Sirius Pintuition Seed, Sirius Pintuition Probe, Sirius Pintuition Base Unit
K200734 · Sirius Medical Systems B.V. · PBY · Feb 24, 2021 · General, Plastic Surgery
Device Facts
Record ID
K200734
Device Name
Sirius Pintuition Seed, Sirius Pintuition Probe, Sirius Pintuition Base Unit
Applicant
Sirius Medical Systems B.V.
Product Code
PBY · General, Plastic Surgery
Decision Date
Feb 24, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4300
Device Class
Class 2
Indications for Use
The Sirius Pintuition Seed is intended to be placed percutaneously in the breast to mark temporarily (< 30 days) a lumpectomy site intended for surgical removal. Using imaging guidance (such as ultrasound or radiography) or aided by non-imaging guidance (Sirius Pintuition Detector) the Sirius Pintuition Seed is located and surgically removed with the target tissue. The Sirius Pintuition Detector is intended for the non-imaging detection and localization of the Sirius Pintuition Seed that has been implanted in a lumpectomy site intended for surgical removal.
Device Story
System consists of implantable Pintuition Seed, handheld Pintuition Probe, and mains-powered Pintuition Base Unit. Seed (1.65 x 5mm, Titanium-encased Neodymium magnet) placed percutaneously in breast tissue via 14G needle. Probe detects magnetic field generated by seed; base unit processes signal to provide real-time audible and visual distance feedback (0-50mm range) to surgeon. Used in clinical/surgical settings to guide lumpectomy procedures. Enables precise localization of target tissue for excision; benefits patient by facilitating accurate removal of breast lesions.
Clinical Evidence
Clinical evaluation included performance data from the subject device, a previous version, and literature review of the predicate and benchmark devices. No specific clinical trial metrics (e.g., sensitivity/specificity) provided in summary; safety and effectiveness supported by design verification, system accuracy/range testing, biocompatibility, MR safety, and sterilization validation.
Technological Characteristics
System uses magnetic sensing. Seed: Grade II Titanium exterior, Neodymium magnet core. Probe: PEEK tissue-contacting material. Delivery: 14G 304 stainless steel needle. Connectivity: Cable-connected probe to console. Sterilization: Ethylene Oxide. Sensing depth: 0-50mm. Visual/audible feedback.
Indications for Use
Indicated for patients requiring temporary (<30 days) marking of a breast lumpectomy site for surgical removal. Used for percutaneous placement of a tissue marker and subsequent intraoperative detection/localization of the marker to guide surgical excision.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 24, 2021
Sirius Medical Systems B.V. Bram Schermers Clinical Application Specialist High Tech Campus 41 Eindhoven, North Brabant 5656AE Netherlands
Re: K200734
Trade/Device Name: Sirius Pintuition Seed, Sirius Pintuition Probe, Sirius Pintuition Base Unit Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: PBY Dated: January 18, 2021 Received: January 25, 2021
Dear Bram Schermers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
## K200734
Device Name
Sirius Pintuition Localization System (Sirius Pintuition Seed and Sirius Pintuition Detector)
Indications for Use (Describe)
The Sirius Pintuition Seed is intended to be placed percutaneously in the breast to mark temporarily (< 30 days) a lumpectomy site intended for surgical removal. Using imaging guidance (such as ultrasound or radiography) or aided by non-imaging guidance (Sirius Pintuition Detector) the Sirius Pintuition Seed is located and surgically removed with the target tissue.
The Sirius Pintuition Detector is intended for the non-imaging detection of the Sirius Pintuition Seed that has been implanted in a lumpectomy site intended for surgical removal.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
or sponsor, and a person is not required information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| | PREMARKET NOTIFICATION 510(k) | | |
|--------------------------------------------|-------------------------------|--|--|
| <img alt="sirius medical logo" src="..."/> | | | |
| Meta | |
|----------------|-----------|
| Identification | 001015 |
| Version | 2.0 |
| Classification | PUBLIC |
| Page | 15 of 121 |
Sirius Pintuition Localization System (Sirius Pintuition Seed and Sirius Pintuition Detector)
### 5 510(k) Summary
#### 5.1 Submitter Information
| Submitter's name: | Sirius Medical Systems B.V. |
|------------------------|-----------------------------------|
| Address: | High Tech Campus 41 |
| | 5656 AE |
| | Eindhoven |
| | The Netherlands |
| Contact Person: | Bram Schermers |
| | Product Owner |
| Telephone: | 0031 6 2011 6299 |
| E-mail: | bram.schermers@sirius-medical.com |
| Date summary prepared: | March 9th, 2020 |
#### 5.2 Device Information
| Trade name: | Sirius Pintuition Seed and Sirius Pintuition Detector |
|-------------------------|-------------------------------------------------------|
| Common name / device: | Temporary Tissue Marker |
| Regulation description: | Implantable Clip |
| Regulation number: | 21 CFR 878.4300 |
| Regulatory Class: | Class II |
| Review Panel: | General & Plastic Surgery |
| Product Code: | PBY |
#### 5.3 Predicate Device
Endomag Magseed and Sentimag (K163541)
#### 5.4 Device Description
The Sirius Pintuition Seed and Sirius Pintuition Detector are part of the Sirius Pintuition Localization System.
The Sirius Pintuition Seed is a small (1.65 x 5mm) Titanium tissue marker that is intended to be placed percutaneously in the breast for temporary (<30 days) marking of a lumpectomy site intended for surgical removal. The device is supplied single-use, sterile and pre-loaded within its delivery needle (12cm or 20cm length).
The Pintuition Detector is designed to detect the presence and proximity of the implanted Pintuition Seed. It consists of a mains-powered, table-top Pintuition Base Unit, and a cableconnected, reusable Pintuition Probe. Using the Pintuition Probe, a user may use the Pintuition Detector prior to and during breast surgery to plan the surgical approach and guide surgery. The location of the seed is fed back to the user using audible and visual cues (distance in mm).
The principle of operation is magnetism, the Pintuition Seed is associated with a magnetic field which the Pintuition Detector utilizes to determine the location of the Pintuition Seed.
{5}------------------------------------------------
| | PREMARKET NOTIFICATION 510(k) | | |
|----------------|-------------------------------|----------------|-----------|
| | Meta | Identification | Version |
| sirius medical | | 001015 | 2.0 |
| | Classification | | Page |
| | PUBLIC | | 16 of 121 |
# Sirius Pintuition Localization System (Sirius Pintuition Seed and Sirius Pintuition Detector)
#### 5.5 Intended Use
The Sirius Pintuition Seed is intended to be placed percutaneously in the breast to mark temporarily (< 30 days) a lumpectomy site intended for surgical removal. Using imaging guidance (such as ultrasound or radiography) or aided by non-imaging guidance (Sirius Pintuition Detector) the Sirius Pintuition Seed is located and surgically removed with the target tissue.
The Sirius Pintuition Detector is intended for the non-imaging detection and localization of the Sirius Pintuition Seed that has been implanted in a lumpectomy site intended for surgical removal.
#### Summary of Technological Characteristics 5.6
| Elements of<br>Comparison | Sirius Pintuition System | Predicate: Magseed Magnetic Marker<br>System | Comparison | | |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|--|--|
| | SYSTEM | | | | |
| 510(k) ID | [TBD] | K163541 | N/A | | |
| Regulation<br>Number | §878.4300 | §878.4300 | Same | | |
| Regulation<br>Description | Implantable Clip | Implantable Clip | Same | | |
| Regulatory<br>Class | Class II | Class II | Same | | |
| Product Code | PBY | PBY | Same | | |
| Intended use | Temporary marking of a breast<br>lumpectomy site for surgical removal | Temporary marking of a breast<br>lumpectomy site for surgical removal | Same | | |
| Indications<br>for use | The Sirius Pintuition Seed is intended to<br>be placed percutaneously in the breast<br>to mark temporarily (< 30 days) a<br>lumpectomy site intended for surgical<br>removal. Using imaging guidance (such<br>as ultrasound or radiography) or aided<br>by non-imaging guidance (Sirius<br>Pintuition Detector) the Sirius Pintuition<br>Seed is located and surgically removed<br>with the target tissue.<br>The Sirius Pintuition Detector is<br>intended for the non-imaging detection<br>and localization of the Sirius Pintuition | The Endomag Magseed Magnetic Marker<br>is intended to be placed percutaneously in<br>the breast to mark temporarily (< 30 days)<br>a lumpectomy site intended for surgical<br>removal. Using imaging guidance (such as<br>ultrasound or radiography) or aided by non-<br>imaging guidance (Endomag Sentimag®<br>System) the Endomag Magseed Magnetic<br>Marker is located and surgically removed<br>with the target tissue.<br>The Endomag Sentimag® System is<br>intended for the non-imaging detection and<br>localization of the "Endomag Magseed | Same | | |
| | Seed that has been implanted in a<br>lumpectomy site intended for surgical<br>removal. | Magnetic Marker" that has been implanted<br>in a lumpectomy site intended for surgical<br>removal | | | |
| Type of Use | Prescription Use | Prescription Use | Same | | |
| Anatomical<br>Locations | Breast | Breast | Same | | |
{6}------------------------------------------------
| | PREMARKET NOTIFICATION 510(k) | | |
|----------------|-------------------------------|----------------|-----------|
| | Meta | Identification | Version |
| sirius medical | | 001015 | 2.0 |
| | Classification | | Page |
| | PUBLIC | | 17 of 121 |
Title
# Sirius Pintuition Localization System (Sirius Pintuition Seed and Sirius Pintuition Detector)
| Elements of<br>Comparison | Sirius Pintuition System | Predicate: Magseed Magnetic Marker<br>System | Comparison |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| Techno-<br>logical<br>Charac-<br>teristics | The Sirius Pintuition System utilizes<br>non-imaging technologies that are<br>comprised of a console that<br>incorporates electronics and a simple<br>user interface, plus a probe handpiece.<br><br>A location marker (Pintuition Seed) is<br>placed percutaneously in situ at the<br>clinical target site by a delivery system<br>and then the detector handpiece is used<br>for the intraoperative detection and<br>localization of the implanted marker.<br><br>The handpiece is connected by a flexible<br>cable to a console unit that provides the<br>user with a visual indication of the<br>presence and proximity of the marker. | The Magseed Magnetic Marker System<br>utilizes non-imaging technologies that are<br>comprised of a console that incorporates<br>electronics and a simple user interface,<br>plus a probe handpiece.<br><br>A location marker (Magseed) is placed<br>percutaneously in situ at the clinical target<br>site by a delivery system and then the<br>detector handpiece is used for the<br>intraoperative detection and localization of<br>the implanted marker.<br><br>The handpiece is connected by a flexible<br>cable to a console unit that provides the<br>user with a visual indication of the<br>presence and proximity of the marker. | Same |
| DETECTOR | | | |
| Probe type | Handheld, flexible, cord-connected,<br>reusable | Handheld, flexible, cord-connected,<br>reusable | Same |
| Probe tissue<br>contacting<br>material | Poly Ether Ether Ketone (PEEK) | Poly Ether Ether Ketone (PEEK) | Same |
| User<br>Feedback | Real-time visual and audible | Real-time visual and audible | Same |
| Sensing<br>Depth | 0-50 mm | 0-30 mm | Similar. Sirius<br>Pintuition System<br>has improved<br>sensing depth. |
| SEED/MARKER | | | |
| Seed/Marker<br>Materials | Commercially Pure Titanium Grade II<br>(Tissue-contacting)<br>Neodymium magnet (Internal) | Surgical Grade Stainless Steel | Similar. Both<br>biocompatible<br>exterior. Internal<br>material different<br>due to different use<br>of magnetism. |
| Seed/Marker<br>diameter | 1.65mm | 1.0mm | Similar. Pintuition<br>Seed is slightly<br>larger to enable<br>greater sensing<br>depth. |
| Seed/Marker<br>length | 5.20mm | 5.0mm | Similar |
| Sterility | Ethylene Oxide | Ethylene Oxide | Same |
| Visibility | X-ray, Ultrasound | X-ray, Ultrasound | Same |
| DELIVERY DEVICE | | | |
| Type | Preloaded, single-use, needle implanter | Preloaded, single-use, needle implanter | Same |
| Material | 304 Stainless Steel | 304 Stainless Steel | Same |
| Elements of<br>Comparison | Sirius Pintuition System | Predicate: Magseed Magnetic Marker<br>System | Comparison |
| Delivery<br>device gauge | 14G | 18G | Similar. Pintuition<br>gauge is slightly<br>larger to enable<br>implantation of<br>larger Seed. |
{7}------------------------------------------------
| | PREMARKET NOTIFICATION 510(k) | | |
|----------------|-------------------------------|----------------|-----------|
| | Meta | Identification | Version |
| sirius medical | | 001015 | 2.0 |
| | Classification | Page | |
| | PUBLIC | | 18 of 121 |
Sirius Pintuition Localization System (Sirius Pintuition Seed and Sirius Pintuition Detector)
#### 5.7 Summary of Non-Clinical Performance Data
Testing was conducted to evaluate and characterize the safety and performance of the Sirius Pintuition Localization System. Pre-clinical testing included:
- Design verification ●
- System accuracy and range verification
- Biocompatibility evaluation .
- MR safety testing ●
- Sterilization validation
- Packaging validation ●
- . Shelf life validation
- . Electrical safety testing
#### 5.8 Summary of Clinical Performance Data
An analysis of available data was conducted to evaluate and characterize the clinical safety and performance of the Sirius Pintuition Localization System. The clinical data support the safety and effectiveness of the device:
- Clinical Evaluation, including clinical safety and performance data with the actual device, a previous version of the device and an extensive evaluation of available literature data pertaining to the previous version device, the predicate device and additional benchmark devices.
#### 5.9 Conclusion
The Sirius Pintuition Localization System has the same Intended Use as the predicate device. The different technological characteristics have not led to additional questions of safety or effectiveness. The data presented in the remainder of this submission show substantial equivalence of the Sirius Pintuition Localization System with the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.