K200382 · Btl · NGX · Jan 6, 2021 · Physical Medicine
Device Facts
Record ID
K200382
Device Name
BTL-703-2
Applicant
Btl
Product Code
NGX · Physical Medicine
Decision Date
Jan 6, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
BTL-703-2 is indicated to be used for: - Improvement of abdominal tone, strengthening of the abdominal muscles, development of firmer abdomen. - Strengthening, toning and firming of buttocks, thighs and calves. - Improvement of muscle tone and firmness, for strengthening muscles in arms.
Device Story
Non-invasive therapeutic device; delivers electromagnetic stimulation and bipolar radiofrequency (RF) energy to targeted tissue. Main unit with two applicators enables hands-free simultaneous treatment; applicators secured by fixation belts. User interface features large color touch-screen for parameter setting and step-by-step therapy guidance. Electromagnetic field interacts with body tissues to achieve muscle stimulation; bipolar RF provides surface warming (below 40°C) to enhance patient comfort. Used in clinical settings by trained operators. Output affects muscle tone and firmness; benefits include improved abdominal tone and muscle strengthening in specified body areas.
Clinical Evidence
Bench testing only. Device evaluated for electrical safety and performance per IEC 60601-1, IEC 60601-1-2, IEC 62304, ISO 14971, and ISO 10993 (cytotoxicity, irritation, sensitization). No clinical data presented.
Technological Characteristics
System console with electromagnetic field and bipolar RF generators, computer, and touch-screen control panel. Two output channels. Electromagnetic field intensity 0.5–2.0 T; pulse repetition 1–150 Hz; pulse duration 190–280 µs; biphasic sinusoidal waveform. RF frequency 27.12 MHz; total power 60 W (2x30 W). Includes overheating protection sensors. Class II, BF electrical protection. Software-controlled.
Indications for Use
Indicated for improvement of abdominal tone, strengthening of abdominal muscles, development of firmer abdomen, and strengthening, toning, and firming of buttocks, thighs, calves, and arms.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The logo is simple and professional, and it is easily recognizable.
January 6, 2021
BTL David Chmel VP of Operations 362 Elm Street Marlborough, Massachusetts 01752
Re: K200382
Trade/Device Name: BTL-703-2 Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: December 10, 2020 Received: December 14, 2020
Dear David Chmel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K200382
Device Name BTL-703-2
Indications for Use (Describe) BTL-703-2 is indicated to be used for:
- Improvement of abdominal tone, strengthening of the abdominal muscles, development of firmer abdomen.
- · Strengthening, toning and firming of buttocks, thighs and calves.
- · Improvement of muscle tone and firmness, for strengthening muscles in arms.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☑</span> |
|----------------------------------------------|-------------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | □ |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of three interconnected blue squares, with the letters "BTL" inside the leftmost square. To the right of the squares is the text "BTL Industries" in a dark gray, sans-serif font. The logo is simple and modern, and the use of blue and gray gives it a professional look.
# 510(k) Summary
# General Information
| Sponsor: | BTL Industries, Inc.<br>362 Elm Street<br>Marlborough, MA 01752<br>Tel: +1-866-285-1656<br>Fax: +1-888-499-2502 |
|------------------------------|-----------------------------------------------------------------------------------------------------------------|
| | Fax: +1-888-499-2502 |
| Applicant: | BTL Industries, Inc.<br>362 Elm Street<br>Marlborough, MA 01752<br>Tel: +1-866-285-1656<br>Fax: +1-888-499-2502 |
| Contact Person: | David Chmel<br>BTL Industries, Inc.<br>chmel@btlnet.com |
| Summary Preparation<br>Date: | 31 December 2020 |
## Device Name
| Trade/Proprietary Name: | BTL-703-2 |
|------------------------------|--------------------------------------------------|
| Primary Classification Name: | Stimulator, Muscle, Powered, Muscle Conditioning |
| Classification Regulation: | 21 CFR 890.5850, Class II |
| Classification Product Code: | NGX |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of three interconnected blue squares with the letters "BTL" inside the leftmost square. To the right of the logo is the text "BTL Industries" in a dark gray sans-serif font. The text is aligned horizontally with the logo.
#### Legally Marketed Predicate Device
The BTL-703-2 is a state-of-the-art device with accessories combining electromagnetic and high frequency energy, and is substantially equivalent to the following product, which is already cleared for distribution in the USA under the following 510(k) Premarket Notification number:
- 799-2L (K190456) ●
#### Reference Device BTL-899
The reference device has been added to this section because it is technologically similar to the BTL-703-2 device. Both devices are tracked internally under a common project number, P899. Hence, all test reports and technical documentation refer to this number.
#### Product Description
The BTL-703-2 is a non-invasive therapeutic device is comprised of a main unit and applicators that deliver electromagnetic and radiofrequency to the targeted tissue. The device's two outputs enable hands-free simultaneous treatment by two applicators.
The BTL-703-2 is equipped with a large color touch-screen that significantly facilitates the use of the device. The on-screen information guides the user step-by-step through the entire therapy procedure. The therapeutic parameters are easily set using the touch-screen of the device. During the therapy the device keeps information about the applied therapy type, remaining therapy time and main therapy parameters on the screen.
#### Indications for Use
BTL-703-2 is indicated to be used for:
- Improvement of abdominal tone, strengthening of the abdominal muscles, development of firmer abdomen.
- Strengthening, toning and firming of buttocks, thighs and calves.
- Improvement of muscle tone and firmness, for strengthening muscles in arms. ●
{5}------------------------------------------------
### Non-clinical Testing
The BTL-703-2 device has been thoroughly evaluated for electrical safety. The device has been found to comply with the following applicable medical device safety standards:
| IEC 60601-1 | Medical electrical equipment – Part 1: General requirements for basic safety<br>and essential performance |
|---------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1-2 | Medical electrical equipment – Part 1-2: General requirements for basic safety<br>and essential performance - Collateral standard: Electromagnetic disturbances<br>– Requirements and tests |
| IEC 62304 | Medical device software - Software lifecycle processes |
| ISO 14971 | Medical devices – Application of risk management to medical devices |
| ISO 10993-1 | Biological evaluation of medical devices – Part 1: Evaluation and testing within<br>a risk management process |
| ISO 10993-5 | Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity |
| ISO 10993-10 | Biological evaluation of medical devices – Part 10: Tests for irritation and skin<br>sensitization |
### Technological Characteristics
The BTL-703-2 device has similar indications for use and technological characteristics and principles of operation to its predicate device. The BTL-703-2 device and its predicate are comprised of a system console and applicators. The system console consists of the electromagnetic field and radiofrequency generators, computer, and a touch-screen control panel.
The generated electromagnetic field is intenact with the tissues of the human body in order to achieve muscle stimulation. The radiofrequency is providing a low degree of warming of the applicators and thus enhance the patient's comfort during the treatment.
The mechanism of action and technological similarities and differences between the BTL-703-2 device and the predicate device are described below in the comparison table. The differences do not raise any new types of safety or effectiveness questions.
{6}------------------------------------------------
# Comparison with the Predicate Devices
| 510(k)<br>number<br>Device name<br>Company<br>name | K200382<br>BTL-703-2<br>BTL Industries, Inc. | K190456<br>BTL 799-2L<br>BTL Industries, Inc. | Reference Device<br>K192224<br>BTL-899<br>BTL Industries, Inc. | Significant<br>difference |
|----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| Product<br>Code<br>and<br>Regulation | Physical Medicine<br>21 CFR 890.5850<br>NGX – Stimulator,<br>Muscle, Powered,<br>Muscle Conditioning | Physical Medicine<br>21 CFR 890.5850<br>NGX - Stimulator,<br>Muscle, Powered,<br>Muscle Conditioning | General & Plastic Surgery<br>21 CFR 878.4400<br>GEI - Electrosurgical,<br>Cutting & Coagulation &<br>Accessories | None |
| Indications<br>for Use | BTL-703-2 is indicated<br>to be used for:<br>Improvement of<br>abdominal tone,<br>strengthening of<br>the abdominal<br>muscles,<br>development of<br>firmer abdomen.Strengthening,<br>toning and firming<br>of buttocks, thighs<br>and calves.Improvement of<br>muscle tone and<br>firmness, for<br>strengthening<br>muscles in arms. | BTL 799-2L is<br>indicated to be used<br>for:<br>Improvement of<br>abdominal tone,<br>strengthening of<br>the abdominal<br>muscles,<br>development of<br>firmer abdomen.Strengthening,<br>toning and firming<br>of buttocks, thighs<br>and calves.Improvement of<br>muscle tone and<br>firmness, for<br>strengthening<br>muscles in arms. | BTL-899 is indicated to<br>be used for:<br>Non-invasive<br>lipolysis (breakdown<br>of fat) of the<br>abdomen.Reduction in<br>circumference of<br>the abdomenThe BTL-899 is intended<br>for use with Skin Type I to<br>Skin Type III. | Not<br>Significantly<br>different |
| Basic<br>Technology | Electromagnetic<br>stimulation<br>accompanied by<br>bipolar radiofrequency | Electromagnetic<br>stimulation | Electromagnetic<br>stimulation<br>accompanied by<br>bipolar radiofrequency | Not<br>Significantly<br>different<br>Please see the<br>discussion and |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for BTL Industries. The logo consists of three interconnected blue squares on the left side of the image. To the right of the squares is the text "BTL Industries" in a dark gray sans-serif font.
| | | | | conclusion<br>below |
|---------------------------------------------------------------------------------|----------------------------------------------------------|-----------------------------------------------------|----------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| Clinical Use | Prescription use | Prescription use | Prescription use | None |
| Electrical<br>Protection | Class II, BF | Class II, BF | Class II, BF | None |
| User<br>Interface | Touch screen | Touch screen | Touch screen | None |
| Firmware<br>Controlled | Yes | Yes | Yes | None |
| Number of<br>outputs<br>channels | 2 | 2 | 2 | Not<br>Significantly<br>different |
| Number of<br>output<br>modes* | 1 | 1 | 1 | None |
| Number of<br>Magnetic<br>coils<br>in the<br>Applicator | 1 | 1 | 1 | None |
| Magnetic<br>Field<br>Intensity (on<br>the coil<br>surface) | 0.5 to 1.8 T<br>for app:<br>BTL-899-AP-C-1 | 0.5 to 1.8 T<br>for app:<br>BTL 299-6 | 0.5 to 1.8 T<br>for app:<br>BTL-899-AP-C-1 | None |
| Magnetic<br>Field<br>Intensity (on<br>the coil<br>surface) | 0.7 to 2.0 T<br>for app:<br>BTL-899-AP-C-2 | 0.7 to 2.0 T<br>for app:<br>BTL 299-7 | N/A | None |
| Maximum<br>magnetic | 908 mT | 980 mT | 908 mT | None |
| field<br>intensity at<br>applicator<br>center<br>surface | for app: BTL-899-AP-<br>C-1 | for app:<br>BTL 299-6 | for app: BTL-899-AP-<br>C-1 | None |
| Maximum<br>magnetic<br>field<br>intensity at<br>applicator<br>center<br>surface | 1238 mT<br>for app: BTL-899-AP-<br>C-2 | 1220 mT<br>for app:<br>BTL 299-7 | 1238 mT | None |
| Pulse<br>Repetition<br>Rate -<br>supported<br>by the<br>device | 1 – 150 Hz | 1 – 150 Hz | 1 – 150 Hz | None |
| Pulse<br>Duration | $280 \mu s \pm 20\% \mu s$<br>for app:<br>BTL-899-AP-C-1 | $280 \mu s \pm 20\% \mu s$<br>for app:<br>BTL 299-6 | $280 \mu s \pm 20\% \mu s$<br>for app:<br>BTL-899-AP-C-1 | None |
| Pulse<br>Duration | $190 \mu s \pm 20\% \mu s$<br>for app:<br>BTL-899-AP-C-2 | $190 \mu s \pm 20\% \mu s$<br>for app:<br>BTL 299-7 | N/A | None |
| Waveform | Biphasic | Biphasic | Biphasic | None |
| Shape | Sinusoidal | Sinusoidal | Sinusoidal | None |
| RF Type | Bipolar | N/A | Bipolar | Not<br>Significantly<br>different<br>Please see<br>discussion on<br>basic technology<br>below |
| Total RF<br>Power | 60 W (2x30) | N/A | 60 W (2x30) | Not<br>Significantly<br>different<br>Please see<br>discussion on<br>basic technology<br>below |
| RF<br>Frequency | 27.12 Mhz | N/A | 27.12 Mhz | Not<br>Significantly<br>different<br>Please see<br>discussion on<br>basic technology<br>below |
| Overheating<br>Protection<br>Sensor | Yes | N/A | Yes | Not<br>Significantly<br>different |
| Selection of<br>parameters | Yes | Yes | Yes | None |
| Application | Hands-free, applicator<br>fixed by fixation belt | Hands-free,<br>applicator fixed by<br>fixation belt | Hands-free, applicator<br>fixed by fixation belt | Not<br>Significantly<br>different |
| Therapy<br>Time | Up to 30 min | Up to 60 min | Up to 30 min | Not<br>Significantly<br>different |
| Energy<br>Source | 100 - 240 V AC, 50-<br>60 Hz | 100 - 240 V AC, 50-<br>60 Hz | 100 - 240 V AC, 50-<br>60 Hz | None |
| Maximum<br>Output<br>Voltage | N/A | N/A | N/A | N/A |
| Maximum<br>Output<br>Current | N/A | N/A | N/A | N/A |
| Maximum<br>Current<br>Density<br>(mA/cm²) | N/A | N/A | N/A | N/A |
| Maximum<br>Phase<br>Charge,<br>(mC) | N/A | N/A | N/A | N/A |
| Maximum<br>Power<br>Density*<br>(W/cm²) | N/A | N/A | N/A | N/A |
| System<br>Dimensions<br>(W×H×D) | 23 x 39 x 29 in<br>(592 x 985 x 730 mm) | 23 x 55 x 23 in<br>(580x1380x580 mm) | 23 x 39 x 29 in<br>(592 x 985 x 730 mm) | Not<br>Significantly<br>different |
| System<br>Weight | 85 kg | 44 kg | 85 kg | Not<br>Significantly<br>different |
| Ambient<br>Storage<br>Temperature | 14 °F to 131 °F (-10°C<br>to +55°C) | 14 °F to 131 °F (-<br>10°C to +55°C) | 14 °F to 131 °F (-10°C<br>to +55°C) | None |
| Relative<br>Storage<br>Humidity | 10% to 85% | 10% to 85% | 10% to 85% | None |
| Environment<br>al<br>Specificatio<br>ns | For indoor use only | For indoor use only | For indoor use only | None |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for BTL Industries. The logo consists of three interconnected blue squares with the letters "BTL" inside. To the right of the logo, the words "BTL Industries" are written in a dark gray sans-serif font.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of three blue squares that are nested inside each other. To the right of the squares is the text "BTL Industries" in a dark gray sans-serif font.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of three nested squares with the letters "BTL" inside the innermost square. To the right of the logo is the text "BTL Industries" in a sans-serif font.
### Substantial Equivalence
The BTL-703-2 is indicated for improvement of abdominal tone, strengthening of the abdominal muscles and development of firmer abdomen, as well as for strengthening, toning and firming of buttocks, thighs and calves. These indications are the same as the predicate device's indications.
The predicate device BTL-799L utilizes electromagnetic stimulation in order to achieve therapeutic effect. The proposed device BTL-703-2 also utilizes electromagnetic stimulation for its main therapeutic effect; however, the device is also equipped with a bipolar radiofrequency generator. The radiofrequency is warming the applicators' surface to below 40℃ thus enhancing the patient's comfort during the treatment.
The BTL-703-2 electromagnetic technology is identical to FDA cleared predicate device BTL 799-2L. The bipolar radiofrequency technology has also been cleared by FDA in a number of devices, for example BTL-899 (K192224).
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of three interconnected blue squares on the left side. To the right of the squares is the text "BTL Industries" in a dark gray, sans-serif font.
Any differences between the predicate device and BTL-703-2 have no significant influence on the safety or effectiveness of the BTL-703-2 device. Therefore, the BTL-703-2 is substantially equivalent to the predicate device.
### Conclusion
Based upon the intended use and known technical information provided in this pre-market notification, the BTL-703-2 device has been shown to be substantially equivalent to the currently marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.