (MRSA3) | light | - | 100.0% (4/4) | 100.0% (4/4) |
| MRSA (MRSA3) | confluent | - | 100.0% (6/6) | 100.0% (6/6) |
| MRSA (MRSA5) | light | - | 100.0% (2/2) | 100.0% (2/2) |
K200839 - Page 18 of 30
{18}
K200839 - Page 19 of 30
| Organism(s) Plated (Organism Code²) | Expected CFU | | Percent Agreement¹ | |
| --- | --- | --- | --- | --- |
| | MRSA | Other | Growth | Color |
| MRSA (MRSA5) | confluent | - | 100.0% (8/8) | 100.0% (8/8) |
| S. haemolyticus (SH6) | - | light | 100.0% (5/5) | 100.0% (5/5) |
| S. haemolyticus (SH6) | - | confluent | 100.0% (5/5) | 80.0% (4/5) |
| S. warneri (STSP4) | - | light | 100.0% (5/5) | 100.0% (5/5) |
| S. warneri (STSP4) | - | confluent | 100.0% (5/5) | 100.0% (5/5) |
| MRSA (MRSA3) and S. haemolyticus (1:1) | confluent | confluent | 100.0% (5/5) | 100.0% (5/5) |
| MRSA (MRSA5) and S. warneri (1:1) | confluent | confluent | 100.0% (5/5) | 100.0% (5/5) |
| MRSA (MRSA3) and S. haemolyticus (1:100) | light | confluent | 100.0% (5/5) | 100.0% (5/5) |
| MRSA (MRSA5) and S. warneri (1:100) | light | confluent | 100.0% (5/5) | 100.0% (5/5) |
¹ Agreement between majority panel microbiologist result (i.e., final panel interpretation) and APAS result.
² Organism code, as noted in Table 4.
## Analytical Specificity / Limit of Blank
A study was conducted to evaluate the potential influence of various external factors on image quality and the ability of the APAS Independence with MRSA Analysis Modules to provide correct results in the presence of known factors. Chromogenic agar plates (BD or TFS/S) were prepared to concurrently simulate a combination of four interference mechanisms (label, applicator, transport media and pen marking). Three blank applicators (flocked swab, cotton swab and 10 µL loop) were used to inoculate chromogenic agar plates using three different transport media (Stuart Transport, Amies Transport and saline) and no transport media (dry). One of four different labels were applied to plates that were either marked with felt-tip pen or not (3 applicators x 4 transport media x 4 labels x 2 marks = 96 plates per chromogenic agar). Although each interference mechanism was evaluated concurrently (i.e., four variables evaluated on each plate), performance is stratified by each interference variable in Table 19.
Table 19. Analytical Specificity (Limit of Blank) Summary
| Interference Mechanism | Interference Variable | Total number of images evaluated | BD Agreement | | TFS/S Agreement | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | Expected Growth (%)¹ | Expected Color (%)² | Expected Growth (%)¹ | Expected Color (%)² |
| Label Type | Paper 1D barcode on base | 24 | 20/24 (83.3) | 24/24 (100) | 58.3 (14/24) | 100.0 (24/24) |
| | Paper 2D barcode on base | 24 | 19/24 (79.2) | 23/24 (95.8) | 66.7 (16/24) | 100.0 (24/24) |
| | Plastic 2D barcode on base | 24 | 18/24 (75.0) | 24/24 (100) | 54.2 (13/24) | 100.0 (24/24) |
| | Paper 1D barcode on side | 24 | 17/24 (70.8) | 24/24 (100) | 58.3 (14/24) | 91.7 (22/24) |
| | Cotton Swab | 32 | 21/32 (65.6) | 31/32 (96.9) | 62.5 (20/32) | 100.0 (32/32) |
{19}
| Interference Mechanism | Interference Variable | Total number of images evaluated | BD Agreement | | TFS/S Agreement | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | Expected Growth (%)1 | Expected Color (%)2 | Expected Growth (%)1 | Expected Color (%)2 |
| Applicator Type | Flocked Swab | 32 | 27/32 (84.3) | 32/32 (100) | 50.0 (16/32) | 93.8 (30/32) |
| | Loop | 32 | 26/32 (81.2) | 32/32 (100) | 65.6 (21/32) | 100.0 (32/32) |
| Transport Media Type | Amies Transport | 24 | 22/24 (91.7) | 24/24 (100) | 62.5 (15/24) | 95.8 (23/24) |
| | Stuart Transport | 24 | 13/24 (54.2) | 23/24 (95.8) | 12.5 (3/24) | 95.8 (23/24) |
| | Saline | 24 | 21/24 (87.5) | 24/24 (100) | 91.7 (22/24) | 100.0 (24/24) |
| | Dry | 24 | 18/24 (75.0) | 24/24 (100) | 70.8 (17/24) | 100.0 (24/24) |
| Pen Markings | No Pen | 48 | 36/48 (75.0) | 47/48 (97.9) | 62.5 (30/48) | 97.9 (47/48) |
| | With Pen | 48 | 38/48 (79.2) | 48/48 (100) | 56.2 (27/48) | 97.9 (47/48) |
1 Value calculated by dividing the number of images with the expected growth (i.e., no growth detected) by the total number of images.
2 Value calculated by dividing the number of images with the expected color (i.e., no color detected) by the total number of images.
In general, the interference variables did interfere with the ability of the APAS Independence with MRSA Analysis Modules to provide accurate "no growth" result, which may produce false positives growth. However, since BD plates with "presumptive MRSA growth" and TFS/S plates with "presumptive MRSA growth" and "presumptive non-MRSA growth" are reviewed by microbiologists before confirmatory testing, the risk for an erroneous result being reported for patient management is mitigated by the review. In addition, plates with "no growth" or "presumptive non-MRSA growth" do not pose a risk for patient management and are therefore discarded.
## Analytical Specificity / Interference
The ability in which the APAS Independence with MRSA Analysis Modules can accurately detect presumptive MRSA and presumptive non-MRSA colonies in the presence of known interferences was evaluated. Three applicators (flocked swab, cotton swab and $10~\mu \mathrm{L}$ loop) were used to inoculate chromogenic agar plates using two different media (Stuart Transport and Amies Transport). Plates were labelled with one of four different labels applied to the base or side of the plate that were either marked with felt-tip pen or not (3 applicators x 2 media x 4 labels x 2 marks = 48 plates per organism, per chromogenic agar). One MRSA (MRSA3) and one non-MRSA (S. haemolyticus, SH6) isolate were prepared to a 0.5 McFarland suspension and diluted accordingly to produce predominantly isolated colonies or heavy growth, respectively, when cultured. Plates were incubated at $36^{\circ}\mathrm{C} \pm 1^{\circ}\mathrm{C}$ for 24 hours then imaged. Performance is summarized in Table 20.
Table 20. Analytical Specificity (Interference) Summary
| Organism (Code1) | Interference Mechanism | Interference Variable | Number of images evaluated | BD Agreement | | TFS/S Agreement | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | Growth detected (%)2 | Color detected (%)3 | Growth detected (%)2 | Color detected (%)3 |
| MRSA (MRSA3) | Label Type | Paper 1D barcode on base | 12 | 12/12 (100) | 12/12 (100) | 12/12 (100) | 12/12 (100) |
| | | Paper 2D barcode on base | 12 | 12/12 (100) | 12/12 (100) | 12/12 (100) | 12/12 (100) |
K200839 - Page 20 of 30
{20}
| Organism (Code1) | Interference Mechanism | Interference Variable | Number of images evaluated | BD Agreement | | TFS/S Agreement | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | Growth detected (%)2 | Color detected (%)3 | Growth detected (%)2 | Color detected (%)3 |
| | | Plastic 2D barcode on base | 12 | 12/12 (100) | 12/12 (100) | 12/12 (100) | 12/12 (100) |
| | | Paper 1D barcode on side | 12 | 12/12 (100) | 12/12 (100) | 12/12 (100) | 12/12 (100) |
| | Applicator Type | Cotton Swab | 16 | 16/16 (100) | 16/16 (100) | 16/16 (100) | 16/16 (100) |
| | | Flocked Swab | 16 | 16/16 (100) | 16/16 (100) | 16/16 (100) | 16/16 (100) |
| | | Loop | 16 | 16/16 (100) | 16/16 (100) | 16/16 (100) | 16/16 (100) |
| | Media Type | Amies Transport | 24 | 24/24 (100) | 24/24 (100) | 24/24 (100) | 24/24 (100) |
| | | Stuart Transport | 24 | 24/24 (100) | 24/24 (100) | 24/24 (100) | 24/24 (100) |
| | Pen Markings | No Pen | 24 | 24/24 (100) | 24/24 (100) | 24/24 (100) | 24/24 (100) |
| | | With Pen | 24 | 24/24 (100) | 24/24 (100) | 24/24 (100) | 24/24 (100) |
| S. haemolyticus (SH6) | Label Type | Paper 1D barcode on base | 12 | 12/12 (100) | 5/12 (41.7) | 12/12 (100) | 12/12 (100) |
| | | Paper 2D barcode on base | 12 | 12/12 (100) | 2/12 (16.7) | 12/12 (100) | 12/12 (100) |
| | | Plastic 2D barcode on base | 12 | 12/12 (100) | 6/12 (50.0) | 12/12 (100) | 12/12 (100) |
| | | Paper 1D barcode on side | 12 | 12/12 (100) | 12/12 (100) | 12/12 (100) | 11/12 (91.7) |
| | Applicator Type | Cotton Swab | 16 | 16/16 (100) | 7/16 (43.8) | 16/16 (100) | 15/16 (93.8) |
| | | Flocked Swab | 16 | 16/16 (100) | 8/16 (50.0) | 16/16 (100) | 16/16 (100) |
| | | Loop | 16 | 16/16 (100) | 10/16 (62.5) | 16/16 (100) | 16/16 (100) |
| | Media Type | Amies Transport | 24 | 24/24 (100) | 12/24 (50.0) | 24/24 (100) | 24/24 (100) |
| | | Stuart Transport | 24 | 24/24 (100) | 13/24 (54.2%) | 24/24 (100) | 23/24 (95.8) |
| | Pen Markings | No Pen | 24 | 24/24 (100) | 13/24 (54.2%) | 24/24 (100) | 23/24 (95.8) |
| | | With Pen | 24 | 24/24 (100) | 12/24 (50.0) | 24/24 (100) | 24/24 (100) |
1 Organism code, as noted in Table 4.
2 Value calculated by dividing the number of images with the expected growth (i.e., growth detected or no growth detected) by the total number of images.
3 Value calculated by dividing the number of images with the expected color (i.e., no color detected) by the total number of images.
In general, the interference variables tested had no impact on the ability of the BD and TFS/S MRSA Analysis Modules to detect presumptive MRSA growth (100% detection). False positive results were obtained with the S. haemolyticus samples; however, the risk of false positive results is mitigated since all plates with "presumptive MRSA growth" are reviewed by a microbiologist prior to confirmatory testing.
K200839 - Page 21 of 30
{21}
4. Assay Reportable Range:
Not applicable
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
The device labeling indicates that the APAS Independence with MRSA Analysis Module requires daily color calibration and a System Check using dedicated tools provided with the instrument. In addition, daily biological Quality Control testing must be performed using reference culture plates inoculated with a MRSA positive strain (e.g., Staphylococcus aureus ATCC 43300). In order to pass, results from the culture media must meet the expected designation (i.e., "presumptive MRSA growth"). There were 22 QC plates tested during the clinical study and 8 QC plates tested during the analytical studies. All QC results passed each day of testing with both the BD and TFS/S MRSA analysis modules.
6. Detection Limit:
Limit of Detection
A study was conducted to determine the smallest colony size that can be detected by the APAS Independence with MRSA Analysis Modules. The effect of colony size on the ability of APAS Independence with MRSA Analysis Modules to detect presumptive MRSA colonies was evaluated. Two strains of MRSA (MRSA3 and MRSA9) used to evaluate presumptive MRSA growth were prepared as 0.5 McFarland suspensions in saline, diluted and plated in six replicates to grow < 100 colonies per plate. Plates were incubated at 36°C ±1°C until pinpoint colonies were observed. Then, plates were imaged and returned to the incubator at 1-hour intervals until the colonies on the plates were approximately 2 mm in diameter (up to 24 hours).
Table 21 summarizes the percent of colonies detected by APAS by colony size (pixels and extrapolated mm). The smallest colony size (SCZ) that was detected as presumptive MRSA >95% was determined to be the LoD. The SCZ detected by the BD analysis module was 0.815 mm for MRSA3 and 0.906 mm for MRSA9. The SCZ detected by the TFS/S analysis module was 0.634 mm for MRSA3 and 0.543 mm for MRSA 9.
Table 21. Limit of Detection (Smallest Colony Size Detected)
| | | APAS-Detected MRSA Colonies / Total Number of Colonies (% of total) | | | |
| --- | --- | --- | --- | --- | --- |
| Colony Size (diameter) | | BD | | TFS/S | |
| Pixels | Extrapolated mm | MRSA31 | MRSA91 | MRSA31 | MRSA91 |
| 1 – 3 | 0.091 – 0.272 | 0/399 (0%) | 1/264 (0.4%) | 0/285 (0%) | 3/163 (1.8%) |
| 4 – 5 | 0.362 – 0.453 | 4/258 (1.6%) | 3/175 (1.7%) | 12/159 (7.5%) | 98/228 (43.0%) |
| 6 | 0.543 | 7/135 (5.2%) | 9/93 (9.7%) | 63/89 (70.8%) | 129/132 (97.7%) |
| 7 | 0.634 | 21/113 (18.6%) | 18/79 (22.8%) | 145/148 (98.0%) | 100/100 (100%) |
| 8 | 0.725 | 66/91 (72.5%) | 36/84 (42.9%) | 197/198 (99.5%) | 71/71 (100%) |
| 9 | 0.815 | 117/123 (95.1%) | 50/74 (67.6%) | 167/167 (100%) | 43/43 (100%) |
| 10 | 0.906 | 147/150 (98.0%) | 84/86 (97.7%) | - | - |
| 11 | 0.996 | 130/130 (100%) | 95/96 (99.0%) | - | - |
| 12 | 1.087 | 156/156 (100% | 90/90 (100%) | - | - |
1 Organism code, as noted in Table 4.
K200839 - Page 22 of 30
{22}
7. Assay Cut-Off:
Not applicable
B Comparison Studies:
1. Method Comparison with Predicate Device:
Not applicable
2. Matrix Comparison:
Not applicable
C Clinical Studies:
1. Clinical Sensitivity:
The clinical performance of the APAS Independence with MRSA Analysis Modules was evaluated in a two-part study. The overall data from both parts were used collectively in the evaluation of performance.
For part 1 of the clinical study, the goal was to evaluate performance of the APAS Independence with MRSA Analysis Modules by inoculating prospective and contrived clinical samples onto chromogenic agar media and comparing APAS-generated results to a panel (i.e., majority) result. This part of the study was performed at a single clinical site (outside United States, oUS) using remnant swab specimens that were leftover from routine, standard of care testing from various healthcare facilities. Each swab specimen was subjected to PCR testing for the detection of MRSA. Positive PCR samples were sent to a confirmatory lab for culture. Both positive and negative PCR samples were inoculated onto chromogenic agar and shipped to a central laboratory for incubation and imaging. A portion of PCR-negative samples were also used to generate contrived-negative and contrived-positive samples to supplement the study. The on-screen (digital) images were interpreted by a panel of three oUS microbiologists. The majority result was considered the truth state (i.e., final panel interpretation) and used as a comparison to the APAS-generated result.
For part 2 of the clinical study, the goal was to assess the ability of a panel of three US microbiologists to interpret the same set of randomized digital images already acquired and reported in part 1. The microbiologists were blinded to the other interpretations and plate source (clinical or contrived). The majority result was compared to the historical APAS result obtained in the first part of the study.
The following 1590 samples were enrolled in the clinical study:
- 1100 PCR-negative clinical samples
- 25 PCR-positive clinical samples
- 60 PCR-negative, contrived-negative samples
- 405 PCR-negative, contrived-positive samples
K200839 - Page 23 of 30
{23}
Thirteen samples were excluded from the BD analysis due to defects in the agar resulting in a total of 1573 BD plates analyzed. Ten samples were excluded from the TFS/S analysis due to defects in the agar, resulting in a total of 1580 TFS/S plates analyzed.
Contrived samples were used due to the low incidence of MRSA in the population. This was acceptable since the APAS Independence with MRSA Analysis Modules only images and interprets microbial colonies (i.e., it is not a selective and differential medium). Leftover PCR-negative swab specimens were spiked (1 - 10,000 CFU/plate) with MRSA organisms known to produce a positive reaction on the agar ("presumptive MRSA growth") or organisms known to grow on the agar but not produce a positive reaction ("presumptive non-MRSA growth"). Some organisms included in the contrived samples are representative of the most common strains from the US, Europe, United King and Australia (Table 22). The other contrived samples were spiked with well-characterized stock clinical isolates obtained by LBT Innovations and identification confirmed using MALDI-TOF with methicillin susceptibility/resistance confirmed using well-accepted methods (i.e., oxacillin MIC assay using an FDA-cleared device or cefoxitin disk diffusion assay).
Table 22. Global MRSA strains used in simulated samples
| MRSA Strain (Organism Number) | Type | Other Name & Information |
| --- | --- | --- |
| NCTC 13395 (MRSA30) | HA | IRISH 2/ IBERIAN ST 247. It is CC8 but different from USA 300. |
| NCTC 13277 (MRSA29) | HA | MRSA-252 |
| NCTC 13656 (MRSA34) | CA | PVL-negative CA-MRSA strain belonging to clonal complex 59, a clone that originated in East Asia. |
| NCTC 13435 (MRSA31) | CA | PVL-positive CA-MRSA strain belonging to clonal complex 80, commonly known as European clone of CA MRSA |
| NCTC 13616 (MRSA32) | HA | EMRSA 15, widely distributed in Europe, this HA strain has now been found in the USA |
| ATCC 1707 (MRSA26) | CA | USA 400, ST1-IV |
| ATCC 1717 (MRSA27) | CA/HA | USA 300, ST8-IV, causes most community-associated MRSA infections and is an increasingly common cause of health care-associated MRSA infections |
| ATCC 1761 (MRSA28) | HA | USA100 MRSA |
HA: Hospital-associated MRSA strain
CA: Community-associated MRSA strain
K200839 - Page 24 of 30
{24}
# Part 1 Clinical Study
Performance of the APAS Independence with MRSA Analysis Modules for the first part of the clinical study (i.e., oUS site) is shown in Table 23 (clinical samples), Table 24 (contrived samples), and Table 25 (clinical and contrived samples combined). The overall agreement of the BD MRSA analysis module (when evaluating combined samples) is $82.6\%$ for "no growth" detection, $84.6\%$ for "presumptive non-MRSA growth", and $100\%$ for "presumptive MRSA growth". The overall agreement of the TFS/S MRSA analysis module (when evaluating clinical and contrived samples, combined) is $81.8\%$ for "no growth" detection, $83.8\%$ for "presumptive non-MRSA growth", and $99.2\%$ for "presumptive MRSA growth".
Table 23. Clinical Study Part 1 - Clinical Samples
| Clinical Samples | Panel of oUS Microbiologists | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | BD (clinical) | | | | TFS/S (clinical) | | | |
| | | No Growth | Presumptive Non-MRSA Growth | Presumptive MRSA Growth | Total | No Growth | Presumptive Non-MRSA Growth | Presumptive MRSA Growth | Total |
| APAS | No Growth | 730 | 11 | - | 741 | 508 | 4 | 1 | 513 |
| | Presumptive Non-MRSA Growth | 148 | 179 | - | 327 | 114 | 316 | 3 | 433 |
| | Presumptive MRSA Growth | 3 | 13 | 34 | 50 | 1 | 62 | 113 | 176 |
| | Total | 881 | 203 | 34 | 1118 | 623 | 382 | 117 | 1122 |
| | | No Growth Agreement: 82.9% (730/881) Presumptive Non-MRSA Agreement: 88.2% (179/203) Presumptive MRSA Agreement: 100% (34/34) | | | | No Growth Agreement: 81.5% (508/623) Presumptive Non-MRSA Agreement: 82.7% (316/382) Presumptive MRSA Agreement: 96.6% (113/117) | | | |
Table 24. Clinical Study Part 1 - Contrived Samples
| Clinical Samples | Panel of oUS Microbiologists | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | BD (contrived) | | | | TFS/S (contrived) | | | |
| | | No Growth | Presumptive Non-MRSA Growth | Presumptive MRSA Growth | Total | No Growth | Presumptive Non-MRSA Growth | Presumptive MRSA Growth | Total |
| APAS | No Growth | 24 | - | - | 24 | 18 | 1 | - | 19 |
| | Presumptive Non-MRSA Growth | 8 | 47 | - | 55 | 2 | 67 | - | 69 |
| | Presumptive MRSA Growth | - | 17 | 359 | 376 | - | 7 | 363 | 370 |
| | Total | 32 | 64 | 359 | 455 | 20 | 75 | 363 | 458 |
| | | No Growth Agreement: 75.0% (24/32) Presumptive Non-MRSA Agreement: 73.4% (47/64) Presumptive MRSA Agreement: 100% (359/359) | | | | No Growth Agreement: 90.0% (18/20) Presumptive Non-MRSA Agreement: 89.3% (67/75) Presumptive MRSA Agreement: 100% (363/363) | | | |
K200839 - Page 25 of 30
{25}
Table 25. Clinical Study Part 1 - Combined Samples
| Combined Samples | Panel of oUS Microbiologists | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | BD (combined) | | | | TFS/S (combined) | | | |
| | | No Growth | Presumptive Non-MRSA Growth | Presumptive MRSA Growth | Total | No Growth | Presumptive Non-MRSA Growth | Presumptive MRSA Growth | Total |
| APAS | No Growth | 754 | 11 | - | 765 | 526 | 5 | 1 | 532 |
| | Presumptive Non-MRSA Growth | 156 | 226 | - | 382 | 116 | 383 | 3 | 502 |
| | Presumptive MRSA Growth | 3 | 30 | 393 | 426 | 1 | 69 | 476 | 546 |
| | Total | 913 | 267 | 393 | 1573 | 643 | 457 | 480 | 1580 |
| | | No Growth Agreement: 82.6% (754/913)
Presumptive Non-MRSA Agreement: 84.6% (226/267)
Presumptive MRSA Agreement: 100% (393/393) | | | | No Growth Agreement: 81.8% (526/643)
Presumptive Non-MRSA Agreement: 83.8% (383/457)
Presumptive MRSA Agreement: 99.2% (476/480) | | | |
Part 2 Clinical Study
Performance of the APAS Independence with MRSA Analysis Modules for the second part of the clinical study (i.e., US site) is shown in Table 26 (clinical samples), Table 27 (contrived samples), and Table 28 (clinical and contrived samples combined). The overall agreement of the BD MRSA analysis module (when evaluating combined samples) is 82.7% for "no growth" detection, 84.0% for "presumptive non-MRSA growth", and 99.7% for "presumptive MRSA growth". The overall agreement of the TFS/S MRSA analysis module (when evaluating clinical and contrived samples, combined) is 79.7% for "no growth" detection, 76.4% for "presumptive non-MRSA growth", and 99.5% for "presumptive MRSA growth".
Table 26. Clinical Study Part 2 - Clinical Samples
| Clinical Samples | Panel of US Microbiologists | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | BD (clinical) | | | | TFS/S (clinical) | | | |
| | | No Growth | Presumptive Non-MRSA Growth | Presumptive MRSA Growth | Total | No Growth | Presumptive Non-MRSA Growth | Presumptive MRSA Growth | Total |
| APAS | No Growth | 731 | 10 | - | 741 | 509 | 3 | 1 | 513 |
| | Presumptive Non-MRSA Growth | 149 | 177 | 1 | 327 | 130 | 302 | 1 | 433 |
| | Presumptive MRSA Growth | 3 | 15 | 32 | 50 | 2 | 102 | 72 | 176 |
| | Total | 883 | 202 | 33 | 1118 | 641 | 407 | 74 | 1122 |
| | | No Growth Agreement: 82.8% (731/883)
Presumptive Non-MRSA Agreement: 87.6% (177/202)
Presumptive MRSA Agreement: 97.0% (32/33) | | | | No Growth Agreement: 79.4% (509/641)
Presumptive Non-MRSA Agreement: 74.2% (302/407)
Presumptive MRSA Agreement: 97.3% (72/74) | | | |
K200839 - Page 26 of 30
{26}
Table 27. Clinical Study Part 2 - Contrived Samples
| Contrived Samples | Panel of US Microbiologists | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | BD (contrived) | | | | TFS/S (contrived) | | | |
| | | No Growth | Presumptive Non-MRSA Growth | Presumptive MRSA Growth | Total | No Growth | Presumptive Non-MRSA Growth | Presumptive MRSA Growth | Total |
| APAS | No Growth | 24 | - | - | 24 | 18 | 1 | - | 19 |
| | Presumptive Non-MRSA Growth | 6 | 49 | - | 55 | 2 | 67 | - | 69 |
| | Presumptive MRSA Growth | - | 18 | 358 | 376 | - | 8 | 362 | 370 |
| | Total | 30 | 67 | 358 | 455 | 20 | 76 | 362 | 458 |
| | | No Growth Agreement: 80.0% (24/30)
Presumptive Non-MRSA Agreement: 73.1% (49/67)
Presumptive MRSA Agreement: 100% (358/358) | | | | No Growth Agreement: 90.0% (18/20)
Presumptive Non-MRSA Agreement: 88.2% (67/76)
Presumptive MRSA Agreement: 100% (362/362) | | | |
Table 28. Clinical Study Part 2 - Combined Samples
| Combined Samples | Panel of US Microbiologists | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | BD (combined) | | | | TFS/S (combined) | | | |
| | | No Growth | Presumptive Non-MRSA Growth | Presumptive MRSA Growth | Total | No Growth | Presumptive Non-MRSA Growth | Presumptive MRSA Growth | Total |
| APAS | No Growth | 755 | 10 | - | 765 | 527 | 4 | 1 | 532 |
| | Presumptive Non-MRSA Growth | 155 | 226 | 1 | 382 | 132 | 369 | 1 | 502 |
| | Presumptive MRSA Growth | 3 | 33 | 390 | 426 | 2 | 110 | 434 | 546 |
| | Total | 913 | 269 | 391 | 1573 | 661 | 483 | 436 | 1580 |
| | | No Growth Agreement: 82.7% (755/913)
Presumptive Non-MRSA Agreement: 84.0% (226/269)
Presumptive MRSA Agreement: 99.7% (390/391) | | | | No Growth Agreement: 79.7% (527/661)
Presumptive Non-MRSA Agreement: 76.4% (369/483)
Presumptive MRSA Agreement: 99.5% (434/436) | | | |
An analysis was conducted to assess concordance of results between part 1 and part 2 of the clinical study. As shown in Table 29, the combined data from part 1 (oUS) was used as the comparator for the combined data from part 2 (US). The overall agreement of the BD MRSA analysis module is 98.6% for "no growth" detection, 94.8% for "presumptive non-MRSA growth", and 99.2% for "presumptive MRSA growth". The overall agreement of the TFS/S MRSA analysis module is 99.1% for "no growth" detection, 94.5% for "presumptive non-MRSA growth", and 90.6% for "presumptive MRSA growth".
K200839 - Page 27 of 30
{27}
Table 29. Clinical Study Part 1 (oUS) vs Part 2 (US) - Combined Samples
| Combined Samples | oUS Results | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | BD (combined) | | | | TFS/S (combined) | | | |
| | | No Growth | Presumptive Non-MRSA Growth | Presumptive MRSA Growth | Total | No Growth | Presumptive Non-MRSA Growth | Presumptive MRSA Growth | Total |
| US Results | No Growth | 900 | 13 | - | 913 | 637 | 24 | - | 661 |
| | Presumptive Non-MRSA Growth | 13 | 253 | 3 | 269 | 6 | 432 | 45 | 483 |
| | Presumptive MRSA Growth | - | 1 | 390 | 391 | - | 1 | 435 | 436 |
| | Total | 913 | 267 | 393 | 1573 | 643 | 457 | 480 | 1580 |
| | | No Growth Agreement: 98.6% (900/913)
Presumptive Non-MRSA Agreement: 94.8% (253/267)
Presumptive MRSA Agreement: 99.2% (390/393) | | | | No Growth Agreement: 99.1% (637/643)
Presumptive Non-MRSA Agreement: 94.5% (432/457)
Presumptive MRSA Agreement: 90.6% (435/480) | | | |
Overall Performance of the APAS Independence with the MRSA Analysis Modules
The overall performances for the BD MRSA and TFS/S MRSA analysis modules were evaluated to determine the possible claims that could be supported and how performance issues could be mitigated. As a result, considering data and supportive metrics from other studies, it was concluded that either a two-designation rule set (i.e., presumptive MRSA growth and negative) or a three-designation rule set (i.e., presumptive MRSA growth, presumptive non-MRSA growth, and negative) would be appropriate.
a) Overall Performance of the APAS Independence with the BD MRSA Analysis Modules
The overall performance of the BD MRSA analysis modules for detection of presumptive MRSA growth (both in the analytical studies and clinical study) is acceptable to support a two-designation rule set (i.e., presumptive MRSA growth and negative). As such, only plates with "presumptive MRSA growth" require review by a microbiologist. The acceptability of the two-designation rule set is based on the following performance metrics:
- >95% agreement for "presumptive MRSA growth" in the digital image quality study (Table 9)
- >98% agreement for "presumptive MRSA growth" in part 1 (Table 25) and part 2 clinical data (Table 28)
- >95% agreement for "presumptive MRSA growth" when comparing part 1 and part 2 clinical data (Table 29)
K200839 - Page 28 of 30
{28}
b) Overall Performance of the APAS Independence with the TFS/S MRSA Analysis Modules The overall performance of the TFS/S MRSA analysis modules for detection of presumptive MRSA growth (both in the analytical studies and clinical study) is acceptable to support a three-designation rule set (i.e., presumptive MRSA growth, presumptive non-MRSA growth, and negative). As such, all plates with growth ("presumptive MRSA growth" or "presumptive non-MRSA growth) require review by a microbiologist. The acceptability of the three-designation rule set is based on the following performance metrics:
- <95% agreement for "presumptive MRSA growth" in the digital image quality study (Table 9)
- >98% agreement for "presumptive MRSA growth" in part 1 (Table 25) and part 2 clinical data (Table 28)
- <95% agreement for "presumptive MRSA growth" when comparing part 1 and part 2 clinical data (Table 29)
2. Clinical Specificity:
Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable.
D Clinical Cut-Off:
Not applicable.
E Expected Values/Reference Range:
A summary of the number and percentage of samples with the different result interpretations observed during parts 1 (oUS) and 2 (US) of the clinical study, as determined by the panel of oUS and US microbiologists (reference), and by the APAS Independence with MRSA Analysis Modules, is described in Table 30.
Table 30. Microbiologists (Reference) and APAS Results from Clinical Study
| Interpretation | BD1 | | | TFS/S | | |
| --- | --- | --- | --- | --- | --- | --- |
| | Microbiologists | | APAS | Microbiologists | | APAS |
| | oUS | US | | oUS | US | |
| No Growth | 913 (58.0%) | 913 (58.0%) | 765 (48.6%) | 643 (40.7%) | 661 (41.8%) | 532 (33.7%) |
| Presumptive Non-MRSA Growth | 267 (17.0%) | 269 (17.1%) | 382 (24.3%) | 457 (28.9%) | 483 (30.6%) | 502 (31.8%) |
| Presumptive MRSA Growth | 393 (25.0%) | 391 (24.9%) | 426 (27.1%) | 480 (30.4%) | 436 (27.6%) | 546 (34.6%) |
| TOTAL | 1573 | 1573 | 1573 | 1580 | 1580 | 1580 |
K200839 - Page 29 of 30
{29}
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K200839 - Page 30 of 30
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.