K200747 · The Lifestyle Company, Inc. · LPN · May 15, 2020 · Ophthalmic
Device Facts
Record ID
K200747
Device Name
PuriLens Plus Preservative Free Saline
Applicant
The Lifestyle Company, Inc.
Product Code
LPN · Ophthalmic
Decision Date
May 15, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5928
Device Class
Class 2
Indications for Use
PuriLens Plus Preservative-Free Saline is indicated for use following proper lens disinfection as recommended by the eye care practitioner. PuriLens Plus Preservative-Free Saline is for rinsing soft (hydrophilic), rigid gas permeable and hard contact lenses prior to lens insertion. This solution may also be used as an insertion solution for large diameter (scleral) contact lenses, as a rinse for contact lens cases, and may be used as needed throughout the day to rinse contact lenses.
Device Story
PuriLens Plus Preservative-Free Saline is a sterile, buffered saline solution; osmolarity and pH are physiological. Used by patients for rinsing and inserting soft, rigid gas permeable, hard, and scleral contact lenses; rinsing lens cases. Product is sold over-the-counter in multi-use plastic containers. Device is identical in formulation, design, and manufacturing to the predicate; this submission expands indications to include rigid gas permeable lenses. No clinical or bench testing required due to identity with predicate.
Clinical Evidence
No clinical data; no bench testing required as the device is identical to the predicate in formulation, manufacturing, and packaging.
Technological Characteristics
Sterile, buffered saline solution; physiological pH and osmolarity. Multi-use plastic resin container with puncture, re-usable cap. No change in design, formulation, or manufacturing from predicate.
Indications for Use
Indicated for rinsing soft (hydrophilic), rigid gas permeable, and hard contact lenses prior to insertion; insertion solution for large diameter (scleral) lenses; rinsing contact lens cases; and as-needed rinsing throughout the day. For use following proper disinfection as recommended by an eye care practitioner.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 15, 2020
The LifeStyle Company, Inc. % Michael A. Siano. MA. RAC Regulatory Affairs Consultant Emergo Global Consulting, LLC 2500 Bee Cave Road, Building 1, Suite 300 Austin, TX 78746
Re: K200747
Trade/Device Name: PuriLens Plus Preservative Free Saline Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (Hydrophilic) Contact Lens Care Products Regulatory Class: Class II Product Code: LPN, MRC Dated: March 18, 2020 Received: March 23, 2020
Dear Michael Siano:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200747
Device Name PuriLens Plus Preservative Free Saline
#### Indications for Use (Describe)
PuriLens Plus Preservative-Free Saline is indicated for use following proper lens disinfection as recommended by the eye care practitioner. PuriLens Plus Preservative-Free Saline is for rinsing soft (hydrophilic), rigid gas permeable and hard contact lenses prior to lens insertion. This solution may also be used as an insertion solution for large diameter (scleral) contact lenses, as a rinse for contact lens cases, and may be used as needed throughout the day to rinse contact lenses.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10pt;"> <input disabled="disabled" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-size: 10pt;"> <input checked="checked" disabled="disabled" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
# PuriLens Plus Preservative Free Saline
### 1. Submission Sponsor
The LifeStyle Company, Inc. 6 Paragon Way Suite 112 Freehold, NJ 07728 USA Tom Seidner, President Email: toms@well.com Phone number: (800) 622-0777
### 2. Submission Correspondent
Emergo Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin, TX 78746 Office Phone: (512) 327-9997 Email: LST.AUS.ProjectManagement@ul.com Contact: Michael A. Siano Title: Regulatory Affairs Consultant
#### 3. Date Prepared
3/18/2020
#### 4. Device Identification
| Trade/Proprietary Name: | PuriLens Plus Preservative Free Saline |
|-------------------------|----------------------------------------------------------------------------------|
| Common/Usual Name: | Accessories, Soft Lens Products Products, Contact Lens Care, Rigid Gas Permeable |
| Regulation Number: | 886.5918, 886.5928 |
| Product Code: | MRC, LPN |
| Class: | Class 2 |
| Classification Panel: | Ophthalmic |
#### 5. Legally Marketed Predicate Device(s)
The LifeStyle Company, Inc., claims substantial equivalence to the Purilens Saline Solution (K002319), Purilens, Inc., in terms of intended use and technological characteristics.
#### 6. Indication for Use
PuriLens Plus Preservative-Free Saline is indicated for use following proper lens disinfection as recommended by the eye care practitioner. PuriLens Plus Preservative-Free Saline is for rinsing soft
{5}------------------------------------------------
(hydrophilic), rigid gas permeable and hard contact lenses prior to lens insertion. This solution may also be used as an insertion solution for large diameter (scleral) contact lenses, as a rinse for contact lens cases, and may be used as needed throughout the day to rinse contact lenses.
#### 7. Device Description
The PuriLens Plus Preservative Free Saline is a sterile, buffered saline solution for rinsing and storing contact lenses. The solution is buffered to roughly physiological pH and has an osmolarity similar to that of human tears. The product is sold over-the-counter in various packaging configurations.
The current generation of the product, Purilens Saline Solution, was cleared under K002319 in 2000, and has been legally marketed in the US since that time. This 510(k) application is to expand the indications for use to include use with rigid gas permeable lenses. There is no change in the device design, formulation, or manufacturing. The labeling will be revised with the new indications.
#### 8. Substantial Equivalence Discussion
The following table compares the PuriLens Plus Preservative Free Saline to the predicate device with respect to intended use, technological characteristics, materials, and performance, forming the basis for the determination of substantial equivalence.
| Attribute | PuriLens Plus Preservative Free Saline | Purilens Saline Solution |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | The LifeStyle Company, Inc. | Purilens, Inc. |
| 510(k) Number | | K002319 |
| Product Code | LPN, MRC | LPN |
| Class | 2 | 2 |
| Classification Panel | Ophthalmic | Ophthalmic |
| Regulation Number | 886.5918<br>886.5928 | 886.5928 |
| Intended Use | Contact lens solution | Contact lens solution |
| Indications for Use | PuriLens Plus Preservative-Free Saline is<br>indicated for use following proper lens<br>disinfection as recommended by the eye<br>care practitioner. PuriLens Plus<br>Preservative-Free Saline is for rinsing soft<br>(hydrophilic), rigid gas permeable and<br>hard contact lenses prior to lens insertion.<br>This solution may also be used as an<br>insertion solution for large diameter<br>(scleral) contact lenses, as a rinse for<br>contact lens cases, and may be used as<br>needed throughout the day to rinse<br>contact lenses. | Purilens Saline Solution is indicated for the<br>cleaning and disinfecting of contact lenses<br>by means of use in a Purilens Electronic<br>Lens Care System. Lenses may by stored in<br>disinfected solution for a period of 24<br>hours following the cleaning and<br>disinfecting. |
| OTC/Rx | OTC | OTC |
| Volume | 120 mL, 60 mL | 120 mL |
Table 1. Substantial Equivalence Comparison
{6}------------------------------------------------
| Preservative-Free | Yes | Yes |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Container Usage | Multi-use | Multi-use |
| Sterility | Sterile | Sterile |
| Materials<br>(container) | Plastic resin container with puncture, re-<br>usable cap | Plastic resin container with puncture, re-<br>usable cap |
| Biocompatibility | Evaluated in accordance with FDA<br>Guidance Premarket Notification (510(K))<br>Guidance Document for Contact Lens Care<br>Products | Evaluated in accordance with FDA<br>Guidance Premarket Notification (510(K))<br>Guidance Document for Contact Lens Care<br>Products |
#### 9. Equivalence Discussion
Some differences exist between the subject and predicate. None of the minor differences noted raise different questions of safety and effectiveness.
#### 10. Non-Clinical Performance Data
Because the subject is identical to the predicate in terms of formulation, manufacturing, and packaging, no bench-testing is required to demonstrate substantial equivalence.
#### 11. Statement of Substantial Equivalence
The PuriLens Plus Preservative-Free Saline has the same intended use and technological characteristics as the Purilens Saline Solution. Therefore, the PuriLens Plus Preservative-Free Saline is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.