← Product Code [BTR](/productcode/BTR) · K200761

# Tenax Laser Resistant Endotracheal Tube (K200761)

_Bryan Medical, Inc. · BTR · Nov 25, 2020 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K200761

## Device Facts

- **Applicant:** Bryan Medical, Inc.
- **Product Code:** [BTR](/productcode/BTR.md)
- **Decision Date:** Nov 25, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5730
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The Tenax® Laser Resistant Endotracheal Tube is intended for endotracheal intubation. It is indicated for use for all types of surgical procedures involving carbon dioxide (10.60 microns) or KTP (532 nm) laser use (normal pulsed or continuous beam delivery in the non-contact mode), when endotracheal intubation is required to administer anesthetic gases or to overcome emergency obstruction of an airway.

## Device Story

Tenax Laser Resistant Endotracheal Tube is a cuffed, single-use airway device designed for use during CO2 or KTP laser surgeries. It features an inner aluminum wrapping for laser resistance and an outer atraumatic silicone covering. The device includes dual inflatable cuffs and a pilot balloon containing blue dye to alert clinicians to potential punctures or damage. It is used by physicians in clinical settings to maintain an airway and deliver anesthetic gases. The device is inserted into the trachea; the laser-resistant construction protects the airway from accidental laser strikes during surgery. It is discarded after a single use.

## Clinical Evidence

Bench testing only. Performance data included dimensional testing, bend radius, cuff inflation/deflation/burst testing, dye visibility, laser resistance, and connector leakage. Biocompatibility testing performed per ISO 10993-1, including cytotoxicity, sensitization, irritation, acute systemic toxicity, and material-mediated pyrogenicity.

## Technological Characteristics

Endotracheal tube with aluminum scattering layer for laser resistance and silicone outer covering. Features dual inflatable cuffs and blue dye pilot balloon indicator. Sterile (EO). Dimensions: 34cm length, 5.0mm-7.5mm ID. Connects to standard anesthesia gas supply. Biocompatibility: Surface device, mucosal membrane, <24h duration.

## Regulatory Identification

A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.

## Predicate Devices

- Laser-Shield® II ([K993582](/device/K993582.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 25, 2020

Bryan Medical, Inc. % Ian Broome Regulatory Affairs Consultant AlvaMed, Inc. 1116 Great Plain Avenue, Suite 1 Needham, Massachusetts 02492

Re: K200761

Trade/Device Name: Tenax Laser Resistant Endotracheal Tube Regulation Number: 21 CFR 868.5730 Regulation Name: Tracheal Tube Regulatory Class: Class II Product Code: BTR, BSK Dated: October 23, 2020 Received: October 26, 2020

Dear Ian Broome:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K200761

Device Name Tenax® Laser Resistant Endotracheal Tube

Indications for Use (Describe)

The Tenax® Laser Resistant Endotracheal Tube is intended for endotracheal intubation. It is indicated for use for all types of surgical procedures involving carbon dioxide (10.60 microns) or KTP (532 nm) laser use (normal pulsed or continuous beam delivery in the non-contact mode), when endotracheal intubation is required to administer anesthetic gases or to overcome emergency obstruction of an airway.

Type of Use (Select one or both, as applicable)

| <span> <b>      ☑     </b>     Prescription Use (Part 21 CFR 801 Subpart D)    </span> |
|----------------------------------------------------------------------------------------|
| <span>     ☐ Over-The-Counter Use (21 CFR 801 Subpart C)    </span>                    |

#### _ Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 5.0 TRADITIONAL 510(K) SUMMARY FOR TENAX® LASER RESISTANT ENDOTRACHEAL TUBE

#### K200761

#### I. SUBMITTER

Bryan Medical, Inc. 5725 Dragon Way, Suite 300 Cincinnati, OH 45227 Phone: (888) 331-3485, x 700 Fax: (617) 249-0955 Contact Person: lan Broome, AlvaMed, Inc. Date Prepared: November 23, 2020

### II. DEVICE

Name of Device: Tenax® Laser Resistant Endotracheal Tube Common or Usual Name: Endotracheal tube (cuffed) Classification Name: Tracheal tube (21 CFR 868.5730) Requlatory Class: II Product Code: BTR, BSK

#### III. PREDICATE DEVICE

Laser-Shield® II, K993582

### IV. DEVICE DESCRIPTION

The Tenax® Laser Resistant Endotracheal Tube is an endotracheal tube with two inflatable cuffs. It is indicated for use in surgical procedures where intubation is required in the presence of KTP or CO2 lasers.

The device is provided sterile (EO). After use, the device is discarded and disposed of in accordance with local regulations. There are no associated device accessories.

{4}------------------------------------------------

### V. INDICATION FOR USE

The Tenax® Laser Resistant Endotracheal Tube is intended for endotracheal intubation. It is indicated for use for all types of surgical procedures involving carbon dioxide (10.60 microns) or KTP (532 nm) laser use (normal pulsed or continuous beam delivery in the non-contact mode), when endotracheal intubation is required to administer anesthetic gases or to overcome emergency obstruction of an airway.

| Intended Population:         | Adults (≥ 22 years old) requiring endotracheal intubation during<br>laser surgery procedures in the upper gastro-respiratory tract.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Environment of Use: | Trachea/Upper GI Tract                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Contraindications:           | The Tenax® Laser Resistant Endotracheal Tube should not be used in patients with narrow airways, which could restrict ventilation, resulting in excessive elevation of intratracheal pressure. Do not use in any patient suffering from conditions that may preclude endotracheal intubation. Do not use this device for any purpose other than its intended use. To avoid device damage, do not exceed a power density of 10,394 W/cm² when used with a CO2 laser. To avoid device damage, do not exceed a power density of 5,305 W/cm² when used with a KTP laser. Do not use with continuous irradiation lasting longer than 30 seconds. |

{5}------------------------------------------------

## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

|                                        | Subject:<br>Tenax® Laser Resistant Endotracheal<br>Tube                                                                                                                                                                                                                                                                                                                                                                                          | Predicate Device:<br>Laser-Shield® II                                                                                                                                                                                                                                                                                                                                                                                       |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                           | Bryan Medical, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                              | Medtronic/Xomed                                                                                                                                                                                                                                                                                                                                                                                                             |
| 510(k) Number                          | K200761                                                                                                                                                                                                                                                                                                                                                                                                                                          | K993582                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Product Code                           | BTR, BSK                                                                                                                                                                                                                                                                                                                                                                                                                                         | BTR                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Regulation                             | 868.5730                                                                                                                                                                                                                                                                                                                                                                                                                                         | 868.5730                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Regulation<br>Description              | Tracheal tube                                                                                                                                                                                                                                                                                                                                                                                                                                    | Tracheal tube                                                                                                                                                                                                                                                                                                                                                                                                               |
| Common Name                            | Endotracheal tube                                                                                                                                                                                                                                                                                                                                                                                                                                | Endotracheal tube                                                                                                                                                                                                                                                                                                                                                                                                           |
| Indications for<br>Use                 | The Tenax® Laser Resistant Endotracheal<br>Tube is intended for endotracheal<br>intubation. It is indicated for use for all<br>types of surgical procedures involving<br>carbon dioxide (10.60 microns) or KTP (532<br>nm) laser use (normal pulsed or continuous<br>beam delivery in the non-contact mode),<br>when endotracheal intubation is required to<br>administer anesthetic gases or to overcome<br>emergency obstruction of an airway. | Xomed Laser-Shield® II is intended for<br>endotracheal intubation. It is indicated for<br>use for all types of surgical procedures<br>involving carbon dioxide (10.60 microns) or<br>KTP (532 nm) laser use (normal pulsed or<br>continuous beam delivery in the non-<br>contact mode), when endotracheal<br>intubation is required to administer<br>anesthesia gases or to overcome<br>emergency obstruction of an airway. |
| Sterility                              | Sterile (ethylene oxide)<br>SAL 10-6                                                                                                                                                                                                                                                                                                                                                                                                             | Sterile (method unknown)                                                                                                                                                                                                                                                                                                                                                                                                    |
| Biocompatibility<br>Classification     | Surface Device, Mucosal Membrane,<br>Limited Duration (< 24h)                                                                                                                                                                                                                                                                                                                                                                                    | Surface Device, Mucosal Membrane,<br>Limited Duration (< 24h)                                                                                                                                                                                                                                                                                                                                                               |
| Design Features                        | ● Endotracheal tube with cuffs at distal<br>end<br>● Aluminum wrapping for laser<br>resistance<br>● Outer atraumatic silicone covering<br>● Blue dye within pilot balloons for<br>leak/damage indication                                                                                                                                                                                                                                         | ● Endotracheal tube with cuffs at distal<br>end<br>● Aluminum wrapping for laser<br>resistance<br>● Outer atraumatic polymer covering<br>● Blue dye within pilot balloons for<br>leak/damage indication                                                                                                                                                                                                                     |
|                                        | Subject:                                                                                                                                                                                                                                                                                                                                                                                                                                         | Predicate Device:                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                        | Tenax® Laser Resistant Endotracheal<br>Tube                                                                                                                                                                                                                                                                                                                                                                                                      | Laser-Shield® II                                                                                                                                                                                                                                                                                                                                                                                                            |
| Connection to<br>Ventilation<br>Source | Standard straight connector to anesthesia<br>gas supply                                                                                                                                                                                                                                                                                                                                                                                          | Standard straight connector to anesthesia<br>gas supply                                                                                                                                                                                                                                                                                                                                                                     |
| Dimensions                             | Length: 34cm<br>ID: (6 sizes)<br>5.0mm (model TG-0050) 5.5mm (TG-0055) 6.0mm (TG-0060) 6.5mm (TG-0065) 7.0mm (TG-0070) 7.5mm (TG-0075)                                                                                                                                                                                                                                                                                                           | Length: 33cm<br>ID: (9 sizes)<br>4.0mm (model 7060100) 4.5mm (7060150) 5.0mm (7060200) 5.5mm (7060250) 6.0mm (7060300) 6.5mm (7060350) 7.0mm (7060400) 7.5mm (7060450) 8mm (7060500)<br>OD:<br>6.6mm (7060100)                                                                                                                                                                                                              |
|                                        | OD:<br>7.9mm (TG-0050) 8.6mm (TG-0055) 9.3mm (TG-0060) 10.0mm (TG-0065) 10.6 (TG-0070) 11.0mm (TG-0075)                                                                                                                                                                                                                                                                                                                                          | 7.3mm (7060150)8.0mm (7060200)8.6mm (7060250)9.0mm (7060300)10.0mm (7060350)10.5mm (7060400)11.0mm (7060450)11.5mm (7060500)                                                                                                                                                                                                                                                                                                |
| Single Use?                            | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                              | Yes                                                                                                                                                                                                                                                                                                                                                                                                                         |

{6}------------------------------------------------

Both devices are laser resistant endotracheal tubes. They share an identical indication for use and are intended for use in the same manner. The subject and predicate devices also share the following technological characteristics:

- Laser resistance using an inner aluminum scattering layer, ●
- . Atraumatic outer layer for protection of patient tissue,
- Inflatable cuffs,
- . Use of internal dye as a warning indicator to alert practitioners of device puncture or damage.

{7}------------------------------------------------

#### VII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination:

- Dimensional testing, ●
- Bend radius,
- . Cuff inflation/deflation/burst testing (time and pressure),
- Dye visibility.
- . Laser resistance,
- Tracheal tube dimensions, ●
- . Connector leakage.
- ●

#### Biocompatibility Testing

The biocompatibility evaluation for the Tenax® Laser Resistant Endotracheal Tube was conducted in accordance with the FDA's guidance "International Standard ISO 10993-1 'Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process." The biocompatibility testing included the following tests:

- Cytotoxicity,
- Sensitization, ●
- Irritation, ●
- Acute Systemic Toxicity, ●
- Material-Mediated Pyrogenicity. .

The Tenax® Laser Resistant Endotracheal Tube is considered tissue-contacting for a duration of less than 24 hours. The device's materials are well-characterized for medical use.

#### VIII. CONCLUSIONS

The non-clinical test data support the safety of the device and the verification and validation testing data demonstrate that the Tenax® Laser Resistant Endotracheal Tube shall perform as intended in the specified use conditions. The subject device and the predicate device have the same intended use, and the technological differences do not raise different questions of safety and effectiveness.

---

**Source:** [https://fda.innolitics.com/device/K200761](https://fda.innolitics.com/device/K200761)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
