Echelon Contour Curved Cutter with blue reload, Echelon Contour Curved Cutter with green reload, Echelon Contour Curved Cutter reload, Blue, Echelon Contour Curved Cutter reload, Green
K200420 · Ethicon Endo-Surgery, LLC · GDW · Jun 29, 2020 · General, Plastic Surgery
Device Facts
Record ID
K200420
Device Name
Echelon Contour Curved Cutter with blue reload, Echelon Contour Curved Cutter with green reload, Echelon Contour Curved Cutter reload, Blue, Echelon Contour Curved Cutter reload, Green
Applicant
Ethicon Endo-Surgery, LLC
Product Code
GDW · General, Plastic Surgery
Decision Date
Jun 29, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Echelon Contour™ Curved Cutter with Reload is intended for transection in colorectal surgical procedures.
Device Story
Multifire, single-patient-use curved cutter stapler; used in colorectal surgery for transection/resection. Device features curved head; delivers four staggered rows of titanium staples with central knife cut. Includes safety interlocks: prevents firing if reload used; prevents firing unless closure trigger latched. Retaining pin holds tissue. Preloaded with blue or green reloads; supports up to six firings per procedure. Operated by surgeons in clinical settings. Provides permanent ligation of colorectal structures. Benefits include consistent staple line integrity and tissue management during resection.
Clinical Evidence
No clinical data. Evidence based on bench testing (leak onset pressure, formed staple height, staple line integrity, force to close/fire) and preclinical animal studies (hemostasis and survival). All testing performed per GLP and ISO 10993-1:2018 standards. Results show performance equivalent to predicate.
Technological Characteristics
Multifire curved cutter stapler; titanium staples. Features GST (Gripping Surface) and 3D staple geometry. Mechanical operation with safety interlocks. Biocompatible materials per ISO 10993-1:2018. MR conditional. Single-patient use.
Indications for Use
Indicated for transection and resection in colorectal surgical procedures.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Echelon Endoscopic Linear Cutter Reload, Black (K131663, K112056)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 29, 2020
Ethicon Endo-Surgery, LLC % Ruth James Senior Regulatory Affairs Program Lead Ethicon Endo-Surgery, Inc. 4545 Creek Rd Cincinnati, Ohio 45252
Re: K200420
Trade/Device Name: Echelon Contour Curved Cutter with blue reload, Echelon Contour Curyed Cutter with green reload, Echelon Contour Curved Cutter reload, Blue, Echelon Contour Curved Cutter reload, Green
Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW, GAG Dated: February 18, 2020 Received: February 20, 2020
Dear Ruth James:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200420
#### Device Name
Echelon Contour™ Curved Cutter Stapler with Reload
Indications for Use (Describe)
The Echelon Contour™ Curved Cutter with Reload is intended for transection in colorectal surgical procedures.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the Ethicon logo in red, with the words "ETHICON" in large, bold letters. Below the logo, in smaller letters, it says "PART OF THE Johnson & Johnson FAMILY OF COMPANIES". The logo is simple and clean, with a focus on the company name.
## 510(k) Summary
| Submitter Information: | Ethicon Endo-Surgery, LLC<br>475 Calle Street<br>Guaynabo, PR 00969 |
|------------------------|---------------------------------------------------------------------|
|------------------------|---------------------------------------------------------------------|
### Application Correspondent
Ruth James MSc, RAC Sr. Regulatory Affairs Program Lead Ethicon Endo-Surgery, LLC. Telephone: (513) 337-3118 Fax: (513) 337-1122 Email: rjames15@its.jnj.com
### Date Prepared February 18, 2020
| Device Trade Name: | Echelon Contour™ Curved Cutter Stapler with Reload |
|--------------------------------|----------------------------------------------------|
| Models: | GCS40B, GCS40G, GCR40B, GCR40G |
| Device Common Name: | Cutter/Stapler |
| Classification Regulation: | 21 CFR 878.4750; Implantable Staple |
| Device Class: | II |
| Panel: | 79, General & Plastic Surgery |
| Classification (Product) Code: | GDW |
Legally Marketed Predicate Device: CONTOUR™ Curved Cutter Stapler and Reload (cleared under K091322, K062869, K040038)
This predicate has not been subject to a design-related recall. Echelon Endoscopic Linear Cutter Reload, Black (K131663, K112056) was used as reference device in this submission for MRI Compatibility.
#### De vice Description
The Echelon Contour™ Curved Cutter Stapler with reload is a multifire, single patient use device with a curved head that cuts and staples. The device will provide ligation of colorectal structures when permanent ligation is required.
The device delivers four staggered rows of titanium staples, with a knife between the second and third row of staples, and staples and creates a 40 mm curved transection. The device is designed with a feature which prevents closing if a used reload is in the instrument.
{5}------------------------------------------------
Another feature is provided to prevent firing unless the closure trigger is latched in the closed position. A retaining pin holds tissue in place and can be positioned either manually or by squeezing the closure trigger.
The instrument is preloaded with a either a blue or green reload. The instrument may be further loaded for a maximum of six firings per instrument during a single procedure.
Each reload includes a knife blade with two staggered rows of staples on each side, an anvil, a yellow cutting washer, a retaining pin, and a staple retainer.
### Indications for Use
The Echelon Contour™ Curved Cutter with Reload is intended for transection and resection in colorectal surgical procedures.
The Indications for Use statement for the new device is not identical to the predicate device; however, the differences do not alter the intended therapeutic use of the device nor do they affect the safety and effectiveness of the device relative to the predicate.
### Comparison of technological characteristics with the predicate de vice
At a high level, the subject and predicate devices are based on the following same technological elements:
- Intended Use
- Indications for use are within the already cleared indications for predicate device
- Device design and operation is the same
- . Same open staple is used in both devices
The following technological differences exist between the subject and predicate devices:
- The reloads are designed with new Gripping Surface (GST) features and a new 3D final . staple shape
{6}------------------------------------------------
### Performance Data
The following performance data were provided in support of the substantial equivalence determination.
Bench Testing
Bench testing included the following tests:
- . Leak Onset Pressure Equivalency
- Formed Staple Height (FSH) Equivalency
- Staple Line Integrity and Staple Form Quality Equivalency .
- . Force to Close
- Force to Fire
- Handle Performance Characterization
The new device was shown to have equivalent performance to the predicate device in all the bench tests conducted. The data generated by each test supports a finding of substantial equivalence of the new device to the predicate device.
### Biocompatibility testing
The biocompatibility of materials used in the Echelon Contour™ Curved Cutter Stapler with Reload was evaluated based on ISO 10993-1:2018, "Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing", and on FDA guidance "Use of International Standard ISO 10993-1. "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", issued on June 16, 2016.
All in vitro and in vivo testing performed on materials was conducted in accordance with applicable requirements in the FDA's Good Laboratory Practice Standard (21 CFR 58).
The battery of testing included the following tests:
- . ISO Cytotoxicity
- . ISO Sensitization
- ISO Irritation (Intracutaneous Reactivity)
- ISO Acute Systemic Toxicity .
- . USP Material Mediated Pyrogenicity
There were no findings of toxicological concern, for the new devices, with testing conducted per ISO 10993-1:2018.
{7}------------------------------------------------
### Accelerated Aging Stability
Testing was performed on new devices that have been exposed to environmental and accelerated aging preconditioning. Device performance was maintained for the proposed shelf life.
### Sterilization
Pyrogen testing was conducted. The pyrogen testing and the risk assessment showed there is minimal to no risk for the staples with respect to the presence of endotoxins.
### Animal Study
The following studies were included in this submission:
- . Hemostasis study
- . Survival study
Both studies were conducted in accordance with applicable requirements in the FDA's Good Laboratory Practice Standard (21 CFR 58).
The new device was shown to have equivalent performance to the predicate device in the preclinical tests conducted. The data generated by each test supports a finding of substantial equivalence of the new device to the predicate device.
Clinical Studies This submission does not include data from Clinical Studies.
### MR Safety Information
Non-clinical testing demonstrated the staples in the device are MR conditional.
### Performance Data summary
The conclusions of the testing demonstrate that the new device performs substantially equivalent to the predicate device and does not raise new questions of safety and effectiveness.
### Overall Conclusions
The non-clinical data included in this submission support the stapler is as safe as the predicate device and demonstrate that the Echelon Contour™ Curved Cutter Stapler with Reload should perform as intended in the specified use conditions and perform comparably to the predicate device that is currently marketed for the same intended use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.