← Product Code [FMF](/productcode/FMF) · K200325

# Orbit Subretinal Delivery System (K200325)

_Orbit Biomedical · FMF · Jul 14, 2020 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K200325

## Device Facts

- **Applicant:** Orbit Biomedical
- **Product Code:** [FMF](/productcode/FMF.md)
- **Decision Date:** Jul 14, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5860
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The Orbit Subretinal Delivery System is indicated for microinjection into the subretinal space.

## Device Story

The Orbit Subretinal Delivery System is a manual ophthalmic surgical instrument used for subretinal microinjections. It consists of a Subretinal Injection Device (SID) set, a Tubing Set, and a Dosing Set. The SID features a flexible ophthalmic cannula with an advanceable needle and an internal magnet. During surgery, a magnetic pad is adhered to a sterile drape on the patient's forehead to provide stability for the SID. The device connects to a vitreoretinal surgical console via the Tubing Set for pneumatic injection or is operated manually. The procedure involves two injections: the first creates an 'entry bleb' using BSS or BSS PLUS, and the second delivers a specific treatment dose into that bleb. The system is used by surgeons in an operating room setting. It eliminates the need for reusable components found in previous versions, providing a fully sterile, single-use solution to facilitate precise subretinal access and fluid delivery.

## Clinical Evidence

Bench testing only. No clinical data provided. Verification and validation included biocompatibility (ISO 10993-5, -10, -11), sterilization validation (ISO 11135, SAL 10^-6), EO/ECH residuals (ISO 10993-7), packaging integrity, accelerated shelf-life testing, and simulated use testing to confirm the ability to form a subretinal bleb. All tests met pre-established acceptance criteria.

## Technological Characteristics

Manual ophthalmic surgical instrument. Components: flexible cannula, magnetic pad with steel core, tubing, and syringes. Materials: biocompatible per ISO 10993-01. Sterilization: Ethylene Oxide (ISO 11135). Connectivity: interfaces with vitreoretinal surgical consoles via CPC connectors for pneumatic fluid control. Single-use, sterile. No software or electronic algorithms.

## Regulatory Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

## Predicate Devices

- Orbit Subretinal Delivery System ([K182274](/device/K182274.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

July 14, 2020

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Orbit Biomedical % Debe Deck Regulatory Consultant ClinReg Consulting Services, Inc. 733 Bolsana Drive Laguna Beach, California 92651

#### Re: K200325

Trade/Device Name: Orbit Subretinal Delivery System Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF, HMX Dated: February 7, 2020 Received: February 10, 2020

#### Dear Debe Deck:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and probibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Elvin Ng Acting Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K200325

Device Name Orbit Subretinal Delivery System

Indications for Use (Describe)

The Orbit Subretinal Delivery System is indicated for microinjection into the subretinal space.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(a).

| APPLICANT:                       | Orbit Biomedical, Inc.<br>300 Brookside Avenue<br>Building 18, Ste 180<br>Ambler, PA 19002<br>(267) 523-2100 |
|----------------------------------|--------------------------------------------------------------------------------------------------------------|
| OFFICIAL CORRESPONDENT:          | Debe Deck<br>ClinReg Consulting Services, Inc.<br>(949) 375-4387<br>debe@clinregconsulting.com               |
| DATE SUMMARY PREPARED:           | February 07, 2020                                                                                            |
| TRADE NAME:                      | Orbit Subretinal Delivery System                                                                             |
| COMMON NAME:                     | Manual Ophthalmic Surgical Instrument                                                                        |
| DEVICE CLASSIFICATION /<br>CODE: | FMF, Syringe, Piston<br>Class II<br>21 CFR 880.5860                                                          |
| SECONDARY PRODUCT CODE:          | HMX, Ophthalmic Cannula<br>Class I<br>21 CFR 886.4350                                                        |
| PREDICATE DEVICE:                | K182274<br>Orbit Subretinal Delivery System<br>FMF, Syringe, Piston<br>Class II<br>21 CFR 880.5860           |

## DEVICE DESCRIPTION

The Orbit Subretinal Delivery System (Figure 1) is designed for microinjection into the subretinal space. The Orbit Subretinal Delivery System (SDS) is comprised of 3 primary component sets including the Subretinal Injection Device Set, the Tubing Set, and the Dosing Set.

{4}------------------------------------------------

Image /page/4/Picture/3 description: The image shows a collection of medical devices and packaging. There are two syringes with tubing attached, a small bottle labeled "orbit SDS", and a white, rounded rectangular package labeled "orbit Subretinal Delivery". One of the syringes has a white and gray handle with the word "dose" printed on the side. The other syringe has a blue band around the tubing connection and also has the word "dose" printed on the side.

Figure 1: Orbit Subretinal Delivery System

Each Orbit SDS Set contains sterile single-use only components. The Subretinal Injection Device set includes the magnetic pad, ophthalmic marker, and subretinal injection device (SID). The Tubing Set includes the components for priming the BSS line and pneumatic BSS control via a vitreoretinal surgical console. The Dosing Set includes the syringe for delivery of a precise dose of Balanced Salt Solution (BSS) or BSS PLUS® infusate. Each of the set components are described below, along with their principles of operation within the context of the intended use.

Subretinal Injection Device (SID) Set: A sterile, flexible ophthalmic cannula with an advanceable needle is used for subretinal access via the suprachoroidal space. The Subretinal Injection Device allows for consecutive injections. The first subretinal injection delivers BSS or BSS PLUS to create an "entry bleb" in the desired subretinal location. The second subretinal injection delivers a specific treatment "dose" (i.e., quantity) of BSS or BSS PLUS into the entry bleb that was created with the first injection. The SID is fitted with an internal magnet designed to allow for secure, simple attachment to the magnetic pad that adheres to a sterile drape that is placed onto the patient's forehead during preparation.

Magnetic Pad: The SID Set includes an adhesive-backed magnet pad with a flexible steel core. The magnet inside the SID is attracted to the steel core, allowing for secure placement and positioning. The magnetic pad is adhered directly to the sterile drape that has been securely placed on the patient's forehead.

Ophthalmic Marker: A custom sterile ophthalmic marker is provided in the SID Set and is intended to be used with a commercially-available sterile gentian violet ink marking pen. The marker is used to stain the sclera for positioning of suture loops and scleral incision.

Tubing Set: A Tubing Set is supplied to allow the user the ability to connect to a commerciallyavailable vitreoretinal surgical console for pneumatic injection (vs. manual injection). The

{5}------------------------------------------------

Tubing Set includes the Tubing Assembly, a sterile 1 mL syringe (i.e., the BSS Syringe) that is used to create an entry bleb and two (2) Snap Collars that connect the sterile Tubing Assembly to the BSS Syringe. The proximal end of the Tubing Assembly is fitted with a CPC connector. A female to male CPC adapter is also included. The CPC connectors interface with the viscous fluid control (VFC) functionality of vitreoretinal surgical consoles.

Dosing Set: The Dosing Set includes the Dose Syringe used to deliver the BSS or BSS PLUS treatment dose, and two (2) Tubing Clamps.

Together, the Subretinal Injection Device, Tubing, and Dosing Sets are used to deliver the infusate (e.g., BSS, BSS PLUS) to the subretinal space as described in the Instructions for Use.

## INDICATIONS FOR USE

The Orbit Subretinal Delivery System is indicated for microinjection into the subretinal space.

## TECHNOLOGICAL CHARACTERISTICS COMPARISON

Technological characteristics of the Orbit Subretinal Delivery System described in this 510(k) are substantially equivalent to the predicate device (Orbit Subretinal Delivery Device, K182274) that share the same mode of operation in the delivery of fluid into the subretinal space.

The predicate device (Orbit Subretinal Delivery System cleared under K182274) included a nonsterile reusable Third Arm Kit that functioned as a "third arm" to hold the subretinal injection cannula housing stable during the procedure. The Third Arm Kit required cleaning and sterilization prior to each use. One of the primary modifications to the Orbit Subretinal Delivery System proposed in this 510(k) is the elimination of the reusable Third Arm Kit, replacing it with components in the single-use sterile Subretinal Injection Device (SID) Set and Dosing Set.

The proposed Orbit SDS eliminates all reusable components and uses an internal magnet (encased within the SID housing) to provide stability. During setup the magnetic pad is adhered directly on a sterile drape, securely positioned on the patient's forehead. The SID is then placed on the magnetic pad.

Other modifications are being made to accommodate the improved device design, support material procurement, and to improve handling during the manufacturing process.

Table 1 on the following page provides additional comparison between the components of the proposed Orbit Subretinal Delivery System and the predicate device.

{6}------------------------------------------------

| 1                             |  |
|-------------------------------|--|
|                               |  |
|                               |  |
| the I should I should all don |  |
|                               |  |
|                               |  |
| 1                             |  |
|                               |  |
|                               |  |
| Separe The Same               |  |
|                               |  |
| 2                             |  |
|                               |  |
|                               |  |
|                               |  |
|                               |  |
|                               |  |
|                               |  |
|                               |  |
|                               |  |

HNOLOGICAL COMPARISON OF THE ORBIT SUBRETINAL DELIVERY SYSTEM TO THE PREDICATE DEVI

| Feature                                  | Orbit Subretinal Delivery System<br>K182274                                                                           | Orbit Subretinal Delivery System<br>Proposed                                                          | Modified<br>Y/N                                                                                                                                                              |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                             | Orbit Biomedical, Inc.                                                                                                | Orbit Biomedical, Inc.                                                                                | No                                                                                                                                                                           |
| Device Classification/Code               | Class II<br>FMF, Syringe, Piston<br>21 CFR 880.5860                                                                   | Class II<br>FMF, Syringe, Piston<br>21 CFR 880.5860                                                   | No                                                                                                                                                                           |
| Secondary Product<br>Classification/Code | Class I<br>HMX, Ophthalmic Cannula<br>21 CFR 886.4350                                                                 | Class I<br>HMX, Ophthalmic Cannula<br>21 CFR 886.4350                                                 | No                                                                                                                                                                           |
| Indications for Use                      | The Orbit Subretinal Delivery System is<br>indicated for microinjection into the<br>subretinal space.                 | The Orbit Subretinal Delivery System is<br>indicated for microinjection into the<br>subretinal space. | No                                                                                                                                                                           |
| Principle of Operation                   | Manual operation of syringe for<br>delivery of infusate                                                               | Manual operation of syringe for<br>delivery of infusate                                               | No                                                                                                                                                                           |
| How Supplied                             | Sterile, single use only<br>(subretinal injection cannula set,<br>tubing set, syringe set)                            | Sterile, single use only<br>(all components)                                                          | Yes, all components of the proposed<br>Orbit SDS are provided sterile and<br>intended for single use only.                                                                   |
| Method of Sterilization                  | Ethylene Oxide Sterilization<br>(syringe set); Gamma Irradiation<br>(subretinal injection cannula set,<br>tubing set) | Ethylene Oxide sterilization<br>(all components)                                                      | The reusable third arm assembly from<br>the predicate device configuration has<br>been eliminated in the proposed device.<br>Yes, all components are now<br>sterilized by EO |

{7}------------------------------------------------

#### PERFORMANCE DATA

The descriptive characteristics of the Orbit Subretinal Delivery System (SDS) are well-defined and adequate to ensure equivalence to the predicate device. Additionally, verification and validation testing of the Orbit SDS represents a comprehensive evaluation for suitability of use.

The following tests were successfully performed with the device components to establish substantial equivalence of the Orbit SDS to the predicate device:

- o Biocompatibility testing in accordance with ISO 10993-01 including Cytotoxicity (per ISO10993-5 and USP<87>), Sensitization (per ISO10993-10), Ocular Irritation (per ISO10993-10), Systemic toxicity (per ISO10993-11) and Material Mediated Pyrogenicity (per ISO10993-11).
- Sterilization conditions were validated for the Orbit Subretinal Delivery 0 System to provide a Sterility Assurance Level of 106, in accordance with ISO11135. EO and ECH residuals testing was also performed in accordance with ISO10993-7. Testing demonstrated product performance met all prior established acceptance criteria.
- Packaging qualification was performed as part of the transportation and 0 environmental conditioning studies for the device to demonstrate that whole package physical integrity requirements and seal integrity requirements for the modified Orbit SDS packaging were met.
- o Accelerated shelf life testing was conducted as part of the transportation and design verification test program for the sterile device to establish the expiration date for the Orbit SDS.
- o Post-distribution, post-accelerated aging performance testing for each set of the Orbit SDS was completed. Testing demonstrated that product performance met all acceptance criteria.
- Simulated use testing of the Orbit SDS was designed to confirm the injection of an infusate to form a subretinal bleb.

All tests passed their respective pre-established test criteria and demonstrate that the Orbit SDS performance is maintained following distribution.

## CONCLUSION

The Orbit Subretinal Delivery System meets all product design requirements and applicable standards. The device shares the same intended use and key technological characteristics as the predicate device. Therefore, the device has been shown to be substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K200325](https://fda.innolitics.com/device/K200325)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
