K200384 · Surgical Planning Associates, Inc. · OSF · Jan 28, 2021 · Neurology
Device Facts
Record ID
K200384
Device Name
HipXpert 3D Display and Anchoring Application
Applicant
Surgical Planning Associates, Inc.
Product Code
OSF · Neurology
Decision Date
Jan 28, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The HipXpert system is a manual surgical instrument and associated software application designed for use in planning surgery and aligning the acetabular components during hip arthroplasty procedures. The HipXpert 3D Display and Anchoring application with the HoloLens2 is indicated for visual alignment of an acetabular cup impactor during hip arthroplasty when pin-based fixation of the HipXpert tool is utilized.
Device Story
System provides patient-specific hip arthroplasty surgical planning and acetabular positioning. Inputs: patient CT scan data. Transformation: software creates 3D pelvic model and calculates instrument settings; mixed reality application (HoloLens2) overlays 3D surgical plan onto physical surgical field using QR target anchoring. Output: visual guidance for surgeon to orient acetabular cup impactor. Used in OR by surgeons. Benefits: improved accuracy of acetabular component orientation via augmented reality visualization of pre-operative plan.
Clinical Evidence
No clinical data. Bench testing only: software validation (file ID, QR recognition, anchoring); system accuracy/registration/tracking verification (analogous to ASTM F2554-10); headset performance verification (FOV, resolution, luminance, distortion, contrast, stability).
Technological Characteristics
Manual mechanical navigation instrument with three-leg pelvic fixation; Microsoft HoloLens2 headset for AR display; QR target for spatial anchoring. Software-based planning and visualization. Sterilization: not specified. Connectivity: standalone/local processing.
Indications for Use
Indicated for patients undergoing hip arthroplasty requiring acetabular component alignment; used for visual alignment of acetabular cup impactors via pin-based fixation.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
January 28, 2021
Surgical Planning Associates, Inc % Kellen Hills Quality and Regulatory Consultant Orchid Design 80 Shelton Technology Center Shelton, Connecticut 06484
Re: K200384
Trade/Device Name: HipXpert System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OSF, OLO Dated: December 23, 2020 Received: December 29, 2020
Dear Kellen Hills:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For; Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200384
Device Name HipXpert System
Indications for Use (Describe)
The HipXpert system is a manual surgical instrument and associated software application designed for use in planning surgery and aligning the acetabular components during hip arthroplasty procedures.
The HipXpert 3D Display and Anchoring application with the HoloLens2 is indicated for visual alignment of an acetabular cup impactor during hip arthroplasty when pin-based fixation of the HipXpert tool is utilized.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K200384
| (a)(1) | Date Prepared: | December 23, 2020 |
|--------|------------------------------------|-----------------------------------------------------------------------------------------------------------|
| | Submitted By: | Surgical Planning Associates, Inc.<br>New England Baptist Hospital,<br>125 Parker Hill Ave Suite 545 |
| | Phone: | Boston, MA 02120<br>617-277-4434 |
| | Contact: | Stephen B Murphy, M.D. |
| | Prepared By: | Orchid Design<br>80 Shelton Technology Center,<br>Shelton, CT 06484 |
| | Phone: | (203) 922 0105 |
| (a)(2) | Proprietary Name: | HipXpert System |
| | Common Name: | Patient specific manual orthopedic<br>stereotaxic system |
| | Classification Name and Reference: | 21CFR 882.4560 - Stereotaxic instrument |
| | Product Codes: | OSF - Patient Specific Manual Orthopedic<br>Stereotaxic System<br>OLO - Orthopedic Stereotaxic Instrument |
| | | |
| (a)(3) | Predicate Devices: | |
| | Primary: | Surgical Planning Associates, Inc - Hip |
| | | Sextant Instrument System (K093491) |
| | Reference: | Augmedics, Ltd - xvision Spine (K190929) |
### (a)(4) Device Description:
The HipXpert system provides a patient-specific hip arthroplasty surgical plan allowing for accurate acetabular positioning using CT mapping of a patient's pelvis using a software application and a reusable, manual, mechanical navigation instrument.
The HipXpert software planning application uses patient image data to create a detailed 3D model of the pelvis as well as the instrument settings necessary for proper acetabular cup orientation.
{4}------------------------------------------------
The HipXpert mechanical instrument has three legs which are secured to the pelvis. The legs form three points which define the sextant plane. The HipXpert mechanical instrument has two protractors that are adjusted to orientate an indicator pin in the direction of the desired orientation of the acetabular component.
The purpose of this submission is to introduce a variant of the device that was cleared in K093491 which uses a different software application to allow the user to see the existing 3D objects provided in the surgery plans in three-dimensions on mixed reality lenses. Additionally, the indications for use have been amended to add clarity and align with the "Physical State" as specified by product code OSF.
The subject HipXpert 3D Display and Anchoring Application utilizes the previously cleared (K093491) HipXpert planning application and HipXpert tools in addition to a mixed reality headset (Microsoft HoloLens2) and QR target. The HoloLens2 is an off-the-shelf component developed and manufactured by Microsoft which is used to view superimposed 3D images from the HipXpert planning application on the real HipXpert tool. In order to properly orient the 3D images displayed by the HoloLens2, a QR target is used to anchor these 3D images in space as they are overlaid on the real HipXpert tool.
- (a)(5) Indications for Use:
The HipXpert system is a manual surgical instrument and associated software application designed for use in planning surgery and aligning the acetabular components during hip arthroplasty procedures.
The HipXpert 3D Display and Anchoring application with the HoloLens2 is indicated for visual alignment of an acetabular cup impactor during hip arthroplasty when pin-based fixation of the HipXpert tool is utilized.
#### (a)(6) Comparison of Technological Characteristics:
The HipXpert system is substantially equivalent to the previously cleared hip sextant system based on similarities in intended use, design, materials, packaging, sterilization and performance. The only difference is an additional software application which allows the user to view existing 3D objects provided in the surgery plans in three-dimensions using mixed reality lenses. The use of a headset for presenting stereoscopic augmented reality (AR) display of patient's anatomy is not a new feature and has been previously cleared under K190929 for the xvision Spine (reference device). The technological characteristics do not raise any new questions of safety and efficacy.
### (b)(1) Non-clinical testing:
The subject device modification was successfully evaluated according to the following:
- Validation of the software used for file identification, QR target recognition and file anchoring
- . Verification of overall system orientation and position accuracy in a cadaveric model using screws placed into the acetabulum
{5}------------------------------------------------
- . Verification of system accuracy, image registration accuracy and tracking accuracy using methods analogous to ASTM F2554-10 under varying conditions of light, motion, distance, and angle
- . Performance of the Headset display was demonstrated by verifying the following elements: Field of View (FOV), resolution, luminance, distortion, contrast ratio, contrast of physical object, location of the virtual image, and stability of the virtual objects due to motion
#### (b)(2) Clinical testing:
Clinical testing was not required to demonstrate substantial equivalence in this premarket notification.
## (b)(3) Conclusions:
Based on the information provided in this premarket notification, we believe that the subject HipXpert system demonstrates substantial equivalence to the predicate Hip Sextant Instrument System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.