← Product Code [FWP](/productcode/FWP) · K200610

# Customized Contour Implant (K200610)

_Implantech Associates, Inc. · FWP · Dec 23, 2020 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K200610

## Device Facts

- **Applicant:** Implantech Associates, Inc.
- **Product Code:** [FWP](/productcode/FWP.md)
- **Decision Date:** Dec 23, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.3550
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Customized Contour Implant is intended for augmentation, reconstructive and cosmetic surgery. The Customized Contour Implant is pre-shaped to the surgeon's specification to meet the needs of a particular patient, for example facial implants, gluteal implants, calf implants, or pectoralis implants.

## Device Story

Patient-matched, single-use silicone elastomer implant; designed for long-term implantation as space-occupying device to form contoured features. Surgeon specifies shape to meet individual patient needs. Used in reconstructive and cosmetic surgery settings. Implant provides physical augmentation of anatomical structures. Benefits include customized fit for specific patient anatomy.

## Clinical Evidence

Bench testing only. Testing included sterilization validation (ISO 17655-1, ANSI/AAMI/ISO 20857), packaging validation (ISO 11607), shelf life validation (accelerated and real-time aging), cytotoxicity testing (ISO 10993-5), and 3D printer validation.

## Technological Characteristics

Medical grade Nusil silicone elastomer; variable durometer range. Patient-matched geometry via 3D printing/custom fabrication. Single-use. Sterilization validated per ISO 17655-1 and ANSI/AAMI/ISO 20857.

## Regulatory Identification

A chin prosthesis is a silicone rubber solid device intended to be implanted to augment or reconstruct the chin.

## Predicate Devices

- Customized Contour Implant ([K191130](/device/K191130.md))
- Gluteal Implant, Models RND 5-X, TRD 6-X ([K052504](/device/K052504.md))
- Calf Implant, Model EC17-X ([K052505](/device/K052505.md))
- Pectoralis Implant ([K952708](/device/K952708.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized image of a human figure, and the text on the right is in blue. The text is arranged in three lines, with "FDA" on the first line, "U.S. FOOD & DRUG" on the second line, and "ADMINISTRATION" on the third line.

December 23, 2020

Implantech Associates Inc. % Pierre Bounaud Senior Consultant AcKnowledge Regulatory Strategies, LLC 2251 San Diego Ave, Suite B-257 San Diego, California 92110

Re: K200610

Trade/Device Name: Customized Contour Implant Regulation Number: 21 CFR 878.3550 Regulation Name: Chin Prosthesis Regulatory Class: Class II Product Code: FWP, KKY, MIB, MIC Dated: March 5, 2020 Received: March 9, 2020

Dear Mr. Bounaud:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K200610

Device Name Customized Contour Implant

#### Indications for Use (Describe)

The Customized Contour Implant is intended for augmentation, reconstructive and cosmetic surgery. The Customized Contour Implant is pre-shaped to the surgeon's specification to meet the needs of a particular patient, for example facial implants, gluteal implants, calf implants, or pectoralis implants.

Type of Use (Select one or both, as applicable)

| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)                 |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for "Implantech". The logo is white text on a black background. Below the main logo is the text "Superior Patient Aesthetics".

# 510(k) Summary

#### DATE PREPARED

March 5, 2020

#### MANUFACTURER AND 510(k) OWNER

Implantech Associates, Inc. 6025 Nicolle St., Suite B, Ventura, CA 93003, USA Telephone: Fax: +1 (805) 339-9414 Official Contact: Craig Arthur, RA Global Compliance Manager

### REPRESENTATIVE/CONSULTANT

Pierre Bounaud, Ph.D. Allison C. Komiyama, Ph.D., R.A.C. AcKnowledge Regulatory Strategies, LLC Telephone: +1 (619) 458-9547 Email: pbounaud@acknowledge-rs.com, akomiyama@acknowledge-rs.com Website: www.AcKnowledge-RS.com

#### DEVICE INFORMATION

| Trade Name/Proprietary Name | Customized Contour Implant |
|-----------------------------|----------------------------|
| Common Name                 | Elastomer, Silicone Block  |

| Classification Name                                                      | Classification Regulations | Device Class | Product Code |
|--------------------------------------------------------------------------|----------------------------|--------------|--------------|
| Chin prosthesis                                                          | 21 CFR 878.3550            | II           | FWP          |
| Polytetrafluoroethylene with carbon fibers<br>composite implant material | 21 CFR 878.3500            | II           | KKY          |
| Ear, nose, and throat synthetic polymer<br>material                      | 21 CFR 874.3620            | II           | MIB          |
|                                                                          |                            | II           | MIC          |

OPEQ/OHT4/Infection Control and Plastic Surgery Devices Premarket Review (DHT4B) General & Plastic Surgery Review Panel

{4}------------------------------------------------

510(k) Summary

# PREDICATE DEVICE IDENTIFICATION

The Customized Contour Implant is substantially equivalent to the following predicates:

| 510(k) Number | Predicate Device Name / Manufacturer                                 | Primary Predicate |
|---------------|----------------------------------------------------------------------|-------------------|
| K191130       | Customized Contour Implant/Implantech Associates, Inc.               | ✓                 |
| K052504       | Gluteal Implant, Models RND 5-X, TRD 6-X/Implantech Associates, Inc. |                   |
| K052505       | Calf Implant, Model EC17-X/Implantech Associates, Inc.               |                   |
| K952708       | Pectoralis Implant/Spirit Ridge Technologies                         |                   |

The predicate devices have not been subject to a design related recall.

# DEVICE DESCRIPTION

The Customized Contour Implant is a patient-matched device intended for augmentation, reconstructive and cosmetic surgery. The device is a single use implant intended for long term implantation as a space occupying device to form a contoured feature. The customized implant is made of medical grade silicone elastomer, in a range of durometers as specified by the surgeon.

# INDICATIONS FOR USE

The Customized Contour Implant is intended for augmentation, reconstructive and cosmetic surgery. The Customized Contour Implant is pre-shaped to the surgeon's specification to meet the needs of a particular patient, for example facial implants, calf implants, or pectoralis implants.

### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

Implantech Associates believes that the Customized Contour Implant is substantially equivalent to the predicate devices based on the information summarized here:

The subject device has the same intended use (facial, gluteal, calf and pectoral implants for reconstructive and augmentation/cosmetic surgery) and patient population as the devices cleared in K191130 (facial implants), K052504 (gluteal implants), K052505 (calf implants), and K952708 (pectoral implants). The subject device has similar design and identical materials as the devices cleared in K191130, K052504, K052505, and K952708. Subject and predicate devices are all made from the same Nusil silicone elastomers with the same durometer range.

# SUMMARY OF NON-CLINICAL TESTING

No FDA performance standards have been established for the Customized Contour Implant. The following non-clinical tests were performed in order to demonstrate safety based on current industry standards:

- Sterilization validation per ISO 17655-1 and ANSI/AAMI/ISO 20857
- Packaging validation per ISO 11607 ●
- Shelf life validation with accelerated and real time aging studies

{5}------------------------------------------------

## 510(k) Summary

Image /page/5/Picture/1 description: The image shows the text "K200610 Page 3 of 3" at the top. Below that is the logo for "Implantech" in a bold, sans-serif font. Underneath the logo is the tagline "Superior Patient Aesthetics" in a smaller font.

- Cytotoxicity testing per ISO 10993-5
- 3D printer validation

The results of these tests indicate that the Customized Contour Implant is substantially equivalent to the predicate devices.

# CONCLUSION

Based on the testing performed, including sterilization validation, packaging validation, shelf life validation, cytotoxicity testing, and 3D printer validation, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate device. The similar indications for use and technological characteristics for the proposed Customized Contour Implant are assessed to be substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K200610](https://fda.innolitics.com/device/K200610)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
