← Product Code [DQD](/productcode/DQD) · K200776

# Eko CORE (K200776)

_Eko Devices, Inc. · DQD · Apr 6, 2020 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K200776

## Device Facts

- **Applicant:** Eko Devices, Inc.
- **Product Code:** [DQD](/productcode/DQD.md)
- **Decision Date:** Apr 6, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1875
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The Eko CORE is an electronic stethoscope that enables amplification, filtering, and transmission of auscultation sound data (heart, lungs, bowel, arteries, and veins), whereby a clinician at one location on network can listen to the auscultation sounds of a patient on site or at a different location on the network. Eko CORE is intended for use on pediatric and adult patients. The Eko CORE is intended to be used by professional users in a clinical environment or by lay users in a nonclinical environment. The device is not intended for self-diagnosis.

## Device Story

Electronic stethoscope attachment for analog stethoscopes; captures heart, lung, bowel, arterial, and venous sounds. Device converts analog audio to digital data; transmits via Bluetooth to mobile application. App provides visualization, storage, playback, and sharing. Used by clinicians for remote monitoring/telemedicine or by lay users to record/share sounds with physicians. Enables transition between traditional and digital auscultation. Benefits include remote specialist consultation and improved patient monitoring. Device includes volume control, power switch, and LED indicator.

## Clinical Evidence

Bench testing only. Includes software verification and validation, electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and wireless coexistence testing. Human factors testing conducted for OTC use with passing results.

## Technological Characteristics

Electronic attachment for analog stethoscopes; digital/analog auscultation modes. Bluetooth connectivity to mobile app. Features volume adjustment, power switch, LED indicator. Standards met: IEC 60601-1 (General Safety), IEC 60601-1-2 (EMC), IEC 60601-1-11 (Home Healthcare).

## Regulatory Identification

Manual stethoscope: A mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs. Electronic stethoscope: An electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- Eko Electronic Stethoscope System (EESS) ([K151319](/device/K151319.md))
- Tyto Stethoscope ([K181612](/device/K181612.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 6, 2020

Eko Devices, Inc. Arezou Azar Sr. Dir. Regulatory Affairs 1212 Broadway, Suite 100 Oakland, California 94612

Re: K200776

Trade/Device Name: Eko CORE Regulation Number: 21 CFR 870.1875 Regulation Name: Stethoscope Regulatory Class: Class II Product Code: DOD Dated: March 23, 2020 Received: March 25, 2020

Dear Arezou Azar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K200776

Device Name Eko CORE

#### Indications for Use (Describe)

The Eko CORE is an electronic stethoscope that enables amplification, filtering, and transmission of auscultation sound data (heart, lungs, bowel, arteries, and veins), whereby a clinician at one location on network can listen to the auscultation sounds of a patient on site or at a different location on the network. Eko CORE is intended for use on pediatric and adult patients. The Eko CORE is intended to be used by professional users in a clinical environment or by lay users in a nonclinical environment. The device is not intended for self-diagnosis.

| Type of Use (Select one or both, as applicable)                                                                                  |                                                                                                                                 |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CER 801 Subpart D)                 </span> </div> | <div> <span> <span style="font-size: 16px;">☑</span> Over-The-Counter Use (21 CER 801 Subpart C)                 </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) SUMMARY

#### Eko Devices, Inc. Eko CORE

Submitter Name Eko Devices, Inc. and Address: 1212 Broadway, Suite 100 Oakland, CA 94612

Contact Person: Arezou Azar, PhD. Sr.Dir of Regulatory Affairs Email: arezou.azar@ekohealth.com Phone: 650-804-0285

Device Trade Name(s): Eko CORE Other Names Used: Eko Electronic Stethoscope System CORE by Eko Common or Usual Name: Electronic Stethoscope Classification Name: Electronic Stethoscope Regulatory Class: Class II Product Code: DQD, 870.1875 Predicate Devices: Eko Electronic Stethoscope System (EESS)(K151319), Eko Devices, Inc. Tyto Stethoscope (K181612) (OTC), Tyto Care Ltd.

#### Device Description

The Eko CORE (OTC) is a digital stethoscope device designed for use by health care professionals as well as lay users in clinical or non-clinical environments. It enables physicians transition between traditional and digital auscultation. It can electronically amplify, filter and transfer sounds to the accompanying mobile application for storage and sharing or can transmit the data for telemedicine use. It also enables regular users to listen to their body sounds (lungs, heart, arteries, veins, Gastrointestinal tract, etc.) and record and share it with their physicians.

lt consists of two primary components: 1) The CORE: an electronic attachment for an analog stethoscope and 2) Eko App, and application that pairs with the CORE attachment.

- . CORE is used to record audio in the stethoscope, convert it to digital data points, and transmit data to a mobile device via Bluetooth". It includes a volume adjustment button, an analog-to-digital power switch, and an LED light indicator.
- . The app captures audio data from the CORE and provides data visualization, secure data storage, audio playback, and sharing features.

These features enable a healthcare professional to monitor patients, seek out second opinions from a specialist or use the device for telemedicine use.

{4}------------------------------------------------

### Intended Use / Indications for Use

The Eko CORE is an electronic stethoscope that enables amplification, filtering, and transmission of auscultation sound data (heart, lungs, bowel, arteries, and veins), whereby a clinician at one location on network can listen to the auscultation sounds of a patient on site or at a different location on the network. Eko CORE is intended for use on pediatric and adult patients. The Eko CORE is intended to be used by professional users in a clinical environment or by lay users in a nonclinical environment. The device is not intended for self-diagnosis.

#### Summary of Technological Characteristics

The Eko CORE is a modified version of the previously-cleared predicate device and has very similar technological characteristics. Minor changes have been made, but the devices have the same overall operational and technological characteristics. Both devices include the ability to attach to a standard analog stethoscope and feature both digital and analog auscultation modes. Additionally, both devices connect via Bluetooth to the Eko App for visualization, recording and transfer of data. Both devices included on/off power buttons and volume adjustment controls. Both devices have the same frequency range and have the same maximum sound level.

A table comparing the key features of the subject and predicate devices is provided below.

#### Performance Data

The device has undergone extensive bench verification, as well as software verification and validation. Electrical safety, EMC and wireless coexistence testing were also successfully conducted.

The OTC use of the device was assessed in human factors testing with passing results.

In all instances, the device functioned as intended. Based on the testing results, the device has a safety and effectiveness profile that is similar to the predicate device.

#### Conclusions

The Eko CORE is as safe and effective as the predicate devices. The Eko CORE has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor differences in indications do not introduce a new intended use and do not raise different questions of safety and effectiveness when used as labeled. In addition, the minor technological differences between the Eko CORE 

{5}------------------------------------------------

and its predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that the Eko CORE is as safe and effective as the predicate device. Thus, the Eko CORE is substantially equivalent.

| Parameters                                              | Eko Electronic<br>Stethoscope System<br>(OTC)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Eko Electronic<br>Stethoscope System<br>(Rx) - K151319                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Tyto Stethoscope<br>(OTC)- K181612                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|---------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use                                  | The Eko CORE is an<br>electronic stethoscope<br>that enables<br>amplification, filtering,<br>and transmission of<br>auscultation sound data<br>(heart, lungs, bowel,<br>arteries, and veins),<br>whereby a clinician at<br>one location on<br>network can listen to<br>the auscultation sounds<br>of a patient on site or at<br>a different location on<br>the network. Eko CORE<br>is intended for use on<br>pediatric and adult<br>patients. The Eko CORE<br>is intended to be used<br>by professional users in<br>a clinical environment<br>or by lay users in a<br>nonclinical<br>environment. The<br>device is not intended<br>for self-diagnosis. | The Eko Electronic<br>Stethoscope System is<br>intended to be used as a<br>part of a physical<br>assessment of a patient<br>by healthcare<br>professionals for<br>diagnostic decision<br>support in clinical<br>settings. Eko is intended<br>for use on pediatric and<br>adult patients. It can<br>electronically amplify,<br>filter and transfer<br>sounds to the<br>accompanying mobile<br>application for storage<br>and sharing. It can used<br>to record heart sounds<br>and cardiac murmurs,<br>bruits, respiratory<br>sounds and abdominal<br>sounds during physical<br>examination in normal<br>patients or those with<br>suspected diseases of<br>the cardiac, vascular,<br>pulmonary or<br>abdominal organ<br>systems. | The Tyto Stethoscope is<br>an electronic<br>stethoscope that<br>enables transmission of<br>auscultation sound data,<br>whereby a<br>clinician at one location<br>on an IP network can<br>listen to the auscultation<br>sounds of a patient on<br>site or at a different<br>location<br>on the IP network with<br>the signal carried on an<br>IP connection between<br>the two locations. The<br>Tyto Stethoscope is<br>intended<br>to be used by<br>professional users in a<br>clinical environment or<br>by lay users in a<br>nonclinical<br>environment. The<br>device is for<br>medical diagnostics<br>purposes only. The<br>device is not intended<br>for self-diagnosis. |
| Standards Met                                           | IEC 60601-1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | IEC 60601-1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | IEC 60601-1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| General Safety                                          | IEC 60601-1-2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | IEC 60601-1-2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | IEC 60601-1-2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| EMC Safety                                              | IEC 60601-1-11                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | IEC 60601-1-11                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Device<br>Classification                                | DQD: Electronic<br>stethoscope                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | DQD: Electronic<br>stethoscope                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | DQD: Electronic<br>stethoscope                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Prescribed                                              | OTC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Rx                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | OTC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Parameters                                              | Eko Electronic<br>Stethoscope System<br>(OTC)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Eko Electronic<br>Stethoscope System<br>(Rx) - K151319                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Tyto Stethoscope<br>(OTC)- K181612                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Classification                                          | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Stethoscope<br>Type                                     | Attachment to an<br>analog stethoscope<br>(Core)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Attachment to an analog<br>stethoscope (Core)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Stand-alone electronic<br>stethoscope                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| User Interface                                          | On/Off Power button<br>Volume adjustment<br>LED status indicator<br>Mobile App                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | On/Off Power button<br>Volume adjustment<br>LED status indicator<br>Mobile App                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | On/Off Power button<br>Volume adjustment<br>LED screen<br>Mobile App                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Analog/ Digital<br>Interoperability                     | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Connectivity                                            | Bluetooth®                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Bluetooth®                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Wi-Fi                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Sound<br>Amplification                                  | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Record and<br>Playback<br>Sounds                        | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Data Transfer<br>to Compatible<br>Computing<br>platform | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |

### Table1. Comparison with Predicate Devices

{6}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/device/K200776](https://fda.innolitics.com/device/K200776)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
