The ADAPT® Tissue is indicated for use as a patch in the repair of cardiac defects including intracardiac defects, septal defects, valve and annulus repair.
Device Story
ADAPT® Tissue is a sterile, bovine pericardial patch; prepared via glutaraldehyde-crosslinking using ADAPT® TEP technology. Device is an acellular collagen sheet; moist; light yellow to beige; 4cm² to 84cm² size; 0.25-0.80mm thickness. Intended for surgical implantation by physicians in hospital settings during cardiac procedures. Functions as a structural patch for defect repair; provides tissue reinforcement. Benefits patient by facilitating repair of cardiac/septal defects and valve/annulus structures.
Clinical Evidence
No new clinical data provided. Substantial equivalence is supported by design verification and validation studies previously performed for predicate devices CardioCel and VascuCel. Biocompatibility meets ISO 10993 requirements.
Technological Characteristics
Bovine pericardium; glutaraldehyde-crosslinked using ADAPT® TEP technology. Acellular collagen sheet. Sterilized via propylene oxide. Biocompatible per ISO 10993.
Indications for Use
Indicated for patients requiring repair of cardiac defects, including intracardiac defects, septal defects, and valve and annulus repair.
Regulatory Classification
Identification
An intracardiac patch or pledget made of polypropylene, polyethylene terephthalate, or polytetrafluoroethylene is a fabric device placed in the heart that is used to repair septal defects, for patch grafting, to repair tissue, and to buttress sutures.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
April 3, 2020
Admedus Regen Pty Ltd Diana Upp Sr. Regulatory Affairs Specialist 860 Blue Gentian Road, Suite 340 Eagan, Minnesota 55121
Re: K200566
Trade/Device Name: ADAPT® Tissue Regulation Number: 21 CFR 870.3470 Regulation Name: Intracardiac Patch Or Pledget Made Of Polypropylene, Polyethylene Terephthalate, Or Polytetrafluoroethylene Regulatory Class: Class II Product Code: PSQ Dated: February 21, 2020 Received: March 4, 2020
Dear Diana Upp:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Rachel Neubrander Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200566
Device Name ADAPT® Tissue
Indications for Use (Describe)
The ADAPT® Tissue is indicated for use as a patch in the repair of cardiac defects including intracardiac defects, septal defects, valve and annulus repair.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K200566 510(k) Summary
#### l. Applicant Information:
| Date Prepared: | March 27, 2020 |
|-----------------------------------|-------------------------------------------------|
| Submitter: | Admedus Regen PTY LTD |
| Address: | 26 Harris Road<br>Malaga Western Australia 6080 |
| Establishment<br>Registration No. | 3010805634 |
| Contact Person: | Kiran Bhirangi, MBBS FRCS (I) |
| Telephone Number: | +1 651-900-2151 |
# II. Device Information:
| Trade Name: | ADAPT® Tissue |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Cardiovascular Patch |
| Classification Name: | Intracardiac Patch or Pledget, Biologically Derived |
| Classification: | Class II, 21 CFR § 870.3470 |
| Product Code: | PSQ |
| Predicate Devices: | CardioCel (K130872) and VascuCel (K162579) |
| Device Description: | The ADAPT® Tissue medical device is a bovine pericardial<br>patch prepared from glutaraldehyde-crosslinked bovine<br>pericardium using the ADAPT® TEP technology. It is a<br>sterile, light yellow to beige colored, moist, pre-cut sheet of<br>acellular collagen. The products range in size from 4cm² to<br>84cm² and can range in thickness from 0.25-0.80mm. |
| Intended Use: | The ADAPT® Tissue is indicated for use as a patch in the<br>repair of cardiac defects including intracardiac defects,<br>septal defects, valve and annulus repair. |
| Comparative<br>Analysis: | The proposed and predicate devices are identical in design<br>and manufacturing. The proposed device represents a<br>labeling change only. Both the subject and predicate devices<br>are manufactured from glutaraldehyde fixed bovine<br>pericardium using the ADAPT® TEP technology. |
{4}------------------------------------------------
Table 5-1 below reveals key similarities and is the basis for substantial equivalence of the ADAPT® Tissue to the predicate devices.
| Description | ADAPT® Tissue<br>(Proposed<br>Device) | VascuCel<br>(Predicate<br>Device) | CardioCel<br>(Primary<br>Predicate Device) | Comparison |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulatory<br>Class | II | II | II | Substantially<br>equivalent |
| 510(k) No. | New | K162579 | K130872 | |
| Classification<br>Name | Intracardiac Patch<br>or Pledget | Intracardiac Patch<br>or Pledget | Intracardiac Patch<br>or Pledget | Substantially<br>equivalent |
| CFR Section | 21 CFR 870.3470 | 21 CFR 870.3470 | 21 CFR 870.3470 | |
| Device Name | ADAPT® Tissue | VascuCel | CardioCel | |
| Trade/Common<br>Name | ADAPT® Tissue | VascuCel | CardioCel | |
| Manufacturer | Admedus Regen<br>PTY LTD | Admedus Regen<br>PTY LTD | Admedus Regen<br>PTY LTD | Same |
| Device<br>Description | The ADAPT®<br>Tissue medical<br>device is a bovine<br>pericardial patch<br>prepared from<br>glutaraldehyde-<br>crosslinked bovine<br>pericardium using<br>the ADAPT® TEP<br>technology. It is a<br>sterile, light yellow<br>to beige colored,<br>moist, pre-cut<br>sheet of acellular<br>collagen. | The VascuCel<br>device is a bovine<br>pericardial patch<br>prepared from<br>glutaraldehyde-<br>crosslinked bovine<br>pericardium using<br>the ADAPT® TEP<br>technology. It is a<br>sterile, light yellow<br>to beige colored,<br>moist, pre-cut<br>sheet of acellular<br>collagen. | The CardioCel<br>device is a<br>cardiovascular<br>patch prepared<br>from<br>glutaraldehyde-<br>crosslinked bovine<br>pericardium using<br>the ADAPT® TEP<br>technology. It is a<br>sterile, light yellow<br>or beige colored,<br>moist, pre-cut, flat<br>sheet of acellular<br>collagen. | Substantially<br>equivalent |
| Intended Use /<br>Indications for<br>Use | The ADAPT®<br>Tissue is indicated<br>for use as a patch<br>in the repair of<br>cardiac defects<br>including<br>intracardiac<br>defects, septal<br>defects, valve and<br>annulus repair. | VascuCel is<br>indicated as a<br>patch in great<br>vessel repair,<br>peripheral vascular<br>reconstruction and<br>suture line<br>buttressing. | CardioCel is<br>indicated for use<br>as a patch in<br>pericardial closure<br>and the repair of<br>cardiac and<br>vascular defects<br>including<br>intracardiac<br>defects; septal<br>defects, valve and<br>annulus repair;<br>great vessel<br>reconstruction,<br>peripheral<br>vascular<br>reconstruction and<br>suture line | Substantially<br>equivalent –<br>the indication<br>for use<br>statement of<br>the proposed<br>device is a<br>subset of the<br>predicate<br>device<br>indication for<br>use |
| Table 5-1: Predicate Device(s) Comparison Chart |
|-------------------------------------------------|
|-------------------------------------------------|
{5}------------------------------------------------
| Description | ADAPT® Tissue<br>(Proposed Device) | VascuCel<br>(Predicate Device) | CardioCel<br>(Primary Predicate Device) | Comparison |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Intended<br>Population | Patients who<br>require repair of<br>cardiac defects<br>including<br>intracardiac<br>defects, septal<br>defects, valve and<br>annulus repair. | Patients who<br>require great<br>vessel repairs;<br>vascular<br>reconstruction and<br>suture-line<br>buttressing | Patients with<br>intracardiac and<br>cardiovascular<br>defects requiring<br>repair (pediatric<br>and adult groups) | Substantially<br>equivalent |
| Clinical Setting | In-hospital<br>(bioimplant that is<br>surgically<br>implanted) | In-hospital<br>(bioimplant that is<br>surgically<br>implanted) | In-hospital<br>(bioimplant that is<br>surgically<br>implanted during<br>open heart<br>surgery) | |
| Anatomical<br>Sites | Cardiovascular | Peripheral<br>Vasculature | Cardiovascular | |
| Materials | Bovine Pericardium | Bovine<br>Pericardium | Bovine<br>Pericardium | |
| Design and<br>Scientific<br>Principles | Glutaraldehyde<br>fixed bovine<br>pericardium using<br>ADAPT® TEP<br>technology | Glutaraldehyde<br>fixed bovine<br>pericardium using<br>ADAPT® TEP<br>technology | Glutaraldehyde<br>fixed bovine<br>pericardium using<br>ADAPT® TEP<br>technology | |
| Performance | A long-term implant<br>for the repair of<br>cardiac defects | A long-term<br>implant for the<br>great vessel repair<br>and peripheral<br>vascular<br>reconstruction | A long-term<br>implant for the<br>repair of<br>cardiovascular<br>defects | Substantially<br>equivalent |
| Sterilization<br>Method | Propylene oxide | Propylene oxide | Propylene oxide | |
| Biocompatibility | Biocompatible;<br>meeting the<br>requirements of<br>ISO 10993 | Biocompatible;<br>meeting the<br>requirements of<br>ISO 10993 | Biocompatible;<br>meeting the<br>requirements of<br>ISO 10993 | Substantially<br>equivalent |
## Performance
Evaluation:
The nonclinical testing performed in support of the general intended use for CardioCel also demonstrates substantial equivalence of the specific ADAPT® Tissue indication for the repair of cardiac defects including intracardiac defects, septal defects, valve and annulus repair. The CardioCel and ADAPT Tissue devices are identical in design and manufacturing. No additional testing was conducted to support the labeling change that is the purpose of this submission.
{6}------------------------------------------------
Summary: The results from design verification and design validation studies performed in support of CardioCel and VascuCel have been found to directly support performance of the ADAPT® Tissue. Therefore, the ADAPT® Tissue is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.