← Product Code [HGX](/productcode/HGX) · K200406

# Evolve Breast Pump (K200406)

_Hygeia II Medical Group, Inc. · HGX · Mar 18, 2020 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K200406

## Device Facts

- **Applicant:** Hygeia II Medical Group, Inc.
- **Product Code:** [HGX](/productcode/HGX.md)
- **Decision Date:** Mar 18, 2020
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 884.5160
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Therapeutic

## Indications for Use

The Evolve breast pump is intended to be used by lactating women to express and collect milk from their breasts.

## Device Story

Evolve Breast Pump is an electrically powered device for lactating women to express and collect milk; supports single or double pumping. Device uses DC-powered motor driving reciprocating vacuum pump and electromechanical solenoid; controlled electronically via soft-touch buttons and LCD display. Features two modes: stimulation and expression; allows adjustment of vacuum speed and strength. Intended for home use by multiple users. Subject device adds internal rechargeable Lithium Ion battery for cordless operation; otherwise identical to predicate. User operates device to extract milk; output is collected via Personal Accessory System (breast shield, tubing, bottles). Benefits include portable, adjustable milk expression for lactating women.

## Clinical Evidence

Bench testing only. Performance testing verified vacuum pressure and cycle rates across all settings (stimulation/expression, single/double) and confirmed battery capacity meets operational specifications. Electrical safety and EMC testing performed per IEC 60601-1 and IEC 60601-1-2 standards.

## Technological Characteristics

Powered by AC-DC adapter or internal rechargeable Lithium Ion battery. Reciprocating vacuum pump; electronic vacuum/cycling regulation. Backlit LCD interface. Multi-user home environment design. Silicone diaphragm for backflow protection. Standards: IEC 60601-1:2005 (3rd Ed), AAMI/ANSI ES60601-1:2005/(R)2012, IEC 60601-1-2:2014 (4th Ed).

## Regulatory Identification

A powered breast pump in an electrically powered suction device used to express milk from the breast.

## Predicate Devices

- Hygeia Evolve Breast Pump ([K190465](/device/K190465.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 18, 2020

Hygeia II Medical Group, Inc. Brett Nakfoor CEO 6241 Yarrow Drive Carlsbad, CA 92011

Re: K200406

> Trade/Device Name: Evolve Breast Pump Regulation Number: 21 CFR 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: February 18, 2020 Received: February 19, 2020

Dear Brett Nakfoor:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Sharon M. Andrews Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K200406

Device Name

Evolve Breast Pump

Indications for Use (Describe)

The Evolve breast pump is intended to be used by lactating women to express and collect milk from their breasts.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|                                                 |  |

| | Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) SUMMARY - K200406

In accordance to CFR Title 21, Sec. 807.92 the following summary is provided.

#### SUBMITTER:

Hygeia II Medical Group, Inc. 6214 Yarrow Drive, Suite A Carlsbad, CA 92011 Phone: (714) 515-7571

#### PRIMARY CONTACT PERSON:

Brett Nakfoor CEO Hygeia II Medical Group (714) 515-7571

#### SECONDARY CONTACT PERSON:

John Conklin Product Development Hygeia II Medical Group (714) 515-7571

DATE PREPARED: March 12, 2020

#### Device Information

Trade Name: Evolve Breast Pump Common Name: Powered Breast Pump Classification Name: Powered Breast Pump Classification Panel: Obstetrics/Gynecology Classification Regulation: 21 CFR 884.5160 Device Class: Class II Product Code: HGX (Pump, Breast, Powered)

#### Predicate Device Information

Hygeia Evolve Breast Pump, K190465, October 18, 2019.

The predicate device has not been subject to a design-related recall.

{4}------------------------------------------------

## Device Description

The Evolve Breast Pump, the subject device, is an electrically powered breast pump from an external AC-DC power supply or internal, rechargeable batteries; the device is provided non-sterile. The subject device is a cordless version of the predicate device.

The Evolve Breast Pump is intended to be used by lactating women to express and collect mills from their breasts. Pumping can be performed on one breast (single pumping) or on both breasts of a lactating woman at the same time (double pumping). The breast pump is designed as a multi-user breast pump and is available as a cordless model powered by an internal, rechargeable Lithium lon battery or an external AC-DC power supply.

The Evolve Breast Pump utilizes a DC-powered motor driving a reciprocating-type vacuum pump and an electromechanical solenoid which are controlled electronically to provide vacuum (suction) capability that extracts and collects milk from breasts of lactating women.

The Evolve Breast Pump has a backlit, LCD display. The subject device also has soft-touch buttons for the user to power the device on/off, switch between two different pumping modes - stimulation and expression – and to control vacuum speed and strength within each of the modes. Multiple vacuum speed and strength settings are available in both pumping modes.

The Evolve Breast Pump is intended for repeated use by multiple users in the home environment.

The Evolve Breast Pump requires Hygeia's Personal Accessory System for users to express and collect milk from their breasts.

#### Indications for Use

The Evolve breast pump is intended to be used by lactating women to express and collect milk from their breasts.

#### Comparison of Technological Characteristics

The Evolve Breast Pump has identical indications for use and the same fundamental technology as the legally marketed predicated device, the Evolve Breast Pump (K190465). The modification to the predicate device, subject to this submission, is the addition of an internal rechargeable Lithium lon battery to allow cordless operation of the breast pump.

A summary of the technological characteristics of the subject and predicate device are shown in the following table.

|                                                       | Evolve<br>K200406                                                                                                                                   | Evolve<br>(Predicate Version)<br>K190465                                                                                                            | Comment                                                                                           |
|-------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Indications for Use                                   | The Evolve powered breast<br>pump is to be used by lactating<br>women to express and collect<br>milk from their breasts.                            | The Evolve powered breast<br>pump is to be used by lactating<br>women to express and collect<br>milk from their breasts.                            | Same                                                                                              |
| Environment of<br>Use                                 | Home                                                                                                                                                | Home                                                                                                                                                | Same                                                                                              |
| User Interface - Controls                             |                                                                                                                                                     |                                                                                                                                                     |                                                                                                   |
| User Control                                          | State-dependent controls:<br>On-Off button<br>Stimulation and Expression<br>Mode button                                                             | State-dependent controls:<br>On-Off button<br>Stimulation and Expression<br>Mode button                                                             | Same                                                                                              |
|                                                       | Performance controls:<br>Strength and Speed adjustment<br>via buttons                                                                               | Performance controls:<br>Strength and Speed adjustment<br>via buttons                                                                               |                                                                                                   |
| Visual Indicator                                      | LCD<br>Backlit, liquid-crystal display                                                                                                              | LCD<br>Backlit, liquid-crystal display                                                                                                              | Same                                                                                              |
| Pumping Options                                       | Single pumping<br>Double pumping                                                                                                                    | Single pumping<br>Double pumping                                                                                                                    | Same                                                                                              |
| Adjustable Suction<br>Level                           | Yes                                                                                                                                                 | Yes                                                                                                                                                 | Same                                                                                              |
| Adjustable Speed<br>Level                             | Yes                                                                                                                                                 | Yes                                                                                                                                                 | Same                                                                                              |
| Accessories                                           | Personal Accessory Set:<br>• breast shield<br>• diaphragm<br>• valve<br>• tubing<br>• bottles<br>AC-DC power-adapter-cord                           | Personal Accessory Set:<br>• breast shield<br>• diaphragm<br>• valve<br>• tubing<br>• bottles<br>AC-DC power-adapter-cord                           | Same<br>All components and<br>accessories have been<br>cleared with predicate<br>device (K190465) |
| Cleaning                                              | Breast pump - wipe with clean,<br>damp cloth<br>Tubing - replace if milk appears<br>in tubing<br>Breast pump kit and bottles -<br>wash and sanitize | Breast pump - wipe with clean,<br>damp cloth<br>Tubing - replace if milk appears<br>in tubing<br>Breast pump kit and bottles -<br>wash and sanitize | Same                                                                                              |
| Specifications                                        |                                                                                                                                                     |                                                                                                                                                     |                                                                                                   |
| Power Source                                          | AC-DC 100 - 240 Vac adapter,<br>50/60 Hz                                                                                                            | AC-DC 100 - 240 Vac adapter,<br>50/60 Hz                                                                                                            | Same                                                                                              |
| Batteries                                             | Internal, rechargeable<br>Lithium Ion battery                                                                                                       | N/A                                                                                                                                                 | Subject device has<br>internal<br>Lithium Ion battery                                             |
| Rechargeable<br>Battery<br>operation/charge<br>time   | Operation time:<br>Approximately 120 Minutes<br>Charge time:<br>Approximately 180 Minutes                                                           | N/A                                                                                                                                                 | Subject device provides<br>ability to operate in<br>cordless configuration by<br>battery          |
| Software/firmware                                     | Yes                                                                                                                                                 | Yes                                                                                                                                                 | Same                                                                                              |
| Vacuum aggregate<br>type                              | Reciprocating pump                                                                                                                                  | Reciprocating pump                                                                                                                                  | Same                                                                                              |
| Vacuum<br>Regulation type                             | Electronic                                                                                                                                          | Electronic                                                                                                                                          | Same                                                                                              |
| Maximum vacuum<br>(mmHg)                              | -283 mmHg<br>(-20 mmHg)                                                                                                                             | -283 mmHg<br>(-20 mmHg)                                                                                                                             | Same                                                                                              |
| Vacuum range –<br>double and single<br>pumping (mmHg) | Stimulation<br>-75 to -175 mmHg<br>(±20 mmHg)<br>Expression<br>-140 to -283 mmHg<br>(±20 mmHg)                                                      | Stimulation<br>-75 to -175 mmHg<br>(±20 mmHg)<br>Expression<br>-140 to -283 mmHg<br>(±20 mmHg)                                                      | Same                                                                                              |
| Cycling Control<br>Mechanism                          | Electronic                                                                                                                                          | Electronic                                                                                                                                          | Same                                                                                              |
| Cycle Speed<br>Range<br>(Cycles/Minute =<br>CPM)      | Stimulation<br>75 to 92 CPM<br>(±8 CPM)<br>Expression<br>32 to 49 CPM<br>(±8 CPM)                                                                   | Stimulation<br>75 to 92 CPM<br>(±8 CPM)<br>Expression<br>32 to 49 CPM<br>(±8 CPM)                                                                   | Same                                                                                              |
| Adjustable<br>Cycle Speed<br>Levels                   | Yes                                                                                                                                                 | Yes                                                                                                                                                 | Same                                                                                              |
| Backflow<br>protection                                | Yes, silicone diaphragm                                                                                                                             | Yes, silicone diaphragm                                                                                                                             | Same                                                                                              |

{5}------------------------------------------------

{6}------------------------------------------------

The differences in technological characteristics do not raise different questions of safety and effectiveness.

## Summary of Performance Data

The Evolve Breast Pump complies with voluntary standards for electrical safety, electromagnetic compatibility, and use in the home environment.

The following performance data is provided in support of the substantial equivalence determination:

#### Risk/Hazard

- . Risk Analysis in accordance with ISO 14971:2007 was used to assess impact of modifications to the device
Electrical Safety and Electromagnetic Compatibility (EMC)

- Electrical safety testing per IEC 60601-1:2005 (3rd Edition) with US deviations per . AAMI/ANSI ES60601-1:2005/(R)2012 and A1:2012 standard
- Electromagnetic compatibility testing per IEC 60601-1-2: 2014 (4th edition) standard, ● Medical Electrical Equipment Part 1 - 2

#### Performance Testing

The Evolve Breast Pump was tested to demonstrate it meets its performance specifications. The testing involved measurement of vacuum and cycles for minimum and maximum settings for both single and double pumping mode, and for stimulation and expression mode 

{7}------------------------------------------------

with the internal, rechargeable Lithium Ion battery to show battery energy capacity is sufficient to assure the subject device performs to its specifications were met and are within the required, acceptable ranges for pump operation, cycle rate, vacuum pressure, and battery operation time.

#### Conclusion

The performance testing demonstrates that the Evolve Breast Pump is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K200406](https://fda.innolitics.com/device/K200406)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
