Published clinical literature; Routine clinical practice data
The sponsor utilized a literature review of clinical studies involving closed-system vitrification devices similar to the subject device to support the safety and effectiveness of the VitriGuard for oocyte cryopreservation.
Literature Review; Clinical Outcomes; Oocyte Vitrification; Closed-system device
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Literature 1-6 (Summary of clinical experience); Retrospective and prospective clinical cohorts/case series; Follow-up/Duration: Varies by study; includes long-term storage (median 3.5 years)
Patients undergoing oocyte vitrification; Sample Size: Aggregated data from multiple studies (e.g., 33, 498, 190, 89 oocytes; 33 patients)
Not applicable for this study
Survival rate, fertilization rate, cleavage rate, blastocyst development rate, implantation rate, clinical pregnancy rate, live birth rate
Indications for Use
VitriGuard is intended for use as a cryopreservation storage device in vitrification procedures and indicated to contain and maintain human oocytes (MII), 4-8 cell embryos and blastocyst stage embryos.
Device Story
VitriGuard is a sterile, single-use polystyrene cryopreservation storage device for vitrification procedures; consists of a hexagonal-shaped stick and cap. Oocytes or embryos are loaded into a trough at the tip of the stick; the stick is inserted into a pre-cooled cap to create a seal. The device is then plunged into liquid nitrogen for storage. Used in clinical laboratory settings by embryologists/technicians. Visual aid provided by black markings for orientation. The device maintains specimen integrity during the freezing and storage process; enables subsequent warming and clinical use in assisted reproduction. Benefits include secure containment of specimens during cryopreservation.
Clinical Evidence
No clinical trials performed on the subject device. Substantial equivalence supported by bench testing (sterilization validation, shelf life, package integrity, endotoxin, MEA, thermal profile, container/closure integrity, durability) and a literature review of six studies involving similar closed-system vitrification devices. Literature reported survival rates ranging from 80.4% to 94.5%, fertilization rates from 57.1% to 80.6%, and clinical pregnancy rates from 38.5% to 62.5%.
Technological Characteristics
Polystyrene construction; hexagonal stick and cap design with taper seal. Dimensions/form factor: two-piece assembly. Energy source: none (passive). Sterilization: Radiation (SAL 10^-6). Connectivity: none. MEA specification: ≥80% embryos developed to expanded blastocyst at 96h. Cooling rate: 36,377°C/min; Warming rate: 2,271°C/min. Endotoxin: ≤2.0 EU/device.
Indications for Use
Indicated for containment and maintenance of human oocytes (MII), 4-8 cell embryos, and blastocyst stage embryos during vitrification and cryopreservation procedures.
Regulatory Classification
Identification
Assisted reproduction labware consists of laboratory equipment or supplies intended to prepare, store, manipulate, or transfer human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures. These include syringes, IVF tissue culture dishes, IVF tissue culture plates, pipette tips, dishes, plates, and other vessels that come into physical contact with gametes, embryos or tissue culture media.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when it is a dish or plate intended for general assisted reproduction technology procedures, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 25, 2020
ORIGIO a/s % Christine Kupchick Regulatory Affairs Associate CooperSurgical, Inc. 95 Corporate Drive Trumbull, CT 06611
Re: K200815 Trade/Device Name: VitriGuard Regulation Number: 21 CFR§ 884.6160 Regulation Name: Assisted Reproduction Labware Regulatory Class: II Product Code: MQK Dated: March 27, 2020 Received: March 30, 2020
Dear Christine Kupchick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200815
Device Name VitriGuard
Indications for Use (Describe)
VitriGuard is intended for use as a cryopreservation storage device in vitrification procedures and indicated to contain and maintain human oocytes (MII), 4-8 cell embryos and blastocyst stage embryos.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# a CooperSurgical company
# 510(k) SUMMARY K200815
#### Submitter Information
| Company Name: | ORIGIO a/s |
|------------------|-------------------------------------|
| Company Address: | Knardrupvej 2<br>2760 Måløv Denmark |
| Telephone: | + 45 46 79 02 00 |
| Fax: | + 45 46 79 03 02 |
#### Submission Correspondent
| Company Name: | CooperSurgical Inc. |
|------------------|------------------------------------------|
| Company Address: | 95 Corporate Drive<br>Trumbull, CT 06611 |
### Submission Contact
| Name: | Christine Kupchick |
|------------|---------------------------------------|
| Telephone: | 203-601-5200 Ext. 3370 |
| Fax: | 203-601-9870 |
| Email: | christine.kupchick@coopersurgical.com |
Date Prepared: June 24, 2020
#### Device Identification
| Trade Name: | VitriGuard |
|--------------------|--------------------------------------|
| Common Name: | Cryopreservation Storage Device |
| Regulation Number: | 21 CFR 884.6160 |
| Regulation Name: | Assisted Reproduction Labware |
| Product Code: | MQK (Labware, Assisted Reproduction) |
| Regulatory Class: | Class II |
| Review Panel: | Obstetrics/Gynecology |
Predicate Device Information Vitrolife Rapid-I (K181461).
The predicate device has not been subject to a design-related recall.
#### Device Description
VitriGuard is a sterile, single-use device that is intended for use as a cryopreservation storage device in vitrification procedures. The device consists of a two-piece polystyrene assembly that includes a hexagonal-shaped stick and cap. As part of the vitrification procedure, the oocytes or embryos are loaded on the tip of the stick is capped prior to plunging the device in liquid nitrogen and for subsequent storage. The tip has a trough area for loading, maintaining and securing oocytes or embryos. The stick and cap include a taper design that create a seal when assembled. Black markings at
{4}------------------------------------------------
the end of the stick and the tip of the device provide visual aid for proper device orientation. All VitriGuard devices have a blue cap and the sticks are available in eight (8) translucent colors: clear, blue, green, yellow, lime green, purple, orange and pink.
#### Indications for Use
VitriGuard is intended for use as a cryopreservation storage device in vitrification procedures and indicated to contain and maintain human oocytes (MII), 4-8 cell embryos and blastocyst stage embryos.
#### Substantial Equivalence Discussion
The table below provides a comparison of the intended use and technological characteristics of the subject and predicate device.
| Attribute | Subject VitriGuard | Predicate Rapid-I (K181461) |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | VitriGuard is intended for use as<br>a cryopreservation storage device<br>in vitrification procedures and<br>indicated to contain and maintain<br>human oocytes (MII), 4-8 cell<br>embryos and blastocyst stage<br>embryos. | Cryopreservation device intended<br>to be used to contain, vitrify and<br>maintain human embryos and/or<br>oocytes (MII). |
| Fundamental<br>Technology | The device is composed of a two-<br>piece assembly that includes a<br>stick and cap. As part of the<br>vitrification procedure,<br>specimens to be stored are<br>loaded on the tip of the stick and<br>the stick is inserted into the pre-<br>cooled cap to seal and to vitrify<br>samples for subsequent storage<br>in LN. | The stick has a tip where the<br>samples are loaded. The stick is<br>sealed within a pre-cooled straw<br>that contains a stainless-steel<br>weight to maintain device<br>orientation in LN. The stick is<br>inserted into the pre-cooled straw<br>(after steel rod removal) to vitrify<br>samples. The end of the straw is<br>sealed, and the device is stored in<br>LN. |
| Warming Rate | -2,271°C/min | -1,400°C/min |
| Cooling Rate | 36,377°C/min | 10,000°C/min |
| Endotoxin | ≤2.0 EU/device | ≤1.0 EU/device |
| MEA (1-Cell) | ≥80% embryos developed to<br>expanded blastocyst at 96h | ≥80% embryos developed to<br>expanded blastocyst at 96h |
| Materials | Polystyrene (clear, blue, green,<br>yellow, orange, pink, lime green,<br>and purple) with black marker<br>bands | Polymethyl methacrylate (PMMA)<br>Mediprene<br>Stainless steel |
| Sterilization<br>Method | Radiation, SAL 10-6 | Ethylene Oxide, SAL 10-6 |
| Number of Uses | Single-use, disposable | Single-use, disposable |
#### Substantial Equivalence Comparison
The subject and predicate device have similar indications for use statements and have the same intended use – the vitrification and storage of human oocytes (MII) and embryos.
The subject and predicate device have comparable technological characteristics: samples are loaded at the tip, the stick is inserted into a pre-cooled device component, samples are vitrified, and the sealed device with the vitrified samples is placed in Liquid Nitrogen for subsequent storage. The subject device is identical to the predicate in MEA specifications, SAL, and number of uses. The subject device differs
{5}------------------------------------------------
from the predicate in cooling/warming rates, endotoxin specifications, materials, and sterilization method, however, these differences do not raise different questions of safety and effectiveness.
#### Non-clinical performance testing
The following performance testing was leveraged from the original VitriGuard (K162833) to support the performance of the subject device.
- . Sterilization
- Validation per ISO 11137-1:2006/(R)2010 and ISO 11137-2:2013 O
- Bioburden per ISO 11737-1:2006/(R)2011 and ISO 11737-2:2009 O
- Shelf Life Testing ●
- Accelerated aging per ASTM F1980-07(2011) o
- Package Integrity Testing following accelerated aging o
- l Dye Penetration Testing per ASTM F2096-11
- . Seal Tensile Strength Testing per ASTM F88/F88M-15
- . Seal Peel Testing per ASTM F1886/F1866M-09(2013)
- Ship Testing per ISTA 3A:2008
- Endotoxin per USP <85> and ANSI/AAMI ST72:2002/(R)2010
- . Mouse Embryo Assay (MEA)
- Thermal Profile Evaluation
- O Assessment of warming and cooling rates
- Comparison of warming and cooling rates to other devices cleared for oocyte and embryo O vitrification
- Mouse Embryo Survival Evaluation
- Mouse Embryo Development Evaluation
- Container and Closure Integrity
- O Bacterial/Immersion
- O Bacterial Contaminated LN2
- . Durability Testing
- Product Evaluation
#### Published literature review
To support the use of the VitriGuard for the vitrification of MII oocytes, published literature was summarized to assess the VitriGuard, as well as other legally marketed, closed-system devices similar to the subject device in technological characteristics and indicated to contain and maintain human oocytes during vitrification procedures. A summary of the results is shown below:
- Literature 1: This study evaluated the efficacy of a closed-system vitrification device for oocyte vitrification as compared to an open-system device. The results of the study show a 93.9% survival rate, 80.6% fertilization rate, 96% cleavage rate and 72% blastocyst development rate after using a closed- system device similar to the subject device. Results are based on 33 vitrified/ warmed donated oocytes using the closed-system device.2
- Literature 2: This study evaluated the efficacy of a closed-system vitrification device for oocyte vitrification as compared to an open-system device. The results of the study show a 94.5% survival rate, 57.1% fertilization rate, 24.2% implantation rate, 40% clinical pregnancy rate and 37.5% live birth rate after using a closed-system vitrification device similar to the subject device.
{6}------------------------------------------------
Results are based on 498 vitrified/warmed donated oocytes using the closed-system device, which were fertilized/ cultured and transferred to 40 recipients.2
- Literature 3: This study analyzed oocyte survival rates after warming, and birth outcome of a vitrification protocol in conjunction with a closed-system device. The results of the study show a 94.2% survival rate, 80.4% fertilization rate, 95.1% cleavage rate, 36.1% blastocyst development rate, and 57.1% live birth rate after using a closed-system vitrification device similar to the subject device. Results are based on 190 vitrified/warmed autologous oocytes using the closedsystem device, which were fertilized/ cultured and transferred back to the mother (N=14). 3
- Literature 4: This study reported on clinical experience after introducing a closed-system device for oocyte vitrification in daily routine. The overall results of the study show a 90.5% survival rate, 64.2% fertilization rate, 90.4% cleavage rate, 32.7% implantation rate, and 40.9% clinical pregnancy rate after using a closed-system vitrification device similar to the subject device; with results varying depending on maternal age (<38 yrs. versus ≥38 yrs.), including clinical pregnancy rates of 41.9% and 38.5% respectively. 4
- Literature 5: This study evaluated the efficacy of a closed-system vitrification device for oocyte vitrification as compared to an open-system device based on donated oocytes. The results of the study show a 92.1% survival rate after using a closed-system vitrification device similar to the subject device. Results are based on 89 vitrified/ warmed donated oocytes using the closed-system device.5
<sup>1</sup> Inoue (2014) Efficiency of a closed vitrification system with oocytes and blastocysts.
² Pujol (2019) Comparison of two different oocyte vitrification methods: a prospective, paired study on the same genetic background and stimulation protocol.
<sup>3</sup> Perez et al (2018) Oocyte vitrification using a new vitrification medium and a new closed vitrification device. A sibling oocyte study. 4 Gook et al (2016) Closed vitrification of human oocytes and blastocysts: outcomes from a series of clinical cases.
<sup>5</sup> Pinasco et al (2012) Oocyte vitrification freeze/thaw survival rates using an open versus closed system.
{7}------------------------------------------------
- Literature 6: This study evaluated if long-term cryopreservation (median 3.5 years storage) of human oocytes affects oocyte developmental competence, blastocyst euploidy, or live-birth rates. The results of the study show an overall 80.4% survival rate, 72.8% fertilization rate, 54.5% blastocyst development rate, 65% implantation rate, 62.5% clinical pregnancy rate and 62.5% live birth rate (including some ongoing pregnancies). Data include the use of a closedsystem vitrification device similar to the subject device. Results are based on total 33 patients ending up with 16 embryo transfer cycles.6
The published data summarized for closed-system vitrification devices similar to the subject device demonstrates the safe and effective use of the closed-system vitrification devices for oocyte cryopreservation and storage. Furthermore, a cooling/warming rate comparison demonstrates that the subject device has similar cooling/warming rates to other devices that have been cleared for oocyte vitrification and storage. Therefore, the clinical information provided above can be leveraged to demonstrate substantial equivalence to the predicate device and supports the safe and effective use of the subject device for oocyte cryopreservation and storage.
#### Conclusion
The results of the performance testing described above demonstrate that the VitriGuard is as safe and effective as the predicate device and supports a determination of substantial equivalence
<sup>6</sup> Goldman et al (2015) Long-term crypreservation of human oocytes does not increase embryonic aneuploidy.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.