The Aer X oxygen concentrator device is used on a prescriptive basis by patients requiring supplemental oxygen. It supplies a high concentration of oxygen and is used with a nasal cannula to channel oxygen from the concentrator to the patient. The Aer X may be used in home, institution, vehicle and various mobile environments.
Device Story
Aer X is a portable oxygen concentrator for patients requiring supplemental oxygen. It uses molecular sieve technology and differential pressure swing adsorption to separate oxygen from ambient air. The device senses inhalation via pressure change (trigger sensitivity 0.13 cm water) and delivers a pulse of oxygen (87-94% purity) through a nasal cannula. It features 5 flow settings (equivalent to 1-5 LPM). Operated by patients in home, institution, or mobile environments. Includes a color LCD, keypad, and magnetic breakaway cannula port. Powered by rechargeable Li-Ion batteries. Output allows patients to maintain oxygen therapy while mobile, improving quality of life.
Clinical Evidence
Bench testing only. Testing included output gas composition (VOCs, particulates, CO/CO2/Ozone), user display/LED functions, oxygen flow rate and concentration, electromagnetic compatibility (IEC 60601-1-2), electrical safety (IEC 60601-1), and functional performance (ISO 80601-2-69).
Indicated for patients requiring supplemental oxygen on a prescriptive basis. No specific age or gender restrictions provided.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
3B Medical Inc. Yasser Estafanous QA/RA Director 203 Avenue A NW, Suite 300 Winter Haven, Florida 33881
Re: K200496
Trade/Device Name: Aer X Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: Class II Product Code: CAW Dated: June 22, 2020 Received: June 24, 2020
Dear Yasser Estafanous:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200496
Device Name Aer X
#### Indications for Use (Describe)
The Aer X oxygen concentrator device is used on a prescriptive basis by patients requiring supplemental oxygen. It supplies a high concentration of oxygen and is used with a nasal cannula to channel oxygen from the concentrator to the patient. The Aer X may be used in home, institution, vehicle and various mobile environments.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------|-------------------------------------------------------------|
| <div> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for 3B Medical. The logo consists of the text "3B Medical" in a blue, sans-serif font. To the left of the text is a stylized graphic of three concentric arcs, also in blue, that appear to be orbiting the "3B" portion of the text. The "TM" symbol is located to the upper right of the "3B".
# 510(k) Summary
| Submitter: | 3B Medical Inc.<br>203 Avenue A NW Suite 300<br>Winter Haven, FL 33881 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Yasser Estafanous<br>Director of QA/RA<br>Telephone: 863-226-6285<br>Fax: 863-226-6284<br>Email: yestafanous@3bproducts.com |
| Date Prepared: | July 17, 2020 |
| Trade Name: | Aer X Portable Oxygen Concentrator |
| Common Name: | Oxygen Concentrator |
| Classification: | Class II |
| Product Code: | CAW, 21 CFR 868.5440 |
| Predicate Device(s): | The subject device is equivalent to the following devices:<br>• VBOX Trooper (K121260) |
| Device Description: | The Aer X oxygen concentrator utilies a molecular sieve and a<br>differential pressure swing adsorption methodology to separate<br>the gases in ambient air. The device takes the room air and<br>concentrates the oxygen portion to produce a pulseof oxygen<br>between 87-94% in purity. When the patient inhales, the device<br>senses the pressure change and is triggered to release the oxygen<br>pulse. In between breaths, the device regenerates an oxygen<br>pulse and waits for the next inhalation breath before dispensing<br>it |
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for 3B Medical. The logo is in blue and features the text "3B Medical". To the left of the text is a circular design with three curved lines that resemble a globe. The logo is simple and modern.
# Comparison to the Predicate Device
# Table 1. Comparison to Predicate Device
| | Aer X Oxygen Concentrator<br>(Subject Device) | VBOX Trooper Oxygen<br>Concentrator<br>(Predicate Device) |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K200496 | K121260 |
| Decision Date | | |
| Manufacturer | 3B Medical Inc. | VBOX |
| Classification | Class II | Class II |
| Product Code | CAW | CAW |
| Regulation | 21 CFR 868.5440 | 21 CFR 868.5440 |
| Indications for Use | The Aer X oxygen concentrator<br>device is used on a prescriptive<br>basis by patients requiring<br>supplemental oxygen. It<br>supplies a high concentration of<br>oxygen and is used with a nasal<br>cannula to channel oxygen from<br>the concentrator to the patient.<br>The Aer X may be used in<br>home, institution, vehicle and<br>various mobile environments. | The Trooper oxygen<br>concentrator device is used on a<br>prescriptive basis by patients<br>requiring supplemental oxygen.<br>It supplies a high concentration<br>of oxygen and is used with a<br>nasal cannula to channel<br>oxygen from the concentrator to<br>the patient. The Trooper may be<br>used in home, institution,<br>vehicle and various mobile<br>environments. |
| Prescription Required | Yes | Yes |
| Patient Interface | Same Standard nasal Cannula | Standard nasal cannula |
| Dimensions (LxWxH) | 7.125"x8.25"x2.75"inches | 6 x 2.5 x 6.25 inches |
| Weight | 4.2 lbs (includes battery) | 3.2 lbs (includes battery) |
| Materials | | |
| Sieve Bed | Synthetic zeolite | Synthetic zeolite |
| Sieve bed cartridge | Adsorbent Cartiradges | Adsorbent Cartiradges |
| Battery | Li-Ion | Li-Ion |
| Performance Specifications | | |
| Method of oxygen<br>concentration | Molecular sieve (mechanical) | Molecular sieve (mechanical) |
| Process by which<br>Oxygen is released | Differential pressure swing<br>adsorption | Differential pressure swing<br>adsorption |
| Flow Rate | 5 settings: 1 to 5 (flow rates<br>equivalent to 1 LPM to 5 LPM) | 5 settings: 1 to 5 (flow rates<br>equivalent to 1 LPM to 5 LPM) |
| Duration of flow | Pulsed | Pulsed |
| Trigger Sensitivity | 0.13 cm water (12.7 Pa) | 0.13 cm water (12.7 Pa) |
| Oxygen concentration<br>(% O2) | 87-94% at all settings | 87-94% at all settings |
| Software/ Hardware | Analog and digital electronics<br>with microprocessor | Analog and digital electronics<br>with microprocessor |
| Rechargeable Battery | Yes | Yes |
| Power Options | Battery, AC | Battery, AC |
| Technological charactartristics Summary | | |
| Produce of high purity<br>O2 | Differential pressure vacuum<br>swing adsorption method | Differential pressure vacuum<br>swing adsorption method |
| Flow | Pulsed | Pulsed |
| Cycles to produce<br>concentrated O2 | 1) Dispense, 2) Evacuate and 3)<br>Pressurize | 1) Dispense, 2) Evacuate and 3)<br>Pressurize |
| Material inside<br>adsorbent cartridge | Synthetic Zeolite | Synthetic Zeolite |
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for 3B Medical. The logo consists of the text "3B Medical" in a blue sans-serif font. To the left of the text is a stylized graphic of three concentric circles, also in blue. The letters "3B" are inside the circles, and the letters "TM" are in superscript to the right of the "B".
Table 1. Comparison to Predicate Device
# A summary of technological Characteristics of the new device in comparison to those of the predicate device.
New Changes to the Aer X as compared to the VBOX TROOPER
1.Exterior Shape- The exterior housing shape was required to change due to the Battery change below and the Keypad/ Display change below. The material for the exterior housing is PolyCarb/ ABS with the same fire-retardant rating as was the original device.
2. Battery Location changed from an exterior position with cord and plug to an integrated separate battery chamber with protected style of plug that does not require user to maneuver the plug into position.
3. LCD display with Keypad was updated to more tactile buttons and a color LCD in place of a segmented display.
4. Magnetic Cannula is a new feature to create a "breakaway" cannula that will protect the user and device in the event of the cannula tubing being "caught" on an object. The shape of the cannula plug is shaped to fit the industry standard cannula profile. The cannula plug was also changed to anodized aluminum to keep with industry standards for that component.
5. new design for the Carry Bag needed to change to fit the new exterior shape, also to have a flap that covers the display keypad.
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for 3B Medical. The logo consists of the text "3B Medical" in a blue sans-serif font. To the left of the text is a circular graphic made of blue lines that appear to be emanating from a central point. The "TM" symbol is located to the upper right of the "3B" text.
## Technology not changed:
- . Battery Power: 18650 Li Ion cells in a battery Pack that is user changeable and user chargeable, as before.
- . O2 Production-Separation uses the same Vacuum Pressure swing, purge and repressurizing algorithm employing LILSX adsorbent with high nitrogen capacity. Valving and timing to provide purge and blow down technology is also employed, as before.
- O2 Product Flow Rate: Oxygen is delivered by the Aer X device via a nasal cannula the same standard 2 prong cannula, the Oxygen is delivered in pulse mode a bolus of oxygen at the beginning of each inhalation. The purity of the oxygen is 90% +4% -3% (87-94%) which is also an industry standard for portable oxygen concentrators. The oxygen flow rate is in 5 settings 1 to 5 to be equivalent to 1 LPM to 5 LPM.
To verify that the modified device design met it's functional and Functional and performance requirements, representative samples of the device Safety Testing: underwent biocompatibility, electrical, and mechanical testing in accordance with applicable industry standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, IEC 60601-1-11, and ISO 80601-2-69.
## Non-clinical testing(Bench testing)
Bench testing was performed to provide assurance that the proposed device confirm to the requirements for its intended use. This include the following
- . Output gas composition (e.g. VOCs, particulate matter, ozone/carbon monoxide/carbon dioxide content).
- User display and LED functions
- Oxygen flow rate and concentration
- Electromagnetic compatibility and electrical safety
- Functional performance ●
- Output gas temperature.
{8}------------------------------------------------
K200496
Image /page/7/Picture/1 description: The image shows the logo for 3B Medical. The logo consists of the text "3B Medical" in blue font. To the left of the text is a circular graphic made of curved blue lines. The "TM" symbol is located to the upper right of the "3B" text.
## Summary of results :
As outlined above, the principles of operation, technology, materials and indications for use of the subject device are equivalent to those of the predicate device. The differences in size, weight, LCD & Keypad, exterior housing, carry bag, and the cannula port do not introduce any different questions of safety and effectiveness over the predicate device. The information demonstrates that the subject device is substantially equivalent to the predicate device.
### 3B Medical Inc. considers the Aer X to be substantially Conclusion: equivalent to the predicate device listed above. This conclusion is based upon the devices' similarities in principles of operation, technology, materials and indications for use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.