Viveve System

K200472 · Viveve Medical, Inc. · GEI · Nov 18, 2020 · General, Plastic Surgery

Device Facts

Record IDK200472
Device NameViveve System
ApplicantViveve Medical, Inc.
Product CodeGEI · General, Plastic Surgery
Decision DateNov 18, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Viveve System is indicated for use in general surgical procedures for electrocoagulation and hemostasis.

Device Story

The Viveve System is an electrosurgical device used for electrocoagulation and hemostasis. It utilizes monopolar radiofrequency (RF) energy to heat tissue beneath the surface while simultaneously delivering cryogen to the treatment tip to cool the surface tissue (reverse thermal gradient). The system consists of an RF generator with an integrated cooling module, a handpiece, a footswitch, and sterile disposable treatment tips (5cm or 8cm). The device is operated by a clinician in a clinical setting. The clinician uses the handpiece to apply RF energy to the target tissue. The system is designed to provide controlled heating for surgical coagulation. The device is non-invasive, with the treatment tip applied to the surface of the skin.

Clinical Evidence

No clinical data provided. Substantial equivalence is supported by bench testing, including electrical safety, electromagnetic compatibility (IEC 60601), software verification/validation, packaging, and shelf-life studies.

Technological Characteristics

Monopolar RF electrosurgical system. Energy: 6 MHz continuous sinusoidal waveform, 240W max power. Impedance range: 25–120 Ω. Cooling: Cryogen-based surface cooling. Components: RF generator, handpiece, footswitch, sterile disposable tips. Sterilization: ETO. Connectivity: Standalone. Software: Embedded firmware in console, display module, and handpiece.

Indications for Use

Indicated for use in general surgical procedures for electrocoagulation and hemostasis. No specific patient population, age, or gender restrictions are defined.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. November 18, 2020 Viveve Medical, Inc. Kevin Robison Regulatory Affairs Specialist 345 Inverness Drive South, Building B. Suite 250 Englewood, Colorado 80112 Re: K200472 Trade/Device Name: Viveve System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: October 22, 2020 Received: October 23, 2020 Dear Kevin Robison: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K200472 Device Name Viveve System Indications for Use (Describe) The Viveve System is indicated for use in general surgical procedures for electrocoagulation and hemostasis. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/3/Picture/2 description: The image shows the word "VIVEVE" in a sans-serif font. The letters are a light green color. The letters are all lowercase except for the first letter, which is uppercase. There is a registered trademark symbol after the last letter. # 510(k) SUMMARY #### 1.1 REGULATORY AUTHORITY Safe Medical Devices Act of 1990, 21 CFR 807.92 ### 1.2 APPLICANT INFORMATION | Applicant: | Viveve® Inc. | |------------|---------------------------| | | 345 Inverness Drive South | | | Building B, Suite 250 | | | Englewood, CO 80112 | | | | | Contact: | Kevin Robison<br>krobison@viveve.com<br>C: 317-435-8898<br>F: 720-696-8199 | |----------|----------------------------------------------------------------------------| |----------|----------------------------------------------------------------------------| - Date Prepared: November 18, 2020 #### 1.3 SUBJECT DEVICE INFORMATION | Trade Name: | Viveve® System | |------------------------|-------------------------------------------------------------------------------------| | Common Name: | Electrosurgical System | | Product Code: | GEI | | Classification Name: | Electrosurgical Cutting and Coagulation Device and Accessories (21<br>CFR 878.4400) | | Device Panel: | General Surgery/Restorative Device | | Device Classification: | Class II | #### 1.4 PREDICATE DEVICES Primary Predicate Device: Viveve® System (K180584) Reference Device: Viveve® 2.0 System (K190422) #### 1.5 DEVICE DESCRIPTION The Viveve System utilizes monopolar radiofrequency (RF) energy to selectively heat a given volume of tissue beneath the surface, while cryogen is delivered to the inside of the treatment tip to cool the surface tissue. The generator delivers energy to the treatment tip to create an electric field under the treatment tip (electrode). The mechanism of action is the application of RF energy to the tissue resulting in coagulation and/or hemostasis. The Viveve® System consists of four (4) primary components: - . An RF Generator to provide the heating energy. The Generator incorporates the Cooling Module to supply coolant which provides the cooling energy. - . A hand piece that couples the cooling and heating energy to the tissue through the treatment tip. - A footswitch that allows the user to turn the RF Energy on or off. - 5cm or 8cm Sterile Disposable Treatment Tips. {5}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the word "viveve" in a sans-serif font. The color of the text is a light green. The word is all lowercase, and there is a small circle with an R in it to the right of the last "e". The Viveve treatment tips contained in this package are a sterile single use component of the Viveve® System and designed to deliver radiofrequency (RF) energy from the Viveve console. The Viveve treatment tips can only be used in conjunction with the dedicated Viveve® System. Accessories include: - Coupling Fluid ● - Cryogen ● - Return Cable - Return Pad ● - . Power Cord ## 1.6 INDICATIONS FOR USE The Viveve System is indicated for use in general surgical procedures for electrocoagulation and hemostasis. ## 1.7 COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The technological characteristics of the subject device Viveve System are identical to those of the predicate devices, Viveve System (K180584) and Viveve 2.0 System (K190422). The Viveve System is an electrosurgical device that delivers radiofrequency (RF) energy to selectively heat a given area of tissue, while cryogen is delivered to the inside of the treatment tip to cool the surface tissue at the end of energy deposition. The application of RF energy causes the tissue to coagulate and/or become hemostatic. This submission application confirms the continued conformance to applicable technical design specifications and performance requirements, including requirements associated with industry safety and performance standards. ## 1.8 BASIS FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE The Viveve System is substantially equivalent to the predicate devices listed in K180584 and K190422. The principle of operation between the predicate devices and the subject device remain identical as do all output parameters to tissue. ## 1.8.1 DESIGN MODIFICATIONS ## 1.8.1.1 RF Console (Generator) - No changes have been made to the Console. . ## 1.8.1.2 Footswitch - No changes have been made to the Footswitch. ● ## 1.8.1.3 Handpiece - . No changes have been made to the Handpiece. ## 1.8.1.4 Cryogen - . No changes have been made to the Cryogen. {6}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the text "Traditional 510(k) Premarket Notification: Viveve System". The text is centered on the page and is written in a simple, sans-serif font. The image is a simple document with text. ## 1.8.1.5 Treatment Tip viveve - The packaging of the sterile Treatment Tip has changed to align with ● the packaging of the secondary predicate Viveve 2.0 System (K190422) packaging. ## 1.8.2 SOFTWARE MODIFICATIONS ## 1.8.2.1 Viveve RF Console Software - No changes have been made to the RF Console Software. . ## 1.8.2.2 Viveve RF Display Module Software - No changes have been made to the RF Display Module Software. ● ## 1.8.2.3 Viveve RF Handpiece Software - The handpiece software has been changed since the previous clearance ● (K180584). The handpiece software has been slightly modified to extend the useful life of the handpiece from 11,000 radiofrequency emitting pulses to a minimum of 65,000 radiofrequency pulses. The change in pulse number was validated and assessed for additional risk to the patient and was deemed non-significant. The change was made to allow the customer to use the handpiece for more treatments improving customer satisfaction and lowering operational costs. Further detail on this change can be found in the Viveve Letter to File dated December 5th, 2019 and is available upon request. ## 1.8.3 HARDWARE MODIFICATIONS ## 1.8.3.1 Operating System - No changes have been made to the Operating System. ● ## 1.8.4 LABELING MODIFICATIONS ## 1.8.4.1 Technical User Manual - No changes have been made to the TUM for the Viveve System. ● ## 1.8.4.2 Instructions for Use - . No changes have been made to the IFU for the Viveve System. ## 1.8.5 TECHNICAL/ENVIRONMENTAL SPECIFICATION MODIFICATIONS ## 1.8.5.1 Environmental and Packaging Specifications - IEC60601, Electrostatic Discharge (ESD) and Voltage Dip are aligned ● with CMO's Quality Management System (QMS) Requirements ## 1.8.5.2 RF Frequency - No changes have been made to the RF Frequency output. ● ## 1.8.5.3 Operation temperature No changes have been made to the Operation Temperature. {7}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the word "viveve" in a sans-serif font. The word is in a light green color. A small circle with an R inside is located to the bottom right of the word. #### 1.8.5.4 Storage pressure - . No changes have been made to the Storage Pressure. ## 1.8.6 CONTRACT MANUFACTURER AND CONTRACT STERILIZER MODIFICATIONS - Cirtec Medical will now be the sole contract manufacturer of the Viveve . System sterile Treatment Tip replacing Stellartech Research Corporation (SRC) upon clearance of this submission. - . STERIS will be the Contract Sterilizer for the Viveve System sterile Treatment Tips upon clearance of this submission. All previously outlined modifications to the Viveve System are discussed in further detailed in Section 12: Substantial Equivalence Discussion of this Premarket Notification. A comparison of the technical characteristics of Viveve System are compared to those of the predicate device, Viveve System, in Table 1-1 below. | Item | Viveve System<br>(Subject Device) | Viveve System<br>(Primary Predicate Device<br>K180584) | |---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Number | Subject device | K180584 | | Legal Manufacturer | Viveve, Inc. | Viveve, Inc. | | Contract Manufacturer | Stellartech Medical Systems (Generator and Handpiece)Cirtec Medical (Treatment Tips) | Stellartech Research Corporation (All 3 system components) | | Indication for Use | The Viveve System is indicated for use in general surgical procedures for electrocoagulation and hemostasis. | The Viveve System is indicated for use in general surgical procedures for electrocoagulation and hemostasis. | | FDA Classification | Class II | Class II | | CFR/Product Code | 21 CFR 878.4400/GEI | 21 CFR 878.4400/GEI | | Invasiveness of Treatment | Non-invasive. Device applies to the surface. | Non-invasive. Device applies to the surface. | | Principles of Operation | Radiofrequency (RF) energy selectively heats a given volume of tissue beneath the surface, while cryogen is delivered to the inside of the Treatment Tip to cool the surface tissue. | Radiofrequency (RF) energy selectively heats a given volume of tissue beneath the surface, while cryogen is delivered to the inside of the Treatment Tip to cool the surface tissue. The Treatment Tip is placed on the | ## Table 1-1: Comparison of Technological Characteristics of Viveve System and Cleared Primary Predicate Viveve System (K180584) {8}------------------------------------------------ Image /page/7/Picture/1 description: The image shows the word "VIVEVE" in a sans-serif font. The letters are a light green color. A small registered trademark symbol is located to the right of the last "E" in the word. | Item | Viveve System<br>(Subject Device) | Viveve System<br>(Primary Predicate Device<br>K180584) | |----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------| | | The Treatment Tip is placed<br>on the surface of the skin<br>and the internal tissues are<br>heated while the surface<br>tissue is protected. (Reverse<br>thermal gradient) | surface of the skin and the<br>internal tissues are heated while<br>the surface tissue is protected.<br>(Reverse<br>thermal gradient) | | Energy | RF | RF | | Treatment Type | Monopolar | Monopolar | | Main Input<br>(Input<br>voltage/Current/Frequency) | 100 - 120 Vac / 10A / 50/60 Hz<br>220 - 240 Vac / 5A / 50/60 Hz | 100 - 120 Vac / 10A / 50/60 Hz<br>220 - 240 Vac / 5A / 50/60 Hz | | Maximum Power<br>(generator) | 240 Watts | 240 Watts | | Operating Frequency | 6 MHz ±2%. | 6 MHz ±2%. | | Voltage Waveform | 6.0 MHz continuous<br>sinusoidal waveforms | 6.0 MHz continuous sinusoidal<br>waveforms | | Electrode Probe | Monopolar | Monopolar | | Impedance Range | 25 – 120 Ω | 25 - 120 Ω | | Tip | 5cm and 8cm Treatment<br>Tips | 5cm and 8cm Treatment Tips | | Packaging | Tyvek pouch | Tyvek pouch | | Sterility | ETO | ETO | | Contract Sterilizer | STERIS | Sterigenics | | Cooling Solution | Cryogen | Cryogen | #### BIOCOMPATIBILITY 1.9 The biocompatibility endpoints for limited contact duration were conducted on the Viveve System sterile Treatment Tip. ## 1.10 PERFORMANCE DATA Design verification testing, including bench performance, electrical safety/electromagnetic compatibility, software verification/validation, packaging and shelf-life studies, provided in the subject premarket notification demonstrate that the Viveve System is substantially equivalent to the predicate device, Viveve System. ## 1.10.1 Performance Data Design development and control is conducted in phases as described in the Cirtec Design and Development Plan PLN-200054 and VIDA-L Quality Plan QP-200027. The Core Team is responsible for executing and adhering to this VIDA-L Design and Development Plan, and for compliance to Cirtec's {9}------------------------------------------------ Image /page/8/Picture/1 description: The image shows the word "viveve" in a sans-serif font. The color of the text is a light green. A small circle with an R inside is located to the right of the last "e" in the word. Quality Management System, unless otherwise stated in the Plan. The Core Team consists of representatives from Cirtec and Viveve. The following document describes the Design Control of the Viveve System Treatment Tips by Cirtec Medical. | Document Title | Document Number | |--------------------------|--------------------| | Design Control Procedure | Cirtec QSP-C110004 | ### 1.11 CONCLUSION The design, technical characteristics, functionality, indications for use, and principle operation of the subject device Viveve System remains unchanged from that of the predicate device, Viveve System (K180584) and reference devicefor K190422. The proposed treatment tip manufacturer and sterilization facility modifications do not raise new questions of the safety or efficacy of the device and the intended use of the Viveve System remains unchanged from that of the cleared predicates.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...