K200272 · Terragene S.A. · JOJ · Jun 30, 2020 · General Hospital
Device Facts
Record ID
K200272
Device Name
Terragene Chemdye
Applicant
Terragene S.A.
Product Code
JOJ · General Hospital
Decision Date
Jun 30, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Intended Use
Terragene Chemdye® Bowie-Dick Test Pack (BD125X/1), Bowie-Dick Test Card (BD8948X/1) and Bowie-Dick Test Sheet (BDA4/1) are designed for testing air removal efficiency for cycles of 132ºC for 3.5 minutes in dynamic-air-removal steam sterilizers.
Device Story
Chemical indicators designed to monitor air removal efficiency in dynamic-air-removal steam sterilizers. Devices (pack, card, sheet) placed in sterilizer chamber; exposed to steam cycle (132ºC for 3.5 minutes). Chemical indicator changes color upon exposure to specific steam conditions; provides visual confirmation of adequate air removal and steam penetration. Used by healthcare personnel in clinical settings to verify sterilizer performance. Results interpreted visually by operator; failure to achieve expected color change indicates potential sterilizer malfunction or inadequate air removal, prompting investigation before further use of the sterilizer.
Clinical Evidence
Bench testing only. Performance verified through testing of air removal efficiency in dynamic-air-removal steam sterilizers at 132ºC for 3.5 minutes. No clinical data required for this device type.
Technological Characteristics
Chemical indicator system consisting of a pack, card, or sheet. Operates via chemical color change reaction upon exposure to steam and temperature. Designed for 132ºC, 3.5-minute dynamic-air-removal steam sterilization cycles. Non-electronic, passive device.
Indications for Use
Indicated for testing air removal efficiency in dynamic-air-removal steam sterilizers using 132ºC for 3.5-minute cycles.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
K093199 — 3M COMPLY BOWIE- DICK TYPE LEAD FREE TEST PACK · 3M Company · Dec 4, 2009
K960441 — STERITEC INTEGRAPH MODEL CI-101 · Steritec Products Mfg Co, Inc. · May 8, 1997
K132372 — STERITEC GREEN CARD PACK BOWIE-DICK TEST · Steritec Products Mfg Co, Inc. · Nov 26, 2013
K122945 — STERITEC GREEN CARD PACK BOWIE-DICK TEST MODEL BD 126 · Steritec Products Mfg Co, Inc. · Apr 30, 2013
K120592 — DANA REUSABLE BOWIE TEST PACK FOR DANA BOWIE DICK INDICATORS · Dana Products, Inc. · Feb 27, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 30, 2020
Terragene SA c/o.Raymond Kelly, Consultant Licensale Inc 3422 Leonard Ln. New Smyrna Beach FL 32168 USA
Re: K200272
Trade/Device Name: Terragene Chemdye® BD125X/1 Bowie-Dick Test Pack, Terragene Chemdye® BD8948X/1 Bowie-Dick Test Card and Terragene Chemdye® BDA4/1 Bowie-Dick Test Sheet. Regulation Number: 21 CFR 880.2800 Regulation Name: Indicator, Physical/Chemical Sterilization Process Regulatory Class: Class II Product Code: JOJ Dated: Mav 29, 2020
Dear Raymond Kelly:
Received: June 01, 2020
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Christopher K. Dugard, M.S. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
### K200272
#### Device Name
Terragene Chemdye® BD125X/1 Bowie-Dick, Terragene Chemdye® BD8948X/1 Bowie-Dick Test Card and Terragene Chemdye® BDA4/1 Bowie-Dick Test Sheet.
#### Indications for Use (Describe)
Terragene Chemdye® Bowie-Dick Test Pack (BD125X/1), Bowie-Dick Test Card (BD8948X/1) and Bowie-Dick Test Sheet (BDA4/1) are designed for testing air removal efficiency for cycles of 132ºC for 3.5 minutes in dynamicair-removal steam sterilizers.
Type of Use (Select one or both, as applicable)
| | Production Use (Part 21 CFR 601 Subpart D) |
|--|-----------------------------------------------|
| | Study Tier Control Use (21 CFR 601 Subpart G) |
__ Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.