K200746 · Shanghai Apolo Medical Technology Co., Ltd. · ONF · May 15, 2020 · General, Plastic Surgery
Device Facts
Record ID
K200746
Device Name
IPL Treatment Systems
Applicant
Shanghai Apolo Medical Technology Co., Ltd.
Product Code
ONF · General, Plastic Surgery
Decision Date
May 15, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The IPL treatment system is intended for medical use in the treatment of the following dermatologic conditions: - Permanent hair reduction- long-term stable reduction in number of hairs re-growing after a treatment regimen; - Moderate inflammatory acne vulgaris; - Benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, ephelides (freckles); - Cutaneous lesions including scars; - Benign cutaneous vascular lesions including port wine stains, hemangiomas, facial truncal and leg telangiectasias, erythema of rosacea, leg veins, spider angiomas and venous malformations.
Device Story
IPL Treatment Systems utilize intense, broad-spectrum pulsed light (420-1200 nm) generated by a Xenon flash lamp. Device consists of main unit, control panel, and interchangeable treatment handpieces. Principle of operation: selective photothermolysis; optical energy absorbed by specific skin chromophores and converted to heat. Interchangeable filters allow selective targeting for specific indications. Cooling system (water, forced-air, copper, TEC) integrated into handpiece to protect epidermis, reduce pain/discomfort, and minimize thermal injury. Used in clinical settings by trained professionals. Output is pulsed light energy; healthcare providers select parameters based on skin type and condition to achieve therapeutic effects like hair reduction or lesion treatment.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and non-laser light source performance (IEC 60601-2-57). Biocompatibility testing (cytotoxicity, skin irritation, skin sensitization) performed per ISO 10993-1.
Technological Characteristics
Intense pulsed light (Xenon flash lamp) source; 420-1200 nm wavelength; 4.1-50.8 J/cm² fluence; 5-50 ms pulse duration. Direct sapphire coupling delivery. Cooling: water, forced-air, copper, and TEC. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-57. Biocompatible materials per ISO 10993-1.
Indications for Use
Indicated for patients requiring treatment of dermatologic conditions including permanent hair reduction, moderate inflammatory acne vulgaris, benign pigmented epidermal lesions (dyschromia, hyperpigmentation, melasma, ephelides), cutaneous lesions (scars), and benign cutaneous vascular lesions (port wine stains, hemangiomas, telangiectasias, rosacea, leg veins, spider angiomas, venous malformations).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
iPulseLight IPL System (HS 300C, HS 650) (K093627)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized image of a caduceus, which is a traditional symbol of medicine. The text on the right is in blue and is arranged in three lines.
Shanghai Apolo Medical Technology Co., Ltd. Felix Li RA Supervisor 4F, Building A, No.388, Yindu Road, Xuhui District Shanghai, Shanghai 200231 China
Re: K200746
Trade/Device Name: IPL Treatment Systems Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: ONF Dated: March 9, 2020 Received: March 23, 2020
Dear Felix Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin Kejing Chen Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) k200746
Device Name IPL Treatment System
Indications for Use (Describe)
The IPL treatment system is intended for medical use in the treatment of the following dermatologic conditions:
- Permanent hair reduction- long-term stable reduction in number of hairs re-growing after a treatment regimen;
- Moderate inflammatory acne vulgaris;
- Benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, ephelides (freckles);
- Cutaneous lesions including scars;
- Benign cutaneous vascular lesions including port wine stains, hemangiomas, facial truncal and leg telangiectasias, erythema of rosacea, leg veins, spider angiomas and venous malformations.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 3-510(k) summary
#### l Submitter
Shanghai Apolo Medical Technology Co., Ltd. 4F, Building A, No.388, Yindu Road, Xuhui District, Shanghai 200231, China
Establishment Registration Number: 3007120647
Contact person: Felix Li Position: Regulatory Affairs Phone: +86-138 4919 0618 Fax: +86-21-34622840 E-mail: liqiang@apolo.com.cn
Date of preparation: Apr 28th, 2020
#### II Proposed Device
| Trade Name of Device: | IPL Treatment Systems |
|-----------------------|----------------------------------------------------------------------------------------|
| Common name: | Laser surgical instrument for use in general and plastic surgery<br>and in dermatology |
| Regulation Number: | 21 CFR 878.4810 |
| Regulatory Class: | Class II |
| Product code: | ONF |
| Review Panel | General & Plastic Surgery |
#### III Predicate Devices
| 510(k) Number: | K093627 |
|-----------------|----------------------------------------------------------------------------------------|
| Trade name: | iPulseLight IPL System (HS 300C, HS 650) |
| Common name: | Laser surgical instrument for use in general and plastic<br>surgery and in dermatology |
| Classification: | Class II |
| Product Code: | ONF |
| Manufacturer | Shanghai APOLO Medical Technology Co., Ltd. |
#### IV Device description
The IPL Treatment Systems utilizes an intense, visible, broad-spectrum pulse of light that when used with interchangeable filters, allowing it to selectively heat the target
{4}------------------------------------------------
for different indications. The optical component is intended to drive the IPL energy to concentrate where the optical energy has selectively heated the target.
The Intense pulsed light (IPL) is a type of intensive, broadband, coherent light source which has a wavelength spectrum of 420 nm -1200 nm. The light-based component delivers optical energy that is absorbed by specific chromophores in the skin, which is converted to heat, according to the principle of selective photothermolysis.
The proposed device consists of the main unit, control panel and the treatment handpiece(s).
#### V Indication for use
The IPL treatment systems is intended for medical use in the treatment of the following dermatologic conditions:
- Permanent hair reduction- long-term stable reduction in number of hairs re-growing after a treatment regimen;
- -Moderate inflammatory acne vulgaris;
- Benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, ephelides (freckles);
- -Cutaneous lesions including scars;
- Benign cutaneous vascular lesions including port wine stains, hemangiomas, facial truncal and leq telangiectasias, erythema of rosacea, leg veins, spider angiomas and venous malformations.
| Item | Proposed device | Predicate device<br>(K093627) | Discussion |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| Product<br>Code | ONF | ONF | Identical |
| Regulation<br>No. | 21 CFR 878.4810 | 21 CFR 878.4810 | Identical |
| Class | Class II | Class II | Identical |
| Indication<br>for use | The IPL treatment<br>systems is intended for<br>medical use in the<br>treatment of the following<br>dermatologic conditions:<br>- Permanent hair<br>reduction- long-term | iPulseLight IPL System (HS<br>300C, HS 650) are identical<br>with regard to indications<br>for use including<br>recommended filters to be<br>used with Fitzpatrick skin<br>type. Both models are | The<br>indications<br>of<br>the<br>proposed<br>device<br>covered by<br>the |
| | stable reduction in<br>number of hairs<br>re-growing after a<br>treatment regimen; | intended for medical use in<br>the treatment of the<br>following dermatologic<br>conditions: | predicate.<br>The cooling<br>system is<br>an integral<br>part of the<br>device. |
| - | Moderate<br>inflammatory acne<br>vulgaris; | Permanent hair<br>reduction- long-term<br>stable reduction in<br>number of hairs<br>re-growing after a<br>treatment regimen. | |
| - | Benign pigmented<br>epidermal lesions<br>including dyschromia,<br>hyperpigmentation,<br>melasma, ephelides<br>(freckles); | Treatment of: | |
| - | Cutaneous lesions<br>including scars; | • Moderate<br>inflammatory acne<br>vulgaris | |
| - | Benign cutaneous<br>vascular lesions<br>including port wine<br>stains, hemangiomas,<br>facial truncal and leg<br>telangiectasias,<br>erythema of rosacea,<br>leg veins, spider<br>angiomas and venous<br>malformations. | • Benign pigmented<br>epidermal lesions<br>including<br>dyschromia,<br>hyperpigmentation,<br>melasma,<br>ephelides<br>(freckles). | |
| | | • Cutaneous lesions<br>including scars | |
| | | • Benign cutaneous<br>vascular lesions<br>including port wine<br>stains,<br>hemangiomas,<br>facial, truncal and<br>leg telangiectasia,<br>erythema of<br>rosacea, leg veins,<br>spider angiomas<br>and venous<br>malformations. | |
| | | The integrated thermal<br>cooling is indicated for use<br>in cooling the epidermis at | |
| | | the treatment site prior to, | |
| | | during, and after light | |
| | | treatment in general | |
| | | aesthetic dermatologic and | |
| | | plastic surgery procedures. | |
| | | - Reduce pain during light | |
| | | treatment (via partial | |
| | | anesthesia from | |
| | | cooling) | |
| | | - Reduce discomfort | |
| | | during and/or | |
| | | associated with light | |
| | | treatment | |
| | | - Minimize thermal injury, | |
| | | including thermal | |
| | | necrosis, to non-target | |
| | | skin and skin structures | |
| | | during and/or | |
| | | associated with light | |
| | | treatment, thus reducing | |
| | | possible complications | |
| | | such as scabbing, | |
| | | scarring, hyper - and/or | |
| | | hypo pigmentation | |
| | | - Allow the use of higher | |
| | | light or laser fluencies | |
| | | for light treatments | |
| | | (such as for hair | |
| | | removal and the | |
| | | treatment of vascular or | |
| | | pigmented lesions) | |
| | | - Reduce potential side | |
| | | effects of light | |
| | | treatments (such as for | |
| | | hair removal and the | |
| | | treatment of lesions) | |
| Light source | Intense pulsed light<br>(Xenon Flash Lamp) | Intense pulsed light (Xenon<br>Flash Lamp) | Identical |
| Wavelength | 420 – 1200 nm | 420 – 1200 nm | Identical |
| | | | |
| range | | | |
| Hanpiece ports | HS-650K&HS-660K: 2<br>HS620K, HS-300CK &<br>HS-310K: 1 | HS 650: 2<br>HS 300C: 1 | Identical |
| Structure | HS-650K&HS-660K:<br>Vertical<br>HS620K, HS-300CK &<br>HS-310K: Table top | HS 650: Vertical<br>HS 300C: Table top | Identical |
| Energy output | 4.1-50.8 J/cm² | 10-50 J/cm² | Equivalent |
| Pulse width | 5-20 ms | 2-20 ms | Equivalent |
| Pulse duration | 5-50 ms | 5-50 ms | Identical |
| Spot size | 12*35mm, 15*50mm | 12*35mm, 15*50mm | Identical |
| Filters | 420 -1200nm: Acne;<br>510 -1200nm: Acne,<br>vascular, pigment;<br>560 -1200nm: Acne,<br>vascular, pigment;<br>610-1200nm: Hair<br>removal;<br>640-1200nm: Hair<br>removal;<br>690-1200nm: Hair<br>removal; | 420 -1200nm: Acne;<br>510 -1200nm: Acne,<br>vascular, pigment;<br>560 -1200nm: Acne,<br>vascular, pigment;<br>610-1200nm: Hair removal;<br>640-1200nm: Hair removal;<br>690-1200nm: Hair removal; | Identical |
| Fluences | 420 -1200nm:<br>4.1-50.8J/cm²;<br>510 -1200nm: 3.8-47<br>J/cm²;<br>560 -1200nm: 3.7-43.3<br>J/cm²;<br>610-1200nm: 3.5-38.7<br>J/cm²;<br>640-1200nm:<br>3.3-37.4J/cm²;<br>690-1200nm:<br>3.1-33.4J/cm²; | 420 -1200nm: 10-50J/cm²;<br>510 -1200nm: 9.2-46.35<br>J/cm²;<br>560 -1200nm: 9-42.6 J/cm²;<br>610-1200nm: 8.5-38.1<br>J/cm²;<br>640-1200nm: 8-36.8J/cm²;<br>690-1200nm:<br>7.6-32.9J/cm²; | The<br>fluences for<br>each filter<br>are at<br>similar<br>energy<br>level,<br>considered<br>equivalent. |
| Output mode | Pulse mode | Pulse mode | Identical |
| Deliver<br>materials | Direct sapphire Coupling | Direct sapphire Coupling | Identical |
| Cooling<br>method | HS-650K&HS-660K:<br>Water cooling, forced-air<br>cooling, copper and TEC;<br><br>HS620K,<br>HS-300CK&HS-310K:<br>Water cooling and<br>forced-air cooling, | TEC and circulating water &<br>air | Equivalent |
# VI Comparison of technological characteristics with the predicate devices
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
# VII Non-Clinical Testing
A series of tests have been performed to verity that the proposed device met all design specification. The test result demonstrated that the proposed device complies with the following standards:
## Electrical safety and electromagnetic compatibility
- IEC 60601-1: 2005+corr.1:2006+Corr.2.2007+A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2:2014 Medical electrical equipment -Part 1-2: General ● requirements for basic safety and essential performance-Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-2-57: 2011 Medical electrical equipment Part 2: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
## Biocompatibility Evaluation:
The contact material of the proposed device is identical to that of the iPulseLight IPL System cleared in K093627.
Per FDA's Biocompatibility Guidance issued on June 16, 2016 and with regard to Table A.1 Evaluation Tests for consideration in ISO, "Use of international Standard ISO 10993-1, Biological evaluation of medical - Part 1: Evaluation and testing within a risk management process," the following tests performed on the material which contacts with human for Biocompatibility:
- Cytotoxicity;
- Skin irritation; ●
- Skin Sensitization .; ●
{9}------------------------------------------------
# VIII Clinical Testing
It is not applicable.
IX Conclusion substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.