StarPore™ implants are intended for the restoration or augmentation of bony contour in craniofacial defects.
Device Story
StarPore™ is a range of preformed, porous high-density polyethylene (HDPE) implants; designed to restore or augment bony contour in craniofacial defects. Unlike the predicate device (PoreStar, K171037), which is patient-specific based on 3D CT scans, StarPore™ is offered as off-the-shelf implants in various sizes and shapes. Surgeons select an appropriate size and cut the implant intra-operatively to achieve an anatomically matched fit. The porous architecture, created by star-shaped particles, mimics trabecular bone. The device is non-conductive, non-magnetic, and MRI-compatible. It is provided sterile via ethylene oxide gas for single-use in a clinical setting.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing and design validation/verification activities previously performed for the PoreStar (now StarPore) product line, including material properties, sterilization, biocompatibility, and shelf-life/stability testing.
Indicated for patients requiring restoration or augmentation of bony contour in craniofacial defects.
Regulatory Classification
Identification
Ear, nose, and throat synthetic polymer material is a device material that is intended to be implanted for use as a space-occupying substance in the reconstructive surgery of the head and neck. The device is used, for example, in augmentation rhinoplasty and in tissue defect closures in the esophagus. The device is shaped and formed by the suregon to conform to the patient's needs. This generic type of device is made of material such as polyamide mesh or foil and porous polyethylene.
Predicate Devices
Anatomics Pty Ltd. PoreStar Patient Specific Implant - K171037
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
June 2, 2020
Anatomics Pty, Ltd. % Christine Scifert Partner MRC Global 9085 East Mineral Circle, Suite 110 Centennial, CO 80112
Re: K200532
Trade/Device Name: StarPore™ Regulation Number: 21 CFR 874.3620 Regulation Name: Ear, Nose, And Throat Synthetic Polymer Material Regulatory Class: Class II Product Code: JOF Dated: March 3, 2020 Received: March 4, 2020
## Dear Christine Scifert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Michael J. Ryan Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200532
Device Name StarPore™
Indications for Use (Describe)
StarPore™ implants are intended for the restoration of bony contour in craniofacial defects.
Type of Use (Select one or both, as applicable)> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary StarPore™ June 1, 2020
| Company: | Anatomics Pty Ltd<br>Warehouse 1, 246 East Boundary Road,<br>Bentleigh East, VIC 3165, Australia<br>Telephone: +0011 61 3 9529 8088<br>Fax: +0011 61 3 9529 8099 |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary Contact: | Christine Scifert<br>Phone: (901) 831-8053 |
| Company Contact: | Robert Thompson<br>VP Product Innovation<br>E-Mail: Robert.thompson@anatomics.com |
| Trade Name: | StarPore™ |
| Common Name: | Polymer, Ent Synthetic, Porous Polyethylene<br>Plate, Cranioplasty, Preformed, Alterable |
| Classification: | Class II |
| Regulation Number: | 21 CFR 874.3620 (Ear, Nose, and Throat Synthetic Polymer Material) |
| Panel: | Ear, Nose, and Throat |
| Product Code: | JOF |
## Predicate Device:
- . Anatomics Pty Ltd. PoreStar Patient Specific Implant - K171037
#### Reference Devices:
- Porex MEDPOR Surg Implant Material: Prefor Cran/Fac Impla - K922489
- Porex MEDPOR Plus Surgical Biomaterial - K012350
- Porex MEDPOR Pterional Surgical Implant - K002586
K922489 and K012350 provide reference devices that have previously been cleared with preformed shapes for craniofacial defects. The geometries of the subject implants are similar to the reference devices previously cleared under K922489 and K002586 for craniofacial defects.
#### Device Description:
StarPore™ (previously cleared as PoreStar™ Patient Specific Implants) is a range of anatomically shaped surgical implants manufactured from porous high-density polyethylene (HDPE). The star ("Star-") shaped particles create an interconnecting porous ("-Pore") architecture resembling trabecular bone. The implants are provided sterile and intended for single-use. StarPore™ is manufactured from medical grade HDPE (an electrically nonconductive and nonmagnetic polymer) and is suitable for use during Magnetic Resonance Imaging (MRI).
The purpose of this Traditional 510(k) submission is to add "off-the-shelt" implant options to the StarPore™ product line. These off-the-shelf implants would provide surgeons the options of various sizes of implants which, unlike the predicate, would not be individually designed from 3D Computed
{4}------------------------------------------------
Tomography (CT) scans provided to Anatomics by the referring surgeon. Instead, the subject implants would be offered in various off-the-shelf sizes and shapes which could then be cut by the prescribing surgeon intra-operatively to create an anatomically matched patient specific design.
StarPore™ is intended for single use only and is provided sterile, using ethylene oxide gas.
#### Indications for Use:
StarPore™ implants are intended for the restoration or augmentation of bony contour in craniofacial defects.
## Technological Characteristics Summary:
The subject device is substantially equivalent to the following predicate:
Anatomics Pty Ltd. PoreStar Patient Specific Implant - K171037 (S.E. 12/27/2017)
The subject and predicate devices are similar in Intended Use, Technological Characteristics, Performance Specifications, and Material. See the table below for details on comparison between the subject and predicate devices:
| Characteristic | StarPore™ (New Device) | PoreStar Patient Specific Implant (K171037) |
|----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | StarPore™ implants are intended for the<br>restoration or augmentation of bony<br>contour in craniofacial defects. | The PoreStar Patient Specific Implant is<br>intended for the augmentation or restoration of<br>bony contour in craniofacial defects. |
| Material | Medical grade, HDPE (high density<br>polyethylene) | Medical grade, HDPE (high density<br>polyethylene) |
| Dimensions | Various shapes consisting of lengths<br>27mm - 80mm, widths 27mm - 110mm,<br>and thicknesses/heights 1mm - 8mm to be<br>further cut by prescribing surgeons<br>intraoperatively to achieve best fit. | Patient specific implants created based on<br>patient CT scans and approved by prescribing<br>physician |
| Packaging | Double steri-pouch and put in shelf box | Double peel pouched and put in a shelf box |
| Sterility | Sterile, by Ethylene Oxide gas | Sterile, by Ethylene Oxide gas |
## Performance Testing Summary:
Previous design validation and verification activities were performed for the PoreStar™ (renamed StarPore™) implants. A new risk analysis assessment for the subject products and design specifications has been performed and show no new worst case has been introduced. Therefore, all previous testing including validation of specifications related to material properties, sterilization, biocompatibility, shelflife, stability, and material properties are applicable to the subject device. Thus, it has been determined that no further performance testing is needed.
#### Substantial Equivalence Conclusions:
In conclusion, the subject StarPore™ device is similar to the predicate device in intended use, technological characteristics, material, and performance. No new questions of safety and effectiveness have been raised. Therefore, the subject device can be considered substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.