The AtriAmp supports pacing by acting as an extension cable between an external pacemaker and cardiac pacing leads. The AtriAmp also supports monitoring of intra-cardiac signals to one of the unipolar precordial chest leads (leads V1-V6) of a Type CF, defibrillation-proof ECG monitor.
Device Story
AtriAmp is a sterile, single-use passive interface cable; acts as a connection hub between an external pacemaker, patient-contacting epicardial pacing lead wires, and a cardiac monitor. Device enables simultaneous real-time, continuous atrial electrogram monitoring and cardiac pacing. Used in clinical settings; operated by healthcare professionals. Input: electrical signals from cardiac pacing leads. Output: transmitted signals to ECG monitor and pacing current to heart. Benefits patient by allowing continuous monitoring of intra-cardiac signals while maintaining pacing capability.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (ANSI/AAMI ES60601-1, IEC 60601-2-31), noise performance (ANSI/AAMI EC53:2013), biocompatibility (ISO 10993-5, ISO 10993-10), and sterile barrier integrity.
Indicated for patients requiring temporary cardiac pacing and monitoring of intra-cardiac signals via an external pacemaker and a Type CF, defibrillation-proof ECG monitor.
Regulatory Classification
Identification
A patient transducer and electrode cable (including connector) is an electrical conductor used to transmit signals from, or power or excitation signals to, patient-connected electrodes or transducers.
Electrode/Extension Cables, Model series ATAR (K070926)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" stacked on top of each other.
July 1, 2020
Atrility Medical, LLC % Gary Syring Principal Consultant Quality & Regulatory Associates, LLC 800 Levanger Lane Stoughton, Wisconsin 53589
Re: K200674
Trade/Device Name: AtriAmp - Pacing Cable Medtronic Connector, AtriAmp - Pacing Cable Oscor Connector Regulation Number: 21 CFR 870.2900 Regulation Name: Patient Transducer And Electrode Cable (Including Connector) Regulatory Class: Class II Product Code: DSA Dated: May 15, 2020 Received: May 15, 2020
Dear Gary Syring:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Hetal Odobasic Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200674
Device Name AtriAmp
Indications for Use (Describe)
The AtriAmp supports pacing by acting as an extension cable between an external pacemaker and cardiac pacing leads. The AtriAmp also supports monitoring of intra-cardiac signals to one of the unipolar precordial chest leads (leads V1-V6) of a Type CF, defibrillation-proof ECG monitor.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Atrility Medical. The word "Atrility" is in red, with the "A" being larger than the other letters. Below "Atrility" is the word "MEDICAL" in gray, smaller font. The logo is simple and modern.
#### 510(k) Summary
This 510(k) summary was prepared to provide an understanding of the determination of substantial equivalence in accordance with the requirements 21 CFR 807.92.
| Submitters Name: | Atrility Medical, LLC<br>313 Price Place, Suite #13<br>Madison, WI 53705 |
|------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Matthew Knoespel, Regulatory Engineer<br>Atrility Medical, LLC |
| Contact Phone: | Phone: 920-619-8737 |
| Date Summary Prepared: | June 22, 2020 |
| Device Trade Name: | AtriAmp |
| Common Name: | Passive interface cable to external pacemaker, epicardial pacing lead wire and<br>cardiac monitor |
| Classification Name: | 21 CFR 870.290 Patient transducer and electrode cable (including connector)<br>Product Code: DSA, Class II |
| Predicate Devices: | K080621, Fastlock Pacing Wire Extension (Primary Predicate)<br>K070926, Electrode/Extension Cables, Model series ATAR |
#### Device Description
The AtriAmp acts as a connection hub for these three systems (external pacemaker, patient contacting external pacemaker lead wires and cardiac monitor) to provide a real-time, continuous atrial electrogram to the cardiac monitor while simultaneously allowing pacing of the atriAmp is a sterile, single patient use device.
#### Intended Use of the Device
The AtriAmp is a single-use device that is intended to connect an electrode/lead from a patient to a Type CF, defibrillation-proof ECG Monitor and/or an external pacemaker.
#### Summary of Technological Characteristics
The following table provides a side-by-side comparison of the predicate devices applied to support this pre-market notification.
| Technical Characteristics Comparison | | | | |
|--------------------------------------|--------------------------------------------|-------------------------------------------------------------------------------|---------------------------------------------------------------------------|------------------------------------|
| Feature | AtriAmp and<br>Accessories<br>Under Review | Fastlock Pacing<br>Wire Extension<br>Cable<br>(Primary Predicate:<br>K080621) | Electrode/Extension<br>Cables, Model series<br>ATAR<br>Predicate: K070926 | Equivalence<br>Comments |
| Product Code,<br>Classification | DSA,<br>21 CFR 870.2900,<br>Class II | DSA,<br>21 CFR 870.2900,<br>Class II | IKD,<br>21 CFR 890.1175<br>Class II | Identical to primary<br>predicate. |
{4}------------------------------------------------
| Technical Characteristics Comparison | | | | |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | AtriAmp and<br>Accessories<br>Under Review | Fastlock Pacing<br>Wire Extension<br>Cable<br>(Primary Predicate:<br>K080621) | Electrode/Extension<br>Cables, Model series<br>ATAR<br>Predicate: K070926 | Equivalence<br>Comments |
| Indications for<br>Use | The AtriAmp supports<br>pacing by acting as an<br>extension cable between<br>an external pacemaker<br>and cardiac pacing<br>leads. The AtriAmp also<br>supports monitoring of<br>intra-cardiac signals by<br>transmitting signals to<br>one of the unipolar<br>precordial chest leads<br>(leads V1-V6) of a Type<br>CF, defibrillation-<br>proof ECG monitor. | Fastlock pacing wire<br>extension cable with<br>connectors is indicated<br>for use as an electrical<br>extension cable used to<br>transmit signal from, or<br>power or excitation<br>signal to patient-<br>connected electrodes. | Electrode/extension<br>cable is intended to<br>connect an<br>electrode/lead from a<br>patient or another<br>cable to a diagnostic<br>machine or an<br>external pacemaker. | Equivalent<br>The same as the<br>predicate devices, the<br>AtriAmp interfaces<br>between an external<br>pacemaker and patient<br>applied intra-cardiac<br>pacing lead wires.<br>Equivalent to the<br>secondary predicate, the<br>AtriAmp interfaces the<br>signal from the<br>epicardial lead wires to<br>a, defibrillation-proof<br>ECG Monitor, one of<br>the leads (V1 to V6) of<br>an ECG monitor. |
| Duration of<br>use | Anticipated to be less<br>than 14 days | < 7 Days | < 7 Days | Similar |
| Single use,<br>disposable | Yes | Yes | Yes | Same |
| Provided sterile | Yes | Yes | Yes | Same |
| Sterilization<br>method | e-beam | ethylene oxide | ethylene oxide | Similar |
| Sterility<br>Assurance Level<br>(SAL) | $ ≤ 10^{-6} $ | Unknown | Unknown | Acceptable for the<br>intended use |
| Sterile barrier<br>package | Flexible Double Pouch | Tyvek Flexible Barrier | Tyvek Flexible Barrier | Similar |
| Biocompatibility<br>evaluation | Cytotoxicity, Irritation,<br>Sensitization | Unknown | Unknown | Similar |
# Non-clinical Performance Tests
To establish the technical equivalency of the AtriAmp with accessories, evaluations were conducted to confirm
compliance with performance requirements, including:
| Test | Test Method Summary | Result |
|-------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| ANSI/AAMI<br>EC53:2013<br>compliance | The AtriAmp complies with applicable requirements of EC53:2013<br>and performs the same as the predicate for Subclause 5.3.2 Cable<br>and leadwire noise. | Pass |
| ANSI/AAMI ES<br>60601-1 and IEC<br>60601-2-31<br>compliance | The AtriAmp with accessory external pacemaker interface cables<br>complies with applicable requirements of ANSI/AAMI ES60601-<br>1:2005 & A1:2012 and IEC 60601-2-31:2011. | Pass |
{5}------------------------------------------------
| Test | Test Method Summary | Result |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| ES60601-1<br>Subclause 8.5.2.3 | In support of 21 CFR Part 898 compliance, the AtriAmp complies<br>with ES60601-1 Subclause 8.5.2.3 Patient leads or patient cables. | Pass |
| Sterility | Sterilization of the subject device was validated according to ISO<br>11137-1:2006(R)2015 & A1:2013 & A2:2018, demonstrating the<br>device is provided with a minimum sterility assumed level (SAL) of<br>$10^{-6}$ . | Pass |
| Sterile barrier<br>integrity | The sterile barrier package is a flexible pouch. The methods applied<br>to evaluate the sterile barrier package integrity included post<br>terminal sterilization simulated distribution, seal peel and bubble<br>emission testing. | Pass |
| Biocompatibility | The ISO 10993-5:2009 and ISO 10993-10:2020 test methods were<br>applied, with passing results: for Cytotoxicity, Sensitization,<br>Irritation. | Pass |
| Shelf life<br>two (2) years | Confirmation of device functional performance and sterile barrier<br>pouch integrity (seal peel per ASTM F88 and bubble leak per ASTM<br>F2096) with accelerated aging and simulated distribution per ASTM<br>4169-16 and bubble leak (ASTM F2096). | Pass |
| Duration of use | Operation use was evaluated to verify use for up to 14 days. | Pass |
Clinical data are not needed.
## Conclusion
The AtriAmp with accessories meet performance requirements equivalent to the predicate device. The intended use and technology of the AtriAmp with accessories is the same as the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.