← Product Code [IYN](/productcode/IYN) · K200339

# HS40 Diagnostic Ultrasound System (K200339)

_Samsung Medison Co., Ltd. · IYN · Apr 7, 2020 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K200339

## Device Facts

- **Applicant:** Samsung Medison Co., Ltd.
- **Product Code:** [IYN](/productcode/IYN.md)
- **Decision Date:** Apr 7, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The HS40 ultrasound diagnostic system and probes are designed to obtain ultrasound images and analyze body fluids. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel. The HS40 diagnostic ultrasound system is intended for use by, or by the order of, and under the supervision of, a licensed physician who is qualified for direct use of medical devices.

## Device Story

HS40 is a mobile, software-controlled diagnostic ultrasound system. It acquires ultrasound data via various probes to display 2D, M-mode, Color/Power Doppler, PW/CW Spectral Doppler, Harmonic imaging, Tissue Doppler, Panoramic, Freehand 3D, and real-time 4D imaging. Used in clinical settings by physicians or under their supervision. The system provides measurement and analysis packages to assist in diagnosis. It displays real-time mechanical and thermal indices. The device integrates software features like S-Detect for Breast, 2D Follicle, LaborAssist, and CrystalVue. It supports Windows 10 OS. The output allows clinicians to visualize anatomical structures and blood flow, aiding in clinical decision-making and patient diagnosis.

## Clinical Evidence

No clinical studies were required to support substantial equivalence. Safety and effectiveness were demonstrated through non-clinical bench testing, including acoustic output measurement, biocompatibility, software validation, and compliance with electrical/electromagnetic safety standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-1, ISO 14971, NEMA UD 2-2004).

## Technological Characteristics

Mobile, software-controlled ultrasound system. Imaging modes: 2D, M-mode, Color/Power Doppler, PW/CW Spectral Doppler, Harmonic, Tissue Doppler, Panoramic, 3D/4D. Connectivity: Windows 10 OS. Safety: Real-time acoustic output display (mechanical/thermal indices). Sterilization: Probes compatible with cleaning/disinfection protocols. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-1, ISO 14971, NEMA UD 2-2004.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- HS40 Diagnostic Ultrasound System ([K191055](/device/K191055.md))
- HS50/HS60 Diagnostic Ultrasound System ([K192505](/device/K192505.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 7, 2020

Samsung Medison Co., LTD % Scully Kim Regulatory Affairs Specialist 3366, Hanseo-ro, Nam-myeon, Hongcheon-gun, Gangwon-du 25108 REPUBLIC OF KOREA

Re: K200339

Trade/Device Name: HS40 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: February 11, 2020 Received: February 11, 2020

Dear Scully Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K200339

Device Name HS40 Diagnostic Ultrasound System

### Indications for Use (Describe)

The HS40 diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids.

The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.

The HS40 diagnostic ultrasound system is intended for use by, or by the order of, and under the supervision of, a licensed physician who is qualified for direct use of medical devices.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

510(k) Premarket Notification - Traditional

Image /page/3/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold, sans-serif font. The word is set against a blue, oval-shaped background. 

### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT

510(k) No.:

Device Name: HS40 Diagnostic Ultrasound System

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               |   | Mode of Operation (*includes simultaneous B-mode) |     |     |                   |                      |                         |
|---------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|-------------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B | M                                                 | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)        |
| Ophthalmic                | Ophthalmic                    |   |                                                   |     |     |                   |                      |                         |
|                           | Fetal/Obstetrics (See Note 3) | P | P                                                 | P   | P   | P                 | Note 1               | Note 2, 4, 7, 8, 9, 11  |
|                           | Abdominal (See Note 10)       | P | P                                                 | P   | P   | P                 | Note 1               | Note 2, 7, 8, 9, 11     |
|                           | Intra-operative (See Note 6)  |   |                                                   |     |     |                   |                      |                         |
|                           | Intra-operative (Neuro.)      |   |                                                   |     |     |                   |                      |                         |
| Fetal Imaging<br>& Other  | Laparoscopic                  |   |                                                   |     |     |                   |                      |                         |
|                           | Pediatric                     | P | P                                                 | P   |     | P                 | Note 1               | Note 2, 7, 8, 9, 11     |
|                           | Small Organ (See Note 5)      | P | P                                                 | P   |     | P                 | Note 1               | Note 2, 7, 9, 11, 12    |
|                           | Neonatal Cephalic             | P | P                                                 | P   |     | P                 | Note 1               | Note 8, 9, 11           |
|                           | Adult Cephalic                | P | P                                                 | P   | P   | P                 | Note 1               | Note 7                  |
|                           | Trans-rectal                  | P | P                                                 | P   |     | P                 | Note 1               | Note 2, 7, 8, 9, 12     |
|                           | Trans-vaginal                 | P | P                                                 | P   |     | P                 | Note 1               | Note 2, 7, 8, 9, 12     |
|                           | Trans-urethral                |   |                                                   |     |     |                   |                      |                         |
|                           | Trans-esoph. (non-Cardiac)    |   |                                                   |     |     |                   |                      |                         |
|                           | Musculo-skel. (Convent.)      | P | P                                                 | P   |     | P                 | Note 1               | Note 2, 7, 8, 9, 11, 12 |
|                           | Musculo-skel. (Superfic.)     | P | P                                                 | P   |     | P                 | Note 1               | Note 2, 7, 9, 11        |
|                           | Intra-luminal                 |   |                                                   |     |     |                   |                      |                         |
|                           | Other (See Note 13)           | P | P                                                 | P   |     | P                 | Note 1               | Note 2, 7, 9, 12        |
|                           | Cardiac Adult                 | P | P                                                 | P   | P   | P                 | Note 1               | Note 4, 7, 14           |
| Cardiac                   | Cardiac Pediatric             | P | P                                                 | P   | P   | P                 | Note 1               | Note 4, 7, 14           |
|                           | Trans-esophageal (Cardiac)    |   |                                                   |     |     |                   |                      |                         |
|                           | Other (spec.)                 |   |                                                   |     |     |                   |                      |                         |
| Peripheral<br>Vessel      | Peripheral vessel             | P | P                                                 | P   | P   | P                 | Note 1               | Note 2, 7, 8, 9, 11     |
|                           | Other (spec.)                 |   |                                                   |     |     |                   |                      |                         |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+C+PW, B+DPD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

- Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development

- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

- Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate Note 14: Tissue Doppler Imaging (TDI)

{4}------------------------------------------------

510(k) Premarket Notification - Traditional

Image /page/4/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The wordmark is set against a blue, rounded, oval-shaped background. 

## DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT

510(k) No.:

## Device Name: LA3-16AD for use with HS40

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                      |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|----------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)     |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                      |
| Fetal Imaging<br>& Other  | Fetal/Obstetrics (See Note 3) |                                                   |   |     |     |                   |                      |                      |
|                           | Abdominal (See Note 10)       | P                                                 | P | P   |     | P                 | Note 1               | Note 2, 7, 9, 11     |
|                           | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                      |
|                           | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                      |
|                           | Laparoscopic                  |                                                   |   |     |     |                   |                      |                      |
|                           | Pediatric                     | P                                                 | P | P   |     | P                 | Note 1               | Note 2, 7, 9, 11     |
|                           | Small Organ (See Note 5)      | P                                                 | P | P   |     | P                 | Note 1               | Note 2, 7, 9, 11, 12 |
|                           | Neonatal Cephalic             |                                                   |   |     |     |                   |                      |                      |
|                           | Adult Cephalic                |                                                   |   |     |     |                   |                      |                      |
|                           | Trans-rectal                  |                                                   |   |     |     |                   |                      |                      |
|                           | Trans-vaginal                 |                                                   |   |     |     |                   |                      |                      |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                      |
|                           | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                      |
|                           | Musculo-skel. (Convent.)      | P                                                 | P | P   |     | P                 | Note 1               | Note 2, 7, 9, 11, 12 |
|                           | Musculo-skel. (Superfic.)     | P                                                 | P | P   |     | P                 | Note 1               | Note 2, 7, 9, 11     |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                      |
|                           | Other (See Note 13)           |                                                   |   |     |     |                   |                      |                      |
| Cardiac                   | Cardiac Adult                 |                                                   |   |     |     |                   |                      |                      |
|                           | Cardiac Pediatric             |                                                   |   |     |     |                   |                      |                      |
|                           | Trans-esophageal (Cardiac)    |                                                   |   |     |     |                   |                      |                      |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                      |
| Peripheral<br>Vessel      | Peripheral vessel             | P                                                 | P | P   |     | P                 | Note 1               | Note 2, 5, 7, 9, 11  |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                      |

N= new indication; P= previously cleared by FDA K191055 ; E= added under Appendix E

#### Additional Comments:

Note 1: B+M, B+PW, B+CW, B+C, B+PD, B+Elastoscan, B+C+PW, B+DD+PW, B+DPD+PW, B+TD+PW, B+C+M,

B+C+CW, B+PPI, B+PPI+PW, Dual/Quad (B, B+C, B+PD, B+TD, B+DPD)

- Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate
- Note 14: Tissue Doppler Imaging (TDI)

Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)

{5}------------------------------------------------

510(k) Premarket Notification - Traditional

Image /page/5/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold, sans-serif font. The word is set against a blue, oval-shaped background. 

### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT

510(k) No.:

Device Name: CA2-8AD for use with HS40

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                    |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)   |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                    |
|                           | Fetal/Obstetrics (See Note 3) | P                                                 | P | P   | P   | P                 | Note 1               | Note 2,4, 7, 9, 11 |
|                           | Abdominal (See Note 10)       | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 7, 9, 11   |
|                           | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                    |
|                           | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                    |
| Fetal Imaging             | Laparoscopic                  |                                                   |   |     |     |                   |                      |                    |
| &<br>Other                | Pediatric                     | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 7, 9, 11   |
|                           | Small Organ (See Note 5)      |                                                   |   |     |     |                   |                      |                    |
|                           | Neonatal Cephalic             |                                                   |   |     |     |                   |                      |                    |
|                           | Adult Cephalic                |                                                   |   |     |     |                   |                      |                    |
|                           | Trans-rectal                  |                                                   |   |     |     |                   |                      |                    |
|                           | Trans-vaginal                 |                                                   |   |     |     |                   |                      |                    |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                    |
|                           | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                    |
|                           | Musculo-skel. (Convent.)      | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 7, 9, 11   |
|                           | Musculo-skel. (Superfic.)     |                                                   |   |     |     |                   |                      |                    |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                    |
|                           | Other (See Note 13)           |                                                   |   |     |     |                   |                      |                    |
|                           | Cardiac Adult                 |                                                   |   |     |     |                   |                      |                    |
| Cardiac                   | Cardiac Pediatric             |                                                   |   |     |     |                   |                      |                    |
|                           | Trans-esophageal (Cardiac)    |                                                   |   |     |     |                   |                      |                    |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                    |
| Peripheral                | Peripheral vessel             | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 7, 9, 11   |
| Vessel                    | Other (spec.)                 |                                                   |   |     |     |                   |                      |                    |

N= new indication; P= previously cleared by FDA K191055; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+C+PW, B+DPD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate Note 14: Tissue Doppler Imaging (TDI)

Note 2: Includes imaging for guidance of biopsy

Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)

{6}------------------------------------------------

510(k) Premarket Notification - Traditional

Image /page/6/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold, sans-serif font. The word is set against a blue, oval-shaped background. 

### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT

510(k) No.:

Device Name: CF4-9 for use with HS40

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                   |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|-------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)  |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                   |
|                           | Fetal/Obstetrics (See Note 3) | P                                                 | P | P   |     | P                 | Note 1               | Notes 7, 8, 9, 11 |
|                           | Abdominal (See Note 10)       | P                                                 | P | P   |     | P                 | Note 1               | Notes 7, 8, 9, 11 |
|                           | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                   |
|                           | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                   |
| Fetal Imaging<br>& Other  | Laparoscopic                  |                                                   |   |     |     |                   |                      |                   |
|                           | Pediatric                     | P                                                 | P | P   |     | P                 | Note 1               | Notes 7, 8, 9, 11 |
|                           | Small Organ (See Note 5)      |                                                   |   |     |     |                   |                      |                   |
|                           | Neonatal Cephalic             | P                                                 | P | P   |     | P                 | Note 1               | Notes 7, 8, 9, 11 |
|                           | Adult Cephalic                |                                                   |   |     |     |                   |                      |                   |
|                           | Trans-rectal                  |                                                   |   |     |     |                   |                      |                   |
|                           | Trans-vaginal                 |                                                   |   |     |     |                   |                      |                   |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                   |
|                           | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                   |
|                           | Musculo-skel. (Convent.)      | P                                                 | P | P   |     | P                 | Note 1               | Notes 7, 8, 9, 11 |
|                           | Musculo-skel. (Superfic.)     |                                                   |   |     |     |                   |                      |                   |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                   |
|                           | Other (See Note 13)           |                                                   |   |     |     |                   |                      |                   |
|                           | Cardiac Adult                 |                                                   |   |     |     |                   |                      |                   |
| Cardiac                   | Cardiac Pediatric             |                                                   |   |     |     |                   |                      |                   |
|                           | Trans-esophageal (Cardiac)    |                                                   |   |     |     |                   |                      |                   |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                   |
| Peripheral<br>Vessel      | Peripheral vessel             | P                                                 | P | P   |     | P                 | Note 1               | Notes 7, 8, 9, 11 |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                   |

N= new indication; P= previously cleared by FDA K191055; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+C+PW, B+DPD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

- Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis

- Note 11: Panoramic imaging
Note 12: ElastoScan

Note 13: Includes Urology/Prostate Note 14: Tissue Doppler Imaging (TDI)

{7}------------------------------------------------

510(k) Premarket Notification - Traditional

Image /page/7/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The background is a blue, rounded oval shape.

## DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT

510(k) No.:

Device Name: PN2-4 for use with HS40

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                  |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                  |
|                           | Fetal/Obstetrics (See Note 3) |                                                   |   |     |     |                   |                      |                  |
|                           | Abdominal (See Note 10)       | P                                                 | P | P   |     | P                 | Note 1               | Note 7           |
|                           | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                  |
| Fetal Imaging<br>& Other  | Laparoscopic                  |                                                   |   |     |     |                   |                      |                  |
|                           | Pediatric                     |                                                   |   |     |     |                   |                      |                  |
|                           | Small Organ (See Note 5)      |                                                   |   |     |     |                   |                      |                  |
|                           | Neonatal Cephalic             |                                                   |   |     |     |                   |                      |                  |
|                           | Adult Cephalic                | P                                                 | P | P   |     | P                 | Note 1               | Note 7           |
|                           | Trans-rectal (See Note 13)    |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-vaginal (See Note 13)   |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Convent.)      |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Superfic.)     |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                  |
|                           | Other (See Note 13)           |                                                   |   |     |     |                   |                      |                  |
|                           | Cardiac Adult                 | P                                                 | P | P   |     | P                 | Note 1               | Note 4, 7, 14    |
| Cardiac                   | Cardiac Pediatric             | P                                                 | P | P   |     | P                 | Note 1               | Note 4, 7, 14    |
|                           | Trans-esophageal (Cardiac)    |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                  |
| Peripheral<br>Vessel      | Peripheral vessel             |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                  |

N= new indication; P= previously cleared by FDA K191055; E= added under Appendix E

#### Additional Comments:

Note 1: B+M, B+PW, B+CW, B+C, B+PD, B+Elastoscan, B+C+PW, B+DD+PW, B+DPD+PW, B+TD+PW, B+C+M,

B+C+CW, B+PPI, B+TD, B+PPI+PW, Dual/Quad (B, B+C, B+PD, B+TD, B+DPD)

- Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

- Note 11: Panoramic imaging
- Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: Tissue Doppler Imaging (TDI)

{8}------------------------------------------------

510(k) Premarket Notification - Traditional

Image /page/8/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The background is a blue, rounded oval shape.

### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT

510(k) No.:

Device Name: EVN4-9 for use with HS40

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                  |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                  |
| Fetal Imaging<br>& Other  | Fetal/Obstetrics (See Note 3) | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 7, 9     |
|                           | Abdominal (See Note 10)       | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 7, 9     |
|                           | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                  |
|                           | Laparoscopic                  |                                                   |   |     |     |                   |                      |                  |
| & Other                   | Pediatric                     |                                                   |   |     |     |                   |                      |                  |
|                           | Small Organ (See Note 5)      |                                                   |   |     |     |                   |                      |                  |
|                           | Neonatal Cephalic             |                                                   |   |     |     |                   |                      |                  |
|                           | Adult Cephalic                |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-rectal                  | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 7, 9, 12 |
|                           | Trans-vaginal                 | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 7, 9, 12 |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Convent.)      |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Superfic.)     |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.) (See Note 13)   | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 7, 9, 12 |
| Cardiac                   | Cardiac Adult                 |                                                   |   |     |     |                   |                      |                  |
|                           | Cardiac Pediatric             |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-esophageal (Cardiac)    |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                  |
| Peripheral<br>Vessel      | Peripheral vessel             |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                  |

N= new indication; P= previously cleared by FDA K191055; E= added under Appendix E

#### Additional Comments:

Note 1: B+M, B+PW, B+CW, B+C, B+PD, B+Elastoscan, B+C+PW, B+DD+PW, B+DPD+PW, B+TD+PW, B+C+M,

B+C+CW, B+PPI, B+TD, B+PPI+PW, Dual/Quad (B, B+C, B+PD, B+TD, B+DPD)

Note 3: Includes infertility monitoring of follicle development

- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

- Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

- Note 12: ElastoScan Note 13: Includes Urology/Prostate
Note 14: Tissue Doppler Imaging (TDI)

Note 2: Includes imaging for guidance of biopsy

{9}------------------------------------------------

510(k) Premarket Notification - Traditional

Image /page/9/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The background is a blue, rounded oval shape.

### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT

510(k) No.:

Device Name: VN4-8 for use with HS40

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                       |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|-----------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)      |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                       |
|                           | Fetal/Obstetrics (See Note 3) | P                                                 | P | P   | P   | P                 | Note 1               | Note 2,4, 7, 8, 9, 11 |
|                           | Abdominal (See Note 10)       | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 7, 8, 9, 11   |
|                           | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                       |
|                           | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                       |
| Fetal Imaging<br>& Other  | Laparoscopic                  |                                                   |   |     |     |                   |                      |                       |
|                           | Pediatric                     | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 7, 8, 9, 11   |
|                           | Small Organ (See Note 5)      |                                                   |   |     |     |                   |                      |                       |
|                           | Neonatal Cephalic             |                                                   |   |     |     |                   |                      |                       |
|                           | Adult Cephalic                |                                                   |   |     |     |                   |                      |                       |
|                           | Trans-rectal                  |                                                   |   |     |     |                   |                      |                       |
|                           | Trans-vaginal                 |                                                   |   |     |     |                   |                      |                       |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                       |
|                           | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                       |
|                           | Musculo-skel. (Convent.)      |                                                   |   |     |     |                   |                      |                       |
|                           | Musculo-skel. (Superfic.)     |                                                   |   |     |     |                   |                      |                       |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                       |
|                           | Other (See Note 13)           |                                                   |   |     |     |                   |                      |                       |
|                           | Cardiac Adult                 |                                                   |   |     |     |                   |                      |                       |
| Cardiac                   | Cardiac Pediatric             |                                                   |   |     |     |                   |                      |                       |
|                           | Trans-esophageal (Cardiac)    |                                                   |   |     |     |                   |                      |                       |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                       |
| Peripheral<br>Vessel      | Peripheral vessel             |                                                   |   |     |     |                   |                      |                       |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                       |

N= new indication; P= previously cleared by K191055; E= added under Appendix E

#### Additional Comments:

Note 1: B+M, B+PW, B+CW, B+C, B+PD, B+Elastoscan, B+C+PW, B+PD+PW, B+DPD+PW, B+TD+PW, B+C+M,

B+C+CW, B+PPI, B+TD, B+PPI+PW, Dual/Quad (B, B+C, B+PD, B+TD, B+DPD)

Note 3: Includes infertility monitoring of follicle development

- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

- Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)

- Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

- Note 12: ElastoScan
Note 13: Includes Urology/Prostate Note 14: Tissue Doppler Imaging (TDI)

Note 2: Includes imaging for guidance of biopsy

{10}------------------------------------------------

510(k) Premarket Notification - Traditional

Image /page/10/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold, sans-serif font. The word is placed inside a blue, oval-shaped background. 

### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT

510(k) No.:

Device Name: V5-9 for use with HS40

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               |   | Mode of Operation (*includes simultaneous B-mode) |     |     |                   |                      |                      |
|---------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|----------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B | M                                                 | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)     |
| Ophthalmic                | Ophthalmic                    |   |                                                   |     |     |                   |                      |                      |
|                           | Fetal/Obstetrics (See Note 3) | P | P                                                 | P   |     | P                 | Note 1               | Note 2, 7, 8, 9      |
|                           | Abdominal (See Note 10)       | P | P                                                 | P   |     | P                 | Note 1               | Note 2, 7, 8, 9, 12  |
|                           | Intra-operative (See Note 6)  |   |                                                   |     |     |                   |                      |                      |
|                           | Intra-operative (Neuro.)      |   |                                                   |     |     |                   |                      |                      |
| Fetal Imaging<br>& Other  | Laparoscopic                  |   |                                                   |     |     |                   |                      |                      |
|                           | Pediatric                     |   |                                                   |     |     |                   |                      |                      |
|                           | Small Organ (See Note 5)      |   |                                                   |     |     |                   |                      |                      |
|                           | Neonatal Cephalic             |   |                                                   |     |     |                   |                      |                      |
|                           | Adult Cephalic                |   |                                                   |     |     |                   |                      |                      |
|                           | Trans-rectal                  | P | P                                                 | P   |     | P                 | Note 1               | Note 2, 7, 8, 9, 12  |
|                           | Trans-vaginal                 | P | P                                                 | P   |     | P                 | Note 1               | Note 2, 7, 8, 9, 12  |
|                           | Trans-urethral                |   |                                                   |     |     |                   |                      |                      |
|                           | Trans-esoph. (non-Cardiac)    |   |                                                   |     |     |                   |                      |                      |
|                           | Musculo-skel. (Convent.)      |   |                                                   |     |     |                   |                      |                      |
|                           | Musculo-skel. (Superfic.)     |   |                                                   |     |     |                   |                      |                      |
|                           | Intra-luminal                 |   |                                                   |     |     |                   |                      |                      |
|                           | Other (See Note 13)           | P | P                                                 | P   |     | P                 | Note 1               | Notes 2, 7, 8, 9, 12 |
|                           | Cardiac Adult                 |   |                                                   |     |     |                   |                      |                      |
| Cardiac                   | Cardiac Pediatric             |   |                                                   |     |     |                   |                      |                      |
|                           | Trans-esophageal (Cardiac)    |   |                                                   |     |     |                   |                      |                      |
|                           | Other (spec.)                 |   |                                                   |     |     |                   |                      |                      |
| Peripheral<br>Vessel      | Peripheral vessel             |   |                                                   |     |     |                   |                      |                      |
|                           | Other (spec.)                 |   |                                                   |     |     |                   |                      |                      |

N= new indication; P= previously cleared by FDA K191055; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+C+PW, B+DPD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

- Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate Note 14: Tissue Doppler Imaging (TDI)

Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)

{11}------------------------------------------------

510(k) Premarket Notification - Traditional

Image /page/11/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold, sans-serif font. The word is placed inside a blue, oval-shaped background. 

### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT

510(k) No.:

#### Device Name: DP2B for use with HS40

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                  |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                  |
| Fetal Imaging<br>& Other  | Fetal/Obstetrics (See Note 3) |                                                   |   |     |     |                   |                      |                  |
|                           | Abdominal (See Note 12)       |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                  |
|                           | Laparoscopic                  |                                                   |   |     |     |                   |                      |                  |
|                           | Pediatric                     |                                                   |   |     |     |                   |                      |                  |
|                           | Small Organ (See Note 5)      |                                                   |   |     |     |                   |                      |                  |
|                           | Neonatal Cephalic             |                                                   |   |     |     |                   |                      |                  |
|                           | Adult Cephalic                |                                                   |   |     | P   |                   |                      |                  |
|                           | Trans-rectal                  |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-vaginal                 |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Convent.)      |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Superfic.)     |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                  |
|                           | Other (See Note 13)           |                                                   |   |     |     |                   |                      |                  |
| Cardiac                   | Cardiac Adult                 |                                                   |   |     | P   |                   |                      |                  |
|                           | Cardiac Pediatric             |                                                   |   |     | P   |                   |                      |                  |
|                           | Trans-esophageal (Cardiac)    |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                  |
| Peripheral<br>Vessel      | Peripheral vessel             |                                                   |   |     | P   |                   |                      |                  |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                  |

N= new indication; P= previously cleared by FDA K191055; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+C+PW, B+DPD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

- Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate Note 14: Tissue Doppler Imaging (TDI)

{12}------------------------------------------------

510(k) Premarket Notification - Traditional

Image /page/12/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold, sans-serif font. The word is set against a blue, oval-shaped background. 

### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT

510(k) No.:

Device Name: C2-8 for use with HS40

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                         |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|-------------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)        |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                         |
|                           | Fetal/Obstetrics (See Note 3) | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 4, 7,8, 9, 11  |
|                           | Abdominal (See Note 10)       | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 6, 7, 8, 9, 11 |
|                           | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                         |
|                           | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                         |
| Fetal Imaging<br>& Other  | Laparoscopic                  |                                                   |   |     |     |                   |                      |                         |
|                           | Pediatric                     | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 7, 8, 9, 11    |
|                           | Small Organ (See Note 5)      |                                                   |   |     |     |                   |                      |                         |
|                           | Neonatal Cephalic             |                                                   |   |     |     |                   |                      |                         |
|                           | Adult Cephalic                |                                                   |   |     |     |                   |                      |                         |
|                           | Trans-rectal                  |                                                   |   |     |     |                   |                      |                         |
|                           | Trans-vaginal                 |                                                   |   |     |     |                   |                      |                         |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                         |
|                           | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                         |
|                           | Musculo-skel. (Convent.)      |                                                   |   |     |     |                   |                      |                         |
|                           | Musculo-skel. (Superfic.)     |                                                   |   |     |     |                   |                      |                         |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                         |
|                           | Other (See Note 13)           |                                                   |   |     |     |                   |                      |                         |
|                           | Cardiac Adult                 |                                                   |   |     |     |                   |                      |                         |
| Cardiac                   | Cardiac Pediatric             |                                                   |   |     |     |                   |                      |                         |
|                           | Trans-esophageal (Cardiac)    |                                                   |   |     |     |                   |                      |                         |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                         |
| Peripheral<br>Vessel      | Peripheral vessel             |                                                   |   |     |     |                   |                      |                         |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                         |

N= new indication; P= previously cleared by FDA K191055; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+C+PW, B+DPD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging Note 12: ElastoScan

Note 12: ElastoScan
Note 12: Includes Ural.

Note 13: Includes Urology/Prostate Note 14: Tissue Doppler Imaging (TDI)

{13}------------------------------------------------

510(k) Premarket Notification - Traditional

Image /page/13/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold, sans-serif font. The word is placed inside a blue, oval-shaped background. 

### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT

510(k) No.:

Device Name: C2-5 for use with HS40

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                         |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|-------------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)        |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                         |
|                           | Fetal/Obstetrics (See Note 3) | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 4, 7, 8, 9, 11 |
|                           | Abdominal (See Note 10)       | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 6, 7, 8, 9, 11 |
|                           | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                         |
|                           | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                         |
| Fetal Imaging             | Laparoscopic                  |                                                   |   |     |     |                   |                      |                         |
| &<br>Other                | Pediatric                     | P                                                 | P | P   | P   | P                 | Note 1               | Notes 2, 7, 8, 9, 11    |
|                           | Small Organ (See Note 5)      |                                                   |   |     |     |                   |                      |                         |
|                           | Neonatal Cephalic             |                                                   |   |     |     |                   |                      |                         |
|                           | Adult Cephalic                |                                                   |   |     |     |                   |                      |                         |
|                           | Trans-rectal                  |                                                   |   |     |     |                   |                      |                         |
|                           | Trans-vaginal                 |                                                   |   |     |     |                   |                      |                         |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                         |
|                           | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                         |
|                           | Musculo-skel. (Convent.)      |                                                   |   |     |     |                   |                      |                         |
|                           | Musculo-skel. (Superfic.)     |                                                   |   |     |     |                   |                      |                         |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                         |
|                           | Other (See Note 13)           |                                                   |   |     |     |                   |                      |                         |
|                           | Cardiac Adult                 |                                                   |   |     |     |                   |                      |                         |
| Cardiac                   | Cardiac Pediatric             |                                                   |   |     |     |                   |                      |                         |
|                           | Trans-esophageal (Cardiac)    |                                                   |   |     |     |                   |                      |                         |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                         |
| Peripheral                | Peripheral vessel             |                                                   |   |     |     |                   |                      |                         |
| Vessel                    | Other (spec.)                 |                                                   |   |     |     |                   |                      |                         |

N= new indication; P= previously cleared by FDA K191055; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+TD, B+C+PW, B+DPD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

- Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development

- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmo…

---

**Source:** [https://fda.innolitics.com/device/K200339](https://fda.innolitics.com/device/K200339)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
