K200798 · GC America, Inc. · EMA · Nov 24, 2020 · Dental
Device Facts
Record ID
K200798
Device Name
G-CEM ONE
Applicant
GC America, Inc.
Product Code
EMA · Dental
Decision Date
Nov 24, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3275
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
1. Cementation of all types of all ceramic, resin and metal-based inlays, onlays, crowns and bridges. 2. Cementation of metal, ceramic, fiber posts, and cast post and cores. 3. Cementation of all ceramic and composite veneers. 4. Final cementation of crowns and bridges on implant abutments.
Device Story
G-CEM ONE is a dual-cured self-adhesive resin cement used in dental clinics by dentists. It consists of a two-chamber syringe (Paste A and B) applied via an automix tip. The device functions through polymerization of methacrylate monomers, initiated chemically or via light irradiation. An optional 'G-CEM ONE ADHESIVE ENHANCING PRIMER' can be applied to the tooth surface to promote surface modification and cement polymerization. The cement is used to bond restorations (inlays, onlays, crowns, bridges, posts, veneers) to tooth structure or implant abutments. The clinician applies the mixed paste directly to the restoration or cavity; the material hardens to provide physical strength and sealing of dentin tubules. The device benefits patients by providing a stable, durable, and radiopaque luting material for dental restorations.
Clinical Evidence
No clinical testing has been performed on this device. Substantial equivalence is supported by bench testing, including film thickness, working time, setting time, flexural strength, water sorption, solubility, and radiopacity, all conforming to ISO 4049:2019. Biocompatibility was assessed per ISO 10993-1, including cytotoxicity (L929 MEM elution), sensitivity (Kligman maximization), and irritation (buccal irritation) tests.
Technological Characteristics
Dual-cured self-adhesive resin cement; methacrylate-based. Supplied in a one-body/two-chamber syringe. Polymerization via chemical initiators and photo-initiators (light-cured). Complies with ISO 4049:2019 for polymer-based restorative materials. Radiopaque. Includes an optional adhesive enhancing primer (light blue transparent liquid).
Indications for Use
Indicated for cementation of ceramic, resin, and metal-based inlays, onlays, crowns, bridges, posts, cores, and veneers, as well as final cementation on implant abutments.
Regulatory Classification
Identification
Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
November 24, 2020
GC America Inc. Mark Heiss Director, Regulatory & Academic Affairs 3737 W. 127th Street Alsip, Illinois 60803
Re: K200798
Trade/Device Name: G-CEM ONE Regulation Number: 21 CFR 872.3275 Regulation Name: Dental Cement Regulatory Class: Class II Product Code: EMA Dated: March 23, 2020 Received: March 26, 2020
Dear Mark Heiss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements,
{1}------------------------------------------------
including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) CDRH and Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See DICE (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatorythe website assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Srinivas "Nandu" Nandkumar, Ph. D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Section 4 - Indications for Use
| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration |
| Form Approved: OMB No. 0910-0120 |
|----------------------------------|
| Expiration Date: 06/30/2020 |
| See PRA Statement below. |
**Indications for Use**
| 510(k) Number (if known) | K200798 |
|--------------------------|-----------|
| Device Name | G-CEM ONE |
Indications for Use (Describe)
1. Cementation of all types of all ceramic, resin and metal-based inlays, onlays, crowns and bridges.
2. Cementation of metal, ceramic, fiber posts, and cast post and cores.
3. Cementation of all ceramic and composite veneers.
4. Final cementation of crowns and bridges on implant abutments.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) Subpart C) | Over-The-Counter Use (21 CFR 801 |
|-----------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> <img alt="Checked" src="checkbox_checked.png"/> </span> </div> | <div> <span> <img alt="Unchecked" src="checkbox_unchecked.png"/> </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text "GC" in a teal color. The text is surrounded by single quotation marks on both sides. The background is white.
Image /page/3/Picture/1 description: The image shows a geometric design with a partial outline of the number 7 on the left side. To the right of the number 7 are seven small, tilted rectangles arranged in a triangular pattern. The color of the number 7 outline and the rectangles is a teal or turquoise shade.
1
# 510(k) Summary - K200798
- 1. Submitter Information: GC America Inc.
3737 W. 127th Street Alsip, IL 60803
| Contact Person: | Mark Heiss, D.D.S. |
|--------------------|--------------------|
| Phone: | (708) 926-3090 |
| Alternate Contact: | Lori Rietman |
| Phone: | (708) 926-3092 |
| Fax: | (708) 925-0373 |
| Date Prepared: | March 23, 2020 |
- 2. Device Name: Proprietary Name: G-CEM ONE Classification Name: Dental cement Device Classification: Class II, 872.3275 Product Code: EMA
- Predicate Devices: 3.
| Product | Applicant | 510(k) No. | Code No | Predicate | Decision<br>Date |
|-------------------------|--------------------|------------|---------|-----------|------------------|
| G-CEM LinkForce | GC America<br>Inc. | K153231 | EMA | Primary | 07/06/2016 |
| G-CEM LINKACE (GAM-200) | GC America<br>Inc. | K120243 | EMA | Reference | 06/27/2012 |
#### 4. Description of Device:
G-CEM ONE is a dual-cured self-adhesive resin cement with a tooth primer, G-CEM ONE ADHESIVE ENHANCING PRIMER. The mixed cement hardens through polymerization. The adhesive component is an acidic monomer which also can polymerize with frame-forming monomers. This copolymerizing reaction provides higher physical strength than resin-modified glass ionomer cements. Using the primer on the prepared tooth, surface modification and cement polymerization are promoted and hardened along with the cement. The ce ment syringe consists of Paste A and B, which are filled in a one-body/two chamber syringe. Both pastes are automixed with a mixing tip and applied directly to restorations or prepared cavity.
Image /page/3/Picture/12 description: The image shows a G-CEM ONE self-adhesive resin cement syringe. The syringe has a brown dispensing tip on the left side and a black plunger on the right side. The label on the syringe reads "G-CEM ONE SELF ADHESIVE RESIN CEMENT A2 4.6g(2.7mL)".
{4}------------------------------------------------
- న. Indications for Use
- 1. Cementation of all types of all ceramic, resin and metal-based inlays, onlays, crowns and bridges.
- 2. Cementation of metal, ceramic, fiber posts, and cast post and cores.
- 3. Cementation of all ceramic and composite veneers.
- 4. Final cementation of crowns and bridges on implant abutments.
#### 6. Package:
- 1. Entrance Kit
G-CEM ONE syringe (4.6 g / 2.7 mL) (1), G-CEM Automix Tip Regular (8), G-CEM Automix Tip for endo with Extension Tip (2), G-CEM ONE ADHESIVE ENHANCING PRIMER (1),
- 2. Twin Refill G-CEM ONE syringe (4.6 g / 2.7 mL) (2), G-CEM Automix Tip Regular (15), G-CEM Automix Tip for endo with Extension Tip (5)
- 3. Single Refill G-CEM ONE syringe (4.6 g / 2.7 mL) (1), G-CEM Automix Tip Regular (8), G-CEM Automix Tip for endo with Extension Tip (2)
- 4. G-CEM ONE ADHESIVE ENHANCING PRIMER G-CEM ONE ADHESIVE ENHANCING PRIMER (1)
- 7. Shades available: A2, AO3, Translucent, BO1, White opaque
- 8. Shelf Life and Storage Conditions:
- Shelf Life 2 years i
- Recommended for optimal performance, store at temperature of 4-25°C (39.2-77.0°F) away from heat, moisture and direct sunlight.
- 9. Performance Bench Tests:
It is confirmed that the device conforms to the required specifications of ISO 4049: 2019 Dentistry -Polymer-based restorative materials and Company Specification: 1AB-1500-3-10666.
Performance testing includes:
| | Property | Test method | Requirement |
|---|-------------------|----------------------------------------------------------------------|-----------------------------------------------------------------------------|
| 1 | Film thickness | ISO 4049: 2019<br>5.2.2 Film thickness of luting materials | No more than 50 µm. |
| 2 | Working time | ISO 4049: 2019<br>5.2.4 Working time, Class 1 and 3 luting materials | No detectable change in the homogeneity. |
| 3 | Setting time | ISO 4049: 2019<br>5.2.6 Setting time, Class 3 materials | Not more than 10 min. |
| 4 | Flexural strength | ISO 4049: 2019<br>5.2.9 Flexural strength | Equal to or greater than 50 MPa. |
| 5 | Water sorption | ISO 4049: 2019<br>5.2.10 Water sorption and solubility | 40 µg/mm³ or less. |
| 6 | Solubility | ISO 4049: 2019<br>5.2.10 Water sorption and solubility | 7.5 µg/mm³ or less. |
| 7 | Radiopacity | ISO 4049: 2019<br>5.5 Radiopacity | Equal to or greater than the radiopacity of the same thickness of aluminum. |
### G-CEM ONE
{5}------------------------------------------------
| | Property | Test method | Requirement |
|---|------------------|------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|
| 1 | Appearance | Company Specification<br>1AB-1500-3-10666 - Visual inspection | Should be homogenous and free<br>from foreign matters |
| 2 | Coat ability | Company Specification<br>1AB-1500-3-10666 - Apply a primer thinly on a<br>glass plate using an applicator. | Form uniform film without<br>unevenness |
| 3 | Color tone | Company Specification<br>1AB-1500-3-10666 - Visual inspection | It must be light blue transparent<br>liquid. |
| 4 | Refractive index | Company Specification<br>1AB-1500-3-10666 - Refraction analysis | 1.3960-1.3990 |
## G-CEM ONE ADHESIVE ENHANCING PRIMER
#### 10. Non-Clinical Performance Testing:
A biocompatibility assessment was completed according to ISO 10993-1:2009, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.
G-CEM ONE is a dual-cured self-adhesive resin cement with a tooth primer "G-CEM ONE ADHESIVE ENHANCING PRIMER" and does come in contact with body tissues (tooth – enamel, dentin) for more than 24 hours.
In conclusion, biocompatibility of G-CEM ONE is acceptable device from the biological evaluation result.
### Cytotoxicity (L929 MEM ELUTION TEST)
Based on the criteria of the protocol of ISO 10993-5
### Sensitivity (KLIGMAN MAXIMIZATION TEST)
Based on the criteria of the protocol of ISO 10993-10
### Irritation (PRIMARY ORAL (BUCCAL) IRRITATION TEST)
Based on the criteria of the protocol of ISO 10993-10
### 11. Clinical Performance Testing No clinical testing has been performed on this device.
- 12. Comparison of Technology:
The specifications that are important to determine substantial equivalency of a dental adhesive system are bond strengths associated to tooth structure (enamel and dentin) and other substrates. It is also important to evaluate the equivalence of sealing property of dentin tubules.
The bond strength of the applicant device to dentin is equivalent to that of the reference predicate device.
The curing mechanism of the applicant device and predicate device are substantially equivalent in principle. Therefore, the applicant device and predicate device are the same in function, and similar in composition and intended use. This supports that the compatibility of the applicant device is substantially equivalent to the predicate devices.
{6}------------------------------------------------
All the components of the applicant device, G-CEM ONE and G-CEM ONE ADHESIVE ENHANCING PRIMER, have already been used in the predicate devices. The bonding mechanism of the predicate device is applying it to tooth structure, then bonding chemically and mechanically by polymerization of uncured methacrylate ester monomers. The subject device has been shown to be substantially equivalent to the predicate device.
The following differences may be noted between G-CEM ONE and the predicate device.
- -The applicant device also has a primer which allows it to have additional indications, similar to that of the primary predicate device.
- -The applicant device is a self-adhesive resin cement, and unlike the G-CEM LinkForce of an adhesive resin cement, it can be bonded to various materials without a primer.
{7}------------------------------------------------
| | Applicant Device | Primary Device | Reference Device |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade name | G-CEM ONE | G-CEM LinkForce<br>K153231 | G-CEM LINKACE (GAM-200)<br>K120243 |
| Manufacturer | GC Corporation | GC Corporation | GC Corporation |
| Product<br>category | Self-adhesive resin cement | Adhesive resin cement | Self-adhesive resin cement |
| Paste/Paste ratio | Paste A / Paste B = 1.3 / 1.0 (w/w) | Paste A / Paste B = 1.0 / 1.0 (w/w) | Paste A / Paste B = 1.3 / 1.0 (w/w) |
| Indications for Use | 1. Cementation of all types of all ceramic,<br>resin and metal-based inlays, onlays,<br>crowns and bridges.<br>2. Cementation of metal, ceramic, fiber posts,<br>and cast post and cores.<br>3. Cementation of all ceramic and composite<br>veneers.<br>4. Final cementation of crowns and bridges<br>on implant abutments. | 1. Cementation of all types of all ceramic,<br>resin and metal-based inlays, onlays,<br>crowns and bridges.<br>2. Cementation of metal, ceramic, fiber posts,<br>and cast post and cores.<br>3. Cementation of all ceramic and composite<br>veneers.<br>4. Permanent cementation of crowns and<br>bridges on implant abutments. | 1. Cementation of all types of all ceramic,<br>resin and metal-based inlays, onlays,<br>crowns and bridges.<br>2. Cementation of metal, ceramic, fiber posts,<br>and cast post and cores. |
| Product<br>description | The components consist of Paste A and B,<br>which are filled in a one-body syringe. Both<br>pastes are automixed with a mixing tip and<br>directly applied to restorations or the<br>prepared cavity. | The components consist of Paste A and B,<br>which are filled in a one-body syringe. Both<br>pastes are automixed with a mixing tip and<br>directly applied to restorations or the<br>prepared cavity. | The components consist of Paste A and B,<br>which are filled in a one-body syringe. Both<br>pastes are automixed with a mixing tip and<br>directly applied to restorations or the<br>prepared cavity. |
| Instruction for use | 1. Tooth preparation<br>2. Application of G-CEM ONE ADHESIVE<br>ENHANCING PRIMER (Optional)<br>3. Restoration preparation<br>4. Dispensing<br>5. Cementation<br>6. Excess cement removal<br>7. Final set<br>8. Final polishing | 1. Try-fit of the restoration<br>2. Pre-treatment of the restoration<br>3. Pre-treatment of the preparation<br>4. Dispensing<br>5. Cementation<br>6. Excess cement removal<br>7. Final set<br>8. Final polishing and adjustments | 1. Tooth preparation<br>2. Restoration preparation<br>3. Dispensing<br>4. Cementation<br>5. Excess cement removal<br>6. Final set<br>7. Final polishing |
| Light curing<br>specification | Light cure using a light curing unit.<br>10 sec. (High Power LED Light)<br>(>1200mW/cm2)<br>20 sec. (Halogen/LED) (700 mW/cm2) | Light cure using a light curing unit.<br>10 sec. (High Power LED Light)<br>(>1200mW/cm2)<br>20 sec. (Halogen/LED) (700 mW/cm2) | Light cure using a light curing unit.<br>10 sec. (High Power LED Light)<br>(>1200mW/cm2)<br>20 sec. (Halogen/LED) (700 mW/cm2) |
| Table 5.1 (Continued) | | | |
| | Applicant Device | Primary Device | Reference Device |
| Trade name | G-CEM ONE | G-CEM LinkForce<br>K153231 | G-CEM LINKACE (G-CEM LINKACE (GAM-<br>200))<br>K120243 |
| Manufacturer | GC Corporation | GC Corporation | GC Corporation |
| Comparison of<br>Technology | Methacrylates contained in Paste A<br>polymerize by polymerization initiators<br>contained in Paste A and Paste B.<br>In addition, they also polymerize by light<br>irradiation with photo polymerization initiators<br>contained in Paste A.<br>Furthermore, using G-CEM ONE ADHESIVE<br>ENHANCING PRIMER for the cavity and<br>abutment tooth, surface modification and<br>cement polymerization are promoted and<br>hardened with the cement.<br>Methacrylates contained in this material are<br>very hydrophobic and set material is stable.<br>Therefore, the ingredients in the cured<br>material are difficult to elute in water. | Methacrylates contained in Paste A<br>polymerize by polymerization initiators<br>contained in Paste A and Paste B.<br>In addition, they also polymerize by light<br>irradiation thanks to photo polymerization<br>initiators contained in Paste A.<br>Furthermore, using the mixture of G-Premio<br>BOND and G-Premio BOND DCA for the<br>cavity and abutment tooth, surface<br>modification and cement polymerization are<br>promoted and hardened with the cement.<br>Methacrylates contained in this material are<br>very hydrophobic and set material is stable.<br>Therefore, the ingredients in the cured<br>material are difficult to elute in water. | Methacrylates contained in Paste A<br>polymerize by polymerization initiators<br>contained in Paste A and Paste B.<br>In addition, they also polymerize by light<br>irradiation thanks to photo polymerization<br>initiators contained in Paste A.<br>Methacrylates contained in this material are<br>very hydrophobic and set material is stable.<br>Therefore, the ingredients in the cured<br>material are difficult to elute in water. |
Table 5.3. Comparison of applicant and predicate
{8}------------------------------------------------
# 13. Conclusion
Based on similarities in intended use, mode of action, and performance testing, G-CEM ONE is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.