RayCare 3.1

K200487 · RaySearch Laboratories AB (PUBL) · MUJ · Jun 10, 2020 · Radiology

Device Facts

Record IDK200487
Device NameRayCare 3.1
ApplicantRaySearch Laboratories AB (PUBL)
Product CodeMUJ · Radiology
Decision DateJun 10, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

RayCare is an oncology information system used to support workflows and scheduling, as well as clinical information, planning and treatment management for oncology care. RayCare may be used in the transfer, storage, conversion and display of patient data in all areas of cancer care including radiation therapy, medical oncology and surgical oncology. RayCare is not intended for use in diagnostic activities.

Device Story

RayCare 3.1 is an oncology information system (OIS) designed to manage clinical and administrative workflows in cancer care. The software-only system consists of a client-server architecture that processes, stores, and displays patient data. It integrates modules for patient clinical data, care administration, clinical notes, workflow/task management, charge capture, scheduling, PACS, image viewing, simulation/imaging, and system administration. Used by healthcare professionals in clinical settings, the system facilitates information flow, treatment planning, and treatment delivery management. It supports electronic patient charts, image review, treatment plan comparison, dose coverage analysis, and treatment setup review. By centralizing patient data and automating administrative and clinical tasks, the device assists providers in coordinating oncology care, potentially improving workflow efficiency and treatment management.

Clinical Evidence

No clinical data. Verification and validation were performed via non-clinical bench testing based on standard oncology information system functionalities and compliance with ISO 14971, IEC 61217, IEC 62304, and IEC 62366 standards.

Technological Characteristics

Software-only oncology information system with client-server architecture. Features modular functionality for data management, scheduling, and image viewing. Connectivity includes hospital system integrations and PACS. Complies with IEC 62304 (software lifecycle), ISO 14971 (risk management), and IEC 62366 (usability).

Indications for Use

Indicated for oncology patients requiring management of clinical information, workflows, scheduling, and treatment planning across radiation, medical, and surgical oncology departments.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. RaySearch Laboratories AB (publ) % Ms. Viktoria Hammarstedt Quality and Regulatory Affairs Specialist Sveavägen 44 Stockholm, 11134 SWEDEN June 10, 2020 ### Re: K200487 Trade/Device Name: RayCare 3.1 Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: MUJ Dated: May 11, 2020 Received: May 13, 2020 Dear Ms. Hammarstedt: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. For Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K200487 Device Name RayCare 3.1 Indications for Use (Describe) RayCare is an oncology information system used to support workflows and scheduling, as well as clinical information, planning and treatment management for oncology care. RayCare may be used in the transfer, storage, conversion and display of patient data in all areas of cancer care including radiation therapy, medical oncology and surgical oncology. RayCare is not intended for use in diagnostic activities. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------| | <span style="font-family: Arial, sans-serif;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-family: Arial, sans-serif;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</span> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 5. 510(k) Summary #### 5.1 510(k) owner RaySearch Laboratories AB (publ) Sveavägen 44 111 34 Stockholm Sweden Tel: +46 8 510 530 00 #### 5.2 Contact person Viktoria Hammarstedt Quality and Regulatory Affairs Specialist RaySearch Laboratories AB (publ) Email: quality@raysearchlabs.com +46 8 51053003 Tel: #### 5.3 Preparation date February 19, 2020 #### 5.4 Trade name The trade name is RayCare. The trade name and version number are written together, i.e. "RayCare 3.1". The marketing name is RayCare 3B. #### 5.5 Common name Radiation therapy treatment planning system #### 5.6 Classification name Medical charged-particle radiation therapy system (21 CFR 892.5050, Product Code MUJ) #### 5.7 Predicate device K191384, RayCare 2.3 Oncology Information System Product code: MUJ, Regulation number: 892.5050 #### 5.8 Device description RayCare is an oncology information system that supports healthcare professionals in managing cancer care. The system is structured in functionality modules that are listed below. RayCare is a software-only system with a client part that allows the user to interact with the system and a server part that performs the necessary processing and storage functions. Selected aspects of RayCare are configurable, such as adapting workflow templates to the specific needs of the clinic. - Patient clinical data ● - . Care administration - Clinical notes and documents - Workflow and task management . - Charge capture ● - Scheduling and calendar functions - . PACS - Image viewer - Simulation and imaging ● - System administration . - . Authentication, authorization and audit logging ## K200487 {4}------------------------------------------------ - . Hospital system integrations #### 5.9 Technological characteristics summary The technological characteristics are the same for RayCare 3.1 as for the predicate device. Comparing RayCare 3.1 with RayCare 2.3, both devices are oncology information systems and are used to manage clinical and administrative workflows for treatment planning and delivery. They both support information flow among healthcare facility personnel and can be used whenever radiotherapy is prescribed. Both devices include management of electronic patient charts, image reviewing and archiving, comparisons of treatment plans and dose coverage and treatment setup review and recording. #### 5.10 Assessment of non-clinical performance data The verification and validation of RayCare 3.1 is based on the intended use and functionalities included in a standard oncology information system, and on the standards ISO 14971, IEC 61217, IEC 62304 and IEC 62366. The successful verification and validation of RayCare 3.1 therefore support the substantial equivalence to the RayCare 2.3 Oncology Information System predicate device. #### 5.11 Test conclusion The summary of the performed non-clinical tests shows RayCare 3.1 is as safe and effective and performs as well as the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...