K200597 · Straumann USA, LLC (On Behalf of Institut Straumann Ag) · NHA · Aug 19, 2020 · Dental
Device Facts
Record ID
K200597
Device Name
Straumann BLX Gold Abutments
Applicant
Straumann USA, LLC (On Behalf of Institut Straumann Ag)
Product Code
NHA · Dental
Decision Date
Aug 19, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
Straumann abutments are intended to be placed into Straumann dental implants to provide support for prosthetic reconstructions such as crowns, bridges, and overdentures.
Device Story
Straumann BLX Gold Abutments are dental prosthetic components used to support crowns, bridges, and overdentures. The device consists of a Ceramicor alloy base secured to a Straumann BLX dental implant via a Ti-6Al-7Nb (TAN) basal screw. A Polyoxymethylene (POM) burn-out modeling aid is friction-fitted to the base for laboratory casting procedures. The system utilizes the TorcFit conical connection. Used in dental clinics by clinicians, the abutment serves as the interface between the endosseous implant and the final prosthetic restoration. The clinician selects the appropriate size and shape to meet individual patient anatomical needs. The device is provided non-sterile and requires steam sterilization by the end-user prior to placement. By providing a stable connection to the implant, the abutment enables the functional and aesthetic restoration of missing teeth.
Clinical Evidence
Bench testing only. No clinical data was provided. Performance was validated through dynamic fatigue, static strength, and loosening torque testing conducted per FDA guidance for Class II Special Controls for Root-form Endosseous Dental Implants and Endosseous Dental Abutments. Biocompatibility was established via material equivalence to the primary predicate.
Indicated for patients requiring dental implant-supported prosthetic reconstructions, including crowns, bridges, and overdentures.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in a larger, bolder font, and "ADMINISTRATION" in a smaller font below.
August 19, 2020
Straumann USA, LLC (On behalf of Institut Straumann AG) Chanrasmey White Regulatory Affairs Specialist 60 Minuteman Road Andover, Massachusetts 01810
Re: K200597
Trade/Device Name: Straumann® BLX Gold Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: July 16, 2020 Received: July 20, 2020
Dear Chanrasmey White:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200597
Device Name Straumann® BLX Gold Abutments
### Indications for Use (Describe)
Straumann abutments are intended to be placed into Straumann dental implants to provide support for prosthetic reconstructions such as crowns, bridges, and overdentures.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <input checked="" type="checkbox"/> | Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------|----------------------------------------------|
| <input type="checkbox"/> | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Straumann® BLX Gold Abutments
510(k) Summary
#### 510(k) Summary 5
#### Submitter 5.1
Straumann USA, LLC (on behalf of Institut Straumann AG)
60 Minuteman Road
Andover, MA 01810
| Phone Number: | 978-747-2614 |
|---------------|--------------|
|---------------|--------------|
Fax Number: 978-747-0023
Chanrasmey White Contact Person:
Date of Submission: August 18, 2020
## 5.2 Device
| Trade Name: | Straumann® BLX Gold Abutments |
|----------------------|-------------------------------------|
| Common Name: | Endosseous Dental Implant Abutments |
| Classification Name: | Endosseous Dental Implant Abutments |
| Regulatory Class: | II (21 CFR 872.3630) |
| Product Code: | NHA (21 CFR 872.3630) |
## 5.3 Predicate Device
Primary Predicate:
K070549 - RC Gold Abutments
Reference Predicate:
K072497 - NC Gold Abutment, crowns
K071888 – Gold Abutment for Bridge
K190662 – MRI Compatibility for Existing Straumann Dental Implant System
K173961 - BLX Implant System
{5}------------------------------------------------
## Straumann® BLX Gold Abutments
510(k) Summary
## 5.4 Device Description
The Straumann prosthetic line consist of abutments which are used for the restoration of Straumann dental implants of different types, endosteal diameters, lengths and platforms. They are available in a variety of shapes and sizes to fit individual patient needs.
The Straumann BLX Gold Abutments consist of a Ceramicor® alloy base, which is fixed to the implant by mean of a basal screw made of Ti-6AI-7Nb, Titanium Aluminum Niobium (TAN). The screw channel is protected during the lab procedure with a modeling aid made of Polyoxymethylene (POM) which is friction fit to the Ceramicor® alloy base. The modeling aid is a burn-out sleeve for casting and the combination of base and modeling aid is designed to be used together with Straumann BLX Implants having the TorcFit™, conical connection.
## 5.5 Indications for Use
Straumann abutments are intended to be placed into Straumann dental implants to provide support for prosthetic reconstructions such as crowns, bridges, and overdentures.
#### Technoloqical Characteristics 5.6
The subject device is equivalent to the primary predicate cleared per K070549 and reference devices cleared per K072497 and K071888 with the exception of the implant to abutment connection which is TorcFit™ to fit the implants of the BLX Implant System cleared per K173961. The reference device K190662 is included to address MRI compatibility only.
The technological characteristics of the proposed BLX Gold Abutments are compared to the primary predicate in Table 1.
{6}------------------------------------------------
# Straumann® BLX Gold Abutments
510(k) Summary
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE<br>DEVICE | REFERENCE PREDICATE<br>DEVICE | REFERENCE PREDICATE<br>DEVICE |
|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | K200597 | K070549 | K072497 | K071888 |
| Indications for<br>Use | Straumann abutments are<br>intended to be placed into<br>Straumann dental implants to<br>provide support for prosthetic<br>reconstructions such as crowns,<br>bridges, and overdentures. | Abutments are intended to be<br>placed into dental implants to<br>provide support for prosthetic<br>reconstructions such as<br>crowns, bridges and<br>overdentures. | Abutments are intended to be<br>placed into dental implants to<br>provide support for prosthetic<br>reconstructions such as<br>crowns, bridges and<br>overdentures. | Abutments are intended to be<br>placed into dental implants to<br>provide support for prosthetic<br>reconstructions such as<br>crowns, bridges and<br>overdentures. |
| Material | Ceramicor® | Ceramicor® | Ceramicor® | Ceramicor® |
| Abutment<br>Shape in<br>Coronal<br>Portion | Axial screw access hole for<br>placement of abutment into<br>implant | Axial screw access hole for<br>placement of abutment into<br>implant | Axial screw access hole for<br>placement of abutment into<br>implant | Axial screw access hole for<br>placement of abutment into<br>implant |
| Design of<br>Basal Portion<br>of Abutment | TorcFit™ Connection for BLX<br>Implants<br>(conical connection) | CrossFit® Connection for<br>BL/BLT Implants<br>(conical connection) | CrossFit® Connection for<br>BL/BLT Implants<br>(conical connection) | CrossFit® Connection for<br>BL/BLT Implants<br>(conical connection) |
| Restoration | Crown and Bridge | Crown | Crown | Bridge |
| Diameter | 3.8 mm, 4.5 mm, and 5.5 mm | 4.5 mm | 3.8 mm | 4.7 mm |
| Gingiva<br>Height | 1.5 mm | 3.7, 3.8, 5.2 mm | 3.7 mm | 5.2 mm |
| Total<br>Abutment<br>Height | 4.5 mm, 6.8 mm | 8.2 mm | 6.5 mm | 6.2 mm |
| Prosthesis<br>Type | Cemented or screw retained | Cemented or screw-retained | Cemented or screw-retained | Cemented or screw-retained |
| Base Screw | Ti-6Al-7Nb<br>Basal Screw | Ti-6Al-7Nb<br>Basal Screw | Ti-6Al-7Nb<br>Basal Screw | Ti-6Al-7Nb<br>Basal Screw |
| Maximum<br>Angulation | 30° | 30° | 30° | 30° |
Table 1 – Comparison Matrix
{7}------------------------------------------------
## Straumann® BLX Gold Abutments
510(k) Summary
### 5.7 Performance Data
The following performance data is provided in support of the substantial equivalence determination.
### Biocompatibility Testing
The subject devices have been assessed for biological safety according to ISO 10993-1:2009 "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" and to the FDA Guidance document "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process', Guidance for Industry and Food and Drug Administration Staff, Document issued on: June 16, 2016" for the subject device.
The subject device is made out of the same material as the primary predicate device, therefore, biocompatibility is represented by the primary predicate. The subject devices are equivalent with regard to nature of body contact, contact duration, material formulation and sterilization methods compared to the primary and reference predicate devices and therefore, no new testing has been performed.
### Sterilization
The proposed BLX Gold Abutments are single patient devices and are provided nonsterile. The recommended end user sterilization via Moist Heat Steam sterilization has been validated according to ISO 17665-1 and ISO 17665-2 and to applicable recommendations in the FDA guidance "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Guidance for Industry and Food and Drug Administration Staff, Document issued on: March 17, 2015".
### Bench Testing
Dynamic fatigue, static strength, and loosening torque tests were conducted according to the FDA guidance document "Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments." Testing was performed on worst-case construct. The worst-case construct is based on several device features such as diameter, wall thickness and orientation and intends to represent the performance behavior of all the devices within
{8}------------------------------------------------
## Straumann® BLX Gold Abutments
510(k) Summary
this submission. The worst-case construct was tested and results demonstrated equivalence to the predicate devices.
## 5.8 Conclusion
The documentation submitted in this premarket notification demonstrates the subject devices are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.