K200439 · Velano Vascular · FPA · Feb 23, 2021 · General Hospital
Device Facts
Record ID
K200439
Device Name
Velano ExT Extension Set
Applicant
Velano Vascular
Product Code
FPA · General Hospital
Decision Date
Feb 23, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
For administration of intravenous fluids to a patient's vascular system utilizing needle-free components and provide term intravascular or subcutaneous access using a variety of infusates (i.e. anesthesia drugs, chemotherapeutics, drugs, antibiotics, blood products, commonly used prep solutions). The device may also be used with low pressure power injectors rated for a maximum setting of 325 psi. The ExT™ Extension Set is compatible with Velano Vascular PIVOTM devices.
Device Story
Sterile, single-use, non-pyrogenic intravenous administration extension set; 5-inch PVC tubing with T-connection port, needle-free connector, and female luer. Includes integrated stabilizer for anchoring to patient. Used in clinical settings for fluid administration, infusion, or aspiration; compatible with power injectors (max 325 psi, 10 ml/s). Operated by healthcare professionals. Output is fluid delivery to patient vascular system. Benefits include secure, needle-free access for various infusates and power injection capability.
Clinical Evidence
Bench testing only. No clinical data. Testing included visual inspection, simulated shipping, priming volume, microbial ingress, particulate matter, backpressure leak, flow rate, spin collar geometry, simulated use, tubing bond strength, multiple/continuous engagement, activation force, and kink resistance.
Technological Characteristics
PVC tubing; polycarbonate luer; polycarbonate/silicone needle-free connector. 5-inch length. Ethylene oxide sterilization (ISO 11135:2014). Biocompatibility per ISO 10993-1:2018. Particulate testing per USP <788>. Packaging per ASTM F1980:2016. Shipping per ASTM D4169:2016. Manual operation; no software or energy source.
Indications for Use
Indicated for single use in patients requiring intravenous fluid administration, intermittent or continuous infusion, direct injection, or aspiration. Compatible with power injector procedures up to 325 psi at 10 ml/sec flow rate.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
Velano Vascular Q2® Low Power Injector Extension Set (K182897)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Velano Vascular Tiffini Wittwer Consulting Director Regulatory Affairs 221 Pine St #200 San Francisco, California 94104
Re: K200439
Trade/Device Name: Velano ExT™ Extension Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA Dated: February 9, 2021 Received: February 22, 2021
Dear Tiffini Wittwer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200439
Device Name Velano ExTTM Extension Set
#### Indications for Use (Describe)
The Velano ExT™ Extension Set with needle-free connector is for single use only. Each port of the Velano ExT™ Extension Set may be used for direction, intermittent infusion or aspiration. The pressure rated port of the Velano ExT™ Extension Set may be used with power injector procedures to a maximum pressure of 325 psi at a flow rate of 10 ml per second.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|-----------------------------------------------------------------------------------|
| <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Submitter: | Velano Vascular |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Tiffini Wittwer<br>Regulatory Affairs Consultant<br>Phone: (707) 799-6732<br>Email: twittwer@mededge.io |
| Trade Name: | Velano ExT™ Extension Set |
| Common Name: | Intravascular Administration Set |
| Classification: | Class II |
| Product Code: | FPA |
| Regulation: | 880.5440 |
| Predicate Device(s): | The subject device is equivalent to the following devices:<br>• Velano Vascular Q2® Low Power Injector Extension Set (K182897) |
| Device Description: | The Velano ExT™ Extension Set is a sterile, single use, non-pyrogenic intravenous administration extension set comprised of an approximate 5" inch tube bonded to a T-connection port on one end and a female luer on the other end with a clamp in between. The T-connection port has a swabable, needle-free port on one side and a locking male luer on the other side. The device contains and integrated stabilizer under the swabable needle-free port. This component (along with IV tape / adhesive) is intended to help anchor the extension set to the patient. This device is not made with the plasticizer Diethylhexylphthalate (DEHP). The Velano ExT™ Extension Set is compatible with Velano Vascular PIVOT™ devices. |
| Indication for Use: | The Velano ExT™ Extension Set with needle-free connector is for single use only. Each port of the Velano ExT™ Extension Set may be used for direct injection, intermittent infusion, continuous infusion or aspiration. The pressure rated port of the Velano ExT™ Extension Set may be used with power injector procedures to a maximum pressure of 325 psi at a flow rate of 10ml per second. |
# February 23, 2021
| | Subject Device: Velano<br>ExT TM Extension Set<br>(K20439) | Velano Vascular Q2®<br>Low Pressure Power<br>Injector Extension Set<br>(K182897) | Comment |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Velano Vascular | Velano Vascular | |
| Product Code | FPA | Same | |
| Regulation | 880.5440 | Same | |
| Classification | II | Same | |
| Subject Device: Velano | Velano Vascular Q2® | Comment | |
| ExT™ Extension Set | Low Pressure Power | | |
| (K20439) | Injector Extension Set | | |
| | (K182897) | | |
| Indications for<br>Use | The Velano ExT™<br>Extension Set with needle-<br>free connector is for single<br>use only. Each port of the<br>Velano ExT™ Extension Set<br>may be used for direct<br>injection, intermittent<br>infusion, continuous infusion<br>or aspiration. The pressure<br>rated port of the Velano<br>ExT™ Extension Set may be<br>used with power injector<br>procedures to a maximum<br>pressure of 325 psi at a flow<br>rate of 10ml per second. | Pressure Rated: The Q2®<br>Low Pressure Power Injector<br>Extension Set with needleless<br>connector is for single use<br>only. The extension set may<br>be used for direct injection,<br>intermittent infusion,<br>continuous infusion or<br>aspiration. This set may be<br>used with power injector<br>procedures to a maximum<br>pressure of 325 psi at a flow<br>rate of 10ml per second.<br>Non Pressure Rated: The<br>Q2® Low Pressure Power<br>Injector Extension Set with<br>needleless connector is for<br>single use only. The<br>extension set may be used for<br>direct injection, intermittent<br>infusion, continuous infusion | Wording changes for<br>clarity and does not<br>change the indication<br>for use. |
| | | or aspiration. | |
| Intended Use | For administration of<br>intravenous fluids to a<br>patient's vascular system<br>utilizing needle-free<br>components and provide<br>term intravascular or<br>subcutaneous access using<br>a variety of infusates (i.e.<br>anesthesia drugs,<br>chemotherapeutics, drugs,<br>antibiotics, blood products,<br>commonly used prep<br>solutions). The device may<br>also be used with low pressure<br>power injectors rated for a<br>maximum setting of 325 psi.<br>The ExT™ Extension Set is<br>compatible with Velano<br>Vascular PIVOTM devices. | For administration of<br>intravenous fluids to a<br>patient's vascular system<br>utilizing needle-free<br>components and provide<br>term intravascular or<br>subcutaneous access using<br>a variety of infusates (i.e.<br>anesthesia drugs,<br>chemotherapeutics, drugs,<br>antibiotics, blood products,<br>commonly used prep<br>solutions). The device may<br>also be used with low pressure<br>power injectors rated for a<br>maximum setting of 325 psi.<br>The Q2® Extension Set is<br>compatible with Velano<br>Vascular PIVOTM devices. | Same |
| | Subject Device: Velano<br>ExTTM Extension Set<br>(K20439) | Velano Vascular Q2®<br>Low Pressure Power<br>Injector Extension Set<br>(K182897) | Comment |
| Components | Tubing, Luer, Needle-less<br>Connector, Male spin lock<br>connector<br>Non-fluid contacting:<br>Stabilizer, pinch clamp,<br>vented female luer lock cap | Tubing, Luer, Needle-less<br>Connector, Male spin lock<br>connector<br>Non-fluid contacting: Pinch<br>clamp, vented female luer<br>lock cap | Flow rate, spin collar<br>height and spin collar<br>angle demonstrate that<br>the difference does not<br>alter the safety or<br>effectiveness of the<br>device |
| Dimensions | Overall Length: 5.0 inch<br>(12.7cm) | Overall Length: 5.75 inch<br>(14.6cm) | Flow rate, torque, and<br>kink resistance testing<br>demonstrate that the<br>difference does not<br>alter the safety or<br>effectiveness of the<br>device |
| Tubing | PVC | PVC | Same |
| Luer | Polycarbonate | Co-polyester | Biocompatibility<br>testing demonstrate that<br>the difference does not<br>alter safety or<br>effectiveness of the<br>device |
| Needle-free<br>Connector | NP Medical<br>Proprietary Design<br>Polycarbonate / silicone | Quest SwabSite Valve –<br>Proprietary Design<br>Polycarbonate / silicone | Microbial ingress, flow<br>rate, pressure, and<br>multiple<br>activation testing<br>demonstrate that the<br>difference does not<br>alter safety or<br>effectiveness of the<br>device |
| Stabilizing<br>component | Yes | No | Flow rate, spin collar<br>height and spin collar<br>angle demonstrate that<br>the difference does not<br>alter the safety or<br>effectiveness of the<br>device |
| Principle of<br>Operation | Swabable luer activation;<br>direct injection, intermittent<br>infusion, continuous<br>infusion, aspiration | Swabable luer activation;<br>direct injection, intermittent<br>infusion, continuous<br>infusion, aspiration | Same |
| | Subject Device: Velano<br>ExTTM Extension Set<br>(K20439) | Velano Vascular Q2®<br>Low Pressure Power<br>Injector Extension Set<br>(K182897) | Comment |
| Use | Use with low pressure power<br>injectors up to 325 psi and<br>maximum flow rate of 10<br>mL/second. | Use with low pressure power<br>injectors up to 325 psi and<br>maximum flow rate of 10<br>mL/second. | Same |
| Priming<br>volume | < 1 mL | < 1 mL | Same |
| Energy Source | User Operated | User Operated | Same |
| Disposable or<br>Reusable | Disposable | Disposable | Same |
| Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide | Same |
| Minimum SAL | 1 x 10-6 | 1 x 10-6 | Same |
| Packaging | Reinforced paper / nylon | Reinforced paper / nylon | Same |
{4}------------------------------------------------
{5}------------------------------------------------
{6}------------------------------------------------
| Functional and<br>Safety Testing: | To verify that the device design meets its functional and<br>performance requirements, representative samples of thedevice underwent mechanical testing. As a result of<br>verification and validation activities and risk assessment,<br>testing ensured the device design meets its functional and<br>performance requirements. The following tests were<br>performed: |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | • Visual inspection |
| | • Simulated shipping |
| | • Priming volume |
| | • Microbial ingress |
| | • Particulate |
| | • Backpressure leak under normal use and power injection |
| | • Flow rate for normal use and power injection |
| | • Spin collar height and spin collar angle |
| | • Simulated use |
| | • Tubing bond strength |
{7}------------------------------------------------
| • | Multiple Engagement |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| • | Continuous Engagement |
| • | Activation Force |
| • | Tubing kink resistance |
| ISO 11135:2014, Sterilization of health-care products -Ethylene<br>oxide - Requirements for the development, validation and routine<br>control of a sterilization process for medical devices | |
| ISO 10993-1, 2018 Biological Evaluation of Medical Devices | |
| USP <788> Particulate Matter in Injections | |
| ASTM D4169:2016 Standard practice for performance testing of<br>shipping containers and systems | |
| ASTM F1980:2016 Standard guide for accelerated aging of sterile<br>medical device packages. | |
| AAMI TIR28: 2016 Product Adoption and Process Equivalence for<br>Ethylene Oxide Sterilization | |
| Conclusion: | Velano Vascular considers the Velano ExT™ Extension Set device<br>to be equivalent to the predicate devices, K182897, listed above.<br>This conclusion is based upon the identical intended use, Indication<br>for use, principles of operation, patient contacting materials, and<br>sterilization processes. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.