The Precice Ankle Salvage System is intended for tibio-talo-calcaneal fusions. When used for TTC fusion, the Precice Ankle Salvage System may be used for open and closed fracture fixation, pseudarthrosis, mal-unions, or bone transport of long bones adjacent to the fusion site. The device may be used for subsequent limb lengthening once tibiotalo-calcaneal fusion has been achieved.
Device Story
Intramedullary fixation system for tibio-talo-calcaneal (TTC) fusion; consists of nail, locking screws, end caps, and external remote controller (ERC). Nail contains internal rare earth magnet, telescoping distraction rod, and planetary gearing. ERC enables non-invasive adjustment of nail length via magnetic coupling; allows post-operative compression across fusion site or limb lengthening. Used in clinical settings by surgeons. Provides mechanical stability for bone healing; enables controlled distraction or compression to address limb length discrepancies or fusion requirements.
Clinical Evidence
Bench testing only. Performance verification included dynamic compression, bending strength, axial pullout strength, insertion torque, torque resistance, and distraction/retraction force testing per ASTM F1264 and ASTM F543. A cadaver lab was conducted to evaluate osteotomy creation around the nail.
Technological Characteristics
Intramedullary nail manufactured from Biodur 108 alloy (ASTM F2229). Features internal rare earth magnet, telescoping rod, and planetary gearing. Energy source: external magnetic field (ERC). Sterilization: gamma radiation for nail/endcap; screws sterile or non-sterile. Connectivity: magnetic coupling between nail and external controller.
Indications for Use
Indicated for patients requiring tibio-talo-calcaneal (TTC) fusion, including those with open/closed fractures, pseudarthrosis, mal-unions, non-unions, or requiring bone transport of long bones adjacent to the fusion site, and subsequent limb lengthening.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services. To the right of the symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
NuVasive Specialized Orthopedics, Inc. Miriam Cervantes Regulatory Affairs Specialist 101 Enterprise, Suite 100 Aliso Viejo, California 92656 USA
# Re: K200430
Trade/Device Name: Precice® Ankle Salvage System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB, HWC Dated: August 17, 2020 Received: August 19, 2020
# Dear Miriam Cervantes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
September 16, 2020
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Michael Owens Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200430
Device Name Precice® Ankle Salvage System
### Indications for Use (Describe)
The Precice Ankle Salvage System is intended for tibio-talo-calcaneal fusions. When used for TTC fusion, the Precice Ankle Salvage System may be used for open and closed fracture fixation, pseudarthrosis, mal-unions, or bone transport of long bones adjacent to the fusion site. The device may be used for subsequent limb lengthening once tibiotalo-calcaneal fusion has been achieved.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Nuvasive Specialized Orthopedics, Inc. The logo features a stylized purple and gray graphic to the left of the company name. The company name is written in gray, with the words "SPECIALIZED ORTHOPEDICS, INC." in smaller letters below the name.
# Precice Ankle Salvage System 510(k) Summary
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR \$807.92, the following summary of information is provided:
#### A. Submitted by:
Miriam Cervantes Regulatory Affairs Specialist NuVasive Specialized Orthopedics, Inc. 101 Enterprise, Suite 100 Aliso Viejo, CA 92656 Telephone: (909) 229-7836
Date Prepared: September 15, 2020
#### B. Device Name
| Trade or Proprietary Name: | Precice® Ankle Salvage System |
|----------------------------|-----------------------------------------------|
| Common or Usual Name: | Rod, Fixation, Intramedullary and Accessories |
| Classification Name: | Intramedullary Fixation Rod |
| Device Class: | Class II |
| Classification: | 21 CFR § 888.3020 and 888.3040 |
| Product Code: | HSB, HWC |
#### C. Predicate Devices
The subject Precice Ankle Salvage System is substantially equivalent to the following predicate devices:
For tibio-talo-calcaneal fusion the Precice Ankle Salvage System is substantially equivalent to the predicate device DynaNail TTC Fusion System (K171376).
For fracture fixation, including opened and closed fracture fixation, pseudarthrosis, mal-unions, non-unions, or bone transport of long bones adjacent to the fusion site, and for subsequent limb lengthening the Precice Ankle Salvage System is substantially equivalent to the predicate device Precice Stryde System (K180503).
In addition, the following reference device was used to establish substantial equivalence, Precice System (K173129).
#### D. Device Description
The Precice Ankle Salvage System is a tibio-talo-calcaneal fusion system that consists of the Precice Ankle Salvage Nail, locking screws, end caps, and associated general instruments. The Precice Ankle Salvage Nail is compatible with an external remote controller (ERC). The Precice Ankle Salvage Nail and endcap are supplied sterile by gamma radiation. The locking screws are offered either sterile or non-sterile.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for "NUVASIVE SPECIALIZED ORTHOPEDICS, INC.". The logo features a stylized, abstract shape on the left, which is partially purple and partially gray. To the right of the shape, the word "NUVASIVE" is written in gray, block letters. Below "NUVASIVE", the words "SPECIALIZED ORTHOPEDICS, INC." are written in a smaller, purple font.
The Precice Ankle Salvage Nail contains an enclosed rare earth magnet, telescoping distraction rod, and planetary gearing which allows the length of the nail to be adjusted non-invasively by the ERC. Retraction of the Precice Ankle Salvage Nail can be utilized to maintain compression across the tibio-talo-calcaneal joints post-operatively. The Precice Ankle Salvage Nail can also be used to correct a limb length discrepancy that may result after the TTC fusion procedure.
The subject device is manufactured from medical grade Biodur 108 alloy per ASTM F2229. The Precice Ankle Salvage Nail is available in various nail styles, diameters, and lengths to accommodate a variety of patient anatomies and surgeon preference. The locking screws are available in a variety of diameters, lengths, and thread styles. The ERC is available in several compatible models.
#### Indications for Use E.
The Precice Ankle Salvage System is intended for tibio-talo-calcaneal fusions. When used for TTC fusion, the Precice Ankle Salvage System may be used for open and closed fracture fixation, pseudarthrosis, mal-unions, non-unions, or bone transport of long bones adjacent to the fusion site. The device may be used for subsequent limb lengthening once tibio-talo-calcaneal fusion has been achieved.
#### F. Comparison of Indications for use with the Predicate Device
The Precice Ankle Salvage System has the same intended use as the predicate device DynaNail TTC Fusion System (K171376) for tibio-talo-calcaneal fusion. Both the subject and predicate device achieve TTC fusion by providing sustained compression across the fusion site postoperatively.
Additionally, the Precice® Ankle Salvage System has the same intended use as the Precice System (K173129) for fracture fixation and limb lengthening.
Therefore, the differences in indications for use between the predicate devices and subject device do not create a new intended use.
#### G. Comparison of Technological Characteristics with the Predicate Device
As was established in this submission, the subject Precice Ankle Salvage System is substantially equivalent to the predicates, DynaNail TTC Fusion System (K171376), Precice Stryde System (K180503), and Precice System (K173129), which were previously cleared by the FDA for commercial distribution in the United States. The subject device has been shown to be substantially equivalent and have equivalent technological characteristics to the predicates through comparison in areas including design, principles of operation, labeling/intended use, material composition, and function.
The following table describes the summary comparison of technological characteristics of the subject device with the predicate devices.
| Predicate | <i>DynaNail TTC Fusion System</i> (K171376) | <i>Precice Stryde System</i> (K180503) |
|-----------|---------------------------------------------|----------------------------------------|
|-----------|---------------------------------------------|----------------------------------------|
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Nuvasive Specialized Orthopedics, Inc. The logo features a stylized, abstract shape in purple and gray on the left. To the right of the shape is the company name, "NUVASIVE," in gray, block letters. Below the name, in smaller letters, is the phrase "SPECIALIZED ORTHOPEDICS, INC." also in gray.
| Intended Use | TTC Fusion | Fracture Fixation and Limb Lengthening |
|-------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary of the subject device technology similarities to the predicate device | Principle of Operation:<br>sustained compression across the fusion site<br><br>Similarities in Geometry:<br>similar implant design<br><br>similar screw length and style<br>(headed & headless) | Principle of Operation:<br>distraction osteogenesis<br><br>Material Composition:<br>Stainless Steel |
| Summary of the subject device technology differences to the predicate device | Differences in Geometry:<br>subject nail offered in larger diameter of 13mm and longer length nails<br><br>subject nail offers additional smaller and larger screw diameters of the same length<br><br>subject nail is offered with a threaded distal end<br><br>subject nail uses internal gears and magnetic coupling instead of nitinol compressive elements | Differences in Geometry:<br>subject nail is offered with a threaded distal end<br><br>subject nail offers additional larger screw diameter and lengths<br><br>subject nail offers a shorter total stroke length for limb lengthening |
#### H. Performance Data
Nonclinical performance verification testing was performed to demonstrate that the subject Precice Ankle Salvage System is substantially equivalent to the predicate devices. Additionally, a cadaver lab was performed to address creation of osteotomy around the nail.
The following testing was performed:
| Testing Description | Applicable Standard |
|-----------------------------------------|---------------------------------------------------------------------------------------------|
| Dynamic Compression<br>Bending Strength | ASTM F1264 - Standard Specification and Test<br>Methods for Intramedullary Fixation Devices |
| Axial Pullout Strength | |
| Insertion Torque | ASTM F543 - Standard Specification and Test<br>Methods for Metallic Medical Bone Screws |
| Torque Resistance | |
| Distraction and<br>Retraction Force | N/A |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for Nuvasive Specialized Orthopedics, Inc. The logo features a stylized purple and gray graphic to the left of the company name. The company name is in a sans-serif font, with "NUVASIVE" in a larger font size than "SPECIALIZED ORTHOPEDICS, INC."
The results demonstrate that the subject Precice Ankle Salvage System is substantially equivalent to the predicates.
#### I. Conclusions
The subject device, the Precice Ankle Salvage System, has been shown to be substantially equivalent to the legally marketed predicate devices for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.