The Navbit® Sprint System is a computer-controlled system intended to assist the surgeon in relation to reference axes during orthopedic surgical procedures. The Navbit® Sprint System facilitates the accurate positioning of implants, relative to these alignment axes.
Device Story
Navbit Sprint is a computer-assisted surgical navigation system for hip arthroplasty. It uses a palm-sized, single-use disposable computer module containing rate gyroscopes and accelerometers. During surgery, the device is attached to instruments to register the pelvic coordinate system. It provides real-time angular measurements (inclination and anteversion) to guide acetabular cup implantation. Used in hospital operating theaters by surgeons. The system assists in accurate implant positioning relative to registered pelvic axes. Benefits include improved alignment accuracy during hip arthroplasty. Output is displayed via an integrated graphical user interface on the unit attached to the instrumentation.
Clinical Evidence
No clinical trials were performed. Evidence consists of bench testing and simulated use (cadaver) testing. Bench testing verified angular measurement accuracy per ASTM F2554-18. Simulated use testing with an advising surgeon validated that the system meets user needs and clinical pass/fail criteria.
Technological Characteristics
Computer-assisted navigation system; palm-sized module with rate gyroscopes and accelerometers. Materials: 316L surgical stainless steel (ASTM 138) for bone pins, medical-grade polymers, and internal electronics. Energy source: two internal 3V Lithium/Iron Disulfide (Li/FeS2) batteries. Connectivity: standalone. Sterilization: EO sterilization for navigation unit and pins. Single-use disposable.
Indications for Use
Indicated for patients undergoing Hip Arthroplasty surgical procedures by any approach in lateral decubitus or supine position, using uncemented acetabular cups allowing post-impaction correction where confirmatory measurement checks can be obtained.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
December 10, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
Navbit Pty Ltd % Elizabeth O'Keeffe, PhD Director of Regulatory Affairs Secure BioMed Evaluations 7828 Hickorv Flat Highway Suite 120 Woodstock, Georgia 30188
# Re: K200376
Trade/Device Name: Navbit Sprint Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: November 11, 2020 Received: November 12, 2020
## Dear Elizabeth O'Keeffe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For: Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement on last page.
510(k) Number (if known)
K200376
Device Name Navbit® Sprint
Indications for Use (Describe)
### For Lateral Patient Registration:
The Navbit® Sprint System is a computer-controlled system intended to assist the surgeon in relation to reference axes during orthopedic surgical procedures. The Navbit® Sprint System facilitates the accurate positioning of implants, relative to these alignment axes.
The clinical setting and target population for the Navbit® Sprint System is that of a patient undergoing a Hip Arthroplasty surgical procedure by any approach with the patient in a lateral decubitus position.
The Navbit® Sprint System for lateral patient registration is indicated for use:
- . in Hip Arthroplasty surgical procedures;
- with acetabular cups that are uncemented and allow for post-impaction correction; .
- when post-impaction confirmatory measurement checks can be obtained.
### For Supine Patient Registration:
The Navbit® Sprint System is a computer-controlled system intended to assist the surgeon in relation to reference axes during orthopedic surgical procedures. The Navbit® Sprint System facilitates the accurate positioning of implants, relative to these alignment axes.
The clinical setting and target population for the Navbit® Sprint System is that of a patient undergoing a Hip Arthroplasty surgical procedure by any approach with the patient in a supine position.
The Navbit® Sprint System for supine patient registration is indicated for use:
- in Hip Arthroplasty surgical procedures; .
- with acetabular cups that are uncemented and allow for post-impaction correction; .
- . when post-impaction confirmatory measurement checks can be obtained.
Type of Use (Select one or both, as applicable)
図 Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
| This section applies only to requirements of the Paperwork Reduction Act of 1995. |
|-----------------------------------------------------------------------------------------------------------------|
| *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* |
| The burden time for this collection of information is estimated to average 79 hours per response, including the |
| time to review instructions, search existing data sources, gather and maintain the data needed and complete |
| and review the collection of information. Send comments regarding this burden estimate or any other aspect |
| of this information collection, including suggestions for reducing this burden, to: |
| Department of Health and Human Services |
| Food and Druq Administration |
| Office of Chief Information Officer |
| Paperwork Reduction Act (PRA) Staff |
| PRAStaff(@fda.hhs.gov |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of |
| information unless it displays a currently valid OMB number." |
| |
FORM FDA 3881 (7/17)
{3}------------------------------------------------
# 5. 510(k) Summary
| 510(k) Summary - Navbit® Sprint | | | |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Required | By section 21 CFR 807.92 (c) | | |
| Date Prepared | 14th February, 2020 | | |
| Owners Name | David Thomson QA/RA Manager | | |
| Submitter | Navbit Ltd<br>Suite 1A, Level 1, 136 Willoughby Road<br>Crows Nest 2065, Australia | | |
| Official Contact | Secure BioMed Evaluations<br>Elizabeth O'Keeffe, Ph.D.<br>7828 Hickory Flat Highway<br>Suite 120<br>Woodstock, GA 30188<br>770-837-2681 (direct)<br>855-MED-DEV1 (office)<br>Regulatory@SecureBME.com | | |
| Product Codes | OLO: Orthopedic Stereotaxic Instrument | | |
| Class | II | | |
| Classification Reference | 21 CFR 882.4560 | | |
| Common/Usual Name | Stereotaxic instrument. | | |
| Proprietary Name | Navbit® Sprint | | |
| Predicate Device(s) | OrthAlign Plus® System (K172462) | | |
| Reason for submission | New Device | | |
{4}------------------------------------------------
### 5.1. Intended Use and Indications for Use:
## For Lateral Patient Registration:
The Navbit® Sprint System is a computer-controlled system intended to assist the surgeon in determining alignment in relation to reference axes during orthopedic surgical procedures. The Navbit® Sprint System facilitates the accurate positioning of implants, relative to these alignment axes.
The clinical setting and target population for the Navbit® Sprint System is that of a patient undergoing a Hip Arthroplasty surgical procedure by any approach with the patient in a lateral decubitus position.
The Navbit® Sprint System for lateral patient registration is indicated for use:
- in Hip Arthroplasty surgical procedures;
- . with acetabular cups that are uncemented and allow for post-impaction correction;
- . when post-impaction confirmatory measurement checks can be obtained.
## For Supine Patient Registration:
The Navbit® Sprint System is a computer-controlled system intended to assist the surgeon in determining alignment in relation to reference axes during orthopedic surgical procedures. The Navbit® Sprint System facilitates the accurate positioning of implants, relative to these alignment axes.
The clinical setting and target population for the Navbit® Sprint System is that of a patient undergoing a Hip Arthroplasty surgical procedure by any approach with the patient in a supine position.
The Navbit® Sprint System for supine patient registration is indicated for use:
- . in Hip Arthroplasty surgical procedures;
- . with acetabular cups that are uncemented and allow for post-impaction correction;
- when post-impaction confirmatory measurement checks can be obtained.
#### 5.2. Device Description
The Navbit® Sprint System is a computer assisted surgical navigation system for use in hip arthroplasty procedures. The Navbit® Sprint System utilizes a palm-sized computer module containing sensors including rate gyroscopes and accelerometers to generate real time angular measurements (inclination and anteversion angles) used to guide acetabular cup implantation during orthopedic procedures.
During hip arthroplasty procedures, the Navbit® Sprint System assists the surgeon in registering the pelvic coordinate system and determining the inclination angle and the anteversion angle of the introducer/impactor relative to the registered pelvic coordinate system.
{5}------------------------------------------------
The device requires registration to be performed in one of two acceptable patient positions - supine and lateral, and alignment guidelines are provided for each. Any surgical approach that allows this registration to be performed is appropriate for use with the Navbit® Sprint.
The Navbit® Sprint System is provided terminally sterilized, single-use disposable.
### 5.3. Characteristics between predicate and new device
| Trait | Subject Device: Navbit® Sprint | Predicate Device: OrthAlign Plus® | Comparison |
|--------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | K200376 | K172462<br>(prior clearance as K130387) | N/A |
| FDA Regulation | 822.4560 | 822.4560 | Equivalent |
| Product Code | OLO | OLO | Equivalent |
| Product<br>Classification | Class II | Class II | Equivalent |
| Use | Prescription Use / Rx Only<br>Part 21 CFR 801 Subpart D | Prescription Use / Rx Only<br>Part 21 CFR 801 Subpart D | Equivalent |
| Indications for<br>Use | The Navbit® Sprint System is a<br>computer-controlled system intended<br>to assist the surgeon in determining<br>alignment in relation to reference axes<br>during orthopedic surgical procedures.<br>The Navbit® Sprint System facilitates<br>the accurate positioning of implants,<br>relative to these alignment axes.<br>The clinical setting and target<br>population for the Navbit® Sprint<br>System is that of a patient undergoing<br>a Hip Arthroplasty surgical procedure<br>by any approach with the patient in a<br>lateral decubitus /supine position.<br>The Navbit® Sprint System for<br>lateral/supine patient registration is<br>indicated for use:<br>in Hip Arthroplasty surgical<br>procedures; with acetabular cups that are<br>uncemented and allow for<br>post-impaction correction; when post-impaction<br>confirmatory measurement<br>checks can be obtained. | The OrthAlign Plus® System is a<br>computer-controlled system intended<br>to assist to the surgeon in determining<br>reference alignment axes in relation to<br>anatomical and instrumentation<br>structures during stereotactic<br>orthopedic surgical procedures. The<br>OrthAlign Plus® System facilitates the<br>accurate positioning of implants,<br>relative to these alignment axes. The<br>system aids the surgeon in controlling<br>leg length and offset discrepancies in<br>Total Hip Arthroplasty. Ligament<br>balancing is provided by the OrthAlign<br>Plus® System in primary revision Total<br>Knee Arthroplasty.<br>Example orthopedic surgical<br>procedures include but are not limited<br>to:<br>Total Knee Arthroplasty Total Hip Arthroplasty:<br>Anterior/Posterior Unicompartmental Knee<br>Arthroplasty: Tibial transverse<br>resection Ligament Balancing | Equivalent<br>Navbit does not<br>perform functions<br>associated with<br>knees, leg length<br>and offset |
| Trait | Subject Device: Navbit® Sprint | Predicate Device: OrthAlign Plus® | Comparison |
| Principle of<br>Operation | Computer-assisted navigation system | Computer-assisted navigation system | Equivalent |
| Single Use | Yes | Yes | Equivalent |
| Location of Use | Hospital / Operating Theatre | Hospital / Operating Theatre | Equivalent |
| Biocompatibility | Per ISO 10993-1, Externally<br>Communicating Device,<br>Tissue/Bone/Dentin Communicating,<br>with subsystems that have direct and<br>potential indirect contact for a limited<br>contact duration (< 24 hours) | Per ISO 10993-1, Externally<br>Communicating Device,<br>Tissue/Bone/Dentin Communicating,<br>with subsystems that have direct and<br>potential indirect contact for a limited<br>contact duration (< 24 hours) | Equivalent |
| Materials | Stainless Steel grades common to orthopedic surgical instruments Polymer grades common to orthopedic surgical instruments Internal electronics | Stainless Steel grades common to orthopedic surgical instruments Polymer grades common to orthopedic surgical instruments Internal electronics | Equivalent |
| Control<br>Mechanism | Computer generation of positional<br>information, using inertial sensors,<br>microcontroller, digital signal<br>processor and physical positions of<br>registration instruments. | Computer generation of positional<br>information, using inertial sensors,<br>microcontroller, digital signal<br>processor and physical positions of<br>registration instruments. | Equivalent |
| Operating<br>Principles;<br>Registration of<br>Anatomy | Electronics attached to movable<br>instruments, placed in specified<br>procedural positions, on or in contact<br>with bony anatomy, for recording<br>sensor data. | Electronics attached to movable<br>instruments, placed in specified<br>procedural positions, on or in contact<br>with bony anatomy, for recording<br>sensor data. | Equivalent |
| Measurement<br>and Navigation | Computer derived values. Accelerometer and rate gyroscope measurement of orientation of instrument. | Computer derived values. Accelerometer and rate gyroscope measurement of orientation of instrument. Linear encoder reading of probe position (for distance to determine position). | Equivalent<br>Navbit does not<br>support<br>positional<br>measurements |
| Main System<br>Components | Single-use computer unit Navigation and measurement software Single Use instrument sets (provided with disposable kit) | Single-use computer unit Navigation and measurement software Reusable instrument sets (provided as separate instrument set) | Equivalent |
| User Interface | Integrated graphical user interface, on<br>an electronic unit that attaches to<br>instrumentation | Integrated graphical user interface, on<br>an electronic unit that attaches to<br>instrumentation. | Equivalent |
| Energy Type | Two internal IEC-FR03 (ANSI 24-LF) Chemistry: Lithium/Iron Disulfide (Li/FeS2) Voltage: 3V (series) Storage Temp.: -40°C to 60°C Operating Temp.: -40°C to 60°C | One provided IEC-CR2, installed into Navigation Unit by user. Chemistry: Lithium/Manganese Dioxide (Li/MnO2) Voltage: 3V Storage Temp.: -40°C to 60°C Operating Temp.: -40°C to 60°C | Equivalent |
| Trait | Subject Device: Navbit® Sprint | Predicate Device: OrthAlign Plus® | Comparison |
| Sterilization | • Navigation unit: EO sterilization.<br>• Instruments (Pins): EO sterilization. | • Navigation unit: EO sterilization.<br>• Instruments: autoclave sterilization. | Equivalent<br>Navbit® Sprint<br>Device and Bone<br>Pins are supplied<br>sterile |
| Registration | Registration required using pins fixed to the hip. | Registration required using pins fixed to the hip. | Equivalent |
| Patient<br>Interface<br>(i.e. Bone Pins) | • Instrument fixation to pelvic bone with pins.<br>• Pin diameter of 4.0 mm, pin length of 125.0 mm<br>• 316L surgical Stainless Steel (ASTM 138) | • Instrument fixation to pelvic bone with pins and screws.<br>• Pin diameter 5/32" (4.0 mm), pin length 110.0 mm.<br>• 316L surgical Stainless Steel (ASTM 138) | Equivalent<br>Variations in pin<br>length dictated<br>by different<br>interfacing<br>equipment,<br>equivalent<br>patient/thread<br>engagement |
| Environmental<br>specifications | Specified storage and operating environments for typical transport and surgical environments | Specified storage and operating environments for typical transport and surgical environments. | Equivalent |
{6}------------------------------------------------
{7}------------------------------------------------
Table 1: Comparison between Predicate Device OrthAlign Plus® System (K172462) and Navbit® Sprint System
### 5.4. Substantial Equivalence
As a result of the Risk Analysis review and design input requirements, verification activities were performed on the Navbit® Sprint System. All tests confirmed the product met the predetermined acceptance criteria. This included testing against the FDA consensus standard, ASTM F2554-18 Standard Practice for Measurement of Positional Accuracy of Computer Assisted Surgical Systems. Navbit has determined the Navbit® Sprint System to be as safe and effective as the predicate device [OrthAlign Plus® System (K172462)].
### ર. ર. Performance Data
Device performance testing confirms that the Navbit® Sprint System can be used according to its intended use. The Navbit® Sprint System has been verified and validated according to Navbit procedures for product design and development. Performance testing addressed the functionality and surgical procedure workflows/steps as defined in the Navbit® Sprint user manual and demonstrates that the Navbit® Sprint System is as safe, as effective, and performs as well as the predicate device [OrthAlign Plus® System (K172462)].
### 5.6. Non-Clinical Testing
All non-clinical testing was performed to demonstrate the equivalence of the subject device, and all testing met the predetermined acceptance criteria.
Non-clinical testing included:
- . Software verification and validation to ensure the integrity of the code and functionality including reliability of the software in various use sequences.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "NAVBIT" in all caps. The font is a simple sans-serif font. The letters are evenly spaced and the word is centered.
- . System hardware verification/validation testing to ensure the Navbit® Sprint Device and its instruments meet their mechanical requirements.
- Instrumentation cleaning, sterilization and shipping validations or adoptions for the specified processes.
- . System components biocompatibility assessment per ISO 10993-1:2018.
- System accuracy testing: bench testing with mechanical fixtures to verify angular measurement accuracy using ASTM F2554-18.
- Simulated use (cadaver) testing of the Navbit® Sprint System with an advising surgeon to validate that the system meets the requirements for user needs in a simulated use environment provides validation evidence that the product met the predetermined clinical pass/fail criteria.
### 5.7. Clinical Testing
No clinical trial required. Simulated use testing on cadavers performed.
### 5.8. Animal Studies
No animal studies were performed for the Navbit® Sprint system.
### 5.9. Conclusion
The Navbit® Sprint System is substantially equivalent the predicate device OrthAlign Plus® System (K172462). Testing against the FDA consensus performance standard (ASTM F2554-18) and clinical validation using cadaver models has demonstrated that the Navbit® Sprint is as safe and as effective as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.