The Columbus Guidewire is intended for general intravascular use, including the neuro and peripheral vasculature. The Columbus Guidewire is intended to facilitate the selective placement of diagnostic or therapeutic catheters. This device is not intended for use in coronary arteries.
Device Story
Columbus Guidewire is a 0.014" diameter steerable guidewire featuring an in situ deflectable tip. Device consists of a Nitinol braided distal cable, proximal shaft, inner core wire, and handle. User operates handle to push/pull inner core wire, enabling two-direction tip bending. Available in two models: LR (4mm radius) and SR (2mm radius). Used in neuro and peripheral vasculature to facilitate selective placement of therapeutic/diagnostic catheters. Operated by physicians in clinical settings. Provides real-time tip control during navigation. Benefits include enhanced maneuverability and access to complex vascular anatomy compared to traditional manual-shaping guidewires.
Clinical Evidence
No human clinical data. Evidence includes bench testing (visual/dimensional, flexibility, torque, kink resistance, fracture, tensile, particulate, corrosion) and two GLP animal studies in 5 swine. Animal studies assessed usability, performance, and acute vascular safety (perforation, dissection, thrombus). Results showed no adverse events and performance comparable to predicate.
Indicated for general intravascular use, including neuro and peripheral vasculature, to facilitate selective placement of diagnostic or therapeutic catheters. Not for use in coronary arteries.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym followed by the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a solid blue square and the agency name in a lighter blue. The text reads "FDA U.S. FOOD & DRUG ADMINISTRATION".
Rapid-Medical Ltd. % Janice Hogan Partner Hogan Lovells US LLP 1735 Market Street, Floor 23 Philadelphia, Pennsylvania 19103 December 14, 2020
## Re: K200374
Trade/Device Name: Columbus Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: MOF, DQX Dated: November 12, 2020 Received: November 12, 2020
#### Dear Janice Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200374
Device Name Columbus Guidewire
Indications for Use (Describe)
The Columbus Guidewire is intended for general intravascular use, including the neuro and peripheral vasculature. The Columbus Guidewire is intended to facilitate the selective placement of therapeutic catheters. This device is not intended for use in coronary arteries.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary Device: Columbus Guidewire K200374
#### Submission Sponsor
Rapid-Medical Ltd. Carmel Building, POB 337 Yokneam, 2069205 Israel Company Phone No.: +972-72-250-3331
#### Contacts:
Dr. Orit Yaniv, VP of QA/RA Email: orit@rapid-medical.com
Ronen Eckhouse, CEO Email: ronen@rapid-medical.com
## Date Prepared
December 10, 2020
#### Device Identification
Trade/Proprietary Name: Columbus Guidewire Common/Usual Name: Columbus Guidewire Classification Name: Guide, Wire, Catheter, Neurovasculature Regulation Number: 21 CFR 870.1330 Product Code: MOF, DQX Device Class: II Classification Panel: Neurology
#### Legally Marketed Predicate Device(s)
Predicate Device: Traxcess 14 Guidewire (K133725)
#### Indication for Use Statement
The Columbus Guidewire is intended for general intravascular use, including the neuro and peripheral vasculature. The Columbus Guidewire is intended to facilitate the selective placement of diagnostic or therapeutic catheters. This device is not intended for use in coronary arteries.
{5}------------------------------------------------
## Device Description
The Columbus guidewire is a 0.014" diameter steerable guidewire with a deflectable tip to aid in accessing vasculature. The guidewire is supplied sterile (ETO sterilization) and is for single use only. The Columbus guide wire is comprised of a Nitinol braided flexible distal cable, a proximal shaft, an inner core wire and a handle. The braided cable is attached to the proximal shaft via inner connector. The inner core wire runs inside the shaft and the cable from the distal end of the cable to the handle. The distal end of the inner core wire is flattened, looped around and joined to the tip of the distal cable, forming a deflectable tip. In order to actuate the tip deflection in two directions, the Columbus guidewire handle contains a tube assembly section which enables continuous stroke by a self-locking feature. The handle is assembled to the proximal end of the core wire and controls the movement of the distal tip by pulling/pushing the inner moveable core wire, allowing the bending of the distal tip in two directions. Two models are available, Columbus LR (large radius) PN: GWPP4464 which has radius curvature of 4mm and Columbus SR (small radius) PN: GWPP4463 which has radius curvature of 2mm. The Columbus guidewire is provided with a torque accessory to facilitate use of the guidewire and is not intended to contact the patient's body.
## Substantial Equivalence Discussion
The table below compares the Columbus guidewire to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance testing. The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence. The Columbus guidewire does not raise any new questions compared to the predicate based on this comparison.
| | Columbus Guidewire<br>(Subject device) | Traxcess 14 Guidewire<br>(Predicate device) |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)<br>Number | K200374 | K133725 |
| Regulation | 21 CFR 870.1330 | 21 CFR 870.1330 |
| Product<br>Code | MOF, DQX | MOF, DQX |
| | Columbus Guidewire<br>(Subject device) | Traxcess 14 Guidewire<br>(Predicate device) |
| Indications for Use | The Columbus Guidewire is intended<br>for general intravascular use,<br>including the neuro and peripheral<br>vasculature. The Columbus<br>Guidewire is intended to facilitate the<br>selective placement of diagnostic or<br>therapeutic catheters. This device is<br> | The Traxcess 14 Guidewire is<br>intended for general intravascular<br>use, including the neuro and<br>peripheral vasculature. The<br>guidewire can be steered to<br>facilitate the selective placement of<br>diagnostic or therapeutic catheters.<br>This device is not intended for use<br>in coronary arteries. |
| Anatomical<br>Location | General intravascular use, including<br>the neuro and peripheral vasculature<br>but not coronary arteries | General intravascular use, including<br>the neuro and peripheral<br>vasculature but not coronary<br>arteries |
| Overall<br>Length | 200 cm | 200 cm |
| Proximal<br>Diameter | 0.014" | 0.014" |
| Distal<br>Diameter | 0.014" | 0.012" |
| Distal<br>Section<br>Length | 43 cm | 60 cm |
| Core Wire<br>Material | Nitinol | Nitinol |
| Proximal<br>section<br>Material | Stainless Steel | Stainless Steel |
| Coil Length | 43 cm | 40 cm |
| Coil Material | Nitinol and Nitinol DFT 30%<br>tantalum/ DFT 40% Platinum | Stainless steel and platinum |
| Shapeable<br>Tip Length | 2 cm (Columbus LR GWPP4464)<br>1.5 cm (Columbus SR GWPP4463) | 1.4 cm |
| Radiopaque<br>Tip Length | 2 cm (Columbus LR GWPP4464)<br>1.5 cm (Columbus SR GWPP4463) | 3 cm |
| | Columbus Guidewire<br>(Subject device) | Traxcess 14 Guidewire<br>(Predicate device) |
| Tip shaping | In situ tip deflection mechanism,<br>controlled by the user in the<br>proximal end of the device | Manual shaping of the tip prior to<br>device delivery |
| Coating | No coating | Hydrophilic coating<br>(SLIP-COAT) |
| Sterilization | Sterile | Sterile |
| Sterilization<br>Method | Ethylene oxide | Ethylene oxide |
| Single Use | Yes | Yes |
| Packaging | Placed into a Dispenser hoop, Tyvek<br>pouch, and Carton box | Placed into a Dispenser hoop,<br>Tyvek pouch, and Carton box |
{6}------------------------------------------------
{7}------------------------------------------------
#### Non-Clinical Performance Data
As part of demonstrating substantial equivalence to the predicate devices, Rapid-Medical Ltd. completed a number of non-clinical performance tests. The device meets all the requirements of overall design, sterilization, and biocompatibility. Testing results confirm that the design output meets the design inputs for the device.
#### Biocompatibility
The materials used in the manufacture of the subject device Columbus Guidewire, are identical to those used in the manufacturing of Rapid-Medical Neurovascular device, Comaneci Embolization Assist Device (DEN170064) that was granted on April The two devices share the same manufacture process and same 24th, 2019. manufacture environment. In addition, the two devices are intended to be used in the same anatomical locations, and identical in terms of frequency and duration of exposure.
Biocompatibility testing was completed for Comaneci device in accordance with ISO 10993 and consisted of the following tests: Cytotoxicity, Irritation (Intracutaeous reactivity), Sensitization, Hemocompatibility, Pyrogenicity, Acute Systemic and Toxicity Testing. Direct hemolysis, Direct contact complement activation, Ames Mutagenicity and In vivo Thrombogenicity tests were performed directly with the subject device. All tests confirmed biocompatibility.
#### Sterilization and Shelf life
The device and its accessories, are sterilized by 100% Ethylene Oxide and have been adopted into a validated sterilization process in accordance with the principles of EN ISO 11135: 2014/AC: 2014 ("Sterilization of health-care products - Ethylene oxide -
{8}------------------------------------------------
Requirements for the development, validation and routine control of a sterilization process for medical devices"). This statement is supported by standards assessment conducted by Rapid-Medical Ltd.
The device is non-pyrogenic, a complete endotoxin evaluation demonstrated endotoxin level below 2.15 EU/device supporting non-pyrogenicity, endotoxin testing will be conducted on every batch.
The accelerated shelf life testing for Columbus Guidewire has been conducted (T=2.5 years accelerated aging) with test results confirmed that all acceptance criteria were met.
## Bench Tests
The device passed all performance bench testing in accordance with internal requirements, national standards and international standards as shown in the table below to support substantial equivalence of the device. Non-Clinical Performance Tests
| Performance Bench Testing | | |
|----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Test | Standards | Results |
| Visual and<br>Dimensions<br>Verification | FDA Guidance for Industry and Food and Drug<br>Administration Staff: Coronary, Peripheral, and<br>Neurovascular Guidewires Performance Tests and<br>Recommended Labeling, October 2019 (thereinafter,<br>"FDA Guidewire Guidance"), Section G-3 | Pass |
| Tip Flexibility | FDA Guidewire Guidance, Section G-15 | Pass |
| Simulated Use-<br>Delivery and<br>Retrieval Force | FDA Guidewire Guidance, Section G-5 | Pass |
| Simulated Use<br>Model Testing and<br>Product<br>Compatibility | FDA Guidewire Guidance, Section G-5 | Pass |
| Torqueability | FDA Guidewire Guidance, Section G-9 | Pass |
| Kink Resistance | FDA Guidewire Guidance, Section G-14 | Pass |
| Fracture Test | ISO11070 section 8.4 | Pass |
| Flexing Test | ISO11070 section 8.5 | Pass |
| Tensile Force | FDA Guidewire Guidance, Section G-6 | Pass |
| Torque Strength | FDA Guidewire Guidance, Section G-8 | Pass |
| Particulate | FDA Guidewire Guidance, Section G-11 | Pass |
| Corrosion | FDA Guidewire Guidance, Section G-13 | Pass |
Additional two bench test were designed to support subject device in situ tip bending mechanism.
{9}------------------------------------------------
| Test | Description | Results |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Tip deflection force | The maximum tip deflection force was measured during complete deflection in a simulated vessel diameter range of 0.5-5.5mm to verify it is within justifiable safe range | Pass |
| Deflectable tip cyclic fatigue testing | Testing was conducted to determine the tip deflection mechanism durability to 30 full handle actuations | Pass |
## Pre-Clinical Animal Testing Data
As part of demonstrating the substantial equivalence to the predicate device, Rapid-Medical Ltd. completed two controlled GLP studies in domestic swine with the Columbus device and the predicate device. Usability, performance, and acute vascular safety were assessed in the first study, whereas the second study assessed acute safety and thromboresistance.
Acute procedural safety showed no device perforations or dissections or regional vascular thrombus assessed angiographically in either study. Regional devicecontacting thrombus was evaluated in accordance with recognized standards and guidelines in the second study and was also negative for both devices. Downstream macroscopic assessment of renal capsule, parenchyma, and native renal artery were also absent thrombi at the acute term assessment. In terms of usability and performance the following attributes were assessed: deliverability, visibility, controllability level, general impression, interface with accessory devices, radiopacity, intuitive actuation, device removal, ease of use, ergonomics, and device integrity. The usability and performance results of the first study demonstrate comparable performance for both devices. Therefore in 2 adequate and well controlled GLP studies, utilizing a total of 5 swine, assessments by well qualified experts in accordance with recognized methods, standards, and guidelines show safety, performance, usability, and thromboresistance of the Columbus that guidewire is substantially equivalent to Traxcess 14.
## Statement of Substantial Equivalence
The Columbus device has the same intended use and indications for use, and similar technological characteristics compared to the Traxcess 14 predicate device. The minor differences do not raise different questions compared to the predicate, and the bench
{10}------------------------------------------------
and animal testing discussed above demonstrates that these differences do not adversely impact performance. Therefore, the device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.