← Product Code [IZL](/productcode/IZL) · K200637

# SR-8100 Portable X-ray Unit (K200637)

_Shantou Institute of Ultrasonic Instruments Co., Ltd. (Siui) · IZL · Jul 9, 2020 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K200637

## Device Facts

- **Applicant:** Shantou Institute of Ultrasonic Instruments Co., Ltd. (Siui)
- **Product Code:** [IZL](/productcode/IZL.md)
- **Decision Date:** Jul 9, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1720
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The SR-8100 Portable X-ray Unit is a portable X-ray device, intended for use by a qualified/trained physician or technician for the purpose of acquiring X-ray images of the desired parts of patient's anatomy (including head, cervical spine, chest, abdomen, lumbar spine, pelvis and extremities). The device may be used for handheld diagnostic imaging of body extremities. The device may be used for handheld diagnostic imaging of body extremities. The system is subject to the following limitations of use when stand-mounted: The device may be used for diagnostic imaging of head, cervical spine, lumbar spine, pelvis or extremities. The device may be used for imaging of the chest when used without a grid. This device is not to be used on bariatric patients, unless imaging body extremities. This device is not intended for mammography.

## Device Story

Portable digital X-ray unit; emits X-rays for diagnostic imaging. Components: X-ray main unit, exposure hand switch, charger. Operated by physician or technician in mobile or fixed clinical sites. User adjusts kV and mAs parameters via control panel; system control software manages high-voltage tank to excite X-ray tube. Produces X-ray images for diagnosis of bone and soft tissues. Benefits include high-definition imaging and low-dose radiation exposure. Device is battery-powered (lithium-ion).

## Clinical Evidence

No clinical data. Substantial equivalence demonstrated through bench testing, including compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-54, IEC 60601-1-3, ISO 14971, ISO 10993, and IEC 62304. Image quality verified by radiologist review to meet diagnostic requirements.

## Technological Characteristics

Portable X-ray system; 40-100kVp range; 2.0kW output. High-frequency inverter; stationary anode X-ray tube (D-125SB). Lithium-ion battery (29.6V DC). Manual collimator (SR-8100-19L) with LED light. Connectivity: standalone unit with digital detector interface. Software: system control on ARM platform. Sterilization: N/A (non-invasive).

## Regulatory Identification

A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- MinXray, Inc TR90BH ([K182207](/device/K182207.md))

## Reference Devices

- DRGEM Corporation JADE Mobile X-Ray ([K183388](/device/K183388.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Shantou Institute of Ultrasonic Instruments Co., Ltd. (SIUI) % Flower Cai Liaison Manager 77 Jinsha Road Shantou. Guangdong 515041 CHINA

### Re: K200637

Trade/Device Name: SR-8100 Portable X-ray Unit Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: Class II Product Code: IZL Dated: June 12, 2020 Received: June 15, 2020

Dear Flower Cai:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

July 9, 2020

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K200637

Device Name SR-8100 Portable X-ray Unit

### Indications for Use (Describe)

The SR-8100 Portable X-ray Unit is a portable X-ray device, intended for use by a qualified/trained physician or technician for the purpose of acquiring X-ray images of the desired parts of patient's anatomy (including head, cervical spine, chest, abdomen, lumbar spine, pelvis and extremities). The device may be used for handheld diagnostic imaging of body extremities.

The device may be used for handheld diagnostic imaging of body extremities.

The system is subject to the following limitations of use when stand-mounted: The device may be used for diagnostic imaging of head, cervical spine, lumbar spine, pelvis or extremities. The device may be used for imaging of the chest when used without a grid.

This device is not to be used on bariatric patients, unless imaging body extremities. This device is not intended for mammography.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

This summary of 510(k) safety and effectiveness information is provided in accordance with the requirements of SMDA 1990 and 21 CFR 807.92(c).

The assigned 510(k) number is: K200637

#### Submitter 5.1

Shantou Institute of Ultrasonic Instruments Co., Ltd. (SIUI)

77 Jinsha Road, Shantou, Guangdong 515041, China

Tel: 86-754-88250150

Fax: 86-754-88251499

Contact Person: Flower Cai

Shantou Institute of Ultrasonic Instruments Co., Ltd.

77 Jinsha Road, Shantou, Guangdong 515041, China

Date Prepared: June 12, 2020

### 5.2 Device

Name of Device: SR-8100 Portable X-ray Unit

Classification Name: Mobile x-ray system (21 CFR 892.1720)

Regulatory Class: II

Product Code: IZL

#### Predicate Device 5.3

The SR-8100 Portable X-ray Unit is for X-ray imaging, diagnosis and image acquisition with mobile or fixing sites, and is substantially equivalent to the predicate device:

{4}------------------------------------------------

| Legally Marketed<br>Device  | Manufacturer      | Model             | 510(k) Control<br>Number |
|-----------------------------|-------------------|-------------------|--------------------------|
| Primary<br>predicate device | MinXray, Inc      | TR90BH            | K182207                  |
| Reference device            | DRGEM Corporation | JADE Mobile X-Ray | K183388                  |

The reference device is for comparing the collimator parameters and output rating parameters of the portable X-ray units.

### 5.4 Device Description

This SR-8100 Portable X-ray Unit is a portable digital device developed, designed and manufactured by SIUI. Its high-definition imaging and low-dose X-ray significantly minimize harm to patients and practitioners. A detailed comparison table with an equivalent device is provided below. The device consists of the following major components: an X-ray main unit, an exposure hand switch and a charger. The X-ray main unit is mainly for emitting X-rays required for X-ray exams; the hand switch is for output control of X-ray emitting, and the charger is for charging the built-in battery in the X-ray main unit. See the photos below.

The device can be used with an X-ray detector (see the table below for the basic requirements), a computer for receiving and detecting signal results and image processing software.

The detector that can be used with SR-8100 shall comply with the following basic conditions:

| Acceptable minimum X-ray range | 40kV-100kV                                                                     |
|--------------------------------|--------------------------------------------------------------------------------|
| Image Sensor                   | a-Si (Amorphous Silicon) TFT                                                   |
| Minimum Effective Array        | 2304 x 2800                                                                    |
| Minimum Effective Area (H x V) | 343 x 416 mm                                                                   |
| Software                       | Configured with a basic imaging test tool<br>for image acquisition and display |

The major components of the X-ray main unit include: handle, enclosure, control panel, system control (SYS) board, high-voltage tank, collimator (beam limiter), lithium-ion battery and system control software running on the SYS board.

{5}------------------------------------------------

### SIUI SR-8100 Portable X-ray Unit

The system control software is for real-time interaction and control with various circuit modules inside the portable X-ray unit. The software responds to user operations on the control panel. The user can adjust and control the kV and mAs parameters, and the software will display the parameters or directly load the APR parameters. The software loads the control data from X-ray output into the high-voltage generation control circuit of the system control board, and control the high-voltage tank to generate high-voltage to excite the X-ray tube inside to emit X-rays, control the switch of the collimator indicator, and monitor the working status of the device, the battery power status, and control the display of the status indicators.

Image /page/5/Picture/2 description: The image shows an X-ray main unit with several labeled components. The components labeled are the handle, control panel, enclosure, and collimator. The X-ray main unit is light blue and white and has a handle on top for easy transport.

The system is for X-ray imaging and diagnosis in facilities with mobile or fixing sites. Since the kV range of this device is 40~100kVp, which is not suitable for breast exams, the device is not intended for mammography.

The wavelength of medical X-ray spectrum used by the portable X-ray unit SR-8100 shall have a range 0.001-0.1nm.

{6}------------------------------------------------

#### Indications for Use ર્સ રાજ્યના સાથે તેમની સાથે તેમની સાથે છે. આ ગામનાં પ્રાથમિક શાળા, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં પ્રાથમિક શાળા, પંચાયતઘર, આંગણવાડી તેમ જ દૂ

The SR-8100 Portable X-ray Unit is a portable X-ray device, intended for use by a qualified/trained physician or technician for the purpose of acquiring X-ray images of the desired parts of patient's anatomy (including head, cervical spine, chest, abdomen, lumbar spine, pelvis and extremities).

The device may be used for handheld diagnostic imaging of body extremities.

The system is subject to the following limitations of use when stand-mounted: The device may be used for diagnostic imaging of head, cervical spine, abdomen, lumbar spine, pelvis or extremities.

The device may be used for imaging of the chest when used without a grid.

This device is not to be used on bariatric patients, unless imaging body extremities. This device is not intended for mammography.

#### 5.6 Comparison of Technological Characteristics with the Predicate Device

The comparison between the overall specifications of predicate device (TR90BH) and the new device (SR-8100) is shown in table 1, and the comparison between the Collimator specifications of reference device (JADE Mobile X-ray) and the new device (SR-8100) is shown in table 2. Any differences between the predicate and the new device have no impact on safety or efficacy of the new device and do not raise any new potential or increased safety risks, and the new device is equivalent in performance to existing legally marketed devices.

{7}------------------------------------------------

# Table 1 Subject and Predicate (TR90BH) Device Comparison

{8}------------------------------------------------

| Description               | Subject Device<br>SR-8100                                                     | Predicate Device<br>TR90BH (K182207)                                                                                                                                                                                                        |
|---------------------------|-------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comment on<br>Indications | Original Indications                                                          | Certain limitations apply because of the lower maximum kVp output as compared to the predicate and because of the possibility of handheld use on extremities. See indications above.                                                        |
| Weight                    | 12kgs (26.46lb)<br>(Including Collimator)                                     | 7.5kgs (16.53lb)<br>(Including Collimator)                                                                                                                                                                                                  |
| Size                      | 227mm×522mm×217mm                                                             | 219 mm x 442 mm x 190 mm                                                                                                                                                                                                                    |
| Energy Source             | Lithium-ion Rechargeable<br>Battery (29.6V DC),<br>300 exposures per charge   | Lithium-ion Rechargeable Battery<br>(57.6VDC),<br>300 exposures per charge.                                                                                                                                                                 |
| Use Interface             | Soft touch push buttons                                                       | Soft touch push buttons                                                                                                                                                                                                                     |
| Exposure times            | 0.02 sec – 2.5 sec: R'10 sec<br>Step                                          | 0.01 sec - 1.0 sec :0.01 sec Step<br>High Power Mode<br>0.01 sec – 0.3 sec:0.01 sec Step                                                                                                                                                    |
| mA                        | 25 mA @ 40kVDC – 60kVDC (1kVP steps)<br>20 mA @ 61kVDC – 100kVDC (1kVP steps) | 20 mA @ 40kVDC – 60kVDC (2kVP steps)<br>15 mA @ 62kVDC – 80kVDC (2kVP steps)<br>10 mA @ 82kVDC – 90kVDC (2kVP steps)<br>High Power Mode<br>15 mA @ 82kVDC – 90kVDC (2kVP steps)<br>0.02 Comment: 90 kVP maximum instead of 120 kVP maximum. |

{9}------------------------------------------------

### SIUI SR-8100 Portable X-ray Unit

| Description                    | Subject Device<br>SR-8100 | Predicate Device<br>TR90BH (K182207) |
|--------------------------------|---------------------------|--------------------------------------|
| Memory Settings<br>(technique) | 5 memories via pushbutton | 5 memories via pushbutton            |
| HF Generator                   | High Frequency            | High Frequency                       |
| kW                             | 2.0kW                     | 1.35kW                               |
| kVp                            | 40-100kVp                 | 40-90kVp                             |
| X-ray Tube                     | D-125SB                   | D-0814                               |
| FDA<br>Performance<br>Standard | Complies                  | Complies                             |
| Collimator                     | SIUI SR-8100-19L          | Mikasa BLD34L                        |

## Table 2 Subject and Reference Device (JADE-32) Comparison

| Table 2 Subject and Reference Device (JADE-32) Comparison |                                                      |                                                         |
|-----------------------------------------------------------|------------------------------------------------------|---------------------------------------------------------|
| Description                                               | Subject Device                                       | Reference Device                                        |
|                                                           | SR-8100                                              | JADE-32 (K183388)                                       |
| Output Rating                                             | Max. 2.0kW(20mA@100kV)                               | Max. 3.2kW(40mA@ 80kV,<br>32mA@100kV,25mA@120kV)        |
| Type                                                      | Microprocessor controlled<br>High Frequency inverter | Microprocessor controlled High<br>Frequency inverter    |
| kV Range                                                  | 40~100kV,61 Step (1kV Step)                          | 40~120kV,81 Step (1kV Step)                             |
| mA Range                                                  | 20~25mA, 2 Step<br>(20,25mA)                         | 10~80mA, 10 Step<br>(10,12.5,16,20,25,32,40,50,64,80mA) |
| Exposure Time                                             | 0.02~2.5 seconds,<br>21 Step (in 25% Steps)          | 0.01~10 seconds,<br>21 Step (in 25% Steps)              |
| mAs Range                                                 | 0.4~50mAs,<br>35 Step ( in 25% Steps)                | 0.1~250mAs,<br>35 Step ( in 25% Steps)                  |
| X-ray Tube Type                                           | Stationary Anode                                     | Stationary Anode                                        |
| Focal Spot Size<br>(Small/Large)                          | 1.2mm                                                | 0.5/1.5mm                                               |
| Anode Heat Storage Capacity                               | 50,000HU (35,000J)                                   | 56,000HU (40,000J)                                      |

{10}------------------------------------------------

### SIUI SR-8100 Portable X-ray Unit

| Description                            | Subject Device<br>SR-8100                      | Reference Device<br>JADE-32 (K183388)          |
|----------------------------------------|------------------------------------------------|------------------------------------------------|
| Power Cord<br>Length                   | 3m                                             | 5m                                             |
| Exposure<br>Hand-switch<br>Cord Length | 3m (Max. Length)                               | 5m (Max. Length)                               |
| X-ray switching<br>frequency           | 100kHz                                         | 100kHz                                         |
| Control                                | 2 Point Control (kV, mAs)                      | 2 Point Control (kV, mAs)                      |
| Anatomical<br>Programs                 | Preprogrammed 9 APR data-<br>User Programmable | Preprogrammed 9 APR data-<br>User Programmable |

# Table 3 Collimator Specs Comparison

| Description            | Subject Device<br>SIUI SR-8100                        | Reference Device<br>DRGEM JADE Mobile X-Ray<br>System (K183388) |
|------------------------|-------------------------------------------------------|-----------------------------------------------------------------|
| Model                  | SR-8100-19L                                           | KM1                                                             |
| Manufacturer           | SIUI                                                  | DRGEM                                                           |
| Control                | Manual<br>with 15,<br>30,<br>45,<br>60sec. Lamp timer | Manual with 15, 30, 45, 60sec.<br>Lamp timer                    |
| Field Shape            | Rectangular                                           | Rectangular                                                     |
| Max. Field Size        | 44x44cm (at 100cm SID)                                | 44x44cm (at 100cm SID)                                          |
| Leakage<br>Radiation   | < 40mR/hr. (at SID 1m)                                | < 40mR/hr. (at SID 1m)                                          |
| Max. kVp shield        | 150kV                                                 | 150kV                                                           |
| Inherent<br>Filtration | 1.2mmAl eq.                                           | 2.0mmAl eq.                                                     |
| Light source           | 9W LED                                                | 19W LED                                                         |
| Standard               | Rotating flange                                       | Rotating flange                                                 |

{11}------------------------------------------------

| Description       | Subject Device<br>SIUI SR-8100            | Reference Device<br>DRGEM JADE Mobile X-Ray<br>System (K183388) |
|-------------------|-------------------------------------------|-----------------------------------------------------------------|
| Option            | Tape measure (Max.200cm)                  | Tape measure (Max.200cm)                                        |
| Electrical Rating | 3-12VDC,10W                               | 12 – 45VDC, 20W                                                 |
| Dimension/weight  | 170(W)×180(D)×105(H)mm /<br>1.2kg(2.65lb) | 172(W)×172(D)×97(H)mm/<br>2.5kg(5.51lb)                         |

### 5.7 Non-clinical Testing Summary

The SR-8100 Portable X-ray Unit complies with and/or was tested in accordance with the following FDA guidance and International Standards:

- IEC 60601-1: 2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC60601-1-2: 2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.
- IEC60601-2-54: 2015 Medical electrical equipment Part 2-54: Particular requirements for the basic safety and essential performance of X-Ray equipment for radiography and radioscopy
- IEC60601-1-3: 2013 Medical electrical equipment Part 1-3: General requirements for basic safety and essential performance - Collateral standard: Radiation protection in diagnostic X-Ray equipment
- ISO 14971: 2007 Medical device -Application of risk management to medical devices
- ISO 10993-1: 2018 Biological evaluation of medical devices-Part1: Evaluation and testing
- ISO 10993-5:2009 Tests for in vitro cytotoxicity
- ISO 10993-10:2010 Tests for irritation and delayed-type hypersensitivity
- IEC 62304:2006+AMD1:2015 Medical device software – Software life-cycle processes
- IEC 60601-1-6:2015 Medical electrical equipment Part 1-6: General requirements for basic safety and essential performance – collateral standard: Usability
- IEC 62366-1:2015/COR1:2016 Medical devices - Application of usability

{12}------------------------------------------------

engineering to medical devices

- . ISO 15223-1:2016 Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General requirements
- ISO 13485 Quality management for medical devices
- . Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices

The subject device and the predicate device are comparable in terms of technical features, general functions, applications and intended uses. The test results showed compliance with the above standards. The performance tests of the device shows that the performance specifications of the device comply with the requirements of mobile X-ray system. As confirmed by the radiologist, the results of the image quality test show that the clinical diagnostic requirements of the image quality of each part meet the criteria requirements. Furthermore, the device runs stably during the test without any operating obstacles or failures. The non-clinical tests demonstrate that the device is as safe, as effective, and performs as well as the primary predicate.

#### Software Description 5.8

The system control software of SR-8100 is new software designed by SIUI. The software adopts development languages of keil, the program language is C language, the hardware platform is ARM. The system control software is for real-time interaction and control with various circuit modules inside the portable X-ray unit. The software responds to user operations on the control panel. The user can adjust and control the kV and mAs parameters, and the software will display the parameters or directly load the APR parameters. The software loads the control data from X-ray output into the high-voltage generation control circuit of the system control board, and control the high-voltage tank to generate high-voltage to excite the X-ray tube inside to emit X-rays, control the switch of the collimator indicator, and monitor the working status of the device, the battery power status, and control the display of the status indicators. The Level of Concern for the Software Device is Moderate.

### 5.9 Clinical Testing

Clinical testing is not necessary for the SR-8100 Portable X-ray unit in order to demonstrate substantial equivalence to the predicate device.

{13}------------------------------------------------

### 5.10 Conclusion

The subject device SR-8100 system and the predicate device are comparable in terms of technical features, general functions, applications and indications for use. The subject device SR-8100 system and reference device are comparable in terms of technical features, general functions, and collimator specs comparison.

The SR-8100 can be used for imaging including head, cervical spine, chest, abdomen, lumbar spine, pelvis and extremities, and the predicate device TR90BH can be used for imaging of head, abdomen, extremities and chest. Compared with TR90BH, the SR-8100 can be used for additional imaging of cervical spine, lumbar spine and pelvis. In terms of safety, the exposure conditions and radiation doses used for X-ray exams of cervical spine, lumbar spine and pelvis are equivalent to the parameters and radiation doses required for head, abdomen, extremities and chest exams, which is not intended to increase risks of user safety; In terms of effectiveness, the clinical images of bone and soft tissues of the subject device can address doctor's diagnostic requirements for the sites. Nevertheless, it does not affect the safety and effectiveness of the SR-8100, nor does it change the indications for use of the SR-8100 unit.

The indications for use of the SR-8100 Portable X-ray Unit do not create new potential safety risks, and its performance is comparable as that of the marketed products. Therefore, the device is as safe and effective as the legally marketed predicate device.

---

**Source:** [https://fda.innolitics.com/device/K200637](https://fda.innolitics.com/device/K200637)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
