K200276 · Brain Sentinel, Inc. · POS · Feb 6, 2021 · Neurology
Device Facts
Record ID
K200276
Device Name
SPEAC System
Applicant
Brain Sentinel, Inc.
Product Code
POS · Neurology
Decision Date
Feb 6, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1580
Device Class
Class 2
Indications for Use
The SPEAC® System is intended for use as an adjunct to seizure monitoring in adults in the home or healthcare facilities during periods of rest. The non-EEG Physiological Signal Based Seizure Monitoring System continuously records and stores surface electromyographic (sEMG) data for subsequent review. Trained healthcare professionals may use the electrophysiological sEMG data during a post-hoc review, with other contextual data, to characterize upper-extremity motor activity (UEMA) ipsilateral to the device from other activity. Audio data recorded during seizure monitoring may be available for review by a trained healthcare professional. The device is to be used on the belly of the biceps muscle to analyze sEMG signals. When sEMG signal patterns associated with a unilateral, appendicular, tonic extension that could be associated with a GTC seizure are detected, the SPEAC System sends adjunctive alarms to alert caregivers. Adjunctive alarms may be disabled by a physician order while continuing to record sEMG data for subsequent review.
Device Story
The SPEAC System is a wireless, non-invasive, sEMG-based seizure monitoring and alerting system. It records sEMG signals (1,000 Hz) from the biceps brachii muscle. A proprietary algorithm identifies sustained sEMG contraction patterns (tonic phase and early clonic transition) associated with generalized tonic-clonic (GTC) seizures. The system transmits data to a Laptop Base Station and then to a secure, FedRAMP-certified Microsoft Azure cloud (Data Distribution System) for remote physician review. In the home or hospital, the device provides adjunctive alarms to caregivers when potential GTC-associated patterns are detected. Physicians use the recorded sEMG data, alongside other clinical context, to characterize upper-extremity motor activity. The system aids clinicians in differentiating seizure types and motor events. It is intended for use during rest; alarms can be disabled by physician order. The device benefits patients by providing an objective record of motor activity for post-hoc diagnostic review.
Clinical Evidence
Two prospective clinical trials (Study 1: CLN0055.001; Study 2: CLN0053.001) evaluated the ability of trained clinicians to characterize seizure events using sEMG data. Combined analysis of 243 events from three reviewers showed an overall accuracy of 83% (95% CI [0.78, 0.88]) for individual assessments and 86% (95% CI [0.77, 0.93]) for a committee-based approach. The study compared sEMG-based characterizations against vEEG interpretations by epileptologists. Results support the use of the device for characterizing upper-extremity motor activity.
Technological Characteristics
Wireless, non-invasive sEMG monitor; 1,000 Hz sampling rate; frequency bands 30-40 Hz, 130-240 Hz, 300-400 Hz. Dimensions 3.44" x 2.34" x 1.33"; mass 127g. Rechargeable battery. Connectivity via Wi-Fi to Laptop Base Station and cloud storage. Biocompatible materials (ISO 10993). Proprietary algorithm for GTC seizure pattern detection.
Indications for Use
Indicated for adult patients in home or healthcare facilities during rest for adjunct seizure monitoring. Used to record sEMG data for post-hoc review by trained healthcare professionals to characterize upper-extremity motor activity (UEMA) ipsilateral to the device. Not for use during physical activity; not a standalone seizure monitor; not for pediatric populations.
Regulatory Classification
Identification
A non-electroencephalogram (non-EEG) physiological signal based seizure monitoring system is a noninvasive prescription device that collects physiological signals other than EEG to identify physiological signals that may be associated with a seizure.
Special Controls
In combination with the general controls of the FD&C Act, the Non-EEG physiological signal based seizure monitoring system is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) The technical parameters of the device, hardware and software, must be fully characterized and include the following information:
(i) Hardware specifications must be provided. Appropriate verification, validation, and hazard analysis must be performed.
(ii) Software, including any proprietary algorithm(s) used by the device to achieve its intended use, must be described in detail in the Software Requirements Specification (SRS) and Software Design Specification (SDS). Appropriate software verification, validation, and hazard analysis must be performed.
(2) The patient-contacting components of the device must be demonstrated to be biocompatible.
(3) The device must be designed and tested for electrical, thermal, and mechanical safety and electromagnetic compatibility (EMC).
(4) Clinical performance testing must demonstrate the ability of the device to function as an assessment aid for monitoring for seizure-related activity in the intended population and for the intended use setting. Performance measurements must include positive percent agreement and false alarm rate.
(5) Training must be provided for intended users that includes information regarding the proper use of the device and factors that may affect the collection of the physiologic data.
(6) The labeling must include health care professional labeling and patient-caregiver labeling. The health care professional and the patient-caregiver labeling must include the following information:
(i) A detailed summary of the clinical performance testing, including any adverse events and complications.
(ii) Any instructions technicians and clinicians should convey to patients and caregivers regarding the proper use of the device and factors that may affect the collection of the physiologic data.
(iii) Instructions to technicians and clinicians regarding how to set the device threshold to achieve the intended performance of the device.
Predicate Devices
Brain Sentinel Monitoring and Alerting System (K182180)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.
Brain Sentinel, Inc. Richard Waite Regulatory Consultant 8023 Vantage Drive Suite 216 San Antonio, Texas 78230
Re: K200276
Trade/Device Name: SPEAC System Regulation Number: 21 CFR 882.1580 Regulation Name: Non-Electroencephalogram (EEG) Physiological Signal Based Seizure Monitoring System Regulatory Class: Class II Product Code: POS Dated: January 22, 2021 Received: February 2, 2021
Dear Richard Waite:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200276
Device Name SPEAC System
#### Indications for Use (Describe)
The SPEAC® System is intended for use as an adjunct to seizure monitoring in adults in the home or healthcare facilities during periods of rest.
The non-EEG Physiological Signal Based Seizure Monitoring System continuously records and stores surface electromyographic (sEMG) data for subsequent review.
Trained healthcare professionals may use the electrophysiological sEMG data during a post-hoc review, with other contextual data, to characterize upper-extremity motor activity (UEMA) ipsilateral to the activity.
Audio data recorded during seizure monitoring may be available for review by a trained healthcare professional.
The device is to be used on the belly of the biceps muscle to analyze sEMG signals. When sEMG signal patterns associated with a unilateral, appendicular, tonic extension that could be associated with a GTC seizure are detected, the SPEAC System sends adjunctive alarms to alert caregivers.
Adjunctive alarms may be disabled by a physician order while continuing to record sEMG data for subsequent review.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">☑</span> |
|----------------------------------------------|-----------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | ☐ |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K200276 Traditional 510(k) Summarv
This 510(k) summary is prepared according to the elements outlined in 21 CFR 807.92 Content and format of a 510(k) summary.
| Date Prepared | February 05, 2020 |
|----------------|-------------------------------------------------------------------------------------------------|
| Submitter | Brain Sentinel, Inc<br>8023 Vantage Drive, Suite 216<br>San Antonio, TX 78230<br>(210) 951-8681 |
| Contact Person | Richard Waite, Sr.<br>(214) 662-9277<br>Richard Waite@outlook.com |
#### Subject Device K200276
| Trade Name | SPEAC® System |
|------------------------------|--------------------------------------------------------------------------------------|
| Common Name | Physiological signal-based seizure monitoring system |
| Device Classification | Class II |
| Product Code &<br>Regulation | POS; Non-EEG physiological signal-based seizure monitoring system<br>21 CFR 882.1580 |
| Review Panel | Neurodiagnostic Devices; Neurology |
#### Predicate Device K182180
Manufacturer Brain Sentinel, Inc. Device Name Brain Sentinel Monitoring and Alerting System Regulation Name Non-EEG physiological signal-based seizure monitoring system Product Code POS Regulation Number 21 CFR 882.1580 Device Classification Class II
#### Device Description
The SPEAC System is a wireless, non-invasive, physiological, surface electromyography (sEMG) recording, monitoring, and alerting system to be used as an adjunct to seizure monitoring during periods of rest. The System continuously records and stores surface electromyographic (sEMG) data for subsequent review by a physician. Trained healthcare professionals may use the electrophysiological sEMG data, with other contextual data, to characterize seizures with upper-extremity motor activity ipsilateral to the device from other
{4}------------------------------------------------
activity. SPEAC data gives healthcare professionals another diagnostic tool to characterize seizure events in a home or hospital setting.
The System continuously records and distributes sEMG data at 1,000 Hz (and audio around detected events) for post-hoc review by physicians (or other trained healthcare professionals) for the characterization of seizure events. A physician may perform post-hoc review of the SPEAC System data to characterize motor events that may be associated with seizures.
The seizure monitoring algorithm is able to send alarms to notify patients and caregivers when a pattern that may be associated with a generalized tonic-clonic (GTC) seizure is measured. Physicians may order the System with or without alarms and may order threshold adjustments to customize the level at which the System alarms.
Data collected by the System is uploaded to Brain Sentinel's secure remote storage, the Data Distribution System (DDS), and is remotely accessible for physician review. All patient data is cyber-secured within Microsoft Azure which is FedRAMP certified. Below, Tables 1 and 2 list the functional and operational outputs designed to provide feedback to the patient and caregiver. These modes and outputs are identical to those cleared in the predicate, K182180.
The SPEAC System remains the same with no alterations of any kind. The sEMG-based seizure monitoring algorithm is identical to the predicate SPEAC System. The purpose of this submission is to expand the indications for use based on clinical performance testing that was submitted to support a determination of substantial equivalence. When trained appropriately, clinicians may use the subject device to perform post-hoc analysis of the sEMG data from the device, with other contextual patient data, to characterize seizures with upper-extremity motor activity ipsilateral to the device from other activity.
{5}------------------------------------------------
Image /page/5/Figure/0 description: This image describes the SPEAC system functional modes. The table describes the seizure monitoring mode, seizure alarm mode, record only mode activated, loose electrode alert, sEMG monitor critical battery alert, laptop base station critical battery alert, and lost connection out of Wi-Fi range alert. The table also describes the left and right LED light colors, daily monitoring application screen view, and a description of the functional mode. For example, the seizure monitoring mode has green solid lights for both the left and right LEDs, and the daily monitoring application screen view shows a green checkmark with the text "You are being monitored".
## Table 1: System Functional and Operational Outputs
{6}------------------------------------------------
# Table 2: SPEAC System Outputs Intended for Patients and Caregivers
| Name | Application View | Description |
|-------------------------------|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Settings Button | Image: SETTINGS | Click the Settings button to swap functional modes between Seizure Monitoring Mode and Record Only Mode. |
| Night Mode<br>Button | Image: NIGHT MODE | To help the patient sleep soundly, the Laptop Base Station display screen and lights on the sEMG Monitor may be<br>dimmed. Click the Night Mode button to turn Night Mode ON. |
| Night Mode<br>On Button | Image: NIGHT MODE ON | Click the dark grey Night Mode ON button to disable Night Mode setting and return the SPEAC System<br>display screen to regular brightness. You can alternate between Night Mode OFF and Night Mode ON as<br>often as necessary. |
| Electrode Setup<br>Button | Image: ELECTRODE SETUP | Click the Electrode Setup button every time a new Electrode Patch is placed on a patient. Calibration of a new<br>Electrode Patch should be performed daily when using the System in Seizure Monitoring Mode or Record Only Mode. |
| Alarm History<br>Button | Image: ALARM HISTORY | Click the Alarm History button to display a list of all Seizure Alarms and Operational Alerts that have occurred<br>during the prescribed monitoring period. Use the scroll bar to review the cumulative list. A new or existing<br>Seizure Diary entry can be accessed from this list, as well as using the Seizure Diary button noted below. |
| Seizure Diary<br>Button | Image: SEIZURE DIARY | A yellow exclamation point will appear over the Seizure Diary button to indicate a new Seizure Diary entry has<br>been automatically created. Click the Seizure Diary Button to open the electronic form that will allow the<br>patient or caregiver to record specific details about the event. |
| Wi-Fi-Status<br>Indicator Bar | Image: Wi-Fi signal strength bars | The Laptop Base Station serves as the control hub for the entire SPEAC System. The Wireless Router provides<br>a communications link between the sEMG Monitor in use and the Laptop Base Station. The Wireless Router<br>can communicate with the sEMG Monitor up to a 200-foot range (approximate). The Wi-Fi Status Bars indicate<br>connection signal strength. |
## Intended Use
The SPEAC® System is intended for use as an adjunct to seizure monitoring in adults in the home or healthcare facilities during periods of rest. The non-EEG Physiological Signal Based Seizure Monitoring System continuously records and stores surface electromyographic (sEMG) data for subsequent review. Trained healthcare professionals may use the electrophysiological sEMG data during a post-hoc review, with other contextual data, to characterize upper-extremity motor activity ipsilateral to the device from other activity. Audio data recorded during seizure monitoring may be available for review by a trained healthcare professional.
{7}------------------------------------------------
The device is to be used on the belly of the biceps muscle to analyze sEMG signals. When sEMG signal patterns associated with a unilateral, appendicular, tonic extension that could be associated with a GTC seizure are detected, the SPEAC System sends adjunctive alarms to alert caregivers. Adjunctive alarms may be disabled by a physician order while continuing to record sEMG data for subsequent review.
## Clinical Performance Testing
Prospective clinical trials entitled, "Using sEMG to Identify Epileptic Seizure Semiology" (Study 1), and "Differentiation of Epileptic and Psychogenic Non-Epileptic Seizures Using Single-Channel Surface Electromyography'' (Study 2) evaluated the ability of trained clinicians to differentiate and characterize different types of seizure events using the sEMG data collected by the SPEAC System.
Once characterized, the results of these sEMG-trained healthcare professionals were compared to characterizations of the same seizure events using EEG data interpreted by epileptologists and compared to automated processing for characterization of the seizures.
# Clinical Inclusion and Exclusion Criteria
Enrollment in the studies was limited to patients who met the following selection criteria: Patients with a history of epileptic seizures or psychogenic non epileptic seizures with positive motor involvement of the upper extremities admitted to the EMU for routine seizure characterization were eligible for inclusion. Subjects were not eligible if intracranial EEG monitoring was used.
## Training of Physicians in the Study
Physicians were provided with a brief (less than an hour) sEMG data training session describing characteristics of: voluntary activity, epileptic and non-epileptic seizure events, and artifact from the device. False positives were non-seizure events that could have resulted from either voluntary movement, signal spikes or RF noise from loose electrodes, or ambient background signals with no associated event data. Physicians were asked to evaluate normalized, transformed, and raw sEMG data. Epileptologists reviewing sEMG data were blinded to all other data.
{8}------------------------------------------------
Clinical Endpoints. Trained clinicians were evaluated by determining if they could characterize and differentiate types of seizure events with sEMG data collected by the SPEAC System, and then those results were compared to vEEG data as well as automated event characterization.
#### Study Methods
#### Expert Review
Both Study 1 (CLN0055.001) and Study 2 (CLN0053.001) simultaneously captured vEEG and sEMG data, and three of the total four sEMG reviewers were the same in both studies. The goal in Study 1 was to review epileptic seizures and try to describe the semiology. The goal in Study 2 was to review both epileptic and non-epileptic seizures and to try to properly classify events into those two categories. The similarity between these studies is that epileptic seizures included GTC seizures (referred to as TC seizures in the studies), and focal seizures that contained clonic jerking and complex motor activity (automatisms). Tonic seizures were only recorded during Study 1, and PNES was only recorded during Study 2.
The data from the three common sEMG reviewers was combined from both studies to assess the accuracy for the intended goal in both studies. This analysis combines data that has two different goals but testing the same types of events. The combined accuracy for the studies given a "committee" style approach (majority rules, 2/3) was also analyzed to make the reviews more comparable to the format for used for vEEG.
#### Results
## Expert Reviewer Accuracy
The accuracy from the three reviewers that participated in both studies was combined for each event. The accuracies are presented for each individual reviewer as well as a committee approach for each event. The objective of Study 1 was to categorize ES by semiology. The objective of Study 2 was to categorize seizures as epileptic or non-epileptic. The overall accuracy for individual assessments was 83% (95% CI [0.78 0.88], n = 243), and the overall accuracy for the committee approach was 86% (95% CI [0.77 0.93], n = 81). Table 1 summarizes the accuracies from the reviewers. These results support how the device can be used by clinicians to characterize upper-extremity motor activity (UEMA) ipsilateral to the device from other activity.
{9}------------------------------------------------
| Committee Review of<br>sEMG Classification | Reviewer 1 | Reviewer 2 | Reviewer 3 | Average (Ind.<br>Assess.)<br>(N = n*3) a | Committee |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|------------------------------|------------------------------|------------------------------------------|------------------------------|
| Tonic-Clonic (n = 26) | 88% (95% CI<br>[0.70 0.98]) | 88% (95% CI<br>[0.70 0.98]) | 100% (95% CI<br>[0.87 1.00]) | 92% (95% CI<br>[0.84 0.97]) | 92% (95% CI<br>[0.75 0.99]) |
| Simple Motor ES (Tonic<br>& Clonic, n = 14) | 71% (95% CI<br>[0.42 0.92]) | 79% (95% CI<br>[0.49 0.95]) | 64% (95% CI<br>[0.35 0.87]) | 71% (95% CI<br>[0.55 0.84]) | 79% (95% CI<br>[0.49 0.95]) |
| Complex Motor ES<br>("Other", n = 22) | 73% (95% CI<br>[0.50 0.89]) | 68% (95% CI<br>[0.45 0.86]) | 63% (95% CI<br>[0.40 0.83]) | 68% (95% CI<br>[0.56 0.79]) | 73% (95% CI<br>[0.50 0.89]) |
| All ES (n = 62) | 79% (95% CI<br>[0.67 0.88]) | 79% (95% CI<br>[0.67 0.88]) | 79% (95% CI<br>[0.67 0.88]) | 79% (95% CI<br>[0.72 0.85]) | 82% (95% CI<br>[0.70 0.91]) |
| PNES, whole body<br>involvement (n = 4) | 100% (95% CI<br>[0.40 1.00]) | 100% (95% CI<br>[0.40 1.00]) | 100% (95% CI<br>[0.74 1.00]) | 100% (95% CI<br>[0.67 0.88]) | 100% (95% CI<br>[0.40 1.00]) |
| PNES, arm<br>jerks/hand tremors<br>only (n = 15) | 100% (95% CI<br>[0.78 1.00]) | 93% (95% CI<br>[0.68 1.00]) | 93% (95% CI<br>[0.68 1.00]) | 96% (95% CI<br>[0.85 0.99]) | 100% (95% CI<br>[0.78 1.00]) |
| All PNES (n = 19) | 100% (95% CI<br>[0.82 1.00]) | 95% (95% CI<br>[0.74 1.00]) | 95% (95% CI<br>[0.74 1.00]) | 96% (95% CI<br>[0.88 1.00]) | 100% (95% CI<br>[0.82 1.00]) |
| Overall Accuracy (n =<br>81) | 84% (95% CI<br>[0.74 0.91]) | 83% (95% CI<br>[0.73 0.90]) | 83% (95% CI<br>[0.73 0.90]) | 83% (95% CI<br>[0.78 0.88]) | 86% (95% CI<br>[0.77 0.93]) |
| a - The sample size for this column is: TC: 78, Simple Motor ES: 42, Complex Motor ES: 66, All ES: 186,<br>PNES whole body: 12, PNES arm jerks/hand tremors: 45, PNES: 57, Total: 243 | | | | | |
Table 1– Accuracy of (Common) Expert Reviewers' Classification of Events from Studies 1 and 2
The device used in the studies which represent the data collected were the same as the commercial version of the SPEAC System previously cleared by the FDA. sEMG collected using the SPEAC System has been used by clinicians to characterize seizures. Data provided by the SPEAC System provides healthcare professionals a diagnostic tool based on bicep sEMG signal monitoring to characterize seizure events measured in a home or hospital setting.
### Non-clinical Test Summary
| Test | Summary Results | FDA Recognition Number |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007) (3rd Edition), Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance | Passed | 19-4 |
| IEC 60601-1-2: 2014-02 (4th Edition), Medical Electrical Equipment – Part 1-2: General Requirements for Basic Safety and Essential Performance, Collateral Standard, Electromagnetic Compatibility | Passed | 19-8 |
| Test | Summary Results | FDA Recognition Number |
| IEC 60601-1-6: Medical electrical equipment - Part 1-6:<br>general requirements for basic safety and essential<br>performance - collateral standard: usability | Passed | 5-89 |
| IEC 60601-1-8: Medical electrical equipment - Part 1-8:<br>general requirements for basic safety and essential<br>performance - collateral standard: general requirements,<br>tests, and guidance for alarm systems in medical<br>electrical equipment and medical electrical systems | Passed | 5-76 |
| IEC 60601-1-11: 2010 (1st Edition) General<br>requirements for basic safety and essential performance<br>- Collateral Standard: Requirements for medical<br>electrical equipment and medical electrical systems used<br>in the home healthcare environment. | Passed | This version is no longer FDA<br>recognized but the changes<br>evaluated in this 510(k)<br>submission do not require<br>additional testing to account for<br>this lack of recognition. Changes<br>to the subject device proposed<br>(indications for use) are outside<br>the scope of this test. |
| IEC 60601-2-40: 2016 (2nd Edition) Medical electrical<br>equipment - Part 2-40: Particular requirements for the<br>basic safety and essential performance of<br>electromyographs and evoked response equipment | Passed | N/A Not an FDA recognized<br>standard; This testing has been<br>voluntarily performed by the<br>company to demonstrate<br>performance and validate the<br>system. |
| Biocompatibility Testing<br>ISO 10993-1: Biological evaluation of medical devices -<br>Part 1: Evaluation and testing within a risk management<br>process. | Passed | 2-220 |
| ISO 10993-5: Biological evaluation of medical devices -<br>Part 5: Tests for vitro cytotoxicity | Passed | 2-245 |
| ISO 10993-10: Biological evaluation of medical devices<br>- Part 10: Tests for irritation and skin sensitization | Passed | 2-174 |
| Shipping and Distribution Test (ISTA 3A) | Passed | 5-110 |
| Software Verification and Validation (Appendix H) | Passed | N/A |
{10}------------------------------------------------
# SPEAC System Comparison to Predicate Device and Conclusion:
{11}------------------------------------------------
| Parameter | SPEAC System | SPEAC System | Comparison | |
|-----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|
| 510(k) | K200276 | K182180 | | |
| Indications for<br>Use | The SPEAC® System is intended<br>for use as an adjunct to seizure<br>monitoring in adults in the home<br>or healthcare facilities during<br>periods of rest.<br>The non-EEG Physiological<br>Signal Based Seizure Monitoring<br>System continuously records and<br>stores surface electromyographic<br>(SEMG) data for subsequent<br>review.<br>Trained healthcare professionals<br>may use the electrophysiological<br>sEMG data during a post-hoc<br>review, with other contextual data,<br>to characterize upper-extremity<br>motor activity (UEMA) ipsilateral | The SPEAC® System is indicated<br>for use as an adjunct to seizure<br>monitoring in adults in the home<br>or healthcare facilities during<br>periods of rest.<br>The System records and stores<br>surface electromyographic<br>(sEMG) data for subsequent<br>review by a trained healthcare<br>professional.<br>The device is to be used on the<br>belly of the biceps muscle to<br>analyze sEMG signals that may be<br>associated with generalized tonic-<br>clonic (GTC) seizures. | Similar with new<br>information added;<br>The indications<br>have been modified<br>to clarify the<br>intended use by<br>specifically stating<br>that the device may<br>be used to<br>characterize upper<br>extremity motor<br>activity by a trained<br>healthcare<br>professional. For | |
| | to the device from other activity.<br>Audio data recorded during<br>seizure monitoring may be<br>available for review by a trained<br>healthcare professional.<br>The device is to be used on the<br>belly of the biceps muscle to<br>analyze sEMG signals.<br>When sEMG signal patterns<br>associated with a unilateral,<br>appendicular, tonic extension that<br>could be associated with a GTC<br>seizure are detected, the SPEAC<br>System sends adjunctive alarms to<br>alert caregivers.<br>Adjunctive alarms may be<br>disabled by a physician order<br>while continuing to record sEMG<br>data for subsequent review. | When sEMG signal patterns<br>associated with a unilateral,<br>appendicular, tonic extension that<br>could be associated with a GTC<br>seizure are detected, the SPEAC<br>System sends adjunctive alarms to<br>alert caregivers.<br>Adjunctive alarms may be<br>disabled by a physician order<br>while continuing to record sEMG<br>data for subsequent review. | clarification, Brain<br>Sentinel removed<br>"that may be<br>associated with<br>generalized tonic-<br>clonic (GTC)<br>seizures" in the<br>second paragraph<br>because the<br>SPEAC System<br>analyzes all sEMG<br>to alarm for<br>potential GTC<br>seizures. These<br>changes are<br>supported by<br>clinical<br>performance testing<br>included in the<br>submission. All<br>other claims were<br>preciously cleared<br>in K182180 (the | |
| Parameter | SPEAC System | SPEAC System | Comparison | |
| 510(k) | K200276 | K182180 | | |
| Classification | Class II<br>21 CFR 882.1580<br>Non-EEG physiological<br>signal-based seizure<br>monitoring system | Class II<br>21 CFR 882.1580<br>Non-EEG physiological<br>signal-based seizure<br>monitoring system | Same, No change | |
| Product Code | POS | POS | Same, No change | |
| Principle of<br>Operation | sEMG signal, sampled at the<br>rate of 1000 Hz, from the<br>biceps brachii are processed<br>by a proprietary algorithm to<br>identify sustained sEMG<br>contraction patterns—during<br>the tonic phase and early<br>transition to the clonic<br>phase—that are<br>pathognomonic of GTC<br>seizures. | sEMG signal, sampled at the<br>rate of 1000 Hz, from the<br>biceps brachii are processed<br>by a proprietary algorithm to<br>identify sustained sEMG<br>contraction patterns—during<br>the tonic phase and early<br>transition to the clonic<br>phase—that are<br>pathognomonic of GTC<br>seizures. | Same, No change | |
| Dimensions | 3.44" x 2.34" x 1.33"<br>(H x W x D) | 3.44" x 2.34" x 1.33"<br>(H x W x D) | Same, No change | |
| Mass | 127 g. | 127 g. | Same, No change | |
| Physical<br>Controls | Power On/Off Button<br>(Manual) Alarm Button<br>Cancel Button | Power On/Off Button<br>(Manual) Alarm Button<br>Cancel Button | Same, No change | |
| sEMG sampling<br>rate | 1,000 Hz | 1,000 Hz | Same, No change | |
| sEMG<br>Frequency Bands<br>of Interest | 30-40 Hz, 130-240 Hz, and<br>300-400 Hz | 30-40 Hz, 130-240 Hz, and<br>300-400 Hz | Same, No change | |
| Event Monitored | Potential GTC Seizure | Potential GTC Seizure | Same, No change | |
| Default Alarm<br>Threshold | 135 | 135 | Same, No change | |
| sEMG Monitor<br>Power | Rechargeable Battery | Rechargeable Battery | Same, No change | |
| Parameter | SPEAC System | SPEAC System | Comparison | |
| 510(k) | K200276 | K182180 | | |
| Model | COTS Laptop | COTS Laptop | Same, No change | |
| Input Power | 115/230 VAC 50/60 Hz | 115/230 VAC 50/60 Hz | Same, No change | |
| Graphical User<br>Interface | Laptop Base Station | Laptop Base Station | Same, No change | |
| Software<br>Controls | Alert Mode (adjunctive seizure detection and sEMG recording),<br>Record Only (sEMG recording) | Alert Mode (adjunctive seizure detection and sEMG recording),<br>Record Only (sEMG recording) | Same, No change | |
| sEMG Electrode | sEMG Electrode | sEMG Electrode | Same, No change | |
| Cellular Router | Cradlepoint Router | Cr…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.