LumiCare Caries Diagnostic Rinse

K200601 · GreenMark Biomedical, Inc. · LFC · Mar 29, 2021 · Dental

Device Facts

Record IDK200601
Device NameLumiCare Caries Diagnostic Rinse
ApplicantGreenMark Biomedical, Inc.
Product CodeLFC · Dental
Decision DateMar 29, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.1740
Device ClassClass 2
AttributesPediatric, 3rd-Party Reviewed

Indications for Use

LumiCare™ Caries Detection Rinse is indicated to aid the dental professional in visualization of carious lesions.

Device Story

LumiCare is an oral rinse containing fluorescein-tagged submicron starch particles. Used in a dental operatory by a clinician; the patient swishes the rinse, followed by a water rinse and air drying of teeth. The clinician then exposes the teeth to a blue light source (450-470 nm, 450 mW/cm2) and observes the teeth through light orange filter accessories (including protective eyewear). Areas of tooth decay temporarily stain, fluorescing under the blue light. This visual indicator assists the clinician in identifying carious lesions during examination, potentially improving the accuracy of caries detection and guiding clinical decision-making regarding treatment.

Clinical Evidence

No clinical data. Bench testing on extracted teeth compared the device to a histologic reference standard, demonstrating reproducibility and performance comparable to the predicate. Simulated use testing in a clinical environment using a manikin model confirmed ease of use. Biocompatibility testing (cytotoxicity, sensitization, irritation) was performed per ISO 10993-5 and 10993-10.

Technological Characteristics

Liquid oral rinse containing fluorescein-tagged submicron starch particles in a propylene glycol/water carrier. Requires external blue light source (450-470 nm, 450 mW/cm2) and orange filter accessories for visualization. Biocompatible per ISO 10993-5 and 10993-10. Shelf life of 2 years supported by accelerated testing.

Indications for Use

Indicated for dental professionals to aid in the visualization of carious lesions in adults and children.

Regulatory Classification

Identification

The caries detection device is a device intended to show the existence of decay in a patient's tooth by use of electrical current.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. March 29, 2021 GreenMark Biomedical Inc. % Prithul Bom Responsible Third Party Official Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114 Re: K200601 Trade/Device Name: LumiCare Caries Diagnostic Rinse Regulation Number: 21 CFR 872.1740 Regulation Name: Caries detection device Regulatory Class: Class II Product Code: LFC, EAQ Dated: March 24, 2021 Received: March 26, 2021 Dear Prithul Bom: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K200601 Device Name LumiCare^TM Caries Detection Rinse Indications for Use (Describe) To aid the dental professional in visualization of carious lesions Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Expiration Date: 06/30/2023 See PRA Statement below. Form Approved: OMB No. 0910-0120 . {4}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for GreenMark Biomedical Inc. The word "GREENMARK" is written in a combination of green and black letters, with "GREEN" in green and "MARK" in black. A trademark symbol is located to the right of the word "MARK". Below the word "GREENMARK" is the text "BIOMEDICAL INC." in black. To the right of the text is a green leaf design. K200601 ### 510(k) Summary ### I. SUBMITTER GreenMark Biomedical Inc. Mailing Address: 325 E. Grand River, Suite 314 East Lansing, MI 48823 Offices and lab at: 1600 Huron Parkway, Ann Arbor, MI 48109 Phone: 517-896-3665 Contact Person: Steven Bloembergen: info@greenmark.bio Date Prepared: March 24, 2021 II. DEVICE Name of Device: LumiCare™ Caries Detection Rinse Common or Usual Name: Caries detector Classification Name: Caries Detection Device (21 CFR 872.1740) Regulatory Class: II Product Code: LFC Review Panel: Dental ### III. PREDICATE DEVICE Pulpdent SnoopTM Caries Detecting Dye, K964430 ### IV. DEVICE DESCRIPTION The LumiCare™ device is an oral rinse solution containing starch-based particles tagged with fluorescein. A specific volume is swished in the mouth, followed by a water rinse and air drying of teeth prior to exposure of the teeth to a blue light source with an irradiance of 450 mW/cm2 and wavelength of 450-470 nm. The clinician completes the clinical examination by observing fluorescence through light orange filter accessories, including protective eyewear, to aid the visualization of dental caries. Page 1 of 4 {5}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for GreenMark Biomedical Inc. The logo has the word "GREENMARK" in green and black lettering, with the "GREEN" portion in green and the "MARK" portion in black. To the right of the word is a green leaf design. Below the word "GREENMARK" is the phrase "BIOMEDICAL INC." in black lettering. ## V. INDICATIONS FOR USE / INTENDED USE LumiCare™ Caries Detection Rinse is indicated to aid the dental professional in visualization of carious lesions. The Indications for Use statement for the LumiCare™ device is identical to that of the predicate device, with minor exception. This difference does not alter the intended use of the device nor does it affect the safety and effectiveness of the device relative to the predicate. Both the subject and predicate device have the same intended use of detection of dental caries. ## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE | Characteristic | Subject Device | Predicate Device | |--------------------------------------|-----------------------------------------------------------------------------|----------------------------------------------------------------------------| | Device Name | LumiCare™ Caries<br>Detection Rinse | Pulpdent Snoop™<br>(K964430) | | Regulation Number | 872.1740 | 872.1740 | | Identification | Caries Detection Device | Caries Detection Device | | Intended Use | Detection of dental caries | Detection of dental caries | | Indication for disease | Dental caries | Dental caries | | Indication for patient<br>population | Adults and children | Not specified | | Indication for Use<br>statement | To aid the dental<br>professional in<br>visualization of carious<br>lesions | To assist the dental<br>professional in the<br>detection of carious dentin | | Anatomic Sites | Directly visible tooth<br>surfaces | Directly visible tooth<br>surfaces | | Use Environment | Dental Operatory | Dental Operatory | | Prescription/OTC | Prescription only | Prescription only | | Product Form | Liquid | Liquid | | Primary Carrier Fluids | Propylene Glycol/Water | Propylene Glycol/Water | | Contains Staining<br>Agent | Yes | Yes | | Use | Apply product to teeth | Apply product to teeth | | Detection Method | Areas of tooth decay<br>temporarily stain | Areas of tooth decay<br>temporarily stain | | Shelf Life | 2 years | 2 years | | Staining Agent | Fluorescein | Dark blue dye | ## Table 1. Comparison to Predicate Device Page 2 of 4 {6}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for GreenMark Biomedical Inc. The word "GREENMARK" is written in a combination of green and black letters, with "GREEN" in green and "MARK" in black. To the right of the text is a green leaf design. Below the text, "BIOMEDICAL INC." is written in smaller, black letters. | Characteristic | Subject Device | Predicate Device | |----------------------------|-------------------------------------|---------------------------------------------------------------------| | Carrier for staining agent | Submicron Starch Particle | None | | Viewing Conditions | Blue light/orange filter | Ambient light/white light with the naked eye or under magnification | | Stimulation Wavelength | Blue light (450-470 nm, peak value) | Visible light | The subject and the predicate device are both dye-containing liquids which bind to carious lesions on all (occlusal, facial, lingual, proximal) directly visible surfaces of the teeth, providing a visible indicator of the presence of caries. The subject and predicate devices have similar technological elements: liquid formulation, application directly to teeth and the liquid contains carriers, preservatives, and a staining agent. Technological differences between the subject and predicate device include: - 1. Viewing Conditions: The predicate device's stain is visible under visible light whereas the subject device's stain is visible under blue light conditions when viewed through an orange filter. This difference does not impact safety or performance. - 2. Staining agent and carrier: The subject device contains submicron starch particles as a carrier for the staining agent, fluorescein, whereas the predicate contains blue dye with no carrier. This difference does not impact safety or effectiveness. The fluoresceinated starch particle was demonstrated to be non-cytotoxic, non-irritating and non-sensitizing. ## VII. PERFORMANCE DATA - A. Non-Clinical Performance Testing ## Bench Testing Sensitivity and Specificity testing was completed on extracted teeth using histologic reference standard for the LumiCare™ rinse and the predicate, demonstrating a high degree of reproducibility for the LumiCare™ rinse, and performance comparable to the predicate device. No FDA-recognized standards were available to inform the testing. ## Simulated Use Testing Simulated use testing in a clinical environment was performed by clinicians using a manikin model with extracted teeth placed in a typodont, demonstrating ease and effectiveness of use within the clinical workflow. No FDA-recognized standards were available to inform the testing. {7}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for GreenMark Biomedical Inc. The word "GREENMARK" is written in a combination of green and black letters, with "GREEN" in green and "MARK" in black. To the right of the word "MARK" is a green leaf design. Below the word "GREENMARK" is the phrase "BIOMEDICAL INC." in black letters. ## Biocompatibility A biocompatibility risk assessment was performed (per 2016 FDA Biocompatibility Guidance), along with cytotoxicity, sensitization, and irritation testing (per ISO 10993-5 and 10993-10), indicating the LumiCare™ rinse is biocompatible for use as intended. ## Shelf Life Testing Based on accelerated testing, a shelf life of two years is supported. Real time aging is being performed. No FDA-recognized standards were available to inform the testing. - B. Clinical Performance Testing Clinical performance testing was not conducted. ## VIII. CONCLUSION Based upon the indications for use and comparison of technology, together with the results from non-clinical performance testing, we find the LumiCare™ rinse used together with a blue light source, having an irradiance of 450 mW/cm2 and wavelength of 450-470 nm, is substantially equivalent to the predicate device Pulpdent Snoop™.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...